Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112629

    Validate with FDA (Live)

    Date Cleared
    2011-10-11

    (32 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Onset Medical Access Catheter System is indicated for the introduction of interventional/diagnostic devices into the peripheral, coronary and neuro vasculature.

    Device Description

    The Onset Access Catheter System is a two-catheter system comprised of a Sheath and a Dilator. The Sheath has a distal region which expands when the Dilator is advanced. The Onset Access Catheter System can be used with hydrophilic guidewires up to.038 inches in diameter to access the desired anatomy. The proximal ends of the Sheath and the Dilator have a luer fitting to allow attachment of accessories and infusion of liquids through the catheter. The catheter is offered in various sizes to accommodate physician preferences and patient anatomy. A split sheath is provided in the package to provide support and facilitate the introduction of the Sheath's distal expandable region into the catheter introducer sheath. The Catheter System is provided sterile, non-pyrogenic, and is intended for single use only.

    AI/ML Overview

    The provided document describes the "Onset Access Catheter System," a medical device. However, it does not contain the specific information required to answer the questions about acceptance criteria for an AI/ML powered device, as this is a traditional medical device (a catheter). The document focuses on demonstrating substantial equivalence to a predicate device through biocompatibility, sterilization, and bench-top testing.

    The questions you've asked are typically related to the validation of AI/ML-powered medical devices, which involve performance metrics like accuracy, sensitivity, and specificity, and studies to establish ground truth. This document describes a physical medical device and its general performance and safety.

    Therefore, I cannot extract the following information from the provided text:

    1. A table of acceptance criteria and the reported device performance (for AI/ML metrics)
    2. Sample size used for the test set and the data provenance (for AI/ML models)
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (for AI/ML models)
    4. Adjudication method (for AI/ML models)
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance (for AI/ML models)
    6. If a standalone study was done (for AI/ML models)
    7. The type of ground truth used (for AI/ML models)
    8. The sample size for the training set (for AI/ML models)
    9. How the ground truth for the training set was established (for AI/ML models)

    What the document does provide in relation to "acceptance criteria" and "study":

    The document details a series of bench-top (design verification) tests and biocompatibility tests to show that the Onset Access Catheter System meets established criteria for substantial equivalence to a predicate device.

    Acceptance Criteria and Reported Device Performance (Bench Tests):

    Bench TestsAcceptance Criteria / Result
    Dimensional and Visual InspectionMet established criteria
    Guidewire CompatibilityMet established criteria
    Torque ResponseMet established criteria
    Torque StrengthMet established criteria
    System Deployment Cycle TestMet established criteria
    Kink ResistanceMet established criteria
    Flexibility TestMet established criteria
    Tensile StrengthMet established criteria
    Catheter Leak Test (Liquid Leakage)Met established criteria
    Catheter Leak Test (Air Leakage)Met established criteria
    Dynamic Pressure TestMet established criteria
    Static Burst TestMet established criteria
    Aspiration TestMet established criteria
    Hub GaugingMet established criteria
    Corrosion ResistanceMet established criteria
    USP Particulate TestMet established criteria
    Navigation and Accessibility Capabilities in vitroMet established criteria
    In vivo Tests
    System Deliverability, Compatibility, Visibility and Aspiration PerformanceMet established criteria
    Acute histopathology of treated vesselsMet established criteria
    Biocompatibility testingMet established criteria

    Biocompatibility Testing:

    TestMethodResult
    CytotoxicityL929 MEM Elution TestPassed
    SensitizationKligman MaximizationPassed
    Intracutaneous Reactivity (Irritation)Intracutaneous Injection TestPassed
    Systemic Toxicity (Acute)ISO Acute Systemic Injection TestPassed
    HaemocompatibilityComplement ActivationPassed
    HemolysisPassed
    Inactivated Partial Thromboplastin Time TestPassed
    In vivo thrombogenicityPassed
    PyrogenicityUSP Material Mediated Rabbit Pyrogen TestPassed

    Study Information (for this traditional device):

    • Sample sizes: Not explicitly detailed for each test (e.g., number of catheters tested for torque response), but the document states "All testing was performed on units which were twice sterilized."
    • Data provenance: Not applicable in the context of clinical AI/ML data; these are laboratory and animal (in vivo) tests.
    • Experts/Ground Truth/Adjudication: Not applicable in the context of clinical AI/ML data. The "ground truth" for these tests is defined by established engineering and biological standards.
    • MRMC / Standalone / Training Set: Not applicable as this is not an AI/ML device.

    In summary, the provided text describes the performance validation of a physical medical device through standard engineering and biocompatibility testing, not the performance validation of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1