    K Number
    K102071

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-10-08

    (77 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSTENE®CT is indicated for use as a water-soluble implant material and for use in the control of bleeding from bone surfaces in cardiothoracic surgery surgery following sternotomy .

    Device Description

    Ostene® CT is a water-soluble surgical implant material for use in cardiothoracic surgery. Ostene CT is an odorless, opaque wax-like material designed to be utilized directly out of the package. It is best used immediately following removal from the package, and can be softened and increased in stickiness by warming and by additional handling and manipulation, if so desired.

    Ostene CT is comprised of a sterile mixture of water-soluble alkylene oxide copolymers. Ostene®CT contains no other additives or colorants. The wax-like material is formed into sticks of various weights ranging from .5 to 5 grams each.

    As a bone hemostasis agent, Ostene®CT stops bone bleeding by the creation of a physical barrier along the edges of bones that have been damaged by trauma or cut during a surgical procedure. Ostene CT, when placed on bone under moderate pressure, plugs the vascular openings in the bone. This plug prevents further bleeding.

    Ostene CT is provided sterile by irradiation and must not be resterilized.

    AI/ML Overview

    The provided text is a 510(k) summary for the Ostene®CT Soluble Bone Hemostasis Implant Material. It focuses on establishing substantial equivalence to a predicate device rather than presenting a study with acceptance criteria for device performance. As such, much of the requested information (like specific performance metrics, sample sizes for test/training sets, expert details, adjudication methods, MRMC studies, or standalone algorithm performance) is not available in the provided document.

    However, I can extract the following:

    1. Acceptance Criteria and Reported Device Performance:

    The document states: "The non-clinical evaluations used to determine substantial equivalence included indications, intended use, design, materials, sterilization, and performance. The comparison demonstrates that the device in this substantially equivalent in indications and intended use; and is identical in design, materials, sterilization, and performance to the predicate Ostene®CT (K091636)."

    This indicates that the acceptance criterion was "identical performance" to the predicate device. Specific quantitative performance metrics are not provided in this summary.

    Acceptance CriteriaReported Device Performance
    Identical performance to the predicate device (K091636) in design, materials, sterilization, and performance.The device is identical to the predicate device (K091636) in design, materials, sterilization, and performance.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified.
    • Data Provenance: Not specified. The evaluations were "non-clinical."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not specified, as this appears to be a non-clinical evaluation comparing device characteristics to a predicate, not an evaluation against a clinical ground truth established by experts.

    4. Adjudication method for the test set:

    • Not applicable/Not specified.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/software device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/software device.

    7. The type of ground truth used:

    • For the non-clinical evaluation, the "ground truth" was essentially the established characteristics and performance of the predicate device (Ostene®CT, K091636).

    8. The sample size for the training set:

    • Not applicable. This is not a study involving a training set in the context of machine learning or AI.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K091636

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-06-22

    (18 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSTENE®CT is indicated for use as a water-soluble implant material and for use in the control of bleeding from bone surfaces.

    Device Description

    Ostene®CT is an odorless, opaque wax-like material designed to be utilized directly out of the package. It is best used immediately following removal from the package, and can be softened and increased in stickiness by warming and by additional handling and manipulation, if so desired.

    Ostene®CT is comprised of a sterile mixture of water-soluble alkylene oxide copolymers, derived from ethylene oxide and propylene oxide. * Ostene® CT contains no other additives or colorants. The wax-like material is formed into sticks of various weights ranging from .5 to 5 grams each.

    As a bone hemostasis agent, Ostene®CT stops bone bleeding by the creation of a physical barrier along the edges of bones that have been damaged by trauma or cut during a surgical procedure. Ostene CT, when placed on bone under moderate pressure, plugs the vascular openings in the bone. This plug prevents further bleeding.

    Ostene®CT is provided sterile by irradiation and must not be resterilized. * A substitution has been made in a neutral excipient copolymer.

    AI/ML Overview

    This submission describes a medical device, Ostene®CT, which is a bone hemostasis implant material. The document provided is primarily a 510(k) summary for a Special 510(k) submission, indicating a modification to an already legally marketed device. Therefore, the focus is on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and a comprehensive study report typical for novel devices.

    Given the nature of a Special 510(k) and the provided text, there is no specific performance study with defined acceptance criteria in the way one might expect for a new, high-risk device requiring detailed clinical trial data. The device is for bone hemostasis and is classified as "Unclassified," which typically means less stringent performance study requirements for demonstrating substantial equivalence, especially when it's a modification of a predicate.

    The primary "study" is the demonstration that a "substitution has been made in a neutral excipient copolymer" and that the modified Ostene®CT "has the same intended use and fundamental scientific technology as the legally marketed Ostene®CT Soluble Bone Hemostasis Implant."

    Here's an attempt to answer your questions based on the provided text, noting where specific information is absent due to the type of submission:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a Special 510(k) for a modified device, the "acceptance criteria" are generally that the modified device remains substantially equivalent to the predicate device and that the change (a neutral excipient copolymer substitution) does not adversely affect safety or effectiveness. Specific numerical performance targets are not provided in the document for the new Ostene®CT against the predicate.

    Acceptance Criteria (Implied for Special 510(k))Reported Device Performance (Summary)
    Same intended use as predicateOstene®CT has the "same intended use."
    Same fundamental scientific technology as predicateOstene®CT has the "same fundamental scientific technology."
    No new questions of safety or effectivenessNot explicitly stated as a separate finding, but implied by 510(k) clearance. The modification is a "neutral excipient copolymer."
    Effective in controlling bone bleedingAs a bone hemostasis agent, it "stops bone bleeding by the creation of a physical barrier... This plug prevents further bleeding." (This is a description of its mechanism, not a quantified performance measure from a new study against acceptance criteria).
    Sterile by irradiationProvided "sterile by irradiation."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. Clinical trial data with defined sample sizes and test sets are not presented in this 510(k) summary. The submission focuses on the modification of an excipient in an already cleared device.
    • Data Provenance: Not applicable in the context of a performance study; the submission details the device composition and its mechanism of action, emphasizing its similarity to the predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. There is no mention of a clinical "test set" requiring ground truth established by experts in this 510(k) summary.

    4. Adjudication Method for the Test Set

    • Not applicable. No test set requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • Not applicable. This device is a bone hemostasis implant material, not an AI-powered diagnostic or assistive tool. MRMC studies are irrelevant here.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study Was Done

    • Not applicable. This device is a physical implant material, not an algorithm.

    7. The Type of Ground Truth Used

    • Not applicable in the context of a typical performance study for diagnostic AI or imaging devices. For a medical device like this, "ground truth" often relates to established scientific principles of material compatibility and physiological function, as well as pre-clinical testing (e.g., biocompatibility, hemostatic efficacy in animal models), which are standard for medical device development but not detailed in this 510(k) summary. The primary "ground truth" for this 510(k) is the established safety and effectiveness of the predicate device.

    8. The Sample Size for the Training Set

    • Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. There is no "training set."

    Summary based on the document:

    This 510(k) submission for Ostene®CT is for a modified version of an already cleared device. The "study" proving it meets acceptance criteria is primarily the demonstration of substantial equivalence to the predicate device (Ceremed, Inc. Ostene® Soluble Bone Hemostasis Implant Material, K082491), especially considering a "substitution has been made in a neutral excipient copolymer." The acceptance criteria are implicitly met if the FDA determines that the modified device has the same intended use, fundamental scientific technology, and does not raise new questions of safety or effectiveness compared to the predicate. Details on specific performance metrics, sample sizes, or expert adjudication for clinical performance studies are not provided in this regulatory summary.