    Intended Use

    The OEC Elite™ MiniView™ (mobile mini C-Arm) is designed to provide physicians with real time general fluoroscopic visualization of patients of all ages. It is intended to aid physicians and surgeons during diagnostic or therapeutic treatment/surgical procedures of the limbs/extremities and shoulders including, but not limited to, orthopedics and emergency medicine.

    Device Description

    The OEC Elite™ MiniView™ is a mobile fluoroscopic mini C-arm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment, and surgical procedures of the shoulders, limbs, and extremities. The system consists of a C-arm attached to an image processing workstation. A CsI(TI) -CMOS flat panel detector and the identical X-ray source monoblock are used for image acquisition.

    The C-arm supports the high-voltage generator, X-ray tube, X-ray controls, collimator, and the FPD. The C-arm is capable of performing linear (vertical, horizontal, orbital) and rotational motions that allow the user to position the X-Ray imaging components at various angles and distances with respect to the patient extremity anatomy to be imaged. The C and support arm are mechanically balanced allowing for ease of movement and capable of being "locked" in place using an electronically controlled braking system. The workstation is a stable mobile platform that supports the C-arm, image display monitor(s), image processing equipment/software, recording devices, data input/output devices and power control systems.

    The primary technology change for the subject device compared to the unmodified predicate OEC Elite MiniView, was to add an optional wireless footswitch which provides identical functionalities as the wired footswitch to control X-ray on and off.

    AI/ML Overview

    This document describes the 510(k) summary for the GE OEC Elite™ MiniView™ device, which is a mobile fluoroscopic mini C-arm system. The specific change being submitted for review is the addition of an optional wireless footswitch.

    Here's an analysis of the provided text to answer your questions regarding acceptance criteria and the study proving the device meets them:

    Crucially, this 510(k) submission states that no clinical studies were required to support substantial equivalence due to the nature of the modification (adding a wireless footswitch). Therefore, many of your questions related to clinical study design (sample size, expert ground truth, MRMC studies, standalone performance) will not be applicable or directly answerable from this document.

    The primary method of demonstrating device performance and safety for this submission was through non-clinical testing, including design verification and validation, risk management activities, and compliance with recognized standards.


    1. A table of acceptance criteria and the reported device performance

    Since this is a submission for a modification (wireless footswitch) and relies on non-clinical testing for substantial equivalence, the "acceptance criteria" are primarily related to product safety, effectiveness, and functional equivalence to the wired footswitch and predicate device.

    Acceptance Criteria (Inferred)Reported Device Performance (Summary from Document)
    Functional Equivalence: The wireless footswitch provides identical functionalities to the wired footswitch for controlling X-ray on/off.The wireless footswitch provides "identical functionalities as the wired footswitch to control X-ray on and off." The "functionalities are identical."
    Safety: Device operates safely in conjunction with the wireless footswitch.Developed under GE Healthcare's design controls and quality management system. Risk management activities identified, reviewed, and mitigated potential issues. Mitigations verified and validated.
    Effectiveness: The addition of the wireless footswitch does not negatively affect the overall safety and effectiveness of the device.Design verification and validation testing performed confirmed that "the safety and effectiveness of the devices has not been affected."
    Electromagnetic Compatibility (EMC) & Coexistence: The wireless footswitch operates without causing or being susceptible to electromagnetic interference.EMC and coexistence testing performed per the FDA "Radio Frequency Wireless Technology in Medical Devices" Guidance (August 14, 2013). Results acceptable.
    Compliance with Standards: Device meets applicable electrical, safety, and performance standards.Tested by an NRTL and certified compliant with IEC 60601-1 Ed. 3 series, including IEC60601-2-54 and IEC 60601-2-43. All applicable 21CFR Subchapter J performance standards are met.
    No New Questions of Safety/Effectiveness: The modification should not introduce new safety or effectiveness concerns."The addition of the optional wireless footswitch to the system does not raise new questions on safety and effectiveness."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable in the context of clinical data. The "test set" here refers to the device itself and its components undergoing engineering and functional verification/validation testing. This is not a patient-based test set.
    • Data Provenance: The testing was conducted internally by GE Hualun Medical Systems Co. Ltd. in Beijing, China, as part of their design control processes. This is prospective engineering and quality assurance testing, not data collected from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth in a clinical sense (e.g., disease presence) was not established. The "ground truth" for this engineering submission is functional correctness and safety established by engineering specifications, international standards, and internal quality processes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This was not a human reader adjudication study. The "adjudication" was based on successful completion of predetermined engineering tests and compliance with regulatory standards, as overseen by GE's quality management system.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. The document explicitly states: "The subject of this premarket submission, OEC Elite MiniView with the addition of the wireless footswitch, did not require clinical studies to support substantial equivalence." This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a medical imaging system, and the submission is for a hardware control component (wireless footswitch), not a standalone algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" used for this submission is based on engineering specifications, functional requirements, and compliance with recognized industry and electrical safety standards. For example, the footswitch must reliably transmit an "X-ray ON/OFF" signal, and the device must maintain its safety characteristics as per IEC 60601 series standards. This is not a clinical ground truth.

    8. The sample size for the training set

    • Not applicable. This is not a machine learning/AI device, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI model, there is no ground truth established for one.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160131

    Validate with FDA (Live)

    Date Cleared
    2016-06-21

    (153 days)

    Product Code
    Regulation Number
    892.1650
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OEC Elite MiniView (mobile mini C-Arm) is designed to provide physicians with real time goopic visualization of patients of all ages. It is intended to aid physicians and surgeons during diagnostic or therapeutic treatment/surgical procedures of the limbs/extremities and shoulders including, but not limited to, orthopedics and emergency medicine.

    Device Description

    The OEC Elite™ MiniView™ is a mobile fluoroscopic mini C-arm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment, and surgical procedures of the shoulders, limbs, and extremities. The system consists of a C-arm attached to an image processing workstation. A CsI(TI) - CMOS flat panel detector and the identical X-ray source monoblock are used for image acquisition.

    The C-arm supports the high-voltage generator, X-ray tube, X-ray controls, collimator, and the FPD. The C-arm is capable of performing linear (vertical, horizontal, orbital) and rotational motions that allow the user to position the X-Ray imaging components at various angles and distances with respect to the patient extremity anatomy to be imaged. The C and support arm are mechanically balanced allowing for ease of movement and capable of being "locked" in place using an electronically controlled braking system. The workstation is a stable mobile platform that supports the C-arm, image display monitor(s), image processing equipment/software, recording devices, data input/output devices and power control systems.

    AI/ML Overview

    The OEC Elite™ MiniView™ is a mobile fluoroscopic mini C-arm system. The provided document is a 510(k) Premarket Notification Submission, which focuses on demonstrating substantial equivalence to a predicate device, rather than defining and proving acceptance criteria in the typical sense of a clinical trial for a novel AI device. However, based on the information provided, we can extract details about the performance evaluation done to demonstrate this equivalence.

    Here's an analysis based on the provided text, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC as one might find for an AI diagnostic algorithm. Instead, the evaluation focuses on demonstrating that the performance of the proposed device (OEC Elite™ MiniView™) is at least equivalent to the predicate device (OEC Mini 6800 Digital Mobile C-arm) and reference devices in terms of image quality and clinical capability.

    The "acceptance criteria" here implicitly revolve around ensuring the safety and effectiveness of the updated device, which includes:

    • Meeting design input and user needs.
    • Compliance with regulatory standards (IEC 60601-1 Ed.3 series, IEC 60601-2-54, IEC 60601-2-43, and 21CFR Subchapter J performance standards).
    • Image quality and clinical capability at least equivalent to the predicate device.
    Performance AspectAcceptance Criteria (Implicit)Reported Device Performance
    Overall PerformanceSystem meets design input, user needs, and regulatory standards; image quality and clinical capability at least equivalent to predicate."The system has been NRTL tested and certified compliant... All applicable 21CFR Subchapter J performance standards are met. The OEC Elite™ MiniView™ system was developed under the GE Healthcare's design controls processes... and additional engineering bench testing was performed... to demonstrate system performance."
    Image QualityImage quality metrics (e.g., resolution, noise reduction) are adequate for viewing extremities and are at least equivalent to predicate/reference devices.Pixel Size: Proposed Device: 100 Microns. (Reference Device: 75 Microns). Reported: "larger for reducing image noise. The resolution is higher than the Image Intensifier on the predicate."Array Size: Proposed Device: 1.3k x 1.3k. Reported: "adequate for viewing extremities."Field Size: Proposed Device: Full Field 13 cm circle, Limited Field 10 cm circle. Reported: "appropriate for viewing extremities."
    Clinical CapabilityDemonstrated ability to provide fluoroscopic visualization in diagnostic/therapeutic/surgical procedures of limbs/extremities/shoulders equivalently to the predicate.Cadaver study results: "For all procedures, the study confirmed the clinical capability and overall quality of the images produced by the OEC Elite™ MiniView™ was at least equivalent to that of the Mini 6800 Digital Mobile C-Arm."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document mentions a cadaver study involving two cadavers on which nineteen orthopedic procedures were performed across a variety of extremity anatomies.
    • Data Provenance: The cadaver study was performed as part of the submission process, implying it was a prospective evaluation specifically for this device. The country of origin of the cadavers is not specified in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Two independent physicians were used to evaluate the images.
    • Qualifications of Experts: The document states they were "two independent physicians", and given the nature of the device (fluoroscopic imaging for orthopedic procedures), it's highly probable these were orthopedic surgeons or radiologists with expertise in musculoskeletal imaging and procedures. However, their specific specializations or years of experience are not explicitly stated in the provided text.

    4. Adjudication Method for the Test Set

    The document states: "The performance of the subject device to the predicate was also performed by two independent physicians." It further states that the "study confirmed the clinical capability and overall quality of the images produced by the OEC Elite™ MiniView™ was at least equivalent to that of the Mini 6800 Digital Mobile C-Arm." This implies a consensus or comparative evaluation by the two physicians. However, a specific adjudication method (e.g., 2+1, 3+1, etc.) is not explicitly detailed. It's presented as a direct comparison where both physicians apparently agreed on the equivalence.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • Not a typical MRMC study: The evaluation described is not a traditional MRMC comparative effectiveness study focused on quantifying human reader improvement with AI assistance. This device is an imaging system (hardware and software for image acquisition and processing), not an AI-powered diagnostic aide designed to improve human reader performance for a specific task.
    • Focus on System Equivalence: The study aimed to demonstrate the system's overall clinical capability and image quality equivalence to a predicate device, as evaluated by human readers (the two physicians), rather than measuring the effect size of AI assistance on human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone "Algorithm" Performance: The device itself is an imaging system, not purely an algorithm. Its performance is inherent in the images it produces. Therefore, "standalone" in this context refers to the system's ability to produce diagnostically acceptable images.
    • Bench Testing and Image Quality Tests: The document details extensive "engineering bench testing" and "image quality/performance testing" identified for fluoroscopy. These tests evaluate the system's technical image output without human interpretation as the primary endpoint. This can be considered the equivalent of "standalone" performance for an imaging device. Specifically mentioned are:
      • Demonstration of system performance.
      • Imaging performance evaluation using anthropomorphic phantoms (including a pediatric anthropomorphic phantom).
      • All image quality/performance testing identified for fluoroscopy in FDA's "Information for Industry: X-ray Imaging Devices - Laboratory Image Quality and Dose Assessment, Tests and Standards" was performed.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • For the Cadaver Study (Test Set): The ground truth for evaluating clinical capability and image quality seems to be based on expert consensus (or agreement) between the two independent physicians regarding the equivalence of the images produced by the OEC Elite™ MiniView™ compared to the predicate device for diagnostic and procedural guidance in the cadaveric setting. There is no mention of pathology or outcomes data for this specific evaluation, as it's a technical performance and clinical utility assessment on cadavers.
    • For Bench Testing: The ground truth for bench testing and phantom studies would be defined by known physical properties of the phantoms and established engineering specifications and standards for image quality metrics.

    8. The Sample Size for the Training Set

    The document describes a medical imaging device (C-arm), not an AI algorithm that requires a separate training set. Therefore, the concept of a "training set sample size" as typically applied to machine learning models is not applicable here. The device's underlying technology and software architecture are based on existing, proven designs (predicate and reference devices), with modifications validated through engineering bench tests and the cadaver study.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, there is no explicit "training set" in the context of an AI algorithm described in this document. The device's development involved standard engineering practices, which could be considered an iterative design and testing process that refines the system's performance. The "ground truth" during this development would be based on engineering specifications, performance targets, and established imaging principles, rather than a labeled dataset for training an AI model.