510(k) Data Aggregation

    K Number
    K244044

    Validate with FDA (Live)

    Date Cleared
    2025-03-28

    (88 days)

    Product Code
    Regulation Number
    866.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PBC Separator with Selux AST System is an automated inoculum preparation system that uses lysis, centrifugation and sequential optical density measurements to generate a McFarland-equivalent suspension from positive blood culture samples that can be used for quantitative in vitro antimicrobial susceptibility testing by the Selux AST System. Samples are processed directly from blood culture samples identified as positive by a continuous monitoring blood culture system. Samples should be confirmed as monomicrobial, gram negative rods or gram positive cocci by Gram stain. Organism identification is required for AST result interpretation and reporting, per the Selux AST System Instructions for Use.

    Inoculum preparation by the PBC Separator was evaluated for use with the Selux AST System and the Selux AST Gram Negative Panel. Performance was demonstrated for the antimicrobial agents and organisms identified below:

    • Amikacin: Acinetobacter baumannii complex, Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa

    · Amoxicillin-Clavulanate: Escherichia coli, Klebsiella species (including K. oxytoca, K. pneumoniae), Proteus mirabilis, Proteus vulgaris

    · Ampicillin: Escherichia coli, Proteus mirabilis

    · Ampicillin-Sulbactam: Acinetobacter baumannii complex, Citrobacter koseri, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis

    • Cefazolin: Escherichia coli, Klebsiella pneumoniae

    · Cefepime: Citrobacter freundii complex, Citrobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    • Ceftazidime: Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa

    · Ceftazidime-Avibactam: Citrobacter freundii complex, Citrobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    · Ceftriaxone: Citrobacter freundii complex, Citrobacter cloacae complex, Escherichia coli,

    Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Serratia marcescens

    • Ciprofloxacin: Citrobacter freundii complex, Citrobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    • Ertapenem: Citrobacter freundii complex, Citrobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens

    • Gentamicin: Citrobacter freundii complex, Citrobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    · Imipenem: Acinetobacter baumannii complex, Escherichia coli, Klebsiella pneumoniae

    · Meropenem: Acinetobacter baumannii complex, Citrobacter koser, Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    · Minocycline: Acinetobacter baumannii complex, Escherichia coli, Klebsiella pneumoniae

    • Piperacillin-Tazobactam: Acinetobacter baumannii complex, Citrobacter koseri, Escherichia coli, Klebsiella pneumoniae, Morganella morganii, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    • Tobramycin: Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa

    Inoculum preparation by the PBC Separator was evaluated for use with the Selux AST System and the Selux AST Gram Positive Panel. Performance was demonstrated for the antimicrobial agents and organisms identified below:

    • Ampicillin: Enterococcus faecalis, Enterococcus faecium
    • Ceftaroline: Staphylococcus aureus
    • Daptomycin: Enterococcus faecalis, Enterococcus faecium, Staphylococcus aureus
    • · Linezolid: Enterococcus faecalis, Enterococcus faecium, Staphylococcus aureus
    • Oxacillin: Staphylococcus aureus
    • Vancomycin: Enterococcus faecalis, Enterococcus faecium, Staphylococcus aureus

    The PBC Separator with Selux AST System Gram Positive Panel is a qualitative test for the following antimicrobial agents with the specific target organisms identified below:

    • · Cefoxitin Screen to predict mecA-mediated oxacillin resistance: Staphylococcus aureus
      Susceptibility test results are intended to be used in conjunction with other clinical and laboratory findings. Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive is necessary for the susceptibility testing of organisms present in polymicrobial samples, for testing antimicrobial agents and species not indicated for testing with the device, for epidemiologic testing, and for recovery of organisms present in microbial samples.
    Device Description

    The Positive Blood Culture (PBC) Separator with Selux AST System is an automated sample preparation instrument with associated consumables that uses lysis, centrifugation, and sequential optical density measurements to prepare a tuned McFarland-equivalent inoculum from positive blood culture bottles that have rung positive on a continuous monitoring blood culture system. Inoculums containing monomicrobial, gram negative or gram positive bacteria are used for Antimicrobial Susceptibility Testing (AST) processing with the Selux AST System. The Selux AST System includes a sample prep station (i.e., AST Workbench), an Inoculator, an Analyzer, Workbench Computer, and the reagents and consumables required to perform AST testing. The PBC Separator and all Selux AST System components are connected to a site workstation, which coordinates sample processing on all instruments. The PBC Separator contains embedded software and a graphical user interface that guides users through the PBC Separator workflow. Once processing of the PBC sample is complete, the user transfers the tuned McFarland inoculum to the Selux AST System for further AST processing.

    The PBC Separator with Selux AST System can only provide AST results for monomicrobial samples. Since the PBC Separator with Selux AST System does not perform identification (ID), the monomicrobial nature of the sample under test must be confirmed by an FDA-cleared directfrom-positive blood culture ID system.

    While PBC Separator processing can be performed without species-level ID, this information is required for the Selux AST System to interpret and report susceptibility results. Species ID can be performed by any appropriate method and this information can be either manually input to the Selux AST System or automatically downloaded from the laboratory information system (LIS) at any time, once the sample ID is entered into the LIS.

    The PBC Separator with the Selux AST System utilizes a 384-well panel, either the Selux Gram-Negative Panel or Selux Gram-Positive Panel, that provides parallel results for the antimicrobials indicated for each sample type. The Selux AST System software masks non-indicated results. The average time-to-result for positive blood culture processed with the PBC Separator and Selux AST System is under 7 hours.

    Principle of Operation

    The PBC Separator automatically prepares a tuned bacterial inoculum directly from a blood culture bottle sample that "rang" positive on an FDA-cleared continuous monitoring blood culture system, including the Becton Dickinson BACTEC, the bioMerieux BacT/Alert 3D, and the bioMerieux Virtuo. The PBC Separator removes contaminants through repeated centrifugation-wash cycles and specific chemical lysis of mammalian cells and cell fragments. The PBC Separator utilizes an on-board spectrometer to tune the inoculum for the right cell density to perform AST.

    Tuned inoculums are used with the Selux AST System. The Selux AST System performs AST similarly to the reference broth microdilution method by first incubating samples, then quantifying microbial growth in each well of an antimicrobial dilution series after a growth period, and finally determining the minimum inhibitory concentration (MIC) by comparing growth data in each well,

    AST testing of PBC samples requires that the Gram type (classification) of the organism be known prior to testing on the Selux AST System as the information is necessary to select the proper AST panel to use. Organism identification (ID) is not needed to initiate testing with the Selux AST System. However, the organism ID is necessary for a result to be interpreted and reported because the MIC-determining algorithm is species-specific as is the interpretative Susceptible (S), Susceptible Dose Dependent (SDD), Intermediate (I), or Resistant (R) determination. Any FDAcleared method may be used to provide an ID including biochemical techniques, matrix-assisted laser desorption/ionization mass spectrometry, and multiplex genetic assays.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving device performance, based on the provided FDA 510(k) summary for the PBC Separator with Selux AST System:

    Acceptance Criteria and Device Performance Study

    The document outlines analytical and clinical studies to demonstrate the performance of the PBC Separator with Selux AST System, particularly for gram-positive organisms, compared to a reference method (broth microdilution).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents performance in terms of Essential Agreement (EA), Category Agreement (CA), Very Major Errors (VMJ), Major Errors (MAJ), and Minor Errors (MIN). While explicit "acceptance criteria" are not listed as a single table, the overall performance metrics, particularly the high percentages for EA and CA, and low error rates, indicate the expected and achieved levels of performance. For reproducibility, quantitative acceptance criteria were explicitly stated.

    Reproducibility Acceptance Criteria:

    • Best-case reproducibility: ≥95%
    • Worst-case reproducibility: ≥89% (for inter- and intra-site)

    Essential Agreement (EA) Acceptance Criteria (Implicit from "meets performance criteria"):
    The overall impression from the "Blood Culture Bottle Compatibility Study" and "Interfering Substances Testing" sections is that EA should be >89.9%.
    For the "Clinical Studies" section, while not explicitly stated as a numerical threshold, FDA typically expects high EA and CA (e.g., >90-95%) and very low VMJ/MAJ errors (<1.5% and <3% respectively) for AST devices. The reported values meet these implicit expectations for most combinations.

    Test TypeMetricAcceptance Criteria (Stated/Inferred)Reported Device Performance (Range/Specific Value)
    Reproducibility
    Intra-SiteBest-Case≥95%≥97% (Range: 98.9% - 100%)
    Worst-Case≥89%≥97% (Range: 97.8% - 100%)
    Inter-SiteBest-Case≥95%≥95% (Range: 95.6% - 100%)
    Worst-Case≥89%>94% (Range: 94.8% - 100%)
    Cefoxitin ScreenInter-site (modal)N/A (matched modal result)99.3%
    Intra-site (modal)N/A (matched modal result)100%
    Bottle CompatibilityEssential Agreement (EA)>89.9%Aerobic bottles: 99.4% EA (range ≥96.7%); Anaerobic bottles: 99.3% EA (range ≥96.7%)
    Cefoxitin Screen100% agreement expected100% agreement, except one instance demonstrating 66.7% (2/3) out-of-agreement in BD BACTEC Aerobic Plus bottles.
    Interfering SubstancesEssential Agreement (EA)>89.9%All endogenous interferents: 83.3% - 100% EA. All exogenous interferents: 66.7% - 100% EA.
    Cefoxitin ScreenNo interference (100% agreement)100% agreement.
    Clinical PerformanceEssential Agreement (EA)High (e.g., >90-95%)Range: 94% - 100% (Individual combos: 75% for Ampicillin Enterococci Eval, 0% for Daptomycin E. faecalis Eval, 47.6% for Vancomycin Enterococci Eval)
    Category Agreement (CA)High (e.g., >90-95%)Range: 94.6% - 100%
    Very Major Errors (VMJ)Low (e.g., <1.5%)0% (for all combinations reported)
    Major Errors (MAJ)Low (e.g., <3%)Range: 0% - 2% (for Daptomycin E. faecalis)
    Minor Errors (MIN)LowRange: 0% - 6% (for Ceftaroline Staphylococci)

    2. Sample Size and Data Provenance

    Test Set Sample Size:

    • Reproducibility:
      • Intra-site: For each of 9 antimicrobials, 5 samples tested in triplicate (3 different inoculums) on 3 days at each site yields 9 results per sample, 45 results per antimicrobial at a single site.
      • Inter-site: For each of 9 antimicrobials, 5 samples tested in triplicate (3 different inoculums) on 3 days across 3 sites yields 27 results per sample, 135 results per antimicrobial.
    • Post-Positivity Sample Stability: 30 results (3 replicates of one species from each indicated antimicrobial/organism reporting group at 16-hour timepoints compared to 0-hour).
    • Blood Culture Bottle Compatibility: 3 replicates per bottle type for one isolate from each reporting group with all claimed antimicrobials. Total of 180 results for aerobic bottle types (179 in EA). Total of 150 results for anaerobic bottle types. Specific breakdown by bottle type table provided.
    • Interfering Substances Testing: 3 replicates per interferent type for E. faecalis and S. aureus. Specific breakdown per antimicrobial/interferent in tables provided (e.g., 3/3, 6/6 results).
    • Clinical Studies: 247 clinical isolates (64 fresh, 183 seeded) and 75 challenge isolates. Total datapoints varied per antimicrobial-organism combination, ranging from 100 to 211.

    Data Provenance:

    • Country of Origin: Test sites are stated as "three test sites (2 external, 1 internal)" and "clinical sites to represent geographic diversity across the continental U.S." This indicates United States origin.
    • Retrospective or Prospective:
      • Reproducibility, Post-Positivity, Bottle Compatibility, Interfering Substances: These studies appear to be prospective as they involve specific seeding, incubation, and processing protocols.
      • Clinical Studies: Used "fresh positive blood culture samples left over from routine clinical care" (suggests prospective collection of clinical samples) and "banked frozen isolates seeded" (suggests retrospective isolates used in a prospective study design).

    3. Number of Experts and Qualifications for Ground Truth

    The document does not specify the number of experts or their qualifications for establishing ground truth.

    For susceptibility testing (AST), the "ground truth" or reference method is typically broth microdilution, performed at an independent reference laboratory. The establishment of ground truth for AST is a standardized laboratory procedure, not typically reliant on expert consensus as in image interpretation. The proficiency of the laboratory performing the reference method is assumed, rather than specific qualifications of individuals being listed.

    4. Adjudication Method for the Test Set

    Not applicable. For AST studies, the reference method (broth microdilution) results are considered the ground truth, and there is no human adjudication process involved in comparing the device's MIC results to the reference MIC results. The comparison is quantitative for essential agreement and categorical for category agreement and other error rates.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. The device is an automated in vitro diagnostic (IVD) system for antimicrobial susceptibility testing, not an AI system for image interpretation that would involve human readers. Therefore, an MRMC study is not relevant here.

    6. Standalone (i.e. Algorithm-Only) Performance

    Yes, the studies inherently demonstrate "standalone" performance. The PBC Separator is an automated inoculum preparation system, and the Selux AST System then performs the AST. The "performance" tables (reproducibility, bottle compatibility, interfering substances, clinical performance) show the device's output (MIC values) directly compared to the reference method, without human intervention in the interpretation of the device's results.

    7. Type of Ground Truth Used

    The ground truth used for performance evaluation is reference broth microdilution results (MIC values and categorical interpretations), performed at an independent reference laboratory. This is the gold standard for antimicrobial susceptibility testing.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" or "training data" size for the device's internal algorithms. This is typical for IVD medical devices, where the focus for regulatory submission is on the validation (test) data demonstrating device performance. The algorithms within the Selux AST System for determining MICs and interpretations are based on established microbiological principles (growth kinetics, optical density measurements) and potentially internal development/calibration data, which are not typically disclosed or quantified as a 'training set' in the same way as machine learning models.

    9. How the Ground Truth for the Training Set was Established

    As noted above, a distinct "training set" with ground truth establishment is not typically described for this type of IVD device. The inherent "ground truth" for the development of such a system would involve extensive laboratory work correlating optical density measurements, growth curves, and other internal parameters with known MICs from reference methods (broth microdilution) over various organisms and antimicrobial agents. This is part of the extensive research and development process for such a complex automated system rather than a separately documented "training set" for regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223493

    Validate with FDA (Live)

    Date Cleared
    2024-02-15

    (451 days)

    Product Code
    Regulation Number
    866.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PBC Separator with Selux AST System is an automated inoculum preparation system that uses lysis, centrifugation and sequential optical density measurements to generate a McFarland-equivalent suspension from positive blood culture samples that can be used for quantitative in vitro antimicrobial susceptibility testing by the Selux AST System. Samples are processed directly from blood culture samples identified as positive by a continuous monitoring blood culture system. Samples should be confirmed as monomicrobial, gram negative rods by Gram stain. Organism identification is required for AST result interpretation and reporting, per the Selux AST System instructions for use.

    Inoculum preparation by the PBC Separator was evaluated for use with the Selux AST System and the Selux Gram Negative Panel. Performance was demonstrated for the antimicrobial agents and organisms identified below:

    Amikacin: Acinetobacter baumannii complex, Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa

    Amoxicillin-Clavulanate: Escherichia coli, Klebsiella species (including K. oxytoca, K. pneumoniae), Proteus mirabilis, Proteus vulgaris

    Ampicillin: Escherichia coli, Proteus mirabilis

    Ampicillin-Sulbactam: Acinetobacter baumannii complex, Citrobacter koseri, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis

    Cefazolin: Escherichia coli, Klebsiella pneumoniae

    Cefepime: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa Ceftazidime: Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa

    Ceftazidime-Avibactam: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    Ceftriaxone: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella pneumoniae, Proteus mirabilis, Serratia marcescens

    Ciprofloxacin: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    Ertapenem: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens

    Gentamicin: Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa Imipenem: Acinetobacter baumannii complex, Escherichia coli, Klebsiella pneumoniae

    Meropenem: Acinetobacter baumannii complex, Citrobacter freundii complex, Citrobacter koseri, Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii. Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa Minocycline: Acinetobacter baumannii complex, Escherichia coli, Klebsiella pneumoniae

    Piperacillin-Tazobactam: Acinetobacter baumannii complex, Citrobacter koseri, Escherichia coli, Klebsiella pneumoniae, Morganii, Proteus mirabilis, Proteus vulgaris, Serratia marcescens, Pseudomonas aeruginosa

    Tobramycin: Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa

    Susceptibility test results are intended to be used in conjunction with other clinical and laboratory findings. Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive blood culture is necessary for the susceptibility testing of organisms present in polymicrobial samples, for testing antimicrobial agents and species not indicated for testing with the device, for epidemiologic testing, and for recovery of organisms present in microbial samples.

    Device Description

    The Positive Blood Culture (PBC) Separator with Selux AST System is an automated sample preparation instrument with associated consumables that uses lysis, centrifugation, and sequential optical density measurements to prepare a tuned McFarland-equivalent inoculum from positive blood culture bottles that have rung positive on a continuous monitoring blood culture system. Inoculums containing monomicrobial, gram negative bacteria are used for AST processing with the Selux AST System. The Selux AST System includes a sample prep station (i.e., AST Workbench), an Inoculator, an Analyzer, Workbench Computer, and the reagents and consumables required to perform AST testing. The PBC Separator and all Selux AST System components are connected to a site workstation, which coordinates sample processing on all instruments. The PBC Separator contains embedded software and a graphical user interface that guides users through the PBC Separator workflow. Once processing of the PBC sample is complete, the user transfers the tuned McFarland inoculum to the Selux AST System for further AST processing.

    The PBC Separator with Selux AST System can only provide AST results for monomicrobial samples. Since the PBC Separator with Selux AST System do not perform identification (ID), the monomicrobial nature of the sample under test must be confirmed by an FDA-cleared direct-frompositive blood culture ID system.

    While PBC Separator processing can be performed without species-level ID, this information is required for the Selux AST System to interpret and report susceptibility results. Species ID can be performed by any appropriate method and this information can be either manually input to the Selux AST System or automatically downloaded from the laboratory information system (LIS) at any time, once the sample ID is entered into the LIS.

    The PBC Separator with the Selux AST System utilizes the Selux Gram Negative Panel, a 384well panel that provides parallel results for the antimicrobials indicated for each sample type. The Selux AST System software masks non-indicated results. The average time-to-result for positive blood culture processed with the PBC Separator and Selux AST System is under 7 hours.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the PBC Separator with Selux AST System, based on the provided FDA 510(k) summary:

    Acceptance Criteria and Device Performance

    The core performance criteria for the PBC Separator with Selux AST System relate to its ability to accurately determine antimicrobial susceptibility. The primary metric used is Essential Agreement (EA), which measures how closely the MIC results from the device match the reference method, and Category Agreement (CA), which assesses agreement in susceptibility interpretations (Susceptible, Intermediate, Resistant - SIR).

    Overall Acceptance Criterion: The device must meet performance criteria for each indication, generally interpreted as high percentages of Essential Agreement (EA) and Category Agreement (CA) (typically >90% as seen in the tables for acceptable overall performance, though individual instances below 90% might be deemed acceptable based on the totality of data).

    1. Table of Acceptance Criteria and the Reported Device Performance

    Performance MetricAcceptance Criterion (Implicit)Reported Device Performance (Summary)Notes
    Intra-Site Reproducibility≥ 95% Best- and Worst-Case≥ 95% Best- and Worst-Case for all tested antimicrobials (Min. 98.1%)Achieved for all antimicrobials
    Inter-Site Reproducibility> 95% Best- and Worst-Case> 95% Best- and Worst-Case for all tested antimicrobials (Min. 95.8%)Achieved for all antimicrobials
    Post-Positivity Sample Stability (16 hr)> 95% Essential Agreement99.6% (264/265 results in EA)Achieved overall; individual antimicrobial EA >95%
    Blood Culture Bottle Compatibility (Aerobic)> 89.9% Essential Agreement for all tested bottle types99.3% overall EA (1629/1640); each aerobic bottle type ≥ 98.5% EAExcept for one case (Ciprofloxacin/Enterobacterales in bioMérieux BacT/ALERT SA) which was 89.3%, but deemed acceptable.
    Blood Culture Bottle Compatibility (Anaerobic)> 89.9% Essential Agreement for all tested bottle types99.5% overall EA (974/979); each anaerobic bottle type ≥ 98.5% EAExcept for a few cases below 90% (Ciprofloxacin/Enterobacterales in BD BACTEC Lytic Anaerobic (85.7%) and bioMerieux BacT/ALERT FN Plus (92.3%); Ampicillin-Sulbactam/Enterobacterales in BD BACTEC Lytic Anaerobic (92.9%)), but deemed acceptable.
    Endogenous Interferents (MIC EA)> 89.9% Essential Agreement for every interferent> 89.9% Essential Agreement for every interferent tested (most combinations were 100%)Except for Piperacillin-Tazobactam with K. pneumoniae in all endogenous interferents, which showed <90% EA and very major errors, and Tobramycin with P. aeruginosa (80-92.9%), which were noted as having less than 90% EA but deemed acceptable.
    Exogenous Interferents (MIC EA)> 89.9% Essential Agreement for every interferent> 89.9% Essential Agreement for every interferent tested (most combinations were 100%)Except for Cefepime/Enterobacterales, Imipenem/Enterobacterales, Minocycline/Enterobacterales (with some interferents), Piperacillin-Tazobactam with K. pneumoniae in all exogenous interferents (<90% EA and very major errors), and Tobramycin/Enterobacterales with Cefpodoxime (77.8%), flagged as having less than 90% EA but deemed acceptable.
    Carry-Over/Cross-Contamination> 89.9% Essential Agreement100% for E. coli (60/60); 98.2% for K. pneumoniae (54/55). No contamination on purity plates.Achieved.
    Clinical Performance (EA)Generally high percentage agreement, individual variations noted.Varied by antimicrobial-organism combination (e.g., Amikacin/A. baumannii (complex): 92.1% EA; Ceftazidime/P. aeruginosa: 100% EA)Performance demonstrated for all combinations although some EA values were below 90% in the "Total Eval" category (e.g. Amikacin/Enterobacterales: 66.7%).
    Clinical Performance (CA)Generally high percentage agreement.Varied by antimicrobial-organism combination (e.g., Amikacin/A. baumannii (complex): 94.7% CA; Ceftazidime/P. aeruginosa: 100% CA)Some EA values were below 90% in the "Total Eval" category which might have an impact on the CA, although not explicit in the table.
    QC Testing95% performance criteriaMet for all antimicrobialsAchieved.

    Study Details

    The document primarily describes analytical and clinical studies for the performance evaluation of the PBC Separator with Selux AST System.

    2. Sample Size Used for the Test Set and Data Provenance

    • Reproducibility:
      • Intra-site: Minimum of 45 results per antimicrobial (5 samples * 3 triplicates * 3 days).
      • Inter-site: Minimum of 135 results per antimicrobial (5 samples * 3 triplicates * 3 days * 3 sites).
      • Data Provenance: Not explicitly stated as retrospective or prospective, but involves seeding isolates into fresh human blood and processing, suggesting a controlled laboratory setting. The "3 sites (2 external, 1 internal)" suggests internal and possibly US/international external sites.
    • Post-Positivity Sample Stability: 265 results comparing 16-hour processing to 0-hour processing.
      • Data Provenance: Fresh human blood from a healthy donor.
    • Blood Culture Bottle Compatibility:
      • Aerobic: 1640 results across 11 bottle types.
      • Anaerobic: 979 results across 11 bottle types.
      • Data Provenance: Seeded bacterial samples at clinically relevant concentrations into blood culture bottles with manufacturer-recommended volumes of healthy donor human blood.
    • Interfering Substances Testing:
      • Each interferent tested involved "at least one species for each reporting group for each antimicrobial."
      • Data Provenance: Healthy donor blood used to seed interferents and bacteria.
    • Carry-Over/Cross-Contamination Study: 5 E. coli and 5 K. pneumoniae positive blood culture samples; 5 AST panels for each organism.
      • Data Provenance: Seeded isolates.
    • Clinical Studies:
      • Total Isolates: 469 clinical (162 fresh and 307 seeded) and 87 challenge isolates.
      • Organisms: 12 Enterobacterales species, Acinetobacter baumannii complex, and Pseudomonas aeruginosa.
      • Antimicrobials: 17.
      • Total Data Points: Varied from 38 to 469 per antimicrobial-organism combination.
      • Data Provenance:
        • Fresh clinical samples: "left over from routine clinical care" from two clinical sites in New York City. This indicates retrospective use of fresh samples collected in a clinical setting.
        • Seeded samples: "banked frozen isolates seeded... into blood culture bottles together with approximately 10 mL of fresh human blood from a healthy donor." These seeded samples were chosen "to represent geographic diversity across the continental U.S." This component is laboratory-based but designed to represent real-world diversity.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and their Qualifications

    The document does not explicitly state the number of experts or their qualifications for establishing ground truth. However, it indicates that triplicate broth microdilution results performed at an independent reference laboratory were used as the reference method. This implies that the ground truth was established by laboratory personnel in a CLIA-certified or equivalent reference lab, following a recognized standard for AST.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1, 3+1) for the interpretation of test results or discrepancies. The comparison is made directly between the device's results and the reference broth microdilution results. Any discrepancies would be evaluated against established acceptance criteria, but no formal expert adjudication process is detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is for antimicrobial susceptibility testing, which does not involve human readers interpreting images or data alongside AI. The device provides a direct microbiological result (MIC and SIR category).

    6. If a Standalone Performance Study (algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are primarily standalone performance studies for the device. The "PBC Separator with Selux AST System" directly processes samples and generates MICs and SIR interpretations. While requiring inputs like Gram stain and organism identification before result interpretation, the AST process itself (lysis, centrifugation, optical density measurements, and MIC determination) is automated by the device without real-time human intervention in the result generation. The performance is compared to a reference standard (broth microdilution), reflecting the algorithm's output.

    7. The Type of Ground Truth Used

    The ground truth used for all performance evaluations (reproducibility, stability, compatibility, interferents, clinical performance) was broth microdilution (BMD), which is the gold standard reference method for antimicrobial susceptibility testing. For the clinical studies, it specifies "triplicate broth microdilution results performed at an independent reference laboratory."

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. The provided information focuses entirely on the validation and testing of the device's performance, not its development or training data.

    9. How the Ground Truth for the Training Set Was Established

    Since the document does not mention a training set or its size, it also does not describe how the ground truth for a training set was established.