510(k) Data Aggregation

    K Number
    K152556

    Validate with FDA (Live)

    Device Name
    Oragene Dx
    Manufacturer
    Date Cleared
    2016-05-26

    (261 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Oragene Dx devices are intended for use in the non-invasive collection of saliva samples. Human DNA from the saliva sample is isolated, stabilized, and suitable for use in FDA cleared molecular diagnostic applications. Saliva may be collected by spitting directly into the Oragene Dx container or may be transferred into the Oragene Dx container using a sponge. Saliva samples collected using Oragene Dx are stabilized and can be transported and/or stored long term at ambient conditions.

    Device Description

    The Oragene-Dx family of products offers reliable collection, transportation and long-term room temperature storage of human DNA from saliva. Oragene Dx is a non-invasive alternative for collecting high quality and quantity DNA for use in molecular diagnostic applications. Oragene-Dx is a device family available in multiple device formats or models.

    Oragene-Dx device formats OGD-510, OGD-600, OGD-610 and OGD-675, have the same collection principle and intended use as the FDA cleared Oragene-Dx formats (K110701). All Oragene-Dx formats consist of a collection tube, a DNA stabilizing liquid and optional sponges for assisted collection. In addition, Oragene·Dx device formats are made from the same physical and chemical materials. Oragene Dx formats differ in the amount of DNA stabilizing liquid in the tube and in the difference in the amount of saliva to be collected. The ratio of final sample to stabilizing liquid volume remains the same.

    Saliva collection can take place at home, in a laboratory setting, physician's office, or in the field by untrained (naïve) or professional users. Saliva samples are collected into the device directly by spitting or by using the provided sponges. After saliva is collected, the stabilizing liquid is mixed with the sample. Upon contacting saliva cells, the stabilizing liquid lyses cellular and nuclear membranes to release and stabilize nucleic acids (DNA). Oragene-Dx Samples can be immediately processed, transported or stored for future use. Samples can be shipped at ambient temperature to the laboratory for processing.

    Oragene-Dx device pre-collection shelf life is 24 months at room temperature from the date of manufacture. Post collection, Oragene Dx samples are stable at room temperature for up to 12 months. Oragene Dx device performance and sample integrity are maintained during typical ambient transport and storage conditions.

    AI/ML Overview

    This document (K152556) describes a 510(k) premarket notification for the Oragene®Dx OGD-510, Oragene®Dx OGD-600, Oragene®Dx OGD-610, and Oragene®Dx OGD-675 devices. These devices are intended for the non-invasive collection of saliva samples, from which human DNA can be isolated, stabilized, and used in FDA-cleared molecular diagnostic applications. The submission references studies performed for the predicate device (Oragene®Dx OGD-500, K110701) to support the substantial equivalence of the new devices.

    1. Table of acceptance criteria and reported device performance:

    Since the document bases its claims on substantial equivalence to a predicate device (K110701) and applicable studies mentioned in that predicate, specific numerical acceptance criteria for the new devices are not explicitly stated in a tabular format within this document. Instead, the document states that the performance of the new devices is "the same as predicate" or that "both OGD-510 and OGD-500 samples met acceptance criteria for DNA concentration, total DNA yield, A260/A280 ratio and performance on the eSensor Warfarin Sensitivity Saliva Test."

    The key performance characteristics and their reported outcomes, primarily referencing the predicate K110701, are summarized below:

    Performance CharacteristicAcceptance Criteria (Implied / Referenced from K110701)Reported Device Performance (New Devices OGD-510, OGD-600, OGD-610, OGD-675)
    Reproducibility/PrecisionMeeting study acceptance criteria (as per K110701)Demonstrated (referencing Section 18.4 of K110701). Overall, all samples tested met study acceptance criteria.
    Pre-collection Shelf-Life24 months at room temperatureSupported by studies in K110701 (same physical/chemical components).
    12 months at -20±5°C and 6±4°CSupported by studies in K110701 (same physical/chemical components).
    Post-collection Sample Stability12 months at room temperature, -20±5°C or 6±4°CSupported by studies in K110701 (same physical/chemical components).
    3 months at 50±5°CSupported by studies in K110701 (same physical/chemical components).
    Sample Volume ToleranceNo impact on performance from underfilling by 25-50% or overfilling by 50% of target volume (as per K110701)Demonstrated (referencing Section 18.2 of K110701). Downstream performance was not affected by over or under spitting.
    Interfering SubstancesNo impact on performance from endogenous/exogenous substances, 100% agreement between eSensor® Warfarin Sensitivity Test genotyping and bidirectional sequencing (as per K110701)Demonstrated (referencing Section 18.6 of K110701), with the caveat of donor collection instructions (no eating, drinking, smoking, or chewing gum 30 min prior to collection).
    Analytical Performance (DNA Concentration, Yield, A260/A280, Genotyping Agreement with eSensor Warfarin Sensitivity Saliva Test)Meeting acceptance criteria for DNA concentration, total DNA yield, A260/A280 ratio, and performance on the eSensor Warfarin Sensitivity Saliva Test.For OGD-510 (and applicable to OGD-610): Both OGD-510 and OGD-500 samples met acceptance criteria for these parameters. No significant difference in any performance parameters (except total sample DNA yield).

    2. Sample size used for the test set and the data provenance:

    The document refers to studies in the predicate K110701 for most performance characteristics. For the new comparison study performed for OGD-510 vs. OGD-500, the sample size is not explicitly mentioned. The data provenance is implied to be from Canada, as the submitter, DNA Genotek Inc., is located in Ottawa, Ontario, Canada. The studies are retrospective references to the predicate device or a new comparison study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not explicitly provided in the document. The ground truth for comparative performance (genotyping) would likely be established by the "bidirectional sequencing" method, but details on the experts involved in interpreting this or verifying the ground truth for other studies (e.g., stability) are not given.

    4. Adjudication method for the test set:

    Not applicable or provided for this type of device and performance testing. The "ground truth" for genotyping involved comparison to bidirectional sequencing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a sample collection and stabilization device, not an AI-assisted diagnostic tool requiring human reader interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical device for sample collection, not an algorithm. The "performance" refers to the ability to collect and preserve DNA for downstream molecular diagnostic applications. The comparison studies against the eSensor® Warfarin Sensitivity Saliva Test (K110786) are standalone evaluations of the collected sample's compatibility with a molecular diagnostic test.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the analytical performance (genotyping agreement), the ground truth was established by bidirectional sequencing. For other performance characteristics like stability and reproducibility, the ground truth would be based on established laboratory methods and controls to demonstrate the integrity and functionality of the DNA.

    8. The sample size for the training set:

    Not applicable. This device is a physical sample collection device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. No training set is involved for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141410

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-02-19

    (266 days)

    Product Code
    Regulation Number
    862.1675
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Oragene®●Dx OGD-500.001 is intended for use in the non-invasive collection of saliva samples. DNA from the saliva sample is isolated, stabilized, and suitable for over-the-counter use with FDA cleared, approved, or legally marketed exempt DNA carrier screening genotyping tests. Saliva samples collected using Oragene®Dx OGD-500.001 are stabilized and can be transported and/or stored long term at ambient conditions.

    Device Description

    Oragene Dx family of collection devices offers reliable collection, transportation and longterm ambient temperature storage of human DNA from saliva. Oragene Dx OGD-500.001 is a noninvasive alternative for collecting high quality and quantity DNA and is suitable for use in over-thecounter molecular genotyping applications. Oragene-Dx OGD-500.001 consists of a collection tube with a funnel lid attached (containing a stabilizing liquid). Saliva is delivered directly by spitting into the collection tube. Saliva collection can take place at home, in a laboratory setting, physician's office, or in the field. Untrained (naïve) or professional users can carry out saliva collection. After saliva is collected, the stabilizing liquid is mixed with the sample. A small cap is provided to close the tube for transport and storage (funnel with lid is removed and discarded). Upon contacting saliva cells, the stabilizing liquid lyses cellular and nuclear membranes to release and stabilize nucleic acids (DNA). Samples can be immediately processed, transported or stored for future use. Samples can be shipped at ambient temperature to the laboratory for processing. Oragene-Dx samples are stable at room temperature for up to 12 months. Device and sample integrity are preserved during typical ambient transport and storage conditions.

    AI/ML Overview

    The document describes the Oragene® Dx OGD-500.001 device, a saliva collection kit, and its performance characteristics for substantial equivalence to a predicate device (Oragene® Dx OGD-500).

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly listed in a single table with numerical targets, but rather are inferred from the performance characteristics studies conducted and the claims made. The summary demonstrates that the device performs equivalently or acceptably based on these studies.

    Acceptance Criteria (Inferred)Reported Device Performance (Oragene® Dx OGD-500.001)
    Reproducibility/Precision: Reliable genotyping across multiple operators, days, and reagent lots.Demonstrated reproducibility when used with the 23andMe PGS Carrier Screening Test (details in co-submission DEN140044).
    Pre-collection Shelf-life: Stability for extended periods.24 months at room temperature; 12 months at -20°C ± 5°C and 6°C ± 4°C. (Supported by K110701, as device components are identical to predicate).
    Post-collection Sample Stability: DNA yield, concentration, A260/A280 ratio, and microbial content maintained.12 months at room temperature, -20°C ± 5°C or 6°C ± 4°C; 3 months at 50°C ± 5°C. (Supported by K110701, as device components are identical to predicate).
    Sample Volume Tolerance: Functionality maintained with varying saliva volumes.Underfilling by 25% or 50%, or overfilling by 50% did not impact performance. Mean total sample volume in consumer study: 4.25 ± 0.56 mL (equivalent to 2.25 mL saliva). 97.7% of samples were within acceptable range (2.58 to 5.64 mL total sample).
    Limit of Detection (DNA Concentration): Sufficient DNA for downstream testing.98.3% of samples in the consumer study contained the minimum amount of DNA required for testing. Re-extraction performed if initial concentration is low.
    Interfering Substances (Endogenous): No impact on genotyping from natural saliva components.100% concordant genotype calls for all samples and all endogenous substance conditions tested (salivary α-amylase, hemoglobin, immunoglobulin A, combination).
    Interfering Substances (Exogenous): No impact on genotyping from common oral activities/substances.100% concordant genotype calls for all samples collected immediately after and 30 minutes after consuming exogenous substances (food, drinking, chewing gum, mouthwash) or smoking.
    Consumer Usability/Comprehension (OTC Setting): Ability of naïve users to correctly collect samples and obtain sufficient DNA.98.3% of samples from 302 consumer study participants contained the minimum amount of DNA required for testing. User comprehension of test instructions was assessed.
    Labeling Readability: Instructions are easily understood by general public.Flesch-Kincaid reading grade level of 7.1 obtained for collection device labeling.

    2. Sample Size Used for the Test Set and Data Provenance

    • Reproducibility/Precision: Specific sample size not provided in this document, but referred to co-submission DEN140044 for the 23andMe PGS Carrier Screening Test.
    • Interfering Substances (Endogenous): 10 individuals for saliva samples.
    • Interfering Substances (Exogenous): 5 individuals at 3 time points (baseline, 0 min, 30 min) for each substance; smoking tested in a separate study (sample size not specified for smoking or overall study).
    • Consumer User Study: 302 individuals completed the survey and provided a saliva sample. The majority (96.3%) were naïve users.

    Data Provenance (Retrospective or Prospective / Country of Origin):

    • The studies regarding pre-collection shelf-life and post-collection sample stability and sample volume tolerance refer back to information established in the predicate device's 510(k) (K110701). This suggests a mix of potentially existing data from the predicate and current validation for the proposed device where applicable.
    • The reproducibility, interfering substances, and consumer user study were performed in conjunction with the 23andMe PGS Carrier Screening Test (co-submission DEN140044). The study design (e.g., collecting samples before/after substances) indicates these were prospective studies. The country of origin for the data is not explicitly stated, but 23andMe is a US-based company, which might imply data from the US market. The submitter is DNA Genotek Inc., located in Canada.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (a sample collection kit) does not typically involve expert review for "ground truth" in the same way an imaging or diagnostic algorithm would. The ground truth in these studies is the actual measured performance of the device in yielding DNA, maintaining DNA quality, and ensuring accurate genotyping by a downstream FDA-cleared test. The "experts" would be the laboratory personnel and analytical methods used for DNA extraction, quantification, and genotyping, verifying aspects like DNA yield, purity, and concordance with known genetic markers. No specific number or qualifications of "experts" are provided as a distinct category for establishing ground truth, as the ground truth is based on laboratory measurements and analytical results.

    4. Adjudication Method for the Test Set

    Not applicable for a sample collection device. Adjudication methods like "2+1" (two readers agree, third resolves) are typically used in clinical imaging or diagnostic studies where subjective interpretation is involved. Here, the outcomes are objective laboratory measurements (DNA concentration, genotype concordance).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC comparative effectiveness study is not relevant for this device. This type of study compares the performance of multiple human readers on multiple cases, often with and without AI assistance, to measure the effect of AI on human reader performance. The Oragene Dx OGD-500.001 is a sample collection device, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense, the performance of the device itself (collecting and stabilizing DNA) is assessed in a standalone manner. The studies evaluate the device's ability to provide suitable DNA for downstream testing, independently of human interpretation of that DNA. The "algorithm" here is the collection and stabilization process, and its output (stabilized DNA) is directly measured and assessed against objective criteria (DNA yield, purity, genotyping concordance).

    7. The Type of Ground Truth Used

    The ground truth used in these studies is primarily objective laboratory measurements and genotyping concordance.

    • DNA Yield and Purity: Measured directly using laboratory techniques (e.g., spectrophotometry for A260/A280 ratio, fluorometry for DNA concentration).
    • Genotyping Concordance: Comparison of genotypes obtained from samples collected with the device against reference genotypes (e.g., from unadulterated samples for interference studies, or known genetic profiles if applicable), using an FDA-cleared genotyping test (23andMe PGS Carrier Screening Test). This is a form of outcome data as it assesses the device's ability to enable correct biological outcomes (accurate genotyping).
    • Microbial Content: Laboratory assay for the presence of microbes.

    8. The Sample Size for the Training Set

    No training set is mentioned or applicable for this type of device. Training sets are typically used for machine learning or AI models, which is not the function of this saliva collection device. The device itself is a physical product, and its performance is characterized through validation studies, not by training an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.