510(k) Data Aggregation

    K Number
    K242804

    Validate with FDA (Live)

    Date Cleared
    2024-12-30

    (104 days)

    Product Code
    Regulation Number
    878.5040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aesthetic body contouring

    Device Description

    The MicroAire PAL® Aspiration System provides an adjustable vacuum source that drives the suction of tissue during suction-assisted aesthetic body contouring procedures. The PAL Aspiration System features a modular style PAL® Aspiration Pump (ASP-PUMP-120/240) unit with a separate, stackable PAL® Aspiration Pump Console (ASP-UI) which provides organizational modularity and ease of transport within the medical facility. Communication between the units is achieved via the Aspiration Pump Cable (ASP-CBL). The PAL Aspiration Pump utilizes a dual-piston pump as a vacuum source that delivers 0-29 inHg and drives the suction of tissue. The PAL Aspiration Pump also features a multi-socket outlet (MSO) to provide power to up to four devices using the AC Power Cord Jumper Cable (ASP-CORD), eliminating extra power cords for better cable organization. The PAL Aspiration Pump Console is an electronic control system which allows the operator to digitally set and visualize the vacuum level. It also provides the status of basic pump operations, sends the Pump's ON/OFF signal, and provides receptacles for the connection of the Aspiration Pump Cable and optional accessories. The Aspiration Foot Switch (ASP-FOOT) and the 5020 to Aspiration Pump Cable (5020-ASP) are optional accessories which provide the operator with additional Aspiration Pump ON/OFF control options. The PAL Aspiration Pump requires a Biofilter (ASP-FLTR2), which prevents the ingress of vapor particulate into the motor. The Biofilter is a non-sterile filter assembly with vacuum tubing placed on the vacuum inlet. Connection to a system of generic collection canisters (sold separately) is expected for the PAL Aspiration System to function as intended. Additional Vacuum Tubing (ASP-TUBING) can be used to replace existing Biofilter tubing as needed.

    AI/ML Overview

    The provided text is a 510(k) summary for the MicroAire PAL® Aspiration System, which is a medical device. This document focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria in the way a clinical trial or algorithm validation study would for AI/ML devices.

    Therefore, much of the requested information cannot be found in this type of regulatory submission. The device is a physical suction system, not an AI/ML device, which explains the absence of data related to AI/ML specific criteria.

    Here's an analysis based on the provided text, addressing the points where information is available:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria alongside reported device performance in a quantitative manner for specific benchmarks. Instead, it states that "The PAL Aspiration System passed all non-clinical bench testing, supporting its safety and effectiveness, and its substantial equivalence to the predicate device."

    It lists several voluntary consensus standards and FDA Guidance documents that the device complies with, which implicitly define the acceptance criteria for various aspects:

    Acceptance Criteria Category (Implied by Compliance)Reported Device Performance (Implied by "passed all tests")
    Electrical safety (IEC 60601-1)Met
    Electromagnetic Disturbance (IEC 60601-1-2 and IEC 60601-4-2)Met
    Essential Performance (IEC 60601-1-6)Met
    Software Verification (IEC 62304)Met
    Usability testing (IEC 62366-1)Met
    Safety testing for electrically powered suction pumps (ISO 10079-1)Met
    Vacuum performanceMet specified requirement
    Free air flowMet specified requirement
    NoiseMet specified requirement
    ReliabilityMet specified requirement

    Note on "Maximum Vacuum" and "Maximum Flow Rate": While these are reported as characteristics in Table A, they are presented as comparative figures to the predicate device, not necessarily as acceptance criteria that were quantitatively "met" against a pre-defined threshold.

    CharacteristicPredicate Device PerformanceSubject Device Performance
    Maximum Vacuum27.5 inHg29 inHg
    Maximum Flow Rate6.4 cfm2.12 cfm (60L/min) @ 60Hz 1.77 cfm (50L/min) @ 50Hz
    Sound Level50 dB56.8 dBA

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and not provided. The testing described is non-clinical bench testing of a physical device, not a study involving patient data or a "test set" in the context of AI/ML or clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. As above, there is no "test set" requiring expert-established ground truth in the context of this device's non-clinical bench testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. The device is a suction lipoplasty system, not an AI-assisted diagnostic or treatment system.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable and not provided. The device is a physical system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    This information is not applicable and not provided. For non-clinical bench testing, the "ground truth" would be established by validated measurement equipment and adherence to engineering specifications and international standards.

    8. The sample size for the training set

    This information is not applicable and not provided. There is no "training set" for this type of device.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided.