Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232955

    Validate with FDA (Live)

    Date Cleared
    2024-01-10

    (111 days)

    Product Code
    Regulation Number
    882.5950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Numen™ Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

    Numen™ Coil Embolization System is indicated for endovascular embolization of:

    -Intracranial aneurysms

    -Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae -Arterial and venous embolizations in the peripheral vasculature

    NumenFR™ Detachment System is intended for use with MicroPort NeuroTech Numen™ Coil Embolization System in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature.

    Device Description

    MicroPort NeuroTech has developed the Numen™ Coil Embolization System and NumenFR™ Detachment System. The Numen™ Coil Embolization System is designed to be used in conjunction with the NumenFRTM Detachment System (sold separately) for endovascular embolization of vascular abnormalities described in the intended use.

    The Numen™ Coil Embolization System is composed of two parts as described below:

    • . An introducer sheath: The function of the introducer sheath is to facilitate introduction of the coil into the microcatheter.
    • . The coil system: The coil system is composed of a pusher and coil implant. The coil is a permanent implant intended to occlude blood flow in vascular abnormalities. The pusher is used to deliver the coil implant to the target lesion.

    The MicroPort NeuroTech NumenFRTM Detachment System is a sterile, handheld, single-patient use device designed for use with the MicroPort NeuroTech Numen™ Coil Embolization System. The device is operated by two pre-loaded batteries.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Numen™ Coil Embolization System and NumenFR™ Detachment System. The document details the device, its intended use, comparison to a predicate device, and performance testing. However, it does not describe acceptance criteria for a study proving the device meets those criteria in the context of an Artificial Intelligence/Machine Learning (AI/ML) powered medical device.

    The "Performance Testing" section (page 6) outlines bench testing conducted to evaluate device changes and demonstrate substantial equivalence to a predicate device. This is typical for traditional (non-AI/ML) medical devices, focusing on physical and functional properties, rather than AI performance metrics like sensitivity, specificity, or reader agreement.

    Therefore, I cannot fulfill your request to describe acceptance criteria and a study demonstrating an AI/ML device's performance based on the provided text, as the text pertains to a different type of medical device assessment.

    If you can provide a document that discusses the evaluation of an AI/ML medical device, I would be happy to analyze it according to your requested criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203625

    Validate with FDA (Live)

    Date Cleared
    2021-09-01

    (264 days)

    Product Code
    Regulation Number
    882.5950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Numen™ Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

    Numen™ Coil Embolization System is indicated for endovascular embolization of:

    • Intracranial aneurysms
    • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
    • Arterial and venous embolizations in the peripheral vasculature

    NumenFR™ Detachment System is intended for use with MicroPort NeuroTech Numen™ Coil Embolization System in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature.

    Device Description

    The Numen™ Coil Embolization System is designed to be used in conjunction with the NumenFRTM Detachment System (sold separately) for endovascular embolization of vascular abnormalities described in the intended use. The Numen™ Coil Embolization System is composed of an introducer sheath and the coil system (pusher and coil implant). The coil is a permanent implant intended to occlude blood flow in vascular abnormalities. The pusher is used to deliver the coil implant to the target lesion. The MicroPort NeuroTech NumenFRTM Detachment System is a sterile, handheld, single-patient use device designed for use with MicroPort NeuroTech Numen™ Coil Embolization System. The device is operated by two pre-loaded batteries.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Numen™ Coil Embolization System and NumenFR™ Detachment System submitted to the FDA. This document focuses on demonstrating substantial equivalence to predicate devices through various performance tests and comparisons. It is not an AI/ML device, and thus does not include information typically found in studies for AI/ML devices. Therefore, a direct answer to the request based solely on the provided text, especially regarding AI/ML-specific acceptance criteria, ground truth, and human reader studies, is not possible.

    However, I can extract the general acceptance criteria and performance results as described for this medical device (a coil embolization system), even though they are not related to AI/ML. The provided document details a series of non-clinical performance tests and biocompatibility assessments to establish substantial equivalence.

    Here's an attempt to answer based on the provided text, indicating where information is present and where it is not applicable or unavailable given the nature of the device and the document:


    Acceptance Criteria and Device Performance for Numen™ Coil Embolization System and NumenFR™ Detachment System (Non-AI/ML Medical Device)

    This document describes the validation of a neurovascular embolization device, not an AI/ML system. Therefore, many of the requested criteria related to AI/ML performance metrics, ground truth establishment by experts, and human reader studies are not applicable. The information provided focuses on the physical and functional performance of a medical device (coils and a detachment system).

    1. Table of Acceptance Criteria and Reported Device Performance

    For the Numen™ Coil Embolization System, the acceptance criteria are generally qualitative ("Pass") indicating that the device performed as intended and met pre-specified requirements. The performance results consistently show "Pass" for all tests.

    TestTest Method SummaryAcceptance Criteria (Implicit: "Pass")Reported Device Performance
    Visual inspectionExamine the test sample surface under specific magnification.Visual inspection confirms material quality and absence of defects.Pass
    Dimensional verificationImplant coil, pusher and introducer sheath dimensions are measured to match the specifications.Dimensions meet specifications.Pass
    Simulated useVerify that the coil embolization system performs as intended in a representative tortuous anatomical model.Device performs as intended in simulated use.Pass
    Fatigue testingVerify the durability of the coil embolization system by repeating the simulated use for six times, including coil retraction into microcatheter and re-deployment.Device durable after repeated simulated use.Pass
    Detachment time and detachment reliabilityVerify the reliability of intentional detachment as well as reliability of the coil attachment after fatigue testing of the coil embolization system in a representative tortuous anatomical model.Reliable and timely detachment.Pass
    Ease of deliveryMeasured by max friction force when advancing, deploying or retracting the coil system through the introducer sheath and microcatheter in a relevant tortuous anatomical model.Ease of delivery within acceptable force limits.Pass
    Kink resistanceDemonstrate that kink resistance of the device meets pre-specified acceptance criteria, and could withstand bending forces that the device may encounter during clinical use.Device withstands bending forces without kinking.Pass
    Torque strengthVerify the torque strength by rotating the proximal end of the device for 8 turns.Device maintains integrity under specified torque.Pass
    RadiopacityQualitatively assess the radiopacity of the device under X-ray to demonstrate the equivalence to the control device.Radiopacity comparable to control device.Pass
    Particulate testingVerify that the quantity and size of particulates generated during simulated use in a clinically relevant tortuous anatomical model with all recommended ancillary devices meet the acceptance criteria based on the comparison with the predicate.Particulate generation within acceptable limits compared to predicate.Pass
    Coil deformation forceTo verify that the coil implants are not too stiff to loop adequately and achieve their intended secondary shape.Coils deform adequately to achieve intended shape.Pass
    Tensile strength testingVerify that detachment zone tensile strength, stretch resistance tensile strength, pusher joints tensile strength, and introducer sheath tensile strength meet the acceptance criteria based on a control device.Tensile strengths meet acceptance criteria compared to control.Pass
    Corrosion ResistanceCorrosion resistance testing per ASTM F2129 for coil implant, per ISO 11070 for pusher.Device resists corrosion.Pass
    MR CompatibilityMR testing performed per ASTM F2119, ASTM F2213, ASTM F2052 and additional MRA characterization testing.Device is MR conditional.Testing demonstrated the device is MR conditional.
    Packaging and Shelf LifePackage verification and shelf-life testing performed to demonstrate the integrity of the sterile barrier packaging throughout the proposed shelf life and its ability to protect the packaged devices from damage and maintain sterility during transport and storage conditions.Packaging integrity and sterility maintained for proposed shelf life.Packaging and device testing demonstrate the ability to perform as intended through the labeled 2-year shelf life of the device.
    Sterilization ValidationPer ISO 11135, Annex B Overkill Method.Sterility assurance level of 10-6 achieved.Sterilization process achieves sterility assurance level of 10-6.
    Endotoxin TestingBacterial endotoxin assay validation per USP 85.Endotoxin levels below 2.15 EU/device.The endotoxin levels for the device are below 2.15 EU/device.
    GLP animal studyAnimal testing to evaluate the in vivo performance of the device in a canine model, reporting both acute and chronic time points.In vivo performance is acceptable.Pass
    CytotoxicityMEM Elution per ISO 10993-5 (coil implant, pusher and sheath).Non-cytotoxic.Pass (Non-cytotoxic)
    SensitizationISO Guinea Pig Maximization Sensitization Test per ISO 10993-10 (coil implant, pusher and sheath).Non-sensitizer.Pass (Non-sensitizer)
    IrritationISO Intracutaneous Irritation Test per ISO 10993-10 (coil implant, pusher and sheath).Non-irritant.Pass (Non-irritant)
    Acute ToxicityISO Acute Systemic Injection Test per ISO 10993-11 (coil implant, pusher and sheath).Non-toxic.Pass (Non-toxic)
    PyrogenicityISO Material Mediated Rabbit pyrogen per ISO 10993-11 (coil implant, pusher and sheath).Non-pyrogenic.Pass (Non-pyrogenic)
    Hemocompatibility (Hemolysis)ASTM Hemolysis Study - Direct Contact and Extract method per ISO 10993-4 (coil implant, pusher and sheath).Non-hemolytic.Pass (Non-hemolytic)
    Hemocompatibility (Complement Activation)Complement Activation SC5b-9 Assay per ISO 10993- 4 (coil implant, pusher and sheath).Non-activator.Pass (Non-activator)
    Hemocompatibility (PTT)Partial Thromboplastin Time (PTT) per ISO10993-4 (coil implant, pusher and sheath).Acceptable PTT.Pass
    Hemocompatibility (Thromboresistance)In vivo Thromboresistance Study in Dogs per ISO 10993-4 (pusher and sheath).Thromboresistant.Pass
    Hemocompatibility (Thromboresistance, heparinized)In vivo Thromboresistance Study in Dogs, heparinized model per ISO 10993-4 (pusher and sheath).Thromboresistant in heparinized model.Pass Thromboresistant
    Genotoxicity (Bacterial Mutagenicity)Bacterial Mutagenicity Test per ISO 10993-3 (coil implant).Non-genotoxic.Pass (Non-genotoxic)
    Genotoxicity (Mouse Lymphoma Assay)In vitro Mouse Lymphoma Assay per ISO 10993-3 (coil implant).Non-genotoxic.Pass (Non-genotoxic)
    ImplantationISO Muscle Implantation Study in Rabbits - 1 week, 4 weeks, 13 weeks per ISO 10993-6 (coil implant).Acceptable tissue response.Pass
    Subchronic toxicityISO Systemic Toxicity Study in Rats Following Subcutaneous Implantation, 13 Weeks per ISO 10993-6 and per ISO 10993-11 (coil implant).No evidence of systemic toxicity.Pass (No evidence of systemic toxicity.)
    CarcinogenicityToxicological Risk Assessment (coil implant).Acceptable carcinogenic risk.Pass
    Chronic ToxicityToxicological Risk Assessment (coil implant).Acceptable chronic toxicity risk.Pass

    For the NumenFR™ Detachment System, performance testing included:

    • Functional performance testing to assess proper operation.
    • Software verification in accordance with IEC 62304:2015 Ed. 1.1 and risk assessment per ISO 14971:2007/(R) 2010.
    • Electrical Safety Testing and EMC Testing in accordance with AAMI/ANSI ES 60601-1:2005, IEC 60601-1-2:2014 Ed. 4 and IEC 60601-1-6 Ed.3.1.
    • Sterilization validation per ISO 11135:2014 (SAL of at least 10^-6).
    • Accelerated aging shelf-life study supporting a 2-year shelf life.

    All these tests are reported to have "Pass" results, indicating the device meets its established criteria for safety and performance compared to the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify exact numerical sample sizes for each test (e.g., number of coils tested for fatigue). It generally refers to "test units representative of final finished devices." For the animal study, it mentions "a canine model" but not the number of animals.
    • Data Provenance: The studies are non-clinical (bench testing, in vitro, and in vivo animal studies). There's no mention of human clinical data or geographical origin of such data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the document describes a traditional medical device, not an AI/ML system. Ground truth in this context refers to established engineering specifications, biocompatibility standards (e.g., ISO and ASTM standards), and observed physical/biological performance rather than expert interpretation of medical images or data.

    4. Adjudication Method for the Test Set

    Not Applicable. As this is not an AI/ML diagnostic device involving human interpretation, there is no mention of adjudication methods for a test set. Acceptance is based on meeting predefined engineering and biological performance metrics.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This is a physical medical device, not an AI-assisted diagnostic tool. No MRMC study was conducted or is relevant for this type of device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is a physical medical device. There is no algorithm to test in a standalone manner.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance validation is based on:

    • Established engineering specifications: Pre-defined dimensional, material, and functional criteria.
    • International Standards: Compliance with ISO (e.g., 10993 series for biocompatibility, 11135 for sterilization) and ASTM (e.g., F2129 for corrosion, F2119 for MR compatibility) standards.
    • Predicate Device Performance: Benchmarking against legally marketed predicate devices to demonstrate "substantial equivalence."
    • In-vivo animal model outcomes: Evaluating device performance in a canine model for acute and chronic time points.

    8. The Sample Size for the Training Set

    Not Applicable. This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. As there is no training set for an AI/ML model, this question is not relevant to the provided documentation.


    Ask a Question

    Ask a specific question about this device

    Page 1 of 1