510(k) Data Aggregation

    K Number
    K132134

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-02

    (144 days)

    Product Code
    Regulation Number
    878.4750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OPTIFIX™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

    Device Description

    The OPTIFIX™ Absorbable Fixation System is a sterile single use device that delivers either 15 or 30 synthetic absorbable fasteners via a straight shaft. The shaft of the OPTIFIX™ Absorbable Fixation System is 39 cm in length. The fasteners are designed with a retention feature on the end and are manufactured from Poly (D, L)-lactide.

    AI/ML Overview

    The provided document describes the OPTIFIX™ Absorbable Fixation System, a medical device, and its substantial equivalence to a predicate device, SorbaFix™, based on non-clinical and animal studies. However, the document does not contain information about a study proving the device meets acceptance criteria in the context of an AI/algorithm-driven device. It is a traditional medical device submission for a physical surgical fixation system.

    Therefore, many of the requested fields related to AI/algorithm performance (e.g., test set, ground truth experts, MRMC studies, standalone performance, training set) are not applicable to this document.

    Here's an attempt to extract relevant information from the provided text, acknowledging that the request is geared towards AI/algorithm studies, which isn't the subject of this 510(k) summary.

    1. Table of acceptance criteria and the reported device performance

    The document states that the OPTIFIX™ Absorbable Fixation System passed various non-clinical and animal test requirements, and "All samples tested met the acceptance criteria." However, the specific quantitative acceptance criteria values are not explicitly stated in the provided text. The performance is reported as meeting these (unstated) criteria.

    Acceptance Criteria CategoryReported Device Performance
    Non-Clinical Tests:
    Shear strength testingMet acceptance criteria
    Mass loss determinationMet acceptance criteria
    Dimensional changesMet acceptance criteria
    Inherent viscosityMet acceptance criteria
    Glass transition temperaturesMet acceptance criteria
    Residual monomer contentMet acceptance criteria
    Animal (In-vivo) Studies:
    ContractureMet acceptance criteria
    Tissue IngrowthMet acceptance criteria
    HistologyMet acceptance criteria
    Ex-vivo Test:
    Burst Strength on Pig wallMet acceptance criteria

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated for each test, but implied to be "All samples tested." The specific number of samples for each non-clinical or animal study is not provided.
    • Data Provenance: Not explicitly stated. The studies are described as "non-clinical tests," "animal in-vivo studies," and "Ex-vivo Burst Strength on Pig wall." Country of origin or retrospective/prospective nature is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not Applicable. This document describes the testing of a physical medical device. The "ground truth" for the performance metrics (e.g., shear strength, mass loss, burst strength) would be established by direct physical measurements and observations, not by human expert consensus or interpretation in the way it's done for AI models.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. See explanation above. The assessment of physical and biological properties is based on experimental protocols and quantitative analysis, not expert adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a physical device, not an AI/algorithm. No human "readers" are involved in interpreting its output, and there's no AI assistance being evaluated.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context refers to the measured physical and biological properties of the device and its interaction with tissue. This includes:
      • Direct measurements of shear strength, mass, dimensions, viscosity, glass transition temperatures, and residual monomer content.
      • Observations from animal in-vivo studies (e.g., contracture, tissue ingrowth, histology) which are typically assessed by pathologists or researchers following established protocols.
      • Measurements of burst strength on ex-vivo tissue (pig wall).

    8. The sample size for the training set

    • Not Applicable. This is not an AI/algorithm-driven device; therefore, there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See explanation above.