Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K191672

    Validate with FDA (Live)

    Date Cleared
    2019-10-15

    (113 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NuFACE Mini Device is intended for facial and neck stimulation and is indicated for over-the-counter use.

    Device Description

    The NuFACE® Mini Device is an over- the- counter facial and neck toning device. The chrome plated dual spheres of the NuFACE® Mini are designed to gently glide over the skin to deliver low-level electrical impulses to strategic locations on the face and neck. The NuFACE® Mini electrodes are designed for optimal contact with the face.

    The NuFACE® Mini microcurrent continually alternates between the positive electrodes and allows the user to adjust settings.

    The NuFACE® Mini Device measures approximately 2.5" W x 4.2" D. Its outer case is injection mold thermoplastic resin. The device comes with an external Power Supply to recharge the lithium ion battery of the device when not in use. The external Power Supply is a pre- approved wall adapter. All charging circuitry is contained within the device itself.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the NuFACE Mini Device. It aims to demonstrate substantial equivalence to a predicate device, the NuFACE Trinity Device, rather than proving the device meets specific performance acceptance criteria through clinical studies. Therefore, much of the requested information regarding study design, sample sizes, expert ground truth establishment, and comparative effectiveness studies is not present in this type of submission.

    However, based on the provided text, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    Since this filing is for substantial equivalence, the "acceptance criteria" are primarily established by the characteristics and performance of the predicate device (NuFACE Trinity). The reported device performance for the NuFACE Mini is presented as a comparison to the predicate.

    CharacteristicAcceptance Criteria (Predicate: NuFACE Trinity)Reported Device Performance (NuFACE Mini)Met/Comparable
    Regulatory
    Regulation Number21 C.F.R. § 882.589021 C.F.R. § 882.5890Met (Same)
    Regulation NameTranscutaneous electrical nerve stimulator for pain reliefTranscutaneous electrical nerve stimulator for pain reliefMet (Same)
    Regulatory ClassClass IIClass IIMet (Same)
    Product CodeNFONFOMet (Same)
    Intended UseFacial and neck stimulation for over-the-counter cosmetic useFacial and neck stimulation for over-the-counter cosmetic useMet (Same)
    Indications for UseOver-the-Counter Cosmetic UseOver-the-Counter Cosmetic UseMet (Same)
    Anatomic SitesFace and NeckFace and NeckMet (Same)
    Technological Characteristics
    Form Factor (General)Handheld, dual spherical electrodes, microcurrentHandheld, dual spherical electrodes, microcurrentComparable (Same general design principles)
    WaveformPulsed BiphasicPulsed BiphasicMet (Same)
    ShapeModulated SquareModulated SquareMet (Same)
    Max Output Voltage28 VDC28 VDCMet (Same)
    Max Output Current$400 \mu A$ at 500 Ω$348 \mu A$ at 500 ΩComparable (Slightly lower, but within expected range for similar devices/intended use, and not clinically significant for this type of equivalence claim)
    Max Output Current Density0.419 mA/cm²0.452 mA/cm²Comparable (Slightly higher, but within expected range for similar devices/intended use, and not clinically significant for this type of equivalence claim)
    Output Current (not stimulating)< 1 μΑ< 1 μΑMet (Same)
    Output Tolerance+/- 10%+/- 5%Met (Improved)
    Pulse Width60 msec60 msecMet (Same)
    Frequency (Hz)Approximately 8.3 HzApproximately 8.3 HzMet (Same)
    Beat FrequencyNo Beat FrequencyNo Beat FrequencyMet (Same)
    Burst Mode - Pulses per burst2020Met (Same)
    Burst Mode - Pulses per second8.38.3Met (Same)
    Burst Mode - Burst duration2.4 s2.4 sMet (Same)
    Burst Mode - Duty Cycle50%50%Met (Same)
    ON Time60 msec60 msecMet (Same)
    OFF Time60 msec60 msecMet (Same)
    Power SourceInternal rechargeable Lithium ion batteryInternal rechargeable Lithium ion batteryMet (Same)
    Line Current IsolationType BFType BFMet (Same)
    Patient Leakage CurrentN/A - Battery operatedN/A - Battery operatedMet (Same)
    External Power AdapterNuFACE 7-volt power adapterNuFACE 5-volt power adapterComparable (Different voltage, but functionality and safety verified through standards compliance)
    Number of Output Channels11Met (Same)
    Regulated Current or VoltageBothBothMet (Same)
    Software/Firmware/MicroprocessorYesYesMet (Same)
    Automatic Overload TripNot required due to circuit designNot required due to circuit designMet (Same)
    Automatic No-Load TripYesYesMet (Same)
    Automatic Shut OffYes (20 minutes)Yes (5 minutes)Comparable (Difference in timing, but both include shut-off for safety/usage. Not a critical difference for equivalence)
    Patient Override ControlYesYesMet (Same)
    Indicator DisplayOn/Off, Low Battery, Voltage/Current LevelOn/Off, Low Battery, Voltage/Current LevelMet (Same)
    Compliance with Voluntary StandardsIEC 60601-1, IEC 60601-1-2IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11Met (Same, with additional relevant standards cited for the proposed device)
    Compliance with 21 CFR 898YesYesMet (Same)
    Weight9 oz. without charging base6 oz.Comparable (Different, but lighter weight for a 'mini' version is expected and not adverse to safety/effectiveness for intended use)
    Dimensions2.8" x 5.1" x 1.3"2.5" x 4.2" x 1.2"Comparable (Different, 'mini' implies smaller, not adverse)
    Housing MaterialsThermoplasticThermoplasticMet (Same)
    BiocompatibilityUser contacting materials tested and shown biocompatibleUser contacting materials identical to predicate, previously tested and shown biocompatibleMet (Same)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document does not describe a clinical "test set" in the context of efficacy or clinical performance. The comparison provided is a technical comparison to a predicate device. Performance testing was conducted for safety and EMC standards compliance, but details on sample size for these engineering tests are not provided and typically not part of the 510(k) summary for such devices. There is no information on data provenance related to clinical outcomes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No "ground truth" establishment by medical experts is described, as this is a technical equivalence review against a predicate device and relevant safety standards, not a clinical efficacy study requiring interpretation of outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No adjudication method is described as there is no clinical test set requiring expert consensus.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a transcutaneous electrical nerve stimulator for cosmetic use (facial and neck toning), not an AI-powered diagnostic or interpretive device. Therefore, MRMC studies and AI assistance comparisons are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a hardware product for direct user application, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for this submission is effectively the established safety and performance characteristics of the legally marketed predicate device (NuFACE Trinity) and compliance with recognized safety and EMC standards. The proposed device is demonstrating that its technical specifications and safety parameters are either identical or acceptably similar and safe compared to the predicate.

    8. The sample size for the training set

    Not applicable. This document does not describe a machine learning or AI algorithm that would require a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133823

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-17

    (121 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NuFACE® Mini Device is intended for facial stimulation and is indicated for overthe-counter cosmetic use. (21 CFR 801 Subpart C).

    Device Description

    The NuFACE® Mini Device is a non-invasive at home, over-the-counter facial toning device. The dual spheres of the NuFACE® Mini Device are designed for optimal contact with the face and are identical in size to the predicate device. The NuFACE® Mini Device is extremely responsive as it delivers soft wave micro-current in the millionths of an ampere and has the ability to increase facial contour and firm the skin and muscles.

    The NuFACE® Mini device measures 2.5" W x 4.2" L x 1.2" D. Its outer case is injection molded of thermoplastic resin. The device comes with an external Power Supply to charge the internal batteries of the device when not in use. The external Power Supply provides DC power from a pre-approved wall adapter. All charging circuitry is contained within the device itself. The microcurrent output is disabled when the device is connected to the external Power Supply.

    An ascending sequence of beeps informs the customer the NuFACE® Mini is ready for use. When the user turns the device OFF, a descending tone is emitted. A single control button is used to turn the device on and off and to change the device's output intensity. A long press of the control button toggles the device between on and off.

    To promote proper use, a single audio beep informs the user to relocate the device to treat a new location on the skin. The user can also adjust the output intensity by briefly pressing the control button to step between multiple micro-current output level selections. When the device is set to the maximum output and the control button is pressed, the device sequences to the minimum output.

    AI/ML Overview

    The NuFACE® Mini Device's acceptance criteria and the study proving it meets these criteria are described below.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the NuFACE® Mini Device are based on demonstrating substantial equivalence to its predicate device, the NuFACE® Trinity device (cleared as NuFACE® Plus, K103472). This means that the NuFACE® Mini Device must have identical or nearly identical technological characteristics and performance specifications for its intended use.

    Here's a comparison table summarizing the core electrical output specifications which form the primary acceptance criteria for demonstrating functional equivalence:

    Acceptance Criteria (Predicate Device Specification)NuFACE® Mini Device PerformanceComparison Result
    Waveform Type: Pulsed MonophasicPulsed MonophasicIdentical
    Waveform Shape: Modulated SquareModulated SquareIdentical
    Maximum Output Voltage:
    - 137 mV @ 500 Ω222 mV @ 500 ΩDifferent (Higher) - Note: The document states "identical to the predicate" but the values differ.
    - 769 mV @ 2 kΩ781 mV @ 2 kΩDifferent (Higher) - Note: The document states "identical to the predicate" but the values differ.
    - 3.82 V @ 10 kΩ3.90 V @ 10 kΩDifferent (Higher) - Note: The document states "identical to the predicate" but the values differ.
    Maximum Output Current:
    - 274 μA @ 500 Ω396 μA @ 500 ΩDifferent (Higher) - Note: The document states "identical to the predicate" but the values differ.
    - 387 μA @ 2 kΩ395 μA @ 2 kΩDifferent (Higher) - Note: The document states "identical to the predicate" but the values differ.
    - 383μA @ 10 kΩ391 μA @ 10 kΩDifferent (Higher) - Note: The document states "identical to the predicate" but the values differ.
    Output Tolerance: +/- 2%+/- 2%Identical
    Pulse Width: 119 ms (Total)ON phase: 60.4 ms, OFF phase: 60.4 ms, Total Pulse Width: 120.8 msDifferent (Slightly longer total pulse width) - Note: The document states "identical to the predicate" but the values differ.
    Frequency: 8.40 Hz8.28 HzDifferent (Slightly lower) - Note: The document states "identical to the predicate" but the values differ.
    Maximum Phase Charge: 23.6 µC @ 500 Ω23.7 µC @ 500 ΩDifferent (Slightly higher) - Note: The document states "identical to the predicate" but the values differ.
    Maximum Current Density: 0.512 mA/cm² @ 500 Ω0.514 mA/cm² @ 500 ΩDifferent (Slightly higher) - Note: The document states "identical to the predicate" but the values differ.
    Maximum Power Density: 1991 μW/cm² @ 10k Ω1981 μW/cm² @ 10k ΩDifferent (Slightly lower) - Note: The document states "identical to the predicate" but the values differ.
    Burst Mode:
    - Pulses per burst: 2020Identical
    - Pulses per second: 8.408.28Different (Slightly lower) - Note: The document states "identical to the predicate" but the values differ.
    - Burst duration (s): 2.402.42Different (Slightly longer) - Note: The document states "identical to the predicate" but the values differ.
    - Duty Cycle (on time per burst): 20.220Different (Slightly lower) - Note: The document states "identical to the predicate" but the values differ.
    ON Time: ConstantConstantIdentical

    Note on "Identical" Claims vs. Reported Values:
    The summary prominently states that numerous output characteristics are "identical to the predicate." However, the detailed "Output Specifications" table provided later in the document (Section 3) shows minor numerical differences for Maximum Output Voltage, Maximum Output Current, Pulse Width, Frequency, Maximum Phase Charge, Maximum Current Density, Maximum Power Density, Pulses per second, Burst duration, and Duty Cycle. Despite these small numerical variations in the detailed table, the overall conclusion of "substantial equivalence" implies that these differences were deemed insignificant for the device's safety and effectiveness relative to the predicate by the FDA. The submission likely argues that these slight numerical differences fall within acceptable engineering tolerances or do not alter the fundamental mechanism of action or safety profile.

    Additional acceptance criteria based on compliance standards are:

    • Compliance with FDA's performance Standards set forth in 21 CFR §898.
    • Compliance with IEC 60601-1-2:2010 for radiated and power line conducted emissions.
    • Compliance with IEC 60601-1:2005 (3rd edition) for Electrical Safety.

    The reported performance confirms compliance with these standards (Yes for 21 CFR 898, IEC 60601-1, IEC 60601-1-2).

    2. Sample Size Used for the Test Set and Data Provenance

    The study to prove the device meets acceptance criteria was based on non-clinical performance assessment. This type of assessment typically involves laboratory testing of the device's electrical and physical characteristics rather than testing on human subjects.

    • No specific "test set" in the sense of patient data is mentioned.
    • The provenance is laboratory testing of the device itself. Given the 510(k) submission is for a US market device, the testing would likely be performed in a controlled laboratory environment to meet US regulatory requirements.
    • This was a non-clinical evaluation, not a clinical study with a test set of human subjects.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable as the study was a non-clinical performance assessment focusing on device specifications and compliance with standards, not on clinical outcomes requiring expert consensus or ground truth from patient data.
    • The "ground truth" here is the established specifications and safety profile of the predicate device and relevant industry standards.

    4. Adjudication Method for the Test Set

    • Not applicable, as there was no test set of patient cases requiring adjudication. The assessment was based on direct measurement and comparison of device specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission relies on demonstrating substantial equivalence through comparison of technical specifications and safety standards, not through a clinical evaluation involving human readers or assessment of improved human performance with AI assistance. This device is an over-the-counter cosmetic device, not a diagnostic or therapeutic AI-assisted medical device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    • Not applicable. This is a standalone electrical stimulation device, not an AI algorithm. The device performance was assessed in a standalone capacity (i.e., its intrinsic electrical outputs) without any human-in-the-loop performance being a specific test metric.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance assessment was based on:

    • Predicate Device Specifications: The established electrical output and technological characteristics of the NuFACE® Trinity device (NuFACE® Plus), which was previously cleared.
    • Industry and Regulatory Standards: Compliance with relevant electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and FDA performance standards (21 CFR §898).

    8. The Sample Size for the Training Set

    • Not applicable. This device is an electrical stimulation device, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. There was no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1