Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K081307

    Validate with FDA (Live)

    Device Name
    OMNILUX CLEAR-U
    Date Cleared
    2009-01-16

    (253 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Omnilux Clear-U is intended to emit energy in the red and blue region of the spectrum, specifically indicated to treat mild to moderate acne on the face

    Device Description

    The Omnilux Clear-U is a source of high spectral purity It provides uniform or "hotspot" free illumination The outputs are pre-tuned to a particular wavelength with a narrow spectral bandwidth The red output of the Omnilux Clear-U is 633 + 6 nm, and the blue output is 415 ± 5 nm The Omnilux Clear-U device itself contains a selector switch (red/off/blue), the LED array assembly, and arrangements for aur cooling A separate, universal, power supply converts mains AC power to the DC power requred by the Omnilux Clear-U Treatment time is controlled by the operator

    AI/ML Overview

    The provided text describes the 510(k) summary for the Omnilux Clear-U device, which is an LED array for treating mild to moderate acne. However, it does not include acceptance criteria for device performance or a study that directly proves the device meets specific performance criteria in terms of clinical efficacy.

    Instead, the submission relies heavily on demonstrating substantial equivalence to predicate devices (Omnilux Revive, Omnilux blue, Omnilux revive/Omnilux blue combination) by arguing that "no significant differences exist" and therefore "raises no new issues of safety or effectiveness."

    The "Performance Data" section mentions two tests related to user comprehension and usability, but these are about the user interface and labeling, not the therapeutic efficacy of the light treatment itself.

    Here's a breakdown based on the information available and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided document, there are no specific performance-based acceptance criteria for the clinical efficacy regarding acne treatment explicitly stated, nor is there reported device performance against such criteria.

    The document focuses on usability and safety, as well as substantial equivalence to predicate devices.

    Acceptance CriteriaReported Device Performance
    Clinical Efficacy: (e.g., reduction in acne lesions by X% in Y weeks)Not provided in the document. The submission relies on substantial equivalence to predicate devices for efficacy claims.
    Usability Test Critical Objectives Comprehension RateMinimum comprehension rate of 88% at 95% confidence level
    Self-Selection Test Critical Objectives Comprehension RateMinimum comprehension rate of 93% at 95% confidence level
    Overall Performance Characteristics AnalysisPerformed (stated to be a basis for belief in no significant differences), but specific results or metrics are not detailed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Self-Selection Test: A "sample representative population" was used. The exact sample size is not specified, but it was sufficient to determine a minimum comprehension rate of 93% at a 95% confidence level.
    • Usability Test: A "sample representative population" was used. The exact sample size is not specified, but it was sufficient to determine a minimum comprehension rate of 88% at a 95% confidence level.

    Data Provenance: Not explicitly stated, but given the submitter's address (Carlsbad, California) and the context of a 510(k) submission to the FDA, it is highly likely these tests were conducted in the United States. The document does not specify if these were retrospective or prospective studies; however, usability and self-selection tests are typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable for the reported tests. The self-selection and usability tests evaluate user comprehension and ability to use the device based on labeling, not clinical outcomes that would require expert ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. No expert adjudication methods (like 2+1, 3+1) are described as the tests were user comprehension/usability evaluations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. The document does not mention any MRMC study. The comparison is made to predicate devices based on perceived "no significant differences" rather than a direct comparative effectiveness study involving human readers or clinicians.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. The Omnilux Clear-U is a therapeutic light device, not an AI or algorithm. Therefore, "standalone (algorithm only)" performance is not applicable. The device itself is the standalone therapy.

    7. The Type of Ground Truth Used

    For the usability and self-selection tests, the "ground truth" was likely defined by pre-established correct answers or critical objectives of the product labeling, against which user responses were measured for comprehension. For instance, correctly identifying the treatment duration or safety warnings.

    For the inherent therapeutic claims, the "ground truth" (or basis of equivalence) relied on the established safety and effectiveness profile of the predicate devices, which would have been supported by their own clinical data (not detailed in this specific 510(k) summary).

    8. The Sample Size for the Training Set

    Not applicable as this is not an AI/ML device that requires training.

    9. How the Ground Truth for the Training Set was Established

    Not applicable as this is not an AI/ML device that requires training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1