Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K163423

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-08-31

    (268 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orthopantomograph OP300 panoramic, cephalometric and cone beam computed tomography x-ray device is intended to image the head and neck areas for diagnostic support. This includes temporomandibular Joints (TMJs) and dentomaxillofacial areas, and with the 13x15 cm field of view (FOV) this additionally includes the ear. nose and throat (ENT) regions. The x-ray device produces conventional 2D x-ray images and x-ray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

    Device Description

    Orthopantomograph OP300 x-ray unit is a software controlled diagnostic dental X-ray equipment for producing panoramic, cephalometric and 3D images of the dentomaxillofacial complex, ENT and airway regions of the head and neck. The Orthopantomograph OP300 is available in different configurations (panoramic, cephalometric, 3D) with different 3D Field of View configurations (6X4 cm, 6X8 cm, 5X5 cm, 8X8 cm, 8X15 cm, and 13X15 cm). The components of the device include column, carriage, rotating unit (containing tube head, sensors and collimator), top-level hanger, patient head support, software for image reconstruction, and an optional software package for image reviewing. The proprietary name of the optional image reviewing software is CLINIVIEW.

    Cone Beam Volumetric Tomography is a medical imaging technique that uses X-rays to obtain crosssectional images of the head or neck. The proposed device utilizes cone beam X-ray technology, which generates conical x-ray beams that rotate around the patient's head and incident upon the receptor that generate sufficiently contrasted images. Quality of the images depends on the level and amount of X-ray energy delivered to the tissue. When interpreted by a trained physician, these images provide useful diagnostic information.

    AI/ML Overview

    The provided text describes the Orthopantomograph OP300, a dental X-ray device. The acceptance criteria and the study proving the device meets these criteria are mainly focused on demonstrating substantial equivalence to predicate devices, particularly the Scanora 3D (K110839), for expanded indications including ENT imaging.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria in a table format with corresponding reported performance values for diagnostic accuracy or specific imaging quality metrics. Instead, it focuses on demonstrating substantial equivalence in diagnostic quality.

    Acceptance Criteria (Implied)Reported Device Performance
    Image quality for proposed indications (head, neck, ENT, maxillofacial) is diagnostically acceptable.Clinical images acquired with Orthopantomograph OP300 were reviewed by qualified clinicians and found to be of acceptable quality for the proposed indications for use.
    Diagnostic quality of images is equivalent to predicate device Scanora 3D (K110839).An image comparison study demonstrated that images from the Orthopantomograph OP300 from ENT, airway, and maxillofacial areas are of equivalent diagnostic quality as images from the predicate device Scanora 3D.
    Adherence to relevant safety and performance standards (Biocompatibility, EMC, Electrical Safety).Passed biocompatibility evaluation (ISO 10993-5, ISO 10993-10). Met requirements in IEC 60601-1, IEC 60601-1-6, IEC 62366, IEC 60601-2-63, IEC 60601-2-28, IEC 60601-1-2, and IEC 60601-1-3.
    Minor differences in technical characteristics do not negatively affect device performance or raise new safety/effectiveness concerns.Differences in technical characteristics are "so small that they do not have any effect on the device performance in practice." The slightly bigger Field-of-View "does not negatively affect imaging of the intended anatomical structures and does not affect substantial equivalence of the device."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The text mentions an "image comparison study" and "clinical images acquired using Orthopantomograph OP300." However, it does not specify the sample size for the test set or the provenance (country of origin, retrospective/prospective) of the data used in these studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The text states that "clinical images acquired using Orthopantomograph OP300 were reviewed by qualified clinicians" and that the image comparison study involved "images from ENT, airway and maxillofacial areas." However, it does not specify the number of experts used or their specific qualifications (e.g., number of years of experience, specific board certifications).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The text does not provide any information about the adjudication method used for establishing ground truth or evaluating images in the mentioned studies. It simply states that images were "reviewed by qualified clinicians."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document describes an "image comparison study of predicate device Scanora 3D and proposed device Orthopantomograph OP300 images." This suggests a comparative study between devices rather than a human-in-the-loop study with AI assistance. The device is an X-ray imaging system, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers is not applicable and was not reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This point is not applicable as the Orthopantomograph OP300 is an imaging device, not an AI algorithm. Its performance is evaluated based on the quality of the images it produces for human interpretation, not as a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the "acceptable quality" and "equivalent diagnostic quality" seems to be established through expert clinical review and comparison. There is no mention of pathology or outcomes data being used as ground truth for image quality assessments in the provided text.

    8. The sample size for the training set

    The document makes no mention of a "training set" as it describes an X-ray imaging device, not an AI model that requires training. The studies mentioned are focused on design verification, validation, and clinical image quality assessment.

    9. How the ground truth for the training set was established

    As there is no mention of a training set, the establishment of ground truth for a training set is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093683

    Validate with FDA (Live)

    Date Cleared
    2010-10-06

    (310 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orthopantomograph® OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examinations of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as x-ray projection images of an examined volume for the reconstruction of a 3D view. The device must only be operated and used by dentists and other legally qualified professionals.

    Device Description

    Orthopantomograph® OP300 is an extraoral source dental x-ray device, which produces conventional digital 2D panoramic, cephalometric and TMJ x-ray images as well as digital x-ray projection images taken during cone beam rotations around a patient's head. The projection images are reconstructed to be viewed in 3D by a 3D viewing software. The name Orthopantomograph® OP300 is used when the device can produce panoramic, TMJ and optionally 3D images. The name Orthoceph® OC300 is used when the cephalometric option for cephalometric x-ray images is added to the device. Furthermore, the device can be equipped with the Volumetric Tomography device (K063773).

    AI/ML Overview

    The provided text is a 510(k) summary for the Orthopantomograph® OP300, a dental X-ray device. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study that proves the device meets specific acceptance criteria in terms of performance metrics like sensitivity or specificity.

    Therefore, the requested information regarding acceptance criteria, reported performance, sample sizes, expert qualifications, and adjudication methods for a specific study proving device performance against acceptance criteria is not present in the provided text. The document primarily identifies the device, its intended use, technological characteristics, and claims substantial equivalence to predicate devices.

    Here's a breakdown of why each specific piece of information cannot be extracted:

    • 1. A table of acceptance criteria and the reported device performance: The document does not define specific performance acceptance criteria for the new device (e.g., minimum diagnostic accuracy, image quality scores). It doesn't report any direct performance metrics for the Orthopantomograph® OP300 in a study.
    • 2. Sample size used for the test set and the data provenance: There is no mention of a test set, studies, or clinical data.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: As no test set or study is described, this information is not available.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable as no test set or study is described.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This document is for an X-ray device, not an AI-powered diagnostic tool, and no MRMC studies or AI involvement are mentioned.
    • 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable as this is not an AI algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): No ground truth is mentioned as no study is described.
    • 8. The sample size for the training set: Not applicable as this is not an AI algorithm or a study involving a training set.
    • 9. How the ground truth for the training set was established: Not applicable for the same reasons as above.

    In summary, the provided 510(k) summary (K093683) for the Orthopantomograph® OP300 is a regulatory document focused on establishing substantial equivalence based on intended use and technological characteristics rather than providing detailed clinical study results with performance metrics and acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092591

    Validate with FDA (Live)

    Date Cleared
    2009-11-20

    (88 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orthopantomograph® OP30 dental panoramic x-ray equipment is indented for dental radiographic examinations by producing digital radiographs of teeth. TM-joints and other oral structures by dentists and other legally qualified professionals.

    Device Description

    Orthopantomograph® OP30 is an extraoral source dental x-ray equipment, which produces dental panoramic and TMJ images on its digital x-rav image receptor. The technique factor settings are: 66, 70, 73 or 77 kV, 10 mA and max 10 sec.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving the device meets said criteria. The document is a 510(k) summary for the Orthopantomograph® OP30, focusing on its intended use, technological characteristics, and substantial equivalence to a predicate device. It confirms FDA clearance but does not include details on specific performance metrics, studies, sample sizes, or ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1