    K Number
    K123182

    Validate with FDA (Live)

    Date Cleared
    2012-11-05

    (26 days)

    Product Code
    Regulation Number
    866.3330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OSOM® Influenza A&B Test is an in vitro diagnostic immunochromatographic assay intended for the qualitative detection of influenza A and influenza B viral nucleoprotein antigens from nasal swab specimens in symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and/or B viral infections. This test is not intended for the detection of influenza C viruses. A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.

    Device Description

    The OSOM® Influenza A&B Test consists of a test stick that separately detects influenza A and B. The test procedure requires the solubilization of the nucleoproteins from a swab by mixing the swab in Extraction Buffer. The test stick is then placed in the sample mixture, which then migrates along the membrane surface. If influenza A and/or B viral antigens are present in the sample, it will form a complex with mouse monoclonal IgG antibodies to Influenza A and/or B nucleoproteins conjugated to colloidal gold. The complex will then be bound by another mouse anti-influenza A and/or B antibody coated on the nitrocellulose membrane. A pink to purple control line must appear in the control region of the stick for results to be valid. The appearance of a second and possibly a third light pink to purple line will appear in the test line region indicating an A, B or A and B positive result.

    AI/ML Overview

    The provided text describes an update to the package insert of the OSOM® Influenza A&B Test (K092633) to include additional analytical reactivity information for specific H3N2v Influenza A strains. It is not an original submission for a new device, but rather a modification to an already cleared device. As such, the information typically found in an initial 510(k) for device performance and clinical studies demonstrating efficacy for establishing acceptance criteria and proving they are met is largely absent in this document.

    However, based on the provided text, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here is the ability of the OSOM® Influenza A&B Test to react with and detect specific influenza A strains. The stated performance is that the device does react with these strains.

    Acceptance Criteria (Ability to Detect)Reported Device Performance
    A/WEST VIRGINIA/06/2011 (H3N2v)Reacts with and is detectable (Estimated detectable limit: 1.0E+05 EID50/mL*)
    A/PENNSYLVANIA/14/2010 (H3N2v)Reacts with and is detectable (Estimated detectable limit: 1.0E+08 EID50/mL*)
    A/MINNESOTA/11/2010 (H3N2v)Reacts with and is detectable (Estimated detectable limit: 1.0E+08 EID50/mL*)
    A/KANSAS/13/2009 (H3N2v)Reacts with and is detectable (Estimated detectable limit: 1.0E+05 EID50/mL*)
    A/INDIANA/08/2011 (H3N2v)Reacts with and is detectable (Estimated detectable limit: 1.0E+06 EID50/mL*)
    A/INDIANA/10/2011 (H3N2v)Reacts with and is detectable (Estimated detectable limit: 1.00E+09 EID50/mL*)

    Note: The document specifies that "the performance characteristics of this device with clinical specimens that are positive for these 2009 H1N1 and H3N2v influenza viruses have not been established." This indicates these results are from analytical reactivity studies, not clinical performance studies.

    2. Sample Size Used for the Test Set and Data Provenance

    The test set consisted of cultured strains of the H3N2v Influenza A virus. The specific sample size (i.e., number of replicates for each strain) is not provided.
    The data provenance is from analytical testing (e.g., in vitro laboratory testing) of cultured strains of the H3N2v influenza A virus. The origin of the data is not explicitly stated as a country, but the strains themselves suggest a US origin (West Virginia, Pennsylvania, Minnesota, Kansas, Indiana). The study is retrospective in the sense that the strains were already cultured and tested.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable/provided in the document as the study described is an analytical reactivity study using cultured viral strains. The "ground truth" for the test set is the known presence and concentration of the specific influenza virus strains in the cultured samples.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided as this was an analytical reactivity study, not a clinical study involving human interpretation of results requiring adjudication. The device's reaction (detectable or not) is a direct output.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done for this submission. The submission pertains to updating analytical reactivity information for an already cleared in vitro diagnostic device, not evaluating human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone performance evaluation was done. The OSOM® Influenza A&B Test is an immunochromatographic assay, which is a rapid diagnostic test that provides a visual result (appearance of a pink to purple line). Its performance in detecting the H3N2v strains was evaluated directly, without human interpretation in the loop beyond observing the presence or absence of the test line.

    7. The Type of Ground Truth Used

    The ground truth used was the known presence and estimated concentration (EID50/mL or TCID50/mL) of specific cultured influenza A viral strains, often provided by sources like the CDC. This is a form of analytical truth based on established viral culture and quantification methods.

    8. The Sample Size for the Training Set

    This information is not applicable/provided. The detailed analytical reactivity described is for the test set that demonstrates the device's updated capabilities. For an already cleared device, detailed training set information for its initial development and clearance (K092633) is not part of this specific submission to update labeling. The OSOM® Influenza A&B Test is a lateral flow immunoassay, not a machine learning algorithm, so the concept of a "training set" in the computational sense does not apply. If "training set" refers to samples used during the original development and optimization of the assay, that information is not present here.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/provided for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092633

    Validate with FDA (Live)

    Date Cleared
    2009-09-25

    (29 days)

    Product Code
    Regulation Number
    866.3328
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OSOM® Influenza A&B Test is an in vitro diagnostic immunochromatographic assay intended for the qualitative detection of influenza A and influenza B viral nucleoprotein antigens from nasal swab specimens in symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and/or B viral infections.

    This test is not intended for the detection of influenza C viruses. A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.

    Device Description

    The OSOM Influenza A&B Test consists of a test stick that separately detects influenza A and B. The test procedure requires the solubilization of the nucleoproteins from a swab by mixing the swab in Extraction Buffer. The test stick is then placed in the sample mixture, which then migrates along the membrane surface. If influenza A and/or B viral antigens are present in the sample, it will form a complex with mouse monoclonal IgG antibodies to influenza A and/or B nucleoproteins conjugated to colloidal gold. The complex will then be bound by another mouse anti-influenza A and/or B antibody coated on the nitrocellulose membrane. A pink to purple control line must appear in the control region of the stick for results to be valid. The appearance of a second and possibly a third light pink to purple line will appear in the test line region indicating an A, B or A and B positive result.

    AI/ML Overview

    The provided text describes a 510(k) summary for the OSOM® Influenza A&B Test, primarily focusing on updating the package insert with additional analytical reactivity information for the H1N1 Influenza A strain Mexico/4108/2009.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" for clinical performance. Instead, it demonstrates the device's analytical reactivity with a specific strain of influenza. The acceptance is based on the device showing reactivity to the H1N1 strain.

    Acceptance Criteria (Implied)Reported Device Performance
    Detect the H1N1 Influenza A strain Mexico/4108/2009 in culture.OSOM Influenza A&B Test reacts with a cultured strain of the 2009 H1N1 Influenza A virus (A/Mexico/4108/2009) and is detectable.

    2. Sample Sizes and Data Provenance

    • Test Set Sample Size: The document refers to testing with a "cultured strain" (A/Mexico/4108/2009) of the 2009 H1N1 Influenza A virus. It does not specify a "sample size" in terms of number of patient samples, but rather indicates a single viral strain was used for this analytical reactivity test.
    • Data Provenance: The data is analytical reactivity information, presumably from a laboratory setting. Specific country of origin is not mentioned for this particular test, but the strain name "Mexico/4108/2009" indicates the origin of the virus strain. The study is a prospective analytical study, not a retrospective clinical study.

    3. Number of Experts and Qualifications

    Not applicable. This was an analytical reactivity study, not a study requiring expert interpretation of results. The determination of whether the test "reacts" is a direct laboratory observable.

    4. Adjudication Method

    Not applicable. There was no need for adjudication as this was an analytical reactivity study, not a clinical study involving multiple interpreters.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document describes an analytical reactivity study, not a clinical effectiveness study. There is no mention of human readers or AI assistance.

    6. Standalone Performance Study

    Yes. The study described is a standalone analytical performance study, demonstrating the device's ability to detect the H1N1 strain in a cultured sample without human intervention in the result interpretation (beyond observing the test line).

    7. Type of Ground Truth Used

    The ground truth used was the cultured presence of the specific H1N1 Influenza A virus strain (A/Mexico/4108/2009). This is a laboratory-established ground truth.

    8. Sample Size for the Training Set

    Not applicable. This device is an immunochromatographic assay, not an AI/machine learning algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061508

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-06-12

    (11 days)

    Product Code
    Regulation Number
    866.3328
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OSOM Influenza A&B Test is an in vitro diagnostic immunochromatographic assay intended for the qualitative detection of influenza A and influenza B viral nucleoprotein antigens from nasal swab specimens in symptomatic patients. It is intended to aid in the rapid differential diagnosis of influenza A and/or B viral infections. This test is not intended for the detection of influenza C viruses. A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.

    Device Description

    The OSOM Influenza A&B Test consists of a test stick that separately detects influenza A and B. The test procedure requires the solubilization of the nucleoproteins from a swab by mixing the swab in Extraction Buffer. The test stick is then placed in the sample mixture, which then migrates along the membrane surface. If influenza A and/or B viral antigens are present in the sample, it will form a complex with mouse monoclonal IgG antibodies to influenza A and/or B nucleoproteins conjugated to colloidal gold. The complex will then be bound by another mouse anti-influenza A and/or B antibody coated on the nitrocellulose membrane. A pink to purple control line must appear in the control region of the stick for results to be valid. The appearance of a second and possibly a third light pink to purple line will appear in the test line region indicating an A, B or A and B positive result.

    AI/ML Overview

    This document is a 510(k) summary for the OSOM® Influenza A&B Test. It describes the device, its intended use, and compares it to a legally marketed predicate device (OSOM Influenza A&B Test K051244). The information provided focuses on the device's characteristics and its equivalence to the predicate, rather than a detailed study proving it meets specific acceptance criteria with performance metrics.

    Therefore, much of the requested information regarding acceptance criteria, reported performance, and study details (sample sizes, ground truth establishment, expert involvement, MRMC studies) is not present in the provided text. The document is primarily a regulatory submission for substantial equivalence.

    Here's an analysis of what can be extracted from the provided text based on your request:

    1. A table of acceptance criteria and the reported device performance

    The provided text does not include explicit acceptance criteria or detailed reported device performance (e.g., sensitivity, specificity, accuracy) from a clinical study. It focuses on comparing the new device's technological characteristics to a predicate device. The cross-reactivity data table shows potential interfering substances that were tested and found to have "no affect on the performance," but it doesn't quantify performance metrics.

    Acceptance CriteriaReported Device Performance
    Not specified in the documentNot specified in the document (No detailed performance metrics are provided, such as sensitivity, specificity, or PPV/NPV from a clinical study)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The text describes the device and its intended use but does not detail any clinical study or the sample size used for a test set, nor the provenance of such data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. There is no mention of a test set, ground truth establishment, or experts involved in such a process.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. There is no mention of a test set or any adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. This device is an in vitro diagnostic immunochromatographic assay, not an AI-assisted diagnostic tool, so the concept of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is a lateral flow immunoassay. Its performance is inherent to the chemical reaction and visual interpretation of the test stick, not an algorithm. Therefore, the concept of "standalone (algorithm only)" performance is not applicable. The test provides a direct result (pink to purple line) that is interpreted by the user.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The document states under "Intended Use": "A negative test is presumptive and it is recommended these results be confirmed by cell culture." This implies that cell culture is considered the gold standard or ground truth for confirming negative results of influenza infection. However, it does not explicitly state what was used as ground truth for any performance evaluation in this 510(k) submission.

    8. The sample size for the training set

    This information is not provided in the document. As a lateral flow immunoassay, there wouldn't typically be a "training set" in the machine learning sense. The device's formulation and antibodies are developed through a different process.

    9. How the ground truth for the training set was established

    This information is not provided and is again not applicable in the context of this type of device.