510(k) Data Aggregation

    K Number
    K980070

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-07-23

    (561 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osteonics® HA-AD Coated Devices are single-use, acetabular of femoral hip implant devices intended for cementless applications. The following indications for use are consistent with previous submissions for these acetabular and femoral components intended for cementless application.

    Indications for Osteonics® HA-AD Acetabular Components

    • Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
    • Revision of previous failed femoral head replacement, cup arthroplasty or other procedure.
    • Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
    • Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.

    Indications for Osteonics® HA-AD Femoral Components
    For use as a Bipolar Hip Replacement :

    • Femoral head/neck fractures or non-unions.
    • Aseptic necrosis of the femoral head.
    • Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
    • Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
    • Salvage of failed total hip arthroplasty.

    For use as a Total Hip Replacement:

    • Painful, disabling joint disease of the hip resulting from: degenerative arthritis. rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
    • Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
    • Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
    Device Description

    The devices which are the subject of this 510(k) are commercially-available Osteonics hip acetabular and hip femoral components featuring Osteonics' HA-AD coating.

    AI/ML Overview

    The provided text describes a 510(k) submission for Osteonics HA-Coated Devices, focusing on claims related to the performance of their HA-AD coating. It is noteworthy that this submission is from 1999, and the type of detailed performance criteria, ground truth establishment, and multi-reader studies applied to modern AI/ML devices were not standard practice for medical device submissions at that time.

    Therefore, many of the requested categories for AI/ML device studies will not be present in this document.

    Here's an analysis based on the provided text, highlighting what is (and isn't) present:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the way a modern AI/ML device submission would, with quantifiable thresholds for performance metrics. Instead, it refers to "additional performance claims" demonstrated by referenced literature.

    Acceptance Criteria (Implied)Reported Device Performance
    Increase in bone apposition compared to AD coated samplesPerformance: HA/AD coating increases bone apposition (Ref. 1). Study: "Cortical Remodeling and Bone Apposition to a Textured Canine Hip Implant with and without Hydroxylapatite." Agins HJ, BauerTW, Kudrna JC, Cannestra V, Ming J. Trans, AAOS, February 1997.
    Increase in linear bone contact compared to AD coated samplesPerformance: HA/AD coating increases linear bone contact (Ref. 3). Study: "Biomechanical and Histological Analysis of an HA-Coated, Arc Deposited CPTi Canine Hip Prosthesis." Walenciak MT, Zimmerman MC, Harten RD, Ricci JL, Stamer DT. J. Biomed. Mater. Res. 31 (4):465-474, 1996.
    Increase in degree of bone penetration into implantable chambersPerformance: HA/AD coating increases the degree of bone penetration into implantable chambers (Ref. 2). Study: "Arc Deposited CP Titanium with and without Hydroxyapatite Coating. Preclinical Evaluation Using the Canine Implantable Chamber Model." Hawkins M, Ricci JL. (Book chapter)
    Early evidence of increased bony responsePerformance: Increased bony response may be evidenced as early as 6 weeks after implantation. Study: Implied from the referenced animal studies, likely Ref. 1 and 2.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified in the provided text for the individual studies. The studies are described as using "animal models" (canine hip implant, canine implantable chamber model, canine hip prosthesis).
    • Data Provenance: Animal models (canine). All studies appear to be prospective experimental studies. The original location/country of the animal studies is not stated within this 510(k) summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. Given the nature of these animal studies (histological and biomechanical analyses), the "ground truth" would likely be established by trained researchers and pathologists examining tissue samples and conducting measurements, but explicit details on the number or qualifications of experts are absent.

    4. Adjudication Method for the Test Set

    This information is not provided.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC study was done, as this submission pre-dates the common use of AI in medical devices and the associated study designs.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable, as the device is a physical implant, not an algorithm. Therefore, "standalone" performance testing in the AI/ML sense was not conducted. The "performance data" refers to the biological response to the coating.

    7. The Type of Ground Truth Used

    The ground truth used was histological analysis (e.g., bone apposition, linear bone contact, bone penetration) and biomechanical analysis performed on animal tissue samples. This is analogous to pathology or biological outcomes data in a clinical context.

    8. The Sample Size for the Training Set

    No "training set" in the context of AI/ML algorithms is mentioned or relevant here, as this device is a physical implant, not a software algorithm. The studies referenced are experimental studies on the biological performance of the coating.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for an AI/ML algorithm. The ground truth for the performance studies themselves (as described in point 7) was established through scientific experimental methods, including histological and biomechanical assessments, as detailed in the referenced research papers.