Found 3 results

510(k) Data Aggregation

    K Number
    K092564

    Validate with FDA (Live)

    Date Cleared
    2010-06-17

    (301 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The O-arm® Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm® Imaging System is compatible with certain Image Guided Surgery Systems.

    The O-arm® Imaging System is a portable x-ray system designed for 2D fluoroscopic and 3D imaging for high contrast objects and anatomic structures.

    The O-arm® Imaging System is compatible with certain Image Guided Surgery Systems. The O-arm® Imaging System may be used with medical charged particle radiation therapy systems for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks.

    Device Description

    The O-arm® Imaging System is a mobile x-ray system which provides 3D imaging as well as 2D fluoroscopic imaging. The system consists of two parts: the x-ray O-arm® Stand (comprising x-ray generator, flat dynamic x-ray detector, and the x-ray control user interface) and the mobile view station (comprising the image processors, a user interface for image and patient handling, and viewing monitor).

    AI/ML Overview

    The provided document is a 510(k) summary for the Medtronic O-arm® Imaging System. It describes the device, its indications for use, and a comparison to a predicate device. However, it does not contain a study that proves the device meets specific acceptance criteria.

    The document states: "Based on design characteristics and imaging performance, the modified O-arm® Imaging System is substantially equivalent to the predicate Oarm® Imaging." This suggests that the device was deemed substantially equivalent to a predicate, rather than being subjected to a specific performance study with acceptance criteria.

    Therefore, most of the information requested in your prompt cannot be extracted from this document, as it focuses on substantial equivalence rather than a detailed performance study against acceptance criteria.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    • Cannot be provided. The document does not specify quantitative acceptance criteria or report device performance against such criteria. It relies on a qualitative comparison to a predicate device, stating "Same as predicate O-arm® with image quality improvement."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Cannot be provided. No test set or study data is described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Cannot be provided. No ground truth establishment related to a specific study is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Cannot be provided. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Cannot be provided. This document is for an imaging system (hardware), not an AI-assisted diagnostic device. No MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Cannot be provided. This document is for an imaging system (hardware), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Cannot be provided. No specific ground truth for performance evaluation is mentioned.

    8. The sample size for the training set:

    • Cannot be provided. No training set is mentioned as this is an imaging system, not an AI algorithm.

    9. How the ground truth for the training set was established:

    • Cannot be provided. No training set is mentioned.

    Summary based on the provided text:

    The submission focuses on demonstrating substantial equivalence to existing predicate O-arm® Imaging Systems. The primary method of "proving" its acceptability is through this comparison, noting that the "modified O-arm® Imaging System is substantially equivalent to the predicate Oarm® Imaging" based on "design characteristics and imaging performance." There is no detailed performance study with specific acceptance criteria outlined in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060344

    Validate with FDA (Live)

    Date Cleared
    2006-03-29

    (47 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The O-arm™ Imaging System is designed for 2D fluoroscopic and 3D imaging for intraoperative applications in surgical theaters, particularly for orthopedic applications. The O-arm™ Imaging System is compatible with certain Image Guided Surgical Systems.

    Device Description

    The O-arm™ Imaging System is a mobile x-ray system which provides 3D imaging as well as 2D fluoroscopic imaging. The system consists of two parts: the x-ray O-arm™ Stand (comprising x-ray generator, flat dynamic x-ray detector, and the x-ray control user interface) and the mobile view station (comprising the image processors, a user interface for image and patient handling, and viewing monitors). The 1-D Display Control Mouse is an optional accessory. It is a sterile disposable limited-function wireless mouse. With the 1-D Mouse, a surgeon can remotely control and point at the O-arm™ image display. The 1-D mouse is similar to remote controls for controlling projectors during business presentations. The 1-D Mouse allows the surgeon to control what is displayed and to point out features on the image with a laser pointer.

    AI/ML Overview

    This document is a 510(k) summary for the O-arm™ Imaging System with a 1-D Display Control Mouse. It's a submission to the FDA for market clearance, not a study report demonstrating the device meets acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design, and performance metrics for a robust clinical study is not available in the provided text.

    The core of this submission is to establish substantial equivalence to a previously cleared device (O-arm™ Imaging System, K050996), primarily by demonstrating that the addition of a sterile, wireless 1-D display control mouse does not alter the safety or effectiveness of the system.

    Here's a breakdown of what can be inferred or explicitly stated based on the provided text, and what cannot:

    1. Table of acceptance criteria and reported device performance:

    This information is not provided in the document because it is a 510(k) summary for an accessory, not a performance study. The "performance" discussed primarily relates to the functionality of the new mouse and its impact on the overall system's predicate comparison.

    Acceptance Criteria (Inferred from Substantial Equivalence Logic)Reported Device Performance (from text)
    Safety and Effectiveness Equivalence: The O-arm™ Imaging System with the 1-D Mouse must be as safe and effective as the predicate O-arm™ Imaging System (cleared in K050996)."The O-arm™ Imaging System with 1-D Mouse is virtually identical in safety and effectiveness to the previously cleared O-arm™ Imaging System."
    Functional Equivalence of Mouse: The 1-D Mouse must perform the display control functions similar to the standard mouse."The display control functions provided by the optional sterile disposable wireless 1-D Control Mouse are also provided by console controls and by a non-sterile, hard-wired mouse on the standard O-arm™ Imaging System." "With the 1-D Mouse, a surgeon can remotely control and point at the O-arm™ image display." "The 1-D Mouse allows the surgeon to control what is displayed and to point out features on the image with a laser pointer."
    No Adverse Co-existence/Interference: The wireless nature of the 1-D Mouse should not introduce issues with co-existence, data latency, integrity, or security."The low power, unique communication protocol, and limited function of the 1-D Mouse ensure no co-existence, data latency and integrity, or security issues that might affect safety or effectiveness."
    Maintenance of Sterile Field: The sterile and wireless nature should facilitate maintaining the sterile field."The sterile and wireless characteristics allow use by the surgeon without compromising the sterile field and without cluttering the surgical area with a mouse cable." "The 1-D mouse provides simpler controls, reduced cable clutter, and greater assurance of maintaining the sterile field."
    No Transmission of Clinical Information: The mouse should only transmit control commands."No clinical information is transmitted by mouse, only display control commands."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    This information is not available. The document focuses on demonstrating substantial equivalence based on the nature of the device (an accessory for display control) and its limited function. It does not describe a test set or clinical study in the traditional sense with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not available. No human-reader study or expert evaluation for ground truth is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not available. No adjudication method is described as there's no clinical test set described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a MRMC comparative effectiveness study was not done. This document is for a medical imaging system accessory (a mouse), not an AI algorithm.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    No, a standalone algorithm performance study was not done. This device is an imaging system accessory and not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This information is not available. No ground truth for image interpretation or diagnosis is relevant to the functionality of a display control mouse.

    8. The sample size for the training set:

    This information is not available. This is not an AI or machine learning device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established:

    This information is not available. See point 8.

    In summary, the provided document is a 510(k) premarket notification for an accessory (a control mouse) to an existing imaging system. The primary goal is to demonstrate "substantial equivalence" to the predicate device, arguing that the new accessory does not change the fundamental safety or effectiveness of the imaging system itself. Therefore, detailed clinical study data, acceptance criteria related to diagnostic performance, or ground truth establishment (as would be required for a novel diagnostic device or AI algorithm) are not pertinent to this submission and are not included.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050996

    Validate with FDA (Live)

    Date Cleared
    2005-05-05

    (15 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The O-Arm™ Imaging System is designed for 2D Fluoroscopic and 3D imaging for intraoperative applications in surgical theaters, particularly for orthopedic applications. The O-Arm™ Imaging System is compatible with certain Image Guided Surgery Systems.

    Device Description

    The O-Arm™ Imaging System is a mobile x-ray system which provides 3D imaging as well as 2D fluoroscopic imaging. The system consists of two parts: the x-ray O-Arm™ Stand (comprising x-ray generator, flat dynamic x-ray detector, and the x-ray control user interface) and the mobile view station (comprising the image processors, a user interface for image and patient handling, and viewing monitors).

    AI/ML Overview

    Here's an analysis of the O-Arm™ Imaging System's acceptance criteria and study, based on the provided 510(k) summary:

    O-Arm™ Imaging System Performance Analysis

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Diagnostic EquivalenceFor all image pairs (3-D and fluoroscopic), images from the O-Arm™ and the predicate device (Siemens ISO-C C-Arm Imaging System) were determined to be of equivalent diagnostic capability.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 31 pairs of images (including both 3-D and fluoroscopic images). A "pair" refers to an image from the O-Arm and a corresponding image from the predicate device of the same anatomical structure.
    • Data Provenance: The study was conducted using a cadaver. This suggests the data was from a controlled, experimental setting rather than a clinical environment. The country of origin is not explicitly stated, but the submission is to the US FDA. The study appears to be prospective as it was conducted specifically for the 510(k) submission.

    3. Number and Qualifications of Experts for Ground Truth

    • Number of Experts: One (1) radiologist.
    • Qualifications of Experts: The document simply states "A radiologist." No further details on experience or specialization are provided beyond their profession.

    4. Adjudication Method

    • The document implies a single-reader evaluation by the radiologist. There is no mention of multiple reviewers or an adjudication process like 2+1 or 3+1. The radiologist "determined" the equivalence independently.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed as described. The study involved a single radiologist making a determination of equivalence between the O-Arm and a predicate device. There is no assessment of human readers improving with AI assistance, as this is a device comparison study (equivalence) rather than an AI-assisted diagnostic study.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone study was implicitly done. The "device" in this context is the entire imaging system. The radiologist evaluated the images produced by the O-Arm itself (standalone, without human-in-the-loop assistance for image interpretation) and compared them to those from the predicate device.

    7. Type of Ground Truth Used

    • The ground truth was established through expert consensus/opinion by a single radiologist, comparing images from the investigational device to images from a legally marketed predicate device. The underlying "true status" of the anatomical structures was not explicitly determined by pathology or outcomes data, but rather by the radiologist's assessment of diagnostic capability equivalence between the two imaging systems.

    8. Sample Size for the Training Set

    • The document does not provide any information regarding a training set. This is not an AI/algorithm training study, but rather a device equivalence study. The O-Arm is an imaging system, not an AI software.

    9. How Ground Truth for the Training Set was Established

    • As no training set is mentioned or relevant to this type of device submission, this information is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1