Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K973832

    Validate with FDA (Live)

    Date Cleared
    1997-10-27

    (20 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPUS hLH Controls are quality control materials intended for use in the clinical laboratory to assess or verify the performance of the OPUS hLH assay, run on the OPUS analyzers.

    Device Description

    OPUS hLH Controls are intended for use as quality control material to monitor the accuracy and precision of the OPUS hLH assay run on the OPUS analyzers.

    AI/ML Overview

    The provided document is a 510(k) Notification for the OPUS hLH Controls, a quality control material, not an AI/ML powered device. As such, many of the requested categories in the prompt (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set sample size, how ground truth was established for training set) are not applicable to this type of medical device submission.

    However, I can extract information related to the device performance characteristics, which serve as the acceptance criteria for this quality control material.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Metric)Device Performance
    Intra-assay Precision (Overall %CV)Ranged from 4.51% to 6.48% (for three levels of control material)
    Inter-assay Precision (Overall %CV)Ranged from 2.48% to 6.41% (for three levels of control material)

    2. Sample size used for the test set and the data provenance

    • Sample Size:
      • Intra-assay precision: 20 replicates for each of the three levels of control material (total of 60 measurements).
      • Inter-assay precision: Duplicate measurements over a five-day period, totaling 20 replicates for each of the three levels of control material (total of 60 measurements).
    • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It is a submission for a quality control material, and such studies are typically performed in a controlled laboratory setting by the manufacturer (Dade-Behring Inc., located in Westwood, MA, USA).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a quality control material. Its "ground truth" (i.e., expected LH levels) is established during its manufacturing and assay process, not by expert review of patient data. The study focuses on the precision of the control material when run on an analyzer.

    4. Adjudication method for the test set

    Not applicable. There is no expert adjudication for the performance of a quality control material. The performance is assessed purely based on statistical precision metrics.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a quality control material, not an AI/ML-powered diagnostic tool requiring human reader studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a quality control material for an immunoassay, not an algorithm.

    7. The type of ground truth used

    For quality control materials, the "ground truth" refers to the target or assigned values for the analytes within the control. These are established through:

    • Manufacturer's internal assays: Batch testing during production to determine the concentration of hLH at each control level.
    • Assigned values: The control material is "assayed," meaning it comes with lot-specific values, likely determined by the manufacturer against a reference method or standard.

    The study then assesses how precisely the OPUS hLH assay measures these known levels within the control material.

    8. The sample size for the training set

    Not applicable. Quality control materials do not typically have a "training set" in the context of AI/ML or classification algorithms.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962866

    Validate with FDA (Live)

    Device Name
    OPUS HLH
    Date Cleared
    1996-08-23

    (31 days)

    Product Code
    Regulation Number
    862.1485
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPUS hLH is a fluorogenic enzyme assay for use with the OPUS analyzers used in the quantitative measurement of lutenizing hormone (LH) in serum or plasma.

    Device Description

    OPUS hLH is based on the principle of sandwich binding immunoassay. Each test module contains a solid phase anti-LH polyclonal antibody immobilized onto glass fiber. An anti-LH monoclonal antibody/alkaline phosphatase conjugate solution and a wash/substrate soluiton (4-methylumbellifery|phosphate) are sealed in separate wells within the test module.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the OPUS hLH assay. This document focuses on demonstrating substantial equivalence to a legally marketed device rather than establishing new safety and effectiveness through a rigorous clinical trial with explicit acceptance criteria for diagnostic accuracy metrics. Therefore, some of the requested information (e.g., number of experts, adjudication method, MRMC study, sample size for training set, specific ground truth for training set) is not typically found in this type of submission for in-vitro diagnostic devices.

    However, based on the provided text, I can infer and extract information to address your request as best as possible.

    Here's a breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    For in-vitro diagnostic assays like OPUS hLH, acceptance criteria are often defined by performance characteristics such as precision, accuracy (correlation and recovery), and interference. The submission usually aims to show that the new device performs comparably to or within acceptable ranges relative to the predicate device or established analytical standards.

    Acceptance Criteria CategorySpecific Metric (Implicit)Acceptance Value (Implicit from comparison to predicate/analytical standards)Reported Device Performance
    PrecisionIntra-assay %CVAcceptable low variability (e.g., typically <10-15%)4.56% to 6.32%
    Inter-assay %CVAcceptable low variability (e.g., typically <10-15%)4.55% to 5.64%
    Accuracy (Correlation)Correlation Coefficient (r)Close to 1 (e.g., >0.95 indicating strong agreement)0.98
    SlopeClose to 1 (e.g., 0.9-1.1)1.06
    Y-interceptClose to 0 (e.g., -5 to 5)1.06
    Accuracy (Recovery)Percent RecoveryWithin a generally accepted range (e.g., 90-110%)95.3% to 98.1%
    InterferenceCross-reactivityLow or negligible (e.g., <5-10% depending on analyte)<0.6% for hOG and hGH
    No interference detectionNo significant impact on resultsTSH, FSH, Prolactin

    2. Sample size used for the test set and the data provenance

    • Sample Size for Accuracy (Correlation Study): 143 serum samples.
    • Sample Size for Precision Studies:
      • Intra-assay: 3 serum pools, 20 replicates each.
      • Inter-assay: 3 levels of 2 sets of control sera, assayed in triplicate twice a day for 5 days (total of 30 replicates per level/set).
    • Sample Size for Recovery Study: Samples assayed "in replicates of three" for different LH concentrations added to a normal human serum pool.
    • Data Provenance: The document does not explicitly state the country of origin. It is prospective testing performed to support the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is typically not applicable to an in-vitro diagnostic assay submission of this type. The "ground truth" for the test set in the accuracy correlation study is derived from the results of the already commercially available immunoassay (the predicate device) or by established analytical methods for the recovery and interference studies. It does not involve human expert consensus in the same way an imaging AI algorithm might.

    4. Adjudication method for the test set

    Not applicable for this type of in-vitro diagnostic device performance study. The comparison is between analytical results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in-vitro diagnostic device for measuring a hormone, not an AI-assisted diagnostic imaging or human-in-the-loop system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies described (precision, accuracy, interference) represent the standalone performance of the OPUS hLH assay system (the "algorithm" in a broad sense, referring to the entire test system's analytical capability) without human interpretation affecting the quantitative measurement itself, beyond standard laboratory procedures.

    7. The type of ground truth used

    • Accuracy by Correlation: The "ground truth" for comparison was the results obtained from a "commercially available immunoassay" (the Abbott IMX LH test system, which is the predicate device).
    • Accuracy by Recovery: The "ground truth" was established by adding known concentrations of LH to a serum pool with a known endogenous LH concentration, allowing for calculation of expected versus recovered values.
    • Interference: The "ground truth" was based on established concentrations of potential interfering substances (TSH, FSH, Prolactin, hOG, hGH) and their known impact, or lack thereof, on LH measurements.

    8. The sample size for the training set

    Not applicable. This is an immunoassay, not a machine learning algorithm that requires a "training set" in the conventional sense. The assay is based on chemical and immunological principles, and its parameters are set during its development and manufacturing, not "trained" on data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1