Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K223181

    Validate with FDA (Live)

    Date Cleared
    2023-01-11

    (92 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a pedicle screw fixation system in skeletally mature patients, the NuVasive Reline System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:

    1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
    3. Fracture
    4. Dislocation
    5. Scoliosis
    6. Kyphosis
    7. Spinal tumor and/or
    8. Failed previous fusion (pseudoarthrosis)

    The Nu Vasive Reline System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

    When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline System is also intended for the following indications:

    1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    2. Spinal stenosis
    3. Spondylolisthesis
    4. Spinal deformities
    5. Fracture
    6. Pseudoarthrosis
    7. Tumor resection and/or
    8. Failed previous fusion

    When used for posterior non-cervical screw fixation in pediatric patients, NuVasive Reline System is indicated as an adjunct to fusion to treat:

    1. adolescent idiopathic scoliosis.
    2. progressive spinal deformities (i.e., scoliosis, or lordosis) including infantile and juvenile idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis,
    3. spondylolisthesis,
    4. spondylolysis,
    5. pseudarthrosis,
    6. failed prior fusion and
    7. fracture caused by tumor and/or trauma in pediatric patients.

    Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

    In order to achieve additional levels of fixation, the NuVasive Reline System rods may be connected to the Armada System.

    Device Description

    The NuVasive Reline System is a pedicle screw system that consists of a variety of screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, iliac connectors and associated general instruments. Implant components are available in a variety of sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. The purpose of this submission is to introduce "Dual-Headed Modular Connectors" which will be included in the NuVasive Reline System. Additionally, this submission is intended to specify pediatric progressive deformities in the indications for use statement for NuVasive Reline System. Lastly, this submission is intended to add minor labeling changes to address compatibility updates to the NuVasive Reline system.

    AI/ML Overview

    The provided document describes the NuVasive Reline System, a pedicle screw system, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving that a device meets specific acceptance criteria in the context of diagnostic accuracy or clinical outcomes.

    The "Performance Data" section (G) on page 6 refers to nonclinical testing, which are engineering tests to assess the physical properties and mechanical performance of the device. These are distinct from clinical studies designed to evaluate diagnostic accuracy or the effectiveness of AI algorithms.

    Therefore, I cannot fulfill the request as the document does not provide the necessary information regarding acceptance criteria, device performance, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for a study related to meeting such criteria.

    The information provided is focused on demonstrating substantial equivalence for a medical device (pedicle screw system), primarily through design comparison, material composition, and mechanical testing according to established ASTM standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182974

    Validate with FDA (Live)

    Date Cleared
    2019-02-13

    (110 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a pedicle screw fixation system, the Nu Vasive Reline System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:

    1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
    3. Fracture
    4. Dislocation
    5. Scoliosis
    6. Kyphosis
    7. Spinal tumor and/or
    8. Failed previous fusion (pseudoarthrosis)

    The Nu Vasive Reline System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the LS-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

    When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline System is also intended for the following indications:

    1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    2. Spinal stenosis
    3. Spondylolisthesis
    4. Spinal deformities
    5. Fracture
    6. Pseudoarthosis
    7. Tumor resection and/or
    8. Failed previous fusion

    When used for posterior non-cervical screw fixation in pediatric patients, NuVasive Reline System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline System is intended to treat pediatric patients diagnosed with the following conditions: spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

    In order to achieve additional levels of fixation, the NuVasive Reline System rods may be connected to the Armada System.

    Device Description

    The NuVasive Reline System is a pedicle screw system that consists of a variety screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, iliac connectors and associated general instruments. Implant components are available in a variety sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. The purpose of this submission is to introduce design modifications to previously cleared components and add iliac saddle to previously cleared Reline System.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the NuVasive Reline System, a thoracolumbosacral pedicle screw system.

    Here's an analysis based on your request, but please note that this type of FDA 510(k) summary typically does not include detailed clinical study data with acceptance criteria, sample sizes, expert qualifications, or MRMC studies for traditional orthopedic implants like pedicle screw systems. Instead, it relies heavily on substantial equivalence to predicate devices and mechanical performance testing.

    Therefore, many of the requested points below will be answered with "Not applicable/provided" or explanations about why such details are not present in this type of submission.


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Not explicitly stated as such, but inferred from regulatory requirements and predicate comparison for mechanical testing)Reported Device Performance (as described in the summary)
    Mechanical performance equivalent to predicate devices.Demonstrated substantial equivalence (SE) to predicate devices in areas including design, labeling/intended use, material composition, and function.
    Pass relevant ASTM standards for spinal implant systems. (Specifically mentioned ASTM F1798)ASTM F1798 mechanical testing was performed to evaluate the design modifications, and "engineering rationale was provided to demonstrate that the subject NuVasive Reline System is substantially equivalent to the predicate device."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in the provided document. The testing seems to be based on mechanical (bench) testing, not patient data.
    • Data Provenance: Not applicable, as no human data is described. The engineering rationale and mechanical testing would be conducted in a laboratory setting.
    • Retrospective/Prospective: Not applicable, as no human data is described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/provided. This device is a mechanical implant, and its performance for 510(k) relies on mechanical testing and substantial equivalence to existing devices, not clinical ground truth established by medical experts in an diagnostic or interpretive sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/provided. As there's no clinical test set with human assessments or interpretations, adjudication methods are not relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/provided. This is a spinal implant, not an AI software/diagnostic device. MRMC studies are not relevant to its 510(k) clearance process described here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/provided. This is a physical medical device, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable in the clinical sense. The "ground truth" for this type of device typically refers to engineering specifications, material properties, and the performance standards established by predicate devices and ASTM standards, validated through mechanical testing.

    8. The sample size for the training set

    • Not applicable/provided. This refers to the training of an algorithm or AI model, which is not relevant to this device's submission.

    9. How the ground truth for the training set was established

    • Not applicable/provided. This refers to the ground truth for an AI model's training, which is not relevant to this device's submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143684

    Validate with FDA (Live)

    Date Cleared
    2015-04-23

    (120 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a pedicle screw fixation system, the Nu Vasive Reline System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:

    1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
    3. Fracture
    4. Dislocation
    5. Scoliosis
    6. Kyphosis
    7. Spinal tumor and/or
    8. Failed previous fusion (pseudoarthrosis)

    When used for posterior non-cervical screw fixation in pediatric patients, Nu Vasive Reline System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline System is intended to treat pediatric patients diagnosed with the following conditions: spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

    The Nu Vasive Reline System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

    When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline System is also intended for the following indications:

    1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
    2. Spinal stenosis
    3. Spondylolisthesis
    4. Spinal deformities
    5. Fracture
    6. Pseudoarthosis
    7. Tumor resection and/or
    8. Failed previous fusion

    In order to achieve additional levels of fixation, the Reline System rods may be connected to the Armada System.

    Device Description

    The NuVasive Reline System is a pedicle screw system that consists of a variety screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, iliac connectors and associated general instruments. Implant components are available in a variety sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. This 510(k) is to add new components to the Reline System.

    AI/ML Overview

    I am unable to provide a description of the acceptance criteria and the study that proves the device meets them because the provided text describes a 510(k) premarket notification for the "NuVasive® Reline™ System" which is a pedicle screw spinal system.

    The document is an FDA clearance letter and a 510(k) summary, and it primarily focuses on establishing substantial equivalence to existing predicate devices based on design, intended use, material composition, and function. The "Performance Data" section mentions nonclinical testing (e.g., dynamic compression bending per ASTM F1717) but does not provide quantitative acceptance criteria or detailed results that would allow me to populate a table of acceptance criteria and reported device performance.

    Furthermore, this type of device (spinal fixation system) does not typically involve the kinds of studies (e.g., standalone algorithm performance, MRMC studies, expert consensus on image interpretation, training/test set data provenance, ground truth establishment for AI/diagnostics) that would be needed to answer the specific questions posed about device performance for an AI/diagnostic device. The information requested is relevant to AI/machine learning medical devices, which this product is not.

    Therefore, I cannot fulfill the request as the provided text relates to a different type of medical device lacking the details you are asking for.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1