510(k) Data Aggregation

    K Number
    K230764

    Validate with FDA (Live)

    Date Cleared
    2023-04-18

    (29 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OsteoCentric Integrated Hip Fastener System is indicated for stable and unstable intertrochanteric, subtrochanteric, and basilar neck fractures in which a stable medial buttress can be reconstructed.

    The OsteoCentric Integrated Hip Fastener System is indicated for femoral neck fractures including intracapsular, transcervical, and subcapital fractures.

    Device Description

    The OsteoCentric Integrated Hip Fastener System is internal fixation of intertrochanteric and femoral neck fractures of the proximal femur. The OsteoCentric Integrated Hip Fastener System consists of plates and screws in a variety of lengths and diameters to accommodate different anatomic sizes of patients. Integrated Hip Fasteners which are lag screws implanted through the femoral neck and head are offered in Ø11, Ø13, & Ø15mm sizes with lengths ranging from 50 – 140mm in 5mm increments. Hip side plates are offered in 1-hole to 19-hole configurations (48 – 330mm length). 6.0mm support fasteners are offered in 55 – 125mm lengths. 5.2mm shaft fasteners are offered in 28 – 46mm locking fasteners for the trochanter plate are offered in 16 – 40mm lengths. Compression and attachment screws are also available. Devices are provided non-sterile. All implantable devices are manufactured from stainless steel per ASTM F138.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the OsteoCentric Integrated Hip Fastener System. It does not describe an AI/ML medical device or a study involving human readers and AI assistance for diagnostic purposes. Instead, it concerns a traditional medical device (metallic bone fixation appliances) and its substantial equivalence to previously cleared predicate devices.

    Therefore, the specific information requested about acceptance criteria, study design, expert involvement, and reader performance for an AI/ML device is not present in the provided text. The document focuses on the physical characteristics of the hip fastener system (e.g., sizes, materials) and its intended use for treating hip fractures.

    To answer your request, if this were an AI/ML device submission, the FDA would require detailed information on its performance, typically including:

    1. Acceptance Criteria and Performance Table: This would outline the metrics (e.g., sensitivity, specificity, AUC) the device needed to achieve and the reported performance.
    2. Sample Size and Data Provenance: Details on the number of cases in the test set, their origin (e.g., multi-site, geographical distribution), and whether the data was collected retrospectively or prospectively.
    3. Ground Truth Experts: The number and qualifications of clinicians who established the definitive diagnosis for the test cases.
    4. Adjudication Method: How disagreements among experts were resolved to form the ground truth (e.g., majority vote, senior expert arbitration).
    5. MRMC Study: If applicable, a human-in-the-loop study comparing diagnostic accuracy with and without AI assistance, quantifying the improvement.
    6. Standalone Performance: The algorithm's performance without human interaction.
    7. Type of Ground Truth: Whether the ground truth was based on expert consensus, pathology, long-term outcomes, or other definitive sources.
    8. Training Set Sample Size: The number of cases used to train the AI model.
    9. Training Set Ground Truth: How the ground truth for the training data was established.

    Since this document pertains to a mechanical medical device, none of this information is relevant or provided within the text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210247

    Validate with FDA (Live)

    Date Cleared
    2021-04-30

    (91 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OsteoCentric Integrated Hip Fastener System is intended to treat stable and unstable intertrochanteric, subtrochanteric, and basilar neck fractures in which a stable medial buttress can be reconstructed.

    Device Description

    The OsteoCentric Integrated Hip Fastener System is intended for use in internal fixation of femoral neck fractures. The OsteoCentric Integrated Hip Fastem consist of plates and screws in a variety of lengths and diameters to accommodate different anatomic sizes of patients. The devices are provided non-sterile. All implantable devices are manufactured from Stainless Steel per ASTM F138.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "OsteoCentric Integrated Hip Fastener System." This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with strict acceptance criteria and performance metrics.

    Therefore, the document does not contain the kind of information typically requested for AI/ML device studies, such as:

    • A table of acceptance criteria and reported device performance (in terms of sensitivity, specificity, AUC, etc.)
    • Sample size used for the test set and data provenance (e.g., country of origin, retrospective/prospective)
    • Number of experts and their qualifications used to establish ground truth
    • Adjudication method
    • Multi-reader multi-case (MRMC) comparative effectiveness study
    • Standalone (algorithm-only) performance
    • Type of ground truth used (pathology, outcomes data, etc.)
    • Sample size for the training set
    • How ground truth for the training set was established

    Instead, this document describes an engineering analysis for a mechanical device.

    Here's what the document does provide regarding the "study" that proves the device meets the "acceptance criteria" (using these terms loosely in the context of a mechanical device submission):

    1. Acceptance Criteria and Reported Device Performance (Mechanical Context):

    The acceptance criteria here are implicitly demonstrating substantial equivalence to predicate devices in terms of mechanical properties for the intended use.

    Acceptance Criteria (Implicit)Reported Device Performance (Engineering Analysis Conclusion)
    Plate strength and stiffness sufficient for intended use and comparable to predicates.Plate strength and stiffness was compared by evaluating the material properties and moment of inertia at worst-case locations. The results of the engineering analysis confirmed that the strength, stiffness... of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
    Screw strength sufficient for intended use and comparable to predicates.Screw strength was evaluated and compared to the predicates using the material properties, screw types and sizes. The results of the engineering analysis confirmed that the... strength... of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
    Screw fixation sufficient for intended use and comparable to predicates (e.g., thread profile provides adequate fixation).Screw fixation was evaluated and compared to predicates using information about the thread profile. The results of the engineering analysis confirmed that the... fixation of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
    Overall mechanical performance equivalent to legally marketed predicate devices."The overall technology characteristics and mechanical performance data lead to the conclusion that the OsteoCentric Integrated Hip Fastener System is substantially equivalent to the predicate device." "The results of the engineering analysis confirmed that the strength, stiffness and fixation of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices."
    Materials of manufacture (Stainless Steel per ASTM F138) are appropriate and similar to predicates.Implied by "All implantable devices are manufactured from Stainless Steel per ASTM F138" and "Materials of Manufacture" being identical between subject and predicates.
    Similar technological characteristics to predicates (e.g., modular stainless steel bone fixation system with similar interlocking features, plate angulation, lengths, widths, thicknesses, screw sizes, screw lengths)."The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates: Indications for Use, Materials of Manufacture." Also, "the subject plating/screw construct is also a modular stainless steel bone fixation system with similar interlocking features, plate angulation, plate lengths, plate widths, plate thicknesses, screw sizes and screw lengths."

    2. Sample size used for the test set and the data provenance:

    • Test Set: No discrete "test set" in the context of an AI/ML algorithm is mentioned. The assessment was based on an "engineering analysis" comparing physical properties and design features of the subject device to predicate devices. This involves material properties, moment of inertia, screw types/sizes, and thread profiles.
    • Data Provenance: Not applicable in the context of an engineering analysis of mechanical components. The data would come from specifications, material testing standards (e.g., ASTM F138), and design comparisons.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth for mechanical properties is established through engineering principles, material science, and testing standards, not expert consensus in the medical imaging sense. The submission was prepared by Nathan Wright MS, an Engineer & Regulatory Specialist at Empirical Testing Corp.

    4. Adjudication method for the test set:

    • Not applicable. There was no clinical or interpretative "test set" requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a mechanical bone fixation device, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a mechanical device.

    7. The type of ground truth used:

    • Engineering specifications and standards: The "ground truth" for this device's performance relies on established engineering principles, material science (e.g., Stainless Steel per ASTM F138 specifications), and comparative analysis against the known mechanical properties of legally marketed predicate devices. The aim is to show that the new device's mechanical integrity is equivalent or superior to existing devices for the same indications.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" in the context of this type of mechanical device submission.

    9. How the ground truth for the training set was established:

    • Not applicable.