Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071174

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-25

    (89 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PathFinder - Mini-Open Posterior Approach

    When intended for pedicle screw fixation from L1-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, and failed previous fusion.

    As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.

    After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking info consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.

    Device Description

    The Abbott Spine, Inc. PathFinder® system is a pedicle screw fixation system designed to allow for use of an open or mini open surgical technique. The approved Pathfinder® system consists of various screws and rods and is intended to provide temporary stabilization following surgery to fuse the spine.

    The subject device is the result of modifications to the existing PathFinder® 6.5mm polyaxial pedicle screws resulted in the inclusion of a 4.5mm polyaxial pedicle screw to the Pathfinder System. The subject device shares the same intended use and fundamental scientific technology as the predicate device.

    AI/ML Overview

    The Abbott Spine PathFinder® System is a spinal fixation system. The submission K071174 specifically concerns modifications to the existing PathFinder® 6.5mm polyaxial pedicle screws to include a 4.5mm polyaxial pedicle screw.

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Substantial equivalence to predicate deviceLaboratory and bench testing results demonstrate substantial equivalence to the predicate device.

    Study Details

    1. Acceptance Criteria and Reported Device Performance: See table above.

    2. Sample size used for the test set and the data provenance:

      • Sample Size: Not applicable. The submission states "NONCLINICAL PERFORMANCE AND CONCLUSION: Laboratory and bench testing results demonstrate that the proposed device is substantially equivalent to the predicate device. CLINICAL PERFORMANCE AND CONCLUSIONS: Clinical data and conclusions were not needed for this device." The evaluation relied on nonclinical (laboratory and bench) testing.
      • Data Provenance: Not applicable, as no clinical data was required or provided. The nonclinical testing would have been conducted in a laboratory setting.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical ground truth was established. The device's performance was evaluated through nonclinical laboratory and bench testing against the predicate device.

    4. Adjudication method for the test set: Not applicable, as no clinical ground truth requiring adjudication was established.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a mechanical spinal fixation system, not an AI-assisted diagnostic or therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is a mechanical implant, not an algorithm.

    7. The type of ground truth used: Not applicable for clinical ground truth. For the nonclinical evaluation, the "ground truth" was the
      performance characteristics and regulatory standards applied to the predicate device (PathFinder® System K030625). The new device's
      performance in laboratory and bench tests was compared to these established characteristics to demonstrate substantial equivalence.

    8. The sample size for the training set: Not applicable, as no clinical trials or machine learning models were involved.

    9. How the ground truth for the training set was established: Not applicable, as no clinical trials or machine learning models were involved.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1