Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231290

    Validate with FDA (Live)

    Date Cleared
    2024-01-24

    (265 days)

    Product Code
    Regulation Number
    866.5550
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite Freelite Kappa Free Kit is intended for the quantitative in vitro measurement of Kappa free light chains in serum using the Binding Site Optilite analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE), and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.

    The Optilite Freelite Lambda Free Kit is intended for the quantitative in vitro measurement of Lambda in serum using the Binding Site Optilite analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE), and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.

    Device Description

    The determination of soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Optilite Freelite Kappa Free Kit and Optilite Freelite Lambda Free Kit:

    Context: This submission is for a modification to a previously cleared device (K150658) to extend its indications for use to "aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS)". The core device technology and its principles (immunoturbidimetry for quantitative measurement of free light chains) remain unchanged. Therefore, the performance data provided specifically addresses the new MGUS claim.


    Acceptance Criteria and Reported Device Performance

    Device: Optilite® Freelite® Kappa Free Kit and Optilite® Freelite® Lambda Free Kit

    Intended Use Extension: Aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS).

    Acceptance Criteria (Pre-defined)Reported Device Performance
    Study 1: Clinical Performance - Sensitivity
    Sensitivity for all MGUS samples: at least 30%59.4% (95% CI: 53.0 - 65.5%)
    Sensitivity for Light Chain MGUS (LC-MGUS): Not explicitly listed, but the study highlighted this subgroup.100% (95% CI: 75.8 - 100%)
    Sensitivity for non-LC-MGUS samples: Not explicitly listed, but the study highlighted this subgroup.57.2% (95% CI: 50.6 - 63.5%)
    Study 1: Clinical Performance - Specificity
    Specificity for non-MGUS disease controls: at least 85%86.4% (95% CI: 79.8 - 91.1%)
    Study 2: Evaluation of MGUS Progression (Stable)
    Correct categorization of clinically stable MGUS by FLC testing: at least 80%93.3% of stable patients were categorized as "test positive" (meaning FLC stable based on criteria)
    Study 2: Evaluation of MGUS Progression (Progressive)
    Correct categorization of clinically progressive MGUS by FLC testing: at least 30%50.0% of progressive patients were categorized as "test positive" (meaning FLC progressive based on criteria)

    Study Details:

    1. Sample Sizes and Data Provenance:

    • Study 1 (Sensitivity and Specificity):
      • MGUS Samples (Sensitivity): 234 samples from patients with clinically confirmed MGUS.
      • Disease Controls (Specificity): 140 samples from patients with polyclonal hypergammaglobulinemia (non-MGUS).
      • Data Provenance: Retrospective testing of residual samples. The country of origin is not explicitly stated, but the company (The Binding Site Ltd.) is based in the United Kingdom.
    • Study 2 (MGUS Progression):
      • Total Samples: 185 samples from 49 MGUS patients (45 stable, 4 progressive). Up to 4 individual sample draws for stable, up to 6 for progressive per patient.
      • Data Provenance: Retrospective testing of residual samples. Country of origin not explicitly stated.

    2. Number of Experts and Qualifications for Ground Truth:

    • The document states that the clinical diagnostic criteria that clinicians used to establish the "clinical truth" of MGUS positive samples were confirmed with each site.
    • The ground truth for non-MGUS samples was based on "clinically confirmed polyclonal hypergammaglobulinemia" and "supporting clinical information."
    • For the progression study, "clinically determined stable or progressive status" was the basis.
    • The number of experts and their specific qualifications (e.g., radiologist with 10 years of experience) are not explicitly provided in the summary. It generally refers to "clinicians" and "clinical diagnosis."

    3. Adjudication Method for the Test Set:

    • Not explicitly mentioned. Given that the ground truth relies on "clinical diagnosis" and "clinically confirmed MGUS," it implies that the diagnostic criteria were applied by the treating clinicians at the respective sites prior to the retrospective study. There is no indication of an independent multi-expert adjudication process specifically for this study's test set.

    4. MRMC Comparative Effectiveness Study:

    • No. This submission is for an in-vitro diagnostic (IVD) device (a laboratory test), not an AI-assisted diagnostic tool that would directly assist human readers in image interpretation. Therefore, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study to measure improvement in human readers with AI assistance is not applicable and was not performed.

    5. Standalone Performance (Algorithm Only without Human-in-the-Loop):

    • Yes. The study evaluates the performance of the device (Optilite Freelite Kappa and Lambda Free Kits on the Optilite Analyser) in measuring FLC levels and deriving the kappa:lambda ratio. The performance metrics (sensitivity, specificity, categorization of stable/progressive) are reported for the device's output itself, compared to the clinical ground truth. This is a standalone performance assessment of the IVD test.

    6. Type of Ground Truth Used:

    • Clinical Diagnosis / Clinical Findings:
      • For MGUS positive samples: "clinically confirmed MGUS" based on diagnostic criteria, including those outlined by the 'International Myeloma Working Group (IMWG)' consensus.
      • For non-MGUS samples: "polyclonal hypergammaglobulinemia confirmed by study testing (total IgG/lgA/lgM and serum IFE), with supporting clinical information."
      • For stable/progressive MGUS: "clinically determined stable or progressive status" based on patient follow-up and conversion to MM or stable status over time. This includes evaluation criteria based on IMWG guidelines for multiple myeloma regarding FLC changes.

    7. Sample Size for the Training Set:

    • Not applicable. This submission focuses on the performance of a lab-based immunoassay kit, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The test method is immunoturbidimetry, which relies on chemical reactions and optical detection, not an algorithm that learns from data.

    8. How Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no "training set" for this type of device. The method is a validated laboratory assay.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1