510(k) Data Aggregation

    K Number
    K172067

    Validate with FDA (Live)

    Device Name
    One Mum Pumpset
    Manufacturer
    Date Cleared
    2017-11-29

    (145 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The One Mum Pumpset should be used in combination with Ardo breastpumps and is intended to be used by lactating women to express and collect milk from their breast. The One Mum Pumpset can be used both as a single pumpset and as a double pumpset and is intended for single users only.

    Device Description

    The One Mum Pumpset is a sterile pumpset accessory used by a single user in conjunction with the Ardo breastpumps. The Pumpset has nearly identical technological characteristics and identical principles of operation as the non-sterile pumpset previously cleared in the Ardo Carum and Calypso Powered Breast Pumps (K141742), with the minor difference being that the Pumpset is sterilized using an ethylene oxide gas cycle. Like the predicate device non-sterile pumpset, the One Mum Pumpset consists of multiple components including a bottle, bottle cap, polypropylene breast shell, and silicon tubing kit that connects to the breast pump for the removal, collection, and storage of the mother's breast milk. The Pumpset can be used both as a single pumpset and as a double pumpset and may be used in indoor settings including hospital and home.

    AI/ML Overview

    The provided text is a 510(k) summary for the Ardo Medical AG's One Mum Pumpset. It is a submission to the FDA seeking to demonstrate substantial equivalence to a legally marketed predicate device. This document does not describe the acceptance criteria of an AI/ML powered device, nor does it detail a study proving such a device meets acceptance criteria.

    The information provided pertains to a physical medical device (a breast pump accessory) and its sterilization and biocompatibility testing. Therefore, I cannot extract the requested information regarding acceptance criteria and study details for an AI/ML device from this document.

    However, based on the content related to the One Mum Pumpset, I can extract information relevant to its performance testing, which serves a similar purpose to acceptance criteria for a physical device:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria CategorySpecific Criteria/Standard Adhered ToReported Device Performance
    BiocompatibilityISO 10993-5:2009 (Cytotoxicity)Not cytotoxic
    ISO 10993-10:2010 (Sensitization)Non-sensitizing
    ISO 10993-10:2010 (Intracutaneous)Non-irritating
    ISO 10993-11:2006 (Systemic Toxicity)Not systemically toxic
    Sterilization ValidationISO 11135-1:2007 (SAL)Achieved SAL of 10^-6
    Residual TestingISO 10993-7:2008Residual values within appropriate limits
    Packaging & Shelf LifeISO 11607-1:2009 (Aging Studies)All performance specifications post-aging met; Packaging integrity ensured
    Overall FunctionalitySame as predicate device (K141742)Functioned as intended; Met all same acceptance criteria as predicate

    Regarding the other requested points (2-9), these are entirely focused on AI/ML device studies and are not applicable to the submitted document as it discusses a physical breast pump accessory:

    • No AI/ML device is mentioned.
    • No test set for an algorithm is described.
    • No experts for ground truth establishment for an AI algorithm are mentioned.
    • No adjudication method for an AI test set is discussed.
    • No multi-reader/multi-case comparative effectiveness study for AI assistance is mentioned.
    • No standalone AI algorithm performance is discussed.
    • No AI ground truth type is mentioned.
    • No training set size or ground truth establishment for a training set is relevant to this document.