Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K082749

    Validate with FDA (Live)

    Device Name
    OMNI-TIGHT
    Date Cleared
    2008-12-04

    (76 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The implant for dental purposes, used to replace missing dental organs (teeth). The implant is self-tapping (threads) and is screwed into a pilot bore formed in the jawbone. Upon healing, the implant receives a post, which has a stem, and is adapted to carry dental supra-structures (false teeth).

    Device Description

    The modification is to be able to use stronger titanium because of situations in which this could be beneficial. Since B.A.S.I.C. Dental has been approved as a system using CP Titanium, which is the weakest of the standard titanium available, we would like to be able to use stronger titanium to increase the overall strength of the implant. By using stronger titanium there should be no need for further testing, however the following tests were performed to compile with design controls. The testing revealed as expected that implants produced with stronger titanium had better shear/lateral forces, with no lateral constraints.

    AI/ML Overview

    I'm sorry, but the provided text does not contain the information required to describe the acceptance criteria and the study that proves the device meets them. The document is a 510(k) summary for a dental implant modification, focusing on material changes and substantial equivalence to a predicate device. It discusses:

    • Device Identification and Modification: B.A.S.I.C. Dental Implant System, modified to use stronger titanium.
    • Intended Use: Identical to the predicate device, for replacing missing teeth.
    • Comparison to Predicate Device: Argues substantial equivalence based on identical dimensional characteristics and indications for use. Mentions other predicate devices with similar dimensions and titanium alloy materials.
    • Testing: States that "the following tests were performed to compile with design controls. The testing revealed as expected that implants produced with stronger titanium had better shear/lateral forces, with no lateral constraints."
    • Sterilization and Biocompatibility: Confirms the sterilization process remains the same and references ISO standards.
    • Conclusion: Reaffirms substantial equivalence.
    • FDA Letter and Indications for Use: The official FDA clearance letter and the stated indications for use.

    Crucially, it does not provide details about acceptance criteria, specific study designs, sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness with human readers (MRMC studies). The mention of "testing revealed... better shear/lateral forces" is too high-level to infer specific acceptance criteria or detailed study methodologies.

    Therefore, I cannot populate the requested table or answer the specific questions about the study design, ground truth, or expert involvement based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072595

    Validate with FDA (Live)

    Device Name
    OMNI-TIGHT
    Date Cleared
    2008-05-09

    (238 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The implant for dental purposes, used to replace missing dental organs (teeth). The implant is self-tapping (threads) and is screwed into a pilot bore formed in the jawbone. Upon healing, the implant receives a post, which has a stem, and is adapted to carry dental suprastructures (false teeth).

    Device Description

    The B.A.S.I.C. Dental Implant is an endosseous dental implant. The implant, healing cap and healing screws will now be provided in a sterilized medical grade Tyvek® pouches in accordance with ISO 11137 (SAL 10to). In addition, implants are now provided in diameter of 6.0 mm and lengths of 8.0 mm and 9.0 mm to provide a wider range of options for the dental professional.

    AI/ML Overview

    The provided text is a 510(k) summary for the B.A.S.I.C. Dental Implant System. It describes modifications to an existing device rather than a new device that requires a comprehensive clinical study with acceptance criteria and a detailed study report. Therefore, much of the requested information regarding acceptance criteria, study details, and performance metrics is not explicitly available in this document.

    However, based on the context of a 510(k) submission, the "acceptance criteria" are implied to be substantial equivalence to a predicate device, meaning the modified device performs as safely and effectively as a legally marketed device and does not raise new questions of safety or effectiveness. The "study" proving this is primarily a comparison to predicate devices and adherence to relevant standards, rather than a clinical trial with performance targets.

    Here's an attempt to answer the questions based on the provided text, recognizing the limitations of a 510(k) summary for device modification:


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for device modifications aiming for substantial equivalence rather than a new device with specific performance claims from a clinical trial, the "acceptance criteria" are primarily met through comparison to predicate devices and adherence to standards.

    Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance and Evidence (from 510(k) Summary)
    No change in Intended Use (Safety & Effectiveness maintained for original purpose)"The intended use is identical to the predicate device."
    No alteration of fundamental scientific technology"Per FDA-CDRH criteria for a Special 510(k), the modifications do not affect the intended use or alter the fundamental scientific technology of the device..."
    Equivalent technological characteristics (despite modifications)The updated specifications (new diameters and lengths, sterilized components) are deemed substantially equivalent or improvements, as detailed in the comparison tables.
    No new health risks posed by modifications"Risk analysis which includes health risks related to root form endosseous dental implants... An FMEA report is included in Attachment 9.""The dimensional specification changes do not pose any new health risks to the patient as there are other cleared products on the market with similar dimensions."
    Sterilization effectiveness (for new sterilized components)"B.A.S.I.C. Dental Implant Systems, Inc. has contracted with Steris Isomedix Service to establish and maintain a valid gamma-ray sterilization processes in accordance with ISO 11137 (SAL 10⁻⁶)... Substantiation of a routine sterilization dose was established using the VDmax's method, described in ISO 11137."
    Biocompatibility maintained (as material is unchanged)"Material of Implants: CP Titanium" (unchanged). This implies continued biocompatibility of the material.
    Device integrity and functionality maintained (post-sterilization, new dimensions)"Device integrity and functionality were qualified and/or validated using samples produced under routine manufacturing conditions. The B.A.S.I.C. Dental Implant System meets B.A.S.I.C.'s in-house requirements..."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as a traditional clinical "test set" in the document. The "samples" referenced for sterilization and device integrity validation are "produced under routine manufacturing conditions," but the quantity is not specified.
    • Data Provenance: Not applicable in the context of a clinical study. The data provenance relates to internal testing (sterilization validation, FMEA, in-house requirements) and comparison to existing market data (other cleared devices). This is retrospective in the sense of looking at existing cleared devices and standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. Ground truth in this context is established by adherence to recognized standards (ISO 11137, FDA Guidance) and comparison to legally marketed predicate devices, rather than expert consensus on a specific test set.
    • Qualifications of Experts: N/A.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There was no clinical PIVOTAL trial of read performance to adjudicate.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, an MRMC comparative effectiveness study was not done or required for this 510(k) submission. This type of study is typically associated with new diagnostic devices or AI algorithms where human reader performance is a key outcome.
    • Effect Size: N/A.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    • Standalone Study: No, this is not an AI/algorithm-based device. No standalone performance study in that sense was conducted or described.

    7. Type of Ground Truth Used

    • Ground Truth: For this device modification, the "ground truth" is established by:
      • Regulatory Standards: Compliance with ISO 11137 for sterilization, and internal requirements for device integrity and functionality.
      • Predicate Device Equivalence: The existence and safety/effectiveness of previously cleared devices (both the unmodified B.A.S.I.C. Dental Implant System and other manufacturers' devices) with similar dimensions and characteristics serve as the de facto "ground truth" for the new design's safety and effectiveness.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1