Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K083693

    Validate with FDA (Live)

    Date Cleared
    2009-04-02

    (111 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 840 Ventilator System with Leak Compensation Option is intended to provide continuous ventilation to patients requiring respiratory support. The device is intended for patients with an Ideal Body Weight (IBW) as low as 0.5 kg (with NeoMode option) to adult, and for use in a wide variety of clinical conditions.

    The 840 Ventilator System with Leak Compensation Options is intended for a wide range of patients ranging from neonate to adult (V- 5-2500 mL with NeoMode) or from infant to adult (V-25-2500 mL).

    The 840 Ventilator System with Leak Compensation is intended for use in hospital and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.

    Device Description

    The Leak Compensation Option enables the 840 Ventilator System to automatically detect and characterize gas flow that leaves the breathing circuit through air leak(s) and does not return to the 840's exhalation valve for inclusion in the displayed exhaled spirometry. LC does not makeup for lost delivered volume but does add the necessary additional flow into the breathing circuit to optimize breath triggering and cycling functions as well as maintain the set PEEP, and display clinically valuable leak-related parameters for consideration by the clinician. The Leak Compensation feature is implemented on the 840 Ventilator through additional functionality in software and by use of the existing User Interface panel. No hardware or firmware changes or additions were required. The 840 Ventilator is a dual-microprocessor controlled, critical care ventilator intended to provide continuous ventilation for neonate to adult (with NeoMode Option) or infant to adult (without NeoMode Option) patients who require either invasive ventilation or non-invasive ventilation (via face mask).

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Puritan-Bennett 840 Ventilator System with Leak Compensation Option, focusing on acceptance criteria and study details.

    Important Note: The provided document is a 510(k) summary, which is a high-level overview. It generally does not contain the granular detail of a full study report (like sample sizes for specific tests, expert qualifications, or adjudication methods). Therefore, for several points, the answer will be that the information is not provided in this summary.


    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not provide a table with specific, quantitative acceptance criteria (e.g., "leak compensation accuracy must be within X%"). Instead, it states that the device "passed all tests" against its stated performance specifications and applicable guidance.

    Acceptance Criteria (General)Reported Device Performance
    Meeting stated performance specifications"Functional testing confirms that the 840 Ventilator with Leak Compensation Ventilator is capable of meeting its stated performance specifications. The device passed all tests."
    Compliance with applicable portions of "Draft Reviewer Guidance for Ventilators" (July 1995)"Testing confirms that the 840 Ventilator with Leak Compensation Ventilator complies with the applicable portions of the July 1995 'Draft Reviewer Guidance for Ventilators'... The device passed all applicable tests."
    Compliance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 2005)"All software is tested in accordance with the 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices' dated May 2005. The devices passed all tests."
    Safety and Effectiveness"We conclude that the 840 Ventilator with Leak Compensation meets the stated performance specifications and criteria referenced above and that the device and its accessories will operate safely in its intended environment and will be effective in fulfilling the intended use."

    Study Details

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in this 510(k) summary. The document mentions "functional testing," "testing confirms," and "all software is tested," implying internal verification and validation activities, but does not quantify the sample sizes for these tests (e.g., number of test cases, number of ventilation cycles simulated).
    • Data Provenance: Not specified. The testing described appears to be internal engineering and software verification/validation, not a clinical study with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts & Qualifications: Not applicable/Not specified. The testing described focuses on functional performance and software compliance, not diagnostic accuracy against expert-established ground truth. These types of tests are typically conducted by engineers and quality assurance personnel against predefined performance specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable/Not specified. Adjudication methods are typically used in clinical studies or studies where multiple human readers interpret data, and their decisions need to be reconciled to establish a ground truth. This document describes functional and software testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study Done: No, an MRMC comparative effectiveness study was not described in this 510(k) summary.
    • Effect Size of Human Improvement: Not applicable, as no MRMC study was described. The device is a ventilator with a leak compensation feature, not an AI-assisted diagnostic tool that humans interpret.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, in essence, the "functional testing" and "software testing" described represent the standalone performance of the device's leak compensation algorithm and system. The ventilator itself operates automatically based on its algorithms to detect and compensate for leaks. The summary states: "Functional testing confirms that the 840 Ventilator with Leak Compensation Ventilator is capable of meeting its stated performance specifications." This refers to the device's performance without direct human intervention in the leak compensation process itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: For the functional and software testing, the "ground truth" would be the engineering specifications and design requirements for the leak compensation function, as well as the applicable regulatory standards and guidance documents (e.g., "Draft Reviewer Guidance for Ventilators"). The device was tested against these predefined performance targets.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable/Not specified. This device is a mechanical ventilator with control algorithms, not a machine learning or AI system that requires a "training set" in the conventional sense of supervised learning. Its algorithms are developed through engineering design, mathematical modeling, and control theory rather than statistical training on a dataset.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set Establishment: Not applicable, as there is no "training set" in the context of machine learning. The algorithms were designed and verified against established principles of fluid dynamics, respiratory physiology, and control system engineering, coupled with internal performance specifications and regulatory guidance.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1