Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K211649

    Validate with FDA (Live)

    Device Name
    PCEA Syringe Set
    Date Cleared
    2021-11-23

    (179 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.

    Device Description

    The proposed device is a patient controlled epidural analgesia (PCEA) syringe set. It is a single use disposable device intended for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device. It is a non-pyrogenic, sterile device that can be directly attached to a syringe. The PCEA syringe set consists of non-DEHP PVC tubing/bushing, female NRFit cap, female NRFit lock, anti-siphon valve, on-off clamp, male NRFit lock, and male NRFit cap. It is used to administer anesthetics and/or analgesics from a syringe to the patient's epidural space through an epidural access device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Baxter Healthcare Corporation PCEA Syringe Set (K211649), based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    ISO 80369-6 Tests on Male NRFit (Neuraxial Fit) Lock ConnectorISO 80369-6: 2016, Clause 5 (as applicable), and Clauses 6.1 through 6.6All tests met the acceptance criteria.
    ISO 80369-6 Tests on Female NRFit (Neuraxial Fit) Lock ConnectorISO 80369-6: 2016, Clause 5 (as applicable), and Clauses 6.1 through 6.6All tests met the acceptance criteria.
    Tensile Strength TestISO 8536-9:2015, Clause 5.3All tests met the acceptance criteria.
    Leak TestISO 8536-9:2015, Clause A.4All tests met the acceptance criteria.
    Counter Flow TestISO 8536-12:2007+A1:2013, Clause A.4All tests met the acceptance criteria.
    Blocking Performance TestISO 8536-12:2007+A1:2013, Clause A.6All tests met the acceptance criteria.
    Opening Pressure TestISO 8536-12:2007+A1:2013, Clause A.7.1 and per Baxter Test MethodAll tests met the acceptance criteria.
    Particulate Matter TestUSP Chapter <788>All tests met the acceptance criteria.
    Clamp Activation Force TestPer Baxter Test MethodAll tests met the acceptance criteria.
    Clamp Shut-Off TestISO 8536-14:2018, Clause A.1All tests met the acceptance criteria.
    Non-DEHP Claim VerificationPer Baxter Test Method (as tested in K161808)All tests met the acceptance criteria.
    BiocompatibilityISO-10993-1 categorization (for prolonged contact, external communicating device, tissue/bone/dentin) and FDA-2013-D-0350 Guidance. Specific assays: Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Intracutaneous (Irritation) Reactivity (ISO 10993-10), Acute Systemic Toxicity (ISO 10993-11), Material Mediated Pyrogen (ISO 10993-11), Genotoxicity (ISO 10993-3), Subacute Systemic Toxicity (ISO 10993-11).The device met the designated ISO 10993-1 categorization and is biocompatible and appropriate for the intended use.
    SterilitySterility Assurance Level (SAL) of ≤ 10-6 using ANSI/AAMI/ISO 11137-2. Endotoxin limit of 2.15 EU/device per USP <161>.MSD (Minimum Sterilizing Dose) required to provide a 10-6 SAL was established and validated. Routine periodic pre-sterilization bioburden testing is performed. (Implied to have met criteria, as the conclusion states substantial equivalence).
    Shelf LifeNot explicitly stated as a numerical acceptance criterion in the table, but "supports a shelf-life claim of 3 (three) years."Aging testing performed to support a 3-year shelf-life claim.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document describes "bench tests" and does not specify a separate "test set" in the context of clinical data, human readers, or image analysis. The tests appear to be conducted on the device itself. The provenance of the data is from Baxter Healthcare Corporation's internal testing. There is no information regarding country of origin or retrospective/prospective nature as this refers to device performance testing rather than clinical study data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    Not applicable. The "ground truth" for these tests relates to engineering and manufacturing standards (e.g., ISO standards, USP chapters) rather than human expert interpretation of medical images or conditions.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests are based on objective measurements against established engineering and regulatory standards, not human opinion requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical administration set, not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" is established by:

    • International and national standards bodies (e.g., ISO 80369-6, ISO 8536-9, ISO 8536-12, ISO 8536-14, ISO 10993, ANSI/AAMI/ISO 11137-2)
    • Pharmacopeial standards (e.g., USP Chapter <788>, USP <161>)
    • Baxter Test Methods, which are company-specific internal procedures developed to ensure quality and compliance.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as no training set is used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1