Search Results
Found 1 results
510(k) Data Aggregation
(114 days)
The Nurture Rest (including models 100-1-C) is indicated for premature infants undergoing oxygen therapy to maintain an airway and for whom treatment may be facilitated by this repositioning and stabilization device. The premature infants are being continuously monitored for oxygen saturation levels and heart rate in the Neonatal Intensive Care Unit (NICU) by medical professionals.
Nurture Rest is an infant sleep positioning device used by medical staff to position and secure premature infants in a prone, supine, or side position to limit mobility within an incubator in the neonatal intensive care unit (NICU). The Nurture Rest is shaped like the female anatomy in order to encourage a "Kangaroo Care" style positioning technique that mimics being held against a female body/chest while also keeping the infant's head, neck, and spine aligned. This device gives nurses a useful tool in their daily role of repositioning and caring for these special premature infants. The infants for whom the device is used are undergoing oxygen therapy to maintain an airway (intubation, nasal CPAP, bubble CPAP, high flow nasal cannula, or nasal cannula) and are continuously monitored for oxygen saturation levels and heart rate. The Nurture Rest consists of a washable, flame-retardant cloth material.
There are two models of the device (Models 100-1-O and 100-2-C). Both devices are appropriate for prone, supine and side positioning of premature infants. One model of the device (Model 100-1-O) has customizable opening at the head to allow for prone face-down positioning of micro-preemies for whom positioning with the head turned to the side would increase the chance of hemorrhaging due to increased intracranial pressure. The other model (Model 100-2-C) does not have an opening and is for prone with head turned positioning. The device comes with pillows to help stabilize the infant and swaddling straps to minimize motion of the arms and body of the infant. As with standard NICU practices, the Nurture Rest enables the infant to be re-positioned and treated as needed (typically every 4 to 6 hours), which allows NICU staff to cluster care in order to let the infants rest and grow in between physical and emotional stimulation.
The Nurture Rest device is a pediatric position holder for premature infants in the Neonatal Intensive Care Unit (NICU). Below is a summary of its acceptance criteria and the study that proves the device meets them.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Functionality | Device is able to position an infant within the NICU. | Bench top testing: "In all instances, the Nurture Rest functioned as intended and all test results observed were as expected." & Usability Testing: "This study demonstrated that the device was successfully able to be used to position, align and retain a doll which was representative of actual premature patients, and therefore, meet the intended use." |
| Usability | NICU nurses can safely and effectively use the Nurture Rest as intended, including positioning of infant patients and care/maintenance of the device. | Summative Usability Testing: "The goal of the study was to ensure that NICU nurses can safely and effectively use the Nurture Rest as intended. The simulated use study focused on all anticipated interactions of the intended users with the device including positioning of infant patients (micro-preemie) using the Nurture Rest as well as care and maintenance of the device. This study demonstrated that the device was successfully able to be used to position, align and retain a doll..." |
| Biocompatibility | Materials are non-cytotoxic, non-irritating, and non-sensitizing. | Biocompatibility testing (per ISO 10993-1 surface contact testing): "The Nurture Rest was found to be non-cytotoxic, a non-irritant and did not elicit a sensitization response." |
| Durability | The product's shape shall still enable a user to position a Preemie doll per the instructions for use after 4 weeks of use. | Durability Testing: Demonstrated that the product's shape enabled a user to position a Preemie doll per IFU after 4 weeks of use. |
| Material Properties | Coefficient of Friction is comparable to other products in contact with patient skin in the NICU. Bioburden is comparable to other items in contact with patient skin in the NICU. | Coefficient Friction Testing: Performed to demonstrate comparability. Bioburden Testing: Performed to ensure comparability. |
| Storage & Packaging | The packaged device can withstand worst-case storage and shipping scenarios. | Storage testing: Performed to ensure the packaged device could withstand worst-case storage scenarios. Packaging and ship testing: Performed to ensure the packaged device could withstand worst-case shipping scenarios. |
2. Sample Size and Data Provenance for Test Set
- Sample Size for Test Set: The document does not specify a numerical sample size for "test sets" in the traditional sense, as the studies primarily involve bench, simulated, and material testing.
- For the bench top and usability testing, the 'sample' involved the "Nurture Rest" device itself and a "doll which was representative of actual premature patients." The number of devices or dolls tested is not explicitly stated.
- For biocompatibility, durability, coefficient friction, bioburden, storage, and packaging/ship testing, the 'sample' refers to units of the Nurture Rest device. The specific number of units tested is not provided.
- Data Provenance: Not explicitly stated as retrospective or prospective data from human subjects. The studies are primarily lab-based (bench, simulated, material testing) and do not involve data from patients. The testing was conducted by Neonatal Loving Kare, Inc. or their contracted labs. The country of origin of the data is not specified but implicitly within the US given the FDA submission.
3. Number of Experts and Qualifications for Ground Truth of Test Set
This information is not applicable or not provided in the document. The ground truth for the performance tests was established by objective measurements and functional assessments against predetermined criteria, rather than expert consensus on diagnostic interpretations or outcomes. For usability, the study involved "NICU nurses," but the number and their specific qualifications (e.g., years of experience) beyond being NICU nurses are not detailed.
4. Adjudication Method for Test Set
Not applicable. The reported studies are primarily objective performance, usability, and material tests, not clinical evaluations requiring adjudication of subjective assessments.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. An MRMC comparative effectiveness study was not done. The device is a physical pediatric position holder, not an AI or diagnostic imaging device that would typically undergo such a study. The document describes a comparison to a predicate device primarily based on technological characteristics and intended use, not a comparative clinical effectiveness study against human performance.
6. Standalone (Algorithm Only) Performance Study
No. This is a physical medical device, not an algorithm or software. Therefore, a standalone (algorithm only) performance study is not applicable.
7. Type of Ground Truth Used
The ground truth used for the acceptance criteria was based on:
- Objective functional performance: Whether the device could physically position and retain a representative doll (simulated use).
- Material properties standards: Adherence to biocompatibility standards (ISO 10993-1) and comparison of properties like coefficient of friction and bioburden to existing NICU standards/products.
- Usability observation: Direct observation and assessment of NICU nurses using the device during simulated scenarios to confirm safe and effective use.
- Durability and integrity standards: Confirmation that the device maintains its form and function over time and through transport.
8. Sample Size for the Training Set
Not applicable. This device is a physical product and does not involve machine learning algorithms that require a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1