Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K182616

    Validate with FDA (Live)

    Date Cleared
    2018-11-02

    (42 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NvisionVLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross-sectional, real-time depth visualization and may be used to mark areas of tissue. The software provides segmentation and display of common imaging features, including hyper-reflective surface, layering, and hypo-reflective structures.

    The NvisionVLE® Imaging System is intended to provide an image of tissue microstructure. The safety and effectiveness of this device for diagnostic analysis (i.e. differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.

    Device Description

    The NinePoint Medical NvisionVLE® Imaging System is a high-resolution volumetric imaging system based on optical coherence tomography (OCT). In an analogous fashion to ultrasound imagery, OCT images are formed from the time delay and magnitude of the signal reflected from the tissue of interest. The NvisionVLE Imaging System employs an advanced form of OCT known as sweptsource OCT (SS-OCT), or Optical Frequency Domain Imaging (OFDI), in combination with a scanning optical probe to acquire high-resolution, cross-sectional, real-time imagery of tissue called Volumetric Laser Endomicroscopy (VLE).

    In addition to the imaging capability, the device provides a means of marking areas of tissue with an additionally integrated 1470nm laser. The ability to create temporary laser marks directly on tissue enables a clinician to place visual reference marks on tissue regions of clinical interest immediately following their identification via VLE. The device consists of the following five main components and accessories: (i) a mobile NvisionVLE Console with an integrated computer and two touch-screen interfaces; (ii) proprietary NvisionVLE Software used to acquire, process, and visualize VLE images; (iii) a single-use, sterile NvisionVLE Marking Probe that is inserted through the working channel of an endoscope; (iv) a single-use, sterile NvisionVLE Inflation System that is used to inflate the Marking Probe's balloon to facilitate placement; and (v) a Probe Lock Accessory to prevent longitudinal motion of the Marking Probe within the endoscope.

    The purpose of this 510(k) submission is to add an artificial intelligence software tool referred to as Image and Visualization Enhancements (IVE) to the previously cleared, predicate NvisionVLE Imaging System (K153479). The IVE software module allows enhanced visualization (segmentation and colorized display) of the following commonly observed image features (also referred to as IVE features): (1) hyper-reflective surface, (2) layering and (3) hypo-reflective structures. The segmentation algorithm was developed using an artificial intelligence machine learning technique known as deep learning. Here, an artificial neural network was trained with manually labelled examples of each feature and then locked for realtime inference on new image data acquired by the device. Display of each feature can be toggled via the user interface, where a respective color overlay is presented. The default display of the IVE features is disabled and the standard VLE image data displayed per the cleared NvisionVLE Imaging System. Segmentation of these structures are based on existing image features, and IVE simply increases the conspicuity via the color overlays, thus aiding image review. It is a convenience tool and a resource for the clinician and as such, it does not alter the standard of care or the role of the physician in reviewing and assessing images generated by the system.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study proving the device meets those criteria. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states that "The target true positive and true negative detection fractions were prospectively set." However, the specific target values for these fractions are not explicitly listed in numerical form in the provided text. Instead, it states that the observed results exceeded their target value with a significance level a < 0.05.

    Here's the table of the reported device performance:

    ROITrue Positive Detection Fraction (%)Lower Limit of Exact 95% Confidence IntervalsTrue Negative Detection Fraction (%)Lower Limit of Exact 95% Confidence Intervals
    Tissue Surface92.9 (235/253)89.095.0 (209/220)91.2
    Layering89.6 (205/229)84.897.8 (220/225)94.9
    Hypo-Reflective Structures91.1 (175/192)86.292.9 (247/266)89.1

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size:
      • Hypo-Reflective Structures: 192 positive ROIs and 266 negative ROIs (Total 458 ROIs)
      • Layering: 229 positive ROIs and 225 negative ROIs (Total 454 ROIs)
      • Tissue Surface: 253 positive ROI and 220 negative ROIs (Total 473 ROIs)
    • Data Provenance: Data was obtained from a "sequestered subset (~40%) of the 1000-patient 18-site NvisionVLE Clinical Registry study." This indicates the data is from a prospective clinical registry. The country of origin is not explicitly stated, but the registration of the device with the FDA for a US company (NinePoint Medical, Inc. based in Bedford, Massachusetts) implies that the clinical study data is likely from the United States or at least includes data from the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Number of Experts: Not explicitly stated, but referred to as "trained experts."
    • Qualifications of Experts: They were "trained experts in clinical interpretation of VLE imagery." Specific details like "radiologist with 10 years of experience" are not provided.

    4. Adjudication Method for the Test Set:

    • The document states, "Assessment was performed by comparing ground truth data generated by human observers, with the software-based detection of the IVE features." It does not specify an adjudication method such as 2+1 or 3+1 for resolving discrepancies among multiple human observers. It refers to "trained experts" in the plural, suggesting more than one, but no detail on how consensus was reached or disagreements were handled is provided.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    • No, an MRMC comparative effectiveness study that assesses how much human readers improve with AI vs. without AI assistance was not explicitly stated or described. The study focused on the performance of the algorithm itself in detecting features against a human-generated ground truth, rather than human performance with and without the AI. The document describes the IVE as a "convenience tool" that "does not alter the standard of care or the role of the physician."

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done:

    • Yes, a standalone study was done. The performance testing "evaluated the ability of the software to detect each IVE feature... by comparing ground truth data generated by human observers, with the software-based detection of the IVE features." This describes the algorithm's performance independent of human-in-the-loop interaction for the quantitative metrics presented.

    7. The Type of Ground Truth Used:

    • Expert Consensus: The ground truth consisted of "ROIs labeled by trained experts in clinical interpretation of VLE imagery."

    8. The Sample Size for the Training Set:

    • The document states, "an artificial neural network was trained with manually labelled examples of each feature." However, the sample size for the training set is not explicitly provided in the given text.

    9. How the Ground Truth for the Training Set Was Established:

    • The ground truth for the training set was established through "manually labelled examples of each feature." It can be inferred that these labels were created by experts, similar to how the ground truth for the test set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182261

    Validate with FDA (Live)

    Date Cleared
    2018-10-16

    (56 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NvisionVLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross-sectional, real-time depth visualization.

    Device Description

    The NinePoint Medical NvisionVLE® Imaging System is a high-resolution volumetric imaging system based on optical coherence tomography (OCT). In an analogous fashion to ultrasound imagery, OCT images are formed from the time delay and magnitude of the signal reflected from the tissue of interest. The NvisionVLE® Imaging System employs an advanced form of OCT known as swept-source OCT (SS-OCT), or Optical Frequency Domain Imaging (OFDI), in combination with a scanning optical probe to acquire high-resolution, crosssectional, real-time imagery of tissue called Volumetric Laser Endomicroscopy (VLE).

    The device consists of the following main components and accessories: (i) a mobile NvisionVLE Console with an integrated computer and two touchscreen interfaces; (ii) proprietary NvisionVLE Software used to acquire, process, and visualize VLE images; (iii) a single-use, sterile NvisionVLE Optical Probe that is inserted through the working channel of an endoscope: (iv) a single-use, sterile NvisionVLE Inflation System that is used to inflate the balloon as required, to facilitate placement; and (v) a Probe Lock Accessory to prevent longitudinal motion of the Probe within the endoscope.

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the NvisionVLE Imaging System. It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study results for a new AI/ML-based device.

    Therefore, many of the requested elements for describing acceptance criteria and a study proving a device meets them (especially those related to AI/ML performance metrics like sensitivity, specificity, MRMC studies, or training/test set details for AI) are not present in this type of regulatory document.

    However, I can extract information related to the device's functional and non-clinical performance, which serve as its "acceptance criteria" in the context of this 510(k) pathway.

    Here's a breakdown based on the provided document:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Functional/Non-Clinical)Reported Device Performance (as tested)
    Image Quality: Confirmation of proper image on the ConsoleMet: Confirmed proper image on the Console.
    Optical Performance: Acceptable probe lateral resolution, back reflection, and transmission rateMet: Demonstrated acceptable probe lateral resolution, back reflection, and transmission rate.
    Mechanical Integrity: Mechanical integrity of the probe/sheath tip, demonstrated through tensile testingMet: Passed tensile testing, demonstrating mechanical integrity of the probe/sheath tip.
    Functional Operation: Functional testing to ensure the probe loads and withdraws properlyMet: Performed functional testing, and the probe loads and withdraws properly.
    (Compliance with Voluntary Standards)(Complies with various IEC, ISO, ANSI/AAMI standards for biological evaluation, sterilization, electrical safety, EMC, software, and usability, as listed in Section 7 of the 510(k) summary.)

    Note on "Acceptance Criteria" for this specific 510(k): For this Special 510(k), the "acceptance criteria" are primarily focused on demonstrating that the modification (addition of balloon-less low-profile optical probes) does not adversely affect the fundamental safety and effectiveness of the device, and that it maintains the performance characteristics expected of the original cleared device. There are no clinical diagnostic performance metrics (e.g., sensitivity, specificity for disease detection) because the device's cleared indication for use explicitly states that its safety and effectiveness for diagnostic analysis (differentiating normal vs. specific abnormalities) has not been evaluated.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any specific functional or non-clinical test. The testing described is general "functional testing" and "tensile testing" which implies a sufficient number of samples were tested to ensure design verification, but specific numbers are not disclosed in this summary.
    • Data Provenance: Not applicable in terms of patient data. The testing involves device components and systems (e.g., probes, consoles). This is non-clinical performance data, likely gathered at the manufacturer's testing facilities.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This 510(k) describes non-clinical and functional testing of device characteristics, not clinical performance or diagnostic accuracy assessed by experts.

    4. Adjudication method for the test set:

    • Not applicable. This is not a clinical study involving human interpretation of results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is explicitly stated: "The safety and effectiveness of this device for diagnostic analysis (i.e. differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated." This device is an imaging tool to visualize microstructure, not a diagnostic AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. The device is for human visualization of microstructure. It does not perform standalone diagnostic analysis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests described, the "ground truth" would be engineering specifications and established test methodologies (e.g., correct image display, specified resolution ranges, material strength limits, proper mechanical function). This is not a clinical ground truth like pathology.

    8. The sample size for the training set:

    • Not applicable. This document is for a medical imaging device, not an AI/ML product developed with a training set of data.

    9. How the ground truth for the training set was established:

    • Not applicable. (See #8)

    Summary of what this document does tell us about "acceptance criteria":

    For the NvisionVLE Imaging System, particularly concerning this Special 510(k) for new probe configurations, the acceptance criteria are entirely based on functional and non-clinical performance. The study "proving" the device meets these criteria involved a battery of engineering and bench tests, including:

    • Image quality confirmation on the console.
    • Quantitative measurements of optical performance (lateral resolution, back reflection, transmission rate) to ensure they meet specified levels.
    • Mechanical integrity testing of components (e.g., tensile testing of probes).
    • Verification of proper device operation (e.g., probe loading and withdrawal).
    • Compliance with recognized voluntary consensus standards for quality, safety, and manufacturing.

    The FDA's clearance indicates that these non-clinical tests demonstrate the modified device is "as safe and effective as the predicate device" and does "not raise any new or different questions of safety or effectiveness." It is crucial to reiterate that the device's cleared indication for use specifically states that its diagnostic capability for differentiating abnormalities has not been evaluated, meaning it's primarily a visualization tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143678

    Validate with FDA (Live)

    Date Cleared
    2015-02-06

    (44 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nvision VLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross-sectional, real-time depth visualization.

    Device Description

    The NinePoint Medical NvisionVLE™ Imaging System is a general imaging system comprised of the NvisionVLETM Console, NvisionVLETM Optical Probe and the NvisionVLE™ Inflation Accessory Kit. The NvisionVLE™ Optical Probe is made up of an optical probe subassembly and a quide sheath. The optical probe subassembly is a fiber optic probe assembly secured inside a flexible, stainless steel torque shaft. The distal optics are housed in a stainless steel hypotube which is attached to the torque shaft. The proximal end of the optical fiber and torque shaft terminate in a standard fiber optic connector and catheter connector which interfaces with the system console. The optical probe subassembly transmits the optical signal and detects the reflected optical signal for image reconstruction of the targeted tissue. The guide sheath is a coaxially-designed balloon sheath. The sheath is composed of a PET balloon and a nylon shaft. The inner lumen of the sheath is sealed, enclosing the optical probe subassembly. The guide sheath is positioned within the organ structure of interest and allows the probe to rotate in a helical pattern while positioned in the inner lumen allowing for image reconstruction of the targeted tissue.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Proposed Enhancements)
    Durability / Resistance to Failure in Tortuous Anatomy100% of enhanced probe samples passed the most tortuous curve (Tortuous Curve 4). All 13 enhanced samples passed all four tortuous curves.
    No binding of the torque coil leading to optical fiber fractureNo failures (binding of torque coil, fracture of optical fiber, failure to transmit signal) observed in enhanced samples across tortuous curves.
    Improved device trackabilityStated as a benefit of the torsionally-stiffer torque coil, though not explicitly an acceptance criterion or quantified in the performance data.
    Reduced optical signal disruption and incomplete imaging in tortuous anatomyStated as a benefit of the torsionally-stiffer torque coil, though not explicitly an acceptance criterion or quantified in the performance data.
    Better absorption of vibrational and friction forces in tortuous anatomiesSupported by the pass rates in tortuous curves, indicating improved resilience.
    Reduced stress on the optical fiberSupported by the pass rates, as failure involves optical fiber fracture due to stress.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Initial Engineering Test Samples: 5 K120800 Controls, 5 Proposed Enhancements (Total = 10 samples)
      • Manufacturing Build Samples: 10 K120800 Controls, 7 Proposed Enhancements (Total = 17 samples)
      • Overall Reported (summary): 15 control samples, 13 enhanced samples (This slightly differs from the sum of the detailed tables (15 controls vs 17 controls; 13 enhanced vs 12 enhanced). The summary likely refers to the effective number of control samples that experienced failure across the two tables combined, and the number of enhanced samples that successfully completed the most rigorous testing.)
    • Data Provenance: The data appears to be retrospective in the sense that it evaluates modifications to an already cleared device and compares them against the previously cleared device. It's an internal engineering and manufacturing evaluation. The country of origin is not explicitly stated, but the company is based in Cambridge, Massachusetts, USA, suggesting the testing was likely conducted in the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This study is a performance test on device durability and mechanical integrity, not an evaluation of diagnostic accuracy requiring expert interpretation of images or patient outcomes. The "ground truth" for this test is the physical failure of the device (binding, fiber fracture, signal loss) under simulated tortuosity.

    4. Adjudication Method for the Test Set

    Not applicable. The outcome (pass/fail) is objectively determined by whether the device's optical fiber fractured or the torque coil bound, leading to a loss of signal. This is a direct physical outcome rather than an interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This study assesses the physical durability and mechanical performance of the device's components (optical probe and guide sheath), not its effectiveness in diagnostic tasks with human readers.

    6. Standalone Performance Study (Algorithm Only)

    Yes, in a sense. This is a standalone device performance study focusing on the mechanical durability of the device itself, rather than an algorithm. The "algorithm" aspect of imaging systems (image reconstruction, etc.) is not evaluated here, only the physical integrity of the probe.

    7. Type of Ground Truth Used

    The ground truth used is physical device failure (failure of the torque coil to bind, fracture of the optical fiber, and subsequent failure to transmit signal) under defined simulated conditions of tortuosity.

    8. Sample Size for the Training Set

    Not applicable. This study does not involve a "training set" in the context of machine learning. It's a physical engineering and manufacturing test of device durability.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an algorithm. The "ground truth" for the test set (physical failure) was established by observing the operational state of the device under stress in a test fixture.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1