Search Results
Found 262 results
510(k) Data Aggregation
(29 days)
The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.
N/A
Ask a specific question about this device
(66 days)
This product is intended for delivery of respiratory gases to adult patients in hospitals.
This product is indicated for the delivery of nasal high flow (NHF) and low-flow oxygen by appropriately qualified healthcare professionals.
Qualitative carbon dioxide (CO2) sampling can be used at nasal cannula flow rates from 5 to 50 L/min in operating and procedure rooms.
This product can be used for pre-oxygenation, short-term supplemental oxygenation (up to 10 minutes) during intubation, and allows an anesthesia mask to be placed over the nasal interface to perform mask ventilation, by individuals who perform anesthesia care or airway management.
This product is not intended for apneic ventilation.
This product is not indicated for use during CPR.
The Optiflow™ Switch+ Filtered Nasal Interface with CO2 Sampling interface is intended to deliver respiratory gases to the patient.
The subject device includes a CO2 sampling tube that provides a sample of the patient's exhaled respiratory gases through the CO2 sampling accessory to a CO2 sampling line and CO2 analyzer.
The subject device has been designed to allow for mask ventilation without the need to remove the nasal interface from the patient during therapy. To support this function, a Flow Diverter (K234053) is required.
The product is indicated for the delivery of nasal high flow (NHF) and low flow oxygen by appropriately qualified healthcare professionals.
This is a prescription-only device provided in a non-sterile state. It has a flow range of 5 to 70 L/min. The device will be offered in three sizes, small (S), medium (M), and large (L).
It is intended to be used in combination with a compatible respiratory humidifier, Optiflow Oxygen Kit and the Optiflow Flow Diverter (K234053).
N/A
Ask a specific question about this device
(231 days)
The OptiMesh Multiplanar Expandable Interbody Fusion System is indicated for use as an adjunct to fusion in an intervertebral body fusion at one level in the lumbar spine from L2 to S1 in skeletally mature patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history, physical examination, and radiographic studies. Eligible patients shall have undergone six (6) months of conservative (non-operative) care. The OptiMesh device, along with a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion, is intended for use with supplemental posterior fixation systems intended for use in the lumbar spine.
The OptiMesh Multiplanar Expandable Interbody Fusion System is an intervertebral body graft containment device that is a non-rigid, implanted spinal device designed to contain FDA cleared bone void filler within its internal cavity. The device is inserted into the intervertebral body space of the spine and is intended as an adjunct to intervertebral body fusion.
The OptiMesh implants are available with and without HAnano Surface®, a 20 – 40 nanometer thin hydroxyapatite (HA) surfaces treatment.
N/A
Ask a specific question about this device
(260 days)
Fisher & Paykel Healthcare OptiPAP Junior is for single-patient use and is intended to deliver noninvasive positive airway pressure (PAP) therapy to spontaneously breathing patients who require respiratory support.
OptiPAP Junior is designed for use in hospital environments where the patient is continually monitored and must be prescribed by a physician. It is intended for use by trained medical professionals on a pediatric subpopulation that includes neonates and infants from birth up to two years of age.
The F&P OptiPAP Junior is intended to deliver heated and humidified Positive Airway Pressure (PAP) therapy to neonates and infants requiring respiratory support.
The F&P OptiPAP Junior is a prescription-only device provided in a non-sterile state and intended to be used in a hospital environment and must be prescribed by a physician. The F&P OptiPAP Junior is for single- patient-use up to 14 days. The F&P OptiPAP Junior will be offered in multiple size variants.
N/A
Ask a specific question about this device
(30 days)
The Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The Optima Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum and tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are detached from the delivery pusher in a serial manner until the aneurysm is occluded.
N/A
Ask a specific question about this device
(183 days)
The OptiLUV device is a handheld device intended to emit energy in the RF spectrum to provide topical heating for the purpose of elevating facial tissue temperature for a temporary relief of minor pain and the temporary increase in local blood circulation.
The OptiLUV device is a non-invasive, non-ablative device that delivers RF energy into the skin to generate heat through electrical impedance in the dermis. The device consists of the following components and accessories: the OptiLUV device, the USB Charging Cord, and the OptiLUV Preparation Balm.
N/A
Ask a specific question about this device
(30 days)
Ask a specific question about this device
(175 days)
Optina-4C™ is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying, storing, and transferring images of the retina. The Optina-4C is indicated for the viewing of the posterior segment of the eye at multiple wavelengths (colors) under mydriatic conditions.
Images are intended to aid clinicians in the detection, diagnosis, and management of ocular health and disease.
The Optina-4C™ (MHRC-C1N) is a non-contact mydriatic retinal camera (also called a fundus camera) designed to provide eye care professionals with a series of images of the retina obtained at specific wavelengths (colors). The Optina-4C processes images using algorithms that take advantage of the spectral information to produce 2D en face images of the retina as an aid to clinicians in the diagnosis of age-related macular degeneration (AMD), retinal pigment epithelium (RPE) disruptions and lesions, retinopathies, and other retinal diseases.
Utilizing Hyperspectral Imaging (HSI) technology, the Optina-4C captures a series of 92 high-resolution images at precise 5 nm wavelength increments across the visible and near-infrared (NIR) spectrum on a ~31° field-of-view. Image acquisition takes under two seconds, with only 0.92 seconds of light exposure time. During this time, the camera scans through all wavelengths from 905 to 450 nm, spending 10 milliseconds per 5 nm step.
The Optina-4C processes the images to provide spectrally calibrated, normalized, and registered images in a structured Hyperspectral Cube format for scroll-through viewing. The spectral resolution supports visualization and differentiation of drusenoid deposit types, as well as high contrast viewing of the retinal chromophores hemoglobin, melanin, and carotenoids (combination of lutein and zeaxanthin). Images can be viewed, stored or transferred in their native .h5 format, or as DICOM or PNG formats. Optina-4C images offer sharp, vivid, high-contrast coronal views of the posterior segment of the eye for clinical assessment and evaluation by eye care professionals.
The HSI NormReg Cube is a processed data set of 91 hyperspectral images that have undergone intensity normalization and spatial registration. Normalization corrects for the non-uniform illumination profile of the imaging system and the spectral sensitivity of the camera, while registration compensates for intra-scan eye movement to ensure consistent alignment across frames. The result is a hyperspectral image cube with enhanced spatial and spectral fidelity, referred to as the HIS-NormReg Cube.
N/A
Ask a specific question about this device
(29 days)
OptiVu™ Shoulder is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures (humerus and scapula), humerus resection, and aligning the endoprosthesis with the anatomical structures, provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.
OptiVu™ Shoulder utilizes pre-operative planning files provided by the Zimmer CAS Signature ONE™ System. OptiVu™ Shoulder is compatible with any humeral implants that are supported by the Signature ONE™ System.
OptiVu™ Shoulder is specifically indicated for total shoulder arthroplasty using the Zimmer Biomet Alliance® Glenoid system or reverse shoulder arthroplasty using the Comprehensive® Reverse Shoulder system, to aid the surgeon in locating anatomical structures (humerus and scapula), humerus resection, and aligning the glenoid component with the anatomical structures.
OptiVu Shoulder includes an augmented reality (AR) head-mounted display (HMD) (OptiVu Tilt with HoloLens 2) and trackers to register and optically track anatomical landmarks and surgical instruments in real-time during the procedure. The HMD should not be relied upon solely and should always be used in conjunction with traditional methods.
OptiVu Shoulder is a stereotaxic surgical navigation system designed to aid surgeons in locating anatomical structures and aligning the endoprosthesis in total or reverse shoulder arthroplasty procedures. The system includes an augmented reality (AR) head-mounted display (HMD) (OptiVu Tilt with HoloLens 2) and mixed reality trackers to register and optically track anatomical landmarks and surgical instruments in real-time during the procedure.
The OptiVu Shoulder system is intended to be used specifically with the Zimmer Biomet Alliance® Glenoid or Comprehensive® Reverse Shoulder system for total or reverse shoulder arthroplasty, respectively.
The OptiVu Shoulder system also utilizes pre-operative planning files provided by the Zimmer CAS Signature ONE™ System.
The intended users of the system are surgeons who are trained in performing shoulder arthroplasty procedures.
This 510(k) clearance letter pertains to a Special 510(k) submission for the OptiVu™ Shoulder system. A Special 510(k) is used when changes to a previously cleared device do not alter its fundamental scientific technology or indications for use, but rather involve specific modifications. In this case, the modifications are explicitly stated to be updates to packaging parameters and the site for contract sterilization, not changes to the software, functionality, or performance of the navigation system itself.
Therefore, the provided document does not contain the information requested about the acceptance criteria and study proving the device meets those criteria for the AI/software performance, as this type of study would have been part of the original 510(k) clearance for OptiVu™ Shoulder (K250108), not this Special 510(k) (K252170). The validation testing performed for this submission is limited to sterilization efficacy and packaging safety.
This document explicitly states:
- "The purpose of this Special 510(k) submission is to update both the packaging parameters for the instruments and the site for contract sterilization."
- "There are no differences in technological characteristics that raise questions of safety and efficacy."
- "With the updated packaging and sterilization parameters presented in this submission, there are no technological or design changes in the subcomponents of the OptiVu Shoulder system instruments themselves."
- "Validation testing was conducted according to written protocols with acceptance criteria that were based on established standards. This submission includes or references the following tests in support of a substantial equivalence determination:
- Sterilization validation testing to confirm that there is no impact to the sterilization efficacy of the system
- Packaging validation to confirm that the packaging of OptiVu Shoulder system is safe and effective"
Conclusion based on the provided document:
The provided 510(k) clearance letter for K252170 (OptiVu™ Shoulder) is a Special 510(k), which addresses changes only to the packaging and sterilization of the device, not its core functionality or software performance. As such, the document does not contain any information about acceptance criteria or performance studies related to the artificial intelligence/software accuracy, clinical effectiveness, human reader improvement with AI assistance, or the establishment of ground truth for such studies. This type of information would have been part of the original 510(k) submission (K250108) for the device's fundamental technology.
Therefore, I cannot populate the requested table or answer the specific questions about AI/software performance, ground truth, and study parameters based solely on the provided text.
Ask a specific question about this device
(174 days)
OptiVu Shoulder is intended for use during stereotaxic surgery to aid the surgeon in locating anatomical structures (humerus and scapula), humerus resection, and aligning the endoprosthesis with the anatomical structures, provided that the required anatomical landmarks can be identified on the patient's preoperative CT scan.
OptiVu Shoulder utilizes pre-operative planning files provided by the Zimmer CAS Signature ONE™ System. OptiVu Shoulder is compatible with any humeral implants that are supported by the Signature ONE™ System.
OptiVu Shoulder is specifically indicated for total shoulder arthroplasty using the Zimmer Biomet Alliance® Glenoid system or reverse shoulder arthroplasty using the Comprehensive® Reverse Shoulder system, to aid the surgeon in locating anatomical structures (humerus and scapula), humerus resection, and aligning the glenoid component with the anatomical structures.
OptiVu Shoulder includes an augmented reality (AR) head-mounted display (HMD) (OptiVu Tilt with HoloLens 2) and trackers to register and optically track anatomical landmarks and surgical instruments in real-time during the procedure. The HMD should not be relied upon solely and should always be used in conjunction with traditional methods.
OptiVu Shoulder is a stereotaxic surgical navigation system designed to aid surgeons in locating anatomical structures and aligning the endoprosthesis in total or reverse shoulder arthroplasty procedures. The system includes an augmented reality (AR) head-mounted display (HMD) (OptiVu Tilt with HoloLens 2) and mixed reality trackers to register and optically track anatomical landmarks and surgical instruments in real-time during the procedure.
The OptiVu Shoulder system is intended to be used specifically with the Zimmer Biomet Alliance® Glenoid or Comprehensive® Reverse Shoulder system for total or reverse shoulder arthroplasty, respectively.
The OptiVu Shoulder system also utilizes pre-operative planning files provided by the Zimmer CAS Signature ONE™ System.
The intended users of the system are surgeons who are trained in performing shoulder arthroplasty procedures.
The provided FDA 510(k) clearance letter for OptiVu™ Shoulder describes the device and its intended use, but it does not contain the detailed information necessary to fully answer your request regarding acceptance criteria and the specific study that proves the device meets those criteria.
The document states that non-clinical testing was conducted to demonstrate safety and effectiveness, including "Performance tests to ensure the performance of the implemented features and verify related design inputs" and "Validation lab to validate that the OptiVu Shoulder system is safe and effective and performance of the system is acceptable under full simulated use on cadaveric specimens." However, the specific acceptance criteria values and the quantitative results of these tests are not provided in this public clearance letter. This type of detailed performance data is typically found within the full 510(k) submission, which is a confidential document.
Therefore, I cannot populate the table or provide specific values for most of your questions. I can only infer what kind of studies would have been conducted based on the mention of "Validation lab" with "cadaveric specimens" and "Performance tests."
Here's what I can extract and infer from the provided text, along with clear indications of what information is missing:
While the complete details of the acceptance criteria and the specific study proving the device meets them are not fully contained within the provided FDA 510(k) clearance letter, the document does refer to non-clinical testing that was conducted to demonstrate safety and effectiveness.
The clearance letter mentions:
- "Performance tests to ensure the performance of the implemented features and verify related design inputs"
- "Validation lab to validate that the OptiVu Shoulder system is safe and effective and performance of the system is acceptable under full simulated use on cadaveric specimens"
Based on the nature of a stereotaxic surgical navigation system for shoulder arthroplasty, the acceptance criteria would typically revolve around the accuracy and precision of anatomical structure localization, humerus resection, and endoprosthesis alignment. The "Validation lab" study on cadaveric specimens strongly suggests a performance study was conducted to quantify these aspects.
However, the specific numerical acceptance criteria (e.g., target accuracy in mm and degrees) and the quantitative results (reported device performance) are not provided in this public document.
1. Table of Acceptance Criteria and Reported Device Performance
Note: The specific numerical acceptance criteria and reported performance values are not present in the provided document. The table below represents typical acceptance criteria for such a device, and the "Reported Device Performance" column cannot be filled with concrete data from this document.
| Metric (Inferred) | Acceptance Criteria (Not in document, Typical Example) | Reported Device Performance (Not in document) |
|---|---|---|
| Accuracy of anatomical landmark localization (e.g., bone registration error) | $\le$ 2.0 mm (e.g., RMS error) | Not provided in document |
| Accuracy of humerus resection guidance (e.g., angular and linear deviation from plan) | Angular $\le$ 2.0 degrees, Linear $\le$ 2.0 mm | Not provided in document |
| Accuracy of endoprosthesis (glenoid component) alignment guidance (e.g., angular deviation from plan) | Angular $\le$ 2.0 degrees | Not provided in document |
| Precision (Repeatability) | Not explicitly stated/quantified in document | Not provided in document |
| Latency/Real-time tracking performance | Not explicitly stated/quantified in document | Not provided in document |
| System Reliability/Robustness | Not explicitly stated/quantified in document | Not provided in document |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not specified in the provided document. The document mentions "full simulated use on cadaveric specimens," implying a test set that included multiple cadaveric specimens, but the exact number is missing.
- Data Provenance: The document explicitly states "cadaveric specimens." The geographic origin (country) is not specified. The study was likely prospective as it involved simulated use of the device for testing purposes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Not specified in the provided document.
- Qualifications of Experts: Not specified in the provided document. For a device used in surgical navigation, these would typically be experienced orthopedic surgeons familiar with shoulder arthroplasty.
4. Adjudication Method for the Test Set
- Adjudication Method: Not specified in the provided document. Given the nature of a cadaveric study for surgical navigation accuracy, adjudication might refer to the method used to establish the "true" surgical plan or the "true" anatomical alignment against which the device performance is measured (e.g., independent measurements by a metrology expert, or consensus among multiple surgeons on the optimal plan).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was NOT mentioned or described in the provided document. This type of study typically assesses the impact of an AI algorithm on human reader performance (e.g., radiologists interpreting images). OptiVu™ Shoulder is a surgical navigation system with an Augmented Reality (AR) HMD for real-time guidance, not primarily an image interpretation AI tool. Its effectiveness study would focus on its guidance accuracy during simulated surgery, rather than assisting human readers in diagnostic interpretation.
- Effect Size of Human Reader Improvement: Not applicable, as an MRMC study was not described.
6. Standalone (Algorithm Only) Performance
- Standalone Performance: The document describes "Performance tests" and "Validation lab" studies. While the results are not quantified, these studies would inherently evaluate the "standalone" or "algorithm-only" performance of the navigation system (e.g., its accuracy in tracking, registration, and guidance output) in a controlled environment without active human intervention beyond operating the system as intended. The "HMD should not be relied upon solely and should always be used in conjunction with traditional methods" indicates that the device is intended for human-in-the-loop use, but its core performance metrics (accuracy of localization, resection, alignment) would be evaluated in a standalone manner.
7. Type of Ground Truth Used
- Type of Ground Truth: The document explicitly states "full simulated use on cadaveric specimens." Therefore, the ground truth would likely be established through:
- Pre-operative CT scan planning: The "true" surgical plan (e.g., desired resection angles, implant position) is derived from the pre-operative CT scan using the Zimmer CAS Signature ONE™ System.
- Post-hoc metrology/physical measurements: After using the OptiVu™ Shoulder system on the cadaver, the actual resections and implant placements would be measured using highly accurate metrology tools (e.g., CMM, industrial CT scanning) and compared against the planned values derived from the pre-operative CT. This comparison would quantify the accuracy of the device's guidance.
8. Sample Size for the Training Set
- Training Set Sample Size: Not specified in the provided document. The 510(k) clearance process primarily focuses on the safety and effectiveness of the final device, including its performance characteristics, rather than the details of the training data used for its development (unless it's a specific AI/ML medical device where the training data directly impacts performance).
9. How the Ground Truth for the Training Set Was Established
- Training Set Ground Truth Establishment: Not specified in the provided document. As mentioned above, the details of the training process and data are typically not disclosed in the public 510(k) clearance letter.
Ask a specific question about this device
Page 1 of 27