    Intended Use

    BIOGENNIX osteoSPAN Morpheus is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, posterolateral spine and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process. When used in the posterolateral spine, osteoSPAN Morpheus must be used as a bone graft extender, mixed in a one to one ratio with autogenous bone graft.

    Device Description

    osteoSPAN Morpheus is a moldable, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules suspended in a resorbable organic binder. The osteoSPAN granules used in osteoSPAN Morpheus are a composite of calcium carbonate with a thin calcium phosphate layer, in the form of hvdroxvapatite, coating on all the surfaces of the interconnected porosity. The interconnected pores are approximately 500 microns in diameter and occupy approximately 65% of the volume. The biocompatible, organic binder used in osteoSPAN Morpheus facilitates placement and containment of the implant and is rapidly resorbed in-situ, leaving behind the osteoSPAN granules without affecting their osteoconductive or resorbable properties.

    AI/ML Overview

    The provided document is limited in scope and only describes the OsteoSPAN Morpheus device and its expanded indication for use. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a comprehensive study proving the device meets those criteria.

    Therefore, many of the requested details about acceptance criteria, device performance, sample sizes, expert involvement, and ground truth establishment are not available in the provided text.

    However, I can extract what is available and indicate what information is missing.

    Here's a summary based on the provided text, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria. Instead, it relies on demonstrating substantial equivalence to a predicate device through supplementary preclinical performance testing. The reported performance is a qualitative assessment of this equivalence.

    Performance MetricAcceptance Criteria (Not explicitly stated numerically, but implied as "substantially equivalent" to predicate)Reported Device Performance (Summary of findings)
    Bone IngrowthSubstantially equivalent to predicate osteoSPAN in a posterolateral spine model.Confirmed substantially equivalent to its predicate osteoSPAN when used in a posterolateral spine model.
    Tissue ReactionSubstantially equivalent to predicate osteoSPAN in a posterolateral spine model.Confirmed substantially equivalent to its predicate osteoSPAN when used in a posterolateral spine model.
    Mechanical StrengthSubstantially equivalent to predicate osteoSPAN in a posterolateral spine model.Confirmed substantially equivalent to its predicate osteoSPAN when used in a posterolateral spine model.
    Residual MaterialSubstantially equivalent to predicate osteoSPAN in a posterolateral spine model.Confirmed substantially equivalent to its predicate osteoSPAN when used in a posterolateral spine model.
    OsteoconductiveIdentical to osteoSPAN predicate device (K093342).The osteoSPAN predicate device (K093342), which is the primary component, has identical osteoconductive potential to the current device.
    Resorbable PropertiesUnaffected by organic binder.The organic binder does not affect the osteoSPAN granules' resorbable properties.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Supplementary preclinical performance testing confirmed..." but does not provide details on the sample size for this testing. It only mentions the "posterolateral spine model."

    The data provenance is within the context of a 510(k) submission to the FDA, suggesting the data was generated specifically for supporting this regulatory submission. It does not specify the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable/Not provided. The document describes preclinical performance testing in a "posterolateral spine model," implying animal or in-vitro models were used, rather than human data requiring expert ground truth establishment for diagnostic purposes.

    4. Adjudication Method for the Test Set

    Not applicable/Not provided. As the testing appears to be preclinical performance testing, an adjudication method for human interpretation of results is not relevant or described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention any MRMC study or any study involving human readers with or without AI assistance. This device is a bone void filler, not an imaging or diagnostic AI device.

    6. Standalone (i.e. algorithm only without human-in-the-loop performance) Study

    Not applicable. This device is a physical medical device (bone void filler), not an algorithm or AI. The concept of "standalone performance" for an algorithm does not apply here.

    7. Type of Ground Truth Used

    For the preclinical performance testing, the "ground truth" would likely be based on:

    • Histopathology/Microscopic analysis: To assess bone ingrowth and tissue reaction in the experimental models.
    • Mechanical testing standards: For measuring mechanical strength.
    • Material analysis techniques: For determining residual material.

    However, the specific methods for establishing this "ground truth" are not detailed in the document beyond stating "preclinical performance testing confirmed."

    8. Sample Size for the Training Set

    Not applicable/Not provided. This device is a physical medical device and does not involve AI or machine learning models that require a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable/Not provided. (See point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K132377

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-20

    (293 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOGENNIX osteoSPAN Morpheus is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

    Device Description

    osteoSPAN Morpheus is a moldable, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules suspended in a resorbable organic binder. The osteoSPAN granules used in osteoSPAN Morpheus are a composite of calcium carbonate with a thin calcium phosphate layer, in the form of hydroxyapatite, coating on all the surfaces of the interconnected porosity. The interconnected pores are approximately 500 microns in diameter and occupy approximately 65% of the volume. The biocompatible, organic binder facilitates placement and containment of the implant and is rapidly resorbed in-situ, leaving behind the osteoSPAN granules without affecting their osteoconductive or resorbable properties.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called osteoSPAN Morpheus, a resorbable calcium salt bone void filler. The submission does not include information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML device.

    Instead, this document focuses on demonstrating substantial equivalence to predicate devices based on:

    1. Technological characteristics comparison: A table comparing osteoSPAN Morpheus to a primary predicate (osteoSPAN granules) and a material-reference predicate (Actifuse ABX E-Z-fil Putty) across several characteristics (Calcium Salt, Granule Size, Porosity, Polymer Binder, Osteoconductive).
    2. Safety and performance testing: "Chemical and physical testing confirmed acceptable findings and were consistent with prior osteoSPAN predicate device qualifications. Biogennix osteoSPAN Morpheus was thoroughly evaluated for biocompatibility in accordance with ISO 10993 appropriate parts and for performance, through multiple non-clinical in-vivo and ex-vivo studies."

    Since this is a conventional medical device submission rather than an AI/ML device, the specific questions regarding AI/ML device performance (like sample size for test set, number of experts, adjudication method, MRMC studies, standalone performance, training set details) are not applicable to this document.

    Therefore, I cannot provide the requested information from the given text.