Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K071175

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-17

    (81 days)

    Product Code
    Regulation Number
    880.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time.

    Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

    Device Description

    Giraffe OmniBed is an infant bed, which provides thermal support to infants who are unable to thermo-regulate based on their own physiology. The bed has two modes of operation, enclosed bed operation and open bed operation. In an open mode, the device functions as an incubator and in an enclosed mode it functions as a conventional, infant radiant warmer.

    The Giraffe Uninterruptible Power Supply (UPS) is intended to provide a short term source of electrical power Giraffe OmniBed, thus aiding its intra hospital mobility. The Giraffe UPS does not change the indications for use, control mechanisms, operating principles, performance specifications, or other features of the Giraffe Omnibed.

    The UPS serves as an extension to the Giraffe OmniBed by providing uninterrupted electric power to the device. The UPS comprises a medical grade battery and a shelf.

    AI/ML Overview

    Here's an analysis of the provided text regarding the K071175 510(k) submission, focusing on acceptance criteria and the study proving device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Battery life15 minutes at 37°C and 70% RH at steady state
    Battery recharge time6 hours
    Alarm decibel level reduction when UPS is addedLess than 1 dB reduction
    Minimum lower level of alarm decibel (as per IEC 601-2-21 Amend 1 clause 102.1 and IEC-601-2-19 Amend 1 clause 102.3)Within (or above) the minimum lower level of 50dB
    Performance of OmniBed with UPS against product specifications and consensus standardsEstablished by bench testing (implicitly met, as no failures are reported)

    2. Sample Size Used for the Test Set and Data Provenance

    This submission does not involve a "test set" in the context of AI/ML or clinical efficacy studies with patient data. The performance claims are related to the functional characteristics of a new accessory (UPS) for an existing medical device (Giraffe OmniBed).

    • Sample Size: Not applicable in the traditional sense of patient or image data. The testing was conducted on the Giraffe OmniBed device with the addition of the UPS.
    • Data Provenance: The 'data' are results from bench testing performed on the physical device in a controlled environment. There is no information regarding the country of origin, but generally, such testing would be conducted at the manufacturer's facilities. The testing is prospective in the sense that it's designed to evaluate a new component.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth for clinical outcomes or image interpretation is not relevant for this type of device modification. The performance criteria are objective engineering specifications (e.g., battery life, decibel levels).

    4. Adjudication Method for the Test Set

    Not applicable. There are no subjective assessments or disagreements to adjudicate.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI assistance

    Not applicable. This device is an infant incubator/warmer with an uninterruptible power supply, not an AI-powered diagnostic tool. Hence, no MRMC study was conducted or is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm, but a hardware accessory.

    7. The Type of Ground Truth Used

    The ground truth used for this submission is based on objective engineering and performance specifications derived from:

    • Product specifications of the Giraffe OmniBed and the new UPS.
    • Recognized consensus standards (specifically mentioning IEC 601-2-21 Amend 1 clause 102.1 and IEC-601-2-19 Amend 1 clause 102.3 related to alarm decibel levels).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set mentioned or implied in this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020543

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-05-09

    (79 days)

    Product Code
    Regulation Number
    880.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to ther on user's demand. It cannot be operated in both modes at the same time.

    Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperatures controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uternal environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

    Device Description

    The Ohmeda Medical Giraffe OmniBed is an infant bed which provides thermal support for infants who are unable to provide for their own heat requirements. This bed has two modes of operation, enclosed bed operation and open bed operation.

    In the enclosed bed mode of operation, the bed functions as an incubator, maintaining the infant's temperature by circulating heated air within the enclosed bed compartment. The operator may select either the air or skin temperature control method. Depending on the control method selected, heat is regulated based on either the air temperature or the infant's skin temperature compared to the operator selected control temperature. Physical access to the patient is obtained through the side portholes or by opening one of the side doors. In the enclosed bed mode, humidity and oxygen may also be increased.

    In the open bed mode, this bed operates like a convention open, radiantly heated infant bed. Radiant heat from an infrared heat source is focused on to the bed to warm the patient. The operator may select either the heater power or skin temperature control method. Depending on the control method selected, the heater is either regulated at the operator selected power level or the heater output is modulated to maintain the patient's temperature at the value selected by the operator.

    The optional Giraffe Servo Controlled Oxygen System is a fully integrated option available on the Giraffe OmniBed. The Servo Controlled Oxygen System is capable of oxygenating the entire infant compartment at oxygen concentrations of 21-65% by volume. The device uses fuel cell type sensors that generate specific voltages depending on the oxygen concentrations they contact. The microprocessor stores the sensor output and compares it with the value corresponding to the concentration set by the operator. The valves that supply oxygen to the infant compartment are opened and closed as necessary to maintain the oxygen concentration at the set value. Fluctuations in fuel cell performance due to temperature and humidity are compensated for by the microprocessor.

    AI/ML Overview

    The provided document is a 510(k) summary for the Ohmeda Medical Giraffe® OmniBed™. It describes the device's function, indications for use, and a comparison to predicate devices. However, it explicitly states that no clinical or animal testing was deemed necessary to demonstrate safety and effectiveness for this device. Instead, the submission relies on "extensive bench testing" and software validation.

    Therefore, many of the requested elements for describing acceptance criteria and a study proving device performance (especially those related to clinical studies, human readers, ground truth establishment, and sample sizes for clinical test/training sets) cannot be extracted from this document, as such a study was not conducted or submitted.

    Here's a breakdown based on the information available:

    Acceptance Criteria and Device Performance for Ohmeda Medical Giraffe® OmniBed™

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided document, specific quantitative acceptance criteria or a direct comparison table for performance from a clinical study are not available. The document states: "Since care of newborns in incubators and infant radiant warmers, with or without supplemental oxygen is a well established clinical practice. Ohmeda submits that clinical or animal testing to demonstrate safety and effectiveness is not necessary."

    The performance is implicitly understood to meet the functional requirements for an incubator and warmer, including:

    • Providing thermal support for infants.
    • Operating in enclosed (incubator) and open (warmer) modes.
    • Maintaining infant temperature via air or skin temperature control.
    • Adjusting humidity and oxygen in enclosed mode.
    • Providing stable oxygen concentration (21-65%) with the optional Servo Controlled Oxygen System.

    No quantitative performance metrics (e.g., accuracy of temperature control to within X degrees, stability of oxygen concentration to within Y%) are reported in this 510(k) summary.

    2. Sample size used for the test set and the data provenance

    Not applicable. No clinical or animal test set was used for a study demonstrating safety and effectiveness. The document refers to "extensive bench testing," which implies laboratory-based tests, but no sample sizes for these tests are provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No clinical test set requiring expert-established ground truth was reported.

    4. Adjudication method for the test set

    Not applicable. No clinical test set requiring adjudication was reported.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical apparatus (incubator/warmer), not an AI-assisted diagnostic or interpretive tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not directly applicable in the sense of an "algorithm" for diagnosis or interpretation. The device itself is the "standalone" entity that performs its functions (heating, oxygen control). Its performance was assessed through "extensive bench testing" and "software validation," rather than a clinical trial.

    7. The type of ground truth used

    For the "extensive bench testing" and "software validation," the ground truth would have been established by engineering specifications, calibration standards, and validated measurement equipment. For example, for temperature control, the "ground truth" would be the actual temperature measured by precise instruments, compared against the device's set and reported temperatures. For oxygen concentration, it would involve calibrated gas analyzers.

    8. The sample size for the training set

    Not applicable. This device is hardware with software control, not a machine learning model that requires a training set in the conventional sense. Software validation would involve testing against various inputs and scenarios to ensure correct operation according to specifications.

    9. How the ground truth for the training set was established

    Not applicable. As above, no training set for a machine learning model was used. The "ground truth" for software validation would be the expected output and behavior based on the software design specifications and functional requirements.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1