510(k) Data Aggregation

    K Number
    K192469

    Validate with FDA (Live)

    Device Name
    Nox Sleep System
    Manufacturer
    Date Cleared
    2019-11-13

    (65 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nox Sleep System is used as an aid in the diagnosis of different sleep disorders and for the assessment of sleep.

    The Nox Sleep System is used to measure, record, display, organize, analyze, summarize, and retrieve physiological parameters during sleep and wake.

    The Nox Sleep System allows the user to decide on the complexity of the study by varying the number and types of physiological signals measured.

    The Nox Sleep System allows for generation of user/pre-defined reports based on subject's data.

    The user of the Nox Sleep System are medical professionals who have received training in the areas of hospital/clinical procedures, physiological monitoring of human subjects, or sleep disorder investigation.

    The intended environments are hospitals, institutions, sleep clinics, or other test environments, including patient's home.

    Device Description

    The Nox Sleep System is intended for patients undergoing physiological measurements, for the assessment of sleep quality and the screening for sleep disorders.

    The Nox Sleep System does not provide any alarms and is not intended to be used for continuous monitoring where failure to operate can cause injuries or death of the patient.

    The basic Nox Sleep System consists of two recording/acquisition devices (Nox A1 Recorder and Nox C1 Access Point), a software running on a PC (Noxturnal PSG), an Android application (Noxturnal App) running on mobile platform, along with sensors and accessories. The system supports full Polysomnography (PSG) studies both in ambulatory and online/attended setups but also more simple sleep study setups, recording only few channels. The ambulatory sleep studies may take place in the clinic or in the home environment, but the online/attended sleep studies are only conducted in the clinical environment.

    The Nox A 1 Recorder is a small battery-operated recording unit that is worn by the patient during the study. It records signals from patient applied sensors that connect to the unit but also supports recording of signals from auxiliary devices over Bluetooth. The Nox A1 Recorder allows for communication over Bluetooth with the Noxturnal App during ambulatory setup and with the Nox C1 Access Point during online setup. The recorder is intended to be worn over clothing.

    New accessories and sensors as part of this submission are the Nox A 1 EEG 5 Lead Gold Electrode Cable and Nox A1 EEG Head Cable that are used for recording of EEG/EOG. These components are in direct contact with the patient.

    The Nox C1 Access Point is a separate mains powered unit located remotely from the patient that allows for recording of signals from auxiliary devices. It supports communication over LAN/Ethernet to the Noxturnal PSG, and communication with the Nox A1 Recorder and Noxturnal App over Bluetooth. The Nox C1 Access Point is only used for online study setup and is thus not intended to be used in the home environment.

    The Noxturnal App is used as a mobile interface to the Nox A1 Recorder and Nox C1 Access Point. The communications are via Bluetooth link. The app is normally used in the beginning of a sleep study, for basic tasks such as device configuration, starting a recording, checking the signal quality of signals being recorded and marking events during bio calibration.

    The Noxturnal PSG is used for configuration of the Nox recording/acquisition devices, to download a study from ambulatory recording or to collect an online study. The software supports the viewing, retrieving, storing and processing of data recorded/collected, manual and automatic analysis and reporting on the results from the recorded studies. The purpose with the automatic scoring function in Noxturnal PSG is to assist the trained physician in the diagnosis of a patient. It is not intended to provide the trained physician with a diagnostic results. The type of automatic analysis events scored by Noxturnal PSG include: Sleep Stages (Wake, N1, N2, N3, REM), Apneas, Hypopneas, Apnea Cassification (Obstructive, Mixed and Central Apneas), Limb Movements, Periodic Limb Movements, SpO2 Desaturation Events. and potential Bruxism-Related Events.

    The result of the automatic analysis/scoring must always be manually verified by the trained physician prior to diagnosis.

    AI/ML Overview

    The Nox Sleep System is designed to aid in the diagnosis of sleep disorders and assess sleep quality by measuring, recording, displaying, organizing, analyzing, summarizing, and retrieving physiological parameters during sleep and wake. The system includes automatic scoring functionalities for various sleep events, which are intended to assist trained medical professionals in diagnosis.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document presents separate sections for the performance of different automatic scoring algorithms rather than a single consolidated table. However, the information can be extracted and presented as follows:

    Automatic Scoring AlgorithmAcceptance Criteria (Safety Endpoint/Justification)Reported Device Performance
    Bruxism AnalysisDetect at least 90% of oromandibular movements considered by a human expert to be bruxism-related events with 95% confidence (Sensitivity).Sensitivity: 95.7% (95% CI 93.2% - 97.4%)Specificity: 61.0% (95% CI 58.9% - 63.0%)PPV: 34.6% (95% CI 32.0% - 37.3%)NPV: 98.5% (95% CI 97.7% - 99.1%)
    PLM AnalysisInterclass correlation (ICC) of 0.61 or greater and a bias unlikely to impact a diagnosis for the Periodic Limb Movement Index.ICC for Periodic Limb Movement Index: 0.87
    Respiratory Flow Analysis (AHI)Not classifying patients with an AHI below 5 as having an AHI greater than or equal to 15 (95% confidence), AND Not classifying patients with an AHI greater than or equal to 15 as having an AHI below 5 (95% confidence). Also, Cohen's kappa reported.Cohen's Kappa for AHI (Cannula): 0.78Cohen's Kappa for AHI (RIP flow): 0.62 (95% CI 0.59-0.66)Cohen's Kappa for AHI (cRIP flow): 0.62 (95% CI 0.59-0.66)
    Respiratory Flow Analysis (ODI)(Implicitly similar to AHI, with Cohen's kappa reported)Cohen's Kappa for ODI: 0.87
    Apnea ClassificationICC comparable to what has been reported in scientific literature for Central Apnea Index (0.46).ICC for Central Apnea Index: 0.91Cohen's Kappa for Central Apnea Index: 0.89
    Sleep Staging AnalysisAverage accuracy of at least 60% when scoring wake epochs to ensure total sleep time measurement with 10% error or less (assuming 80% sleep efficiency).Cohen's Kappa: <=0.67Accuracy for Wake: 66.7%Accuracy for N1: 9.8%Accuracy for N2: 87.0%Accuracy for N3: 83.0%Accuracy for REM: 82.5%

    2. Sample size used for the test set and the data provenance

    The document states that the clinical performance testing for the automatic scoring algorithms involved "retrospectively analyzing pre-existing clinical data from sleep studies that had already been collected and manually scored as part of routine clinical care."

    • Bruxism Analysis: "adult Polygraphy recordings, including masseter EMG signal, recorded from multiple U.S. dental clinics."
    • PLM Analysis: "adult Polysomnography recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
    • Respiratory Flow Analysis (AHI & ODI): "adult Polysomnography recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
    • Apnea Classification: "adult Polygraphy recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."
    • Sleep Staging Analysis: "adult recordings recorded at a national hospital serving a general population who seeks medical attention at a sleep clinic."

    The exact sample size (number of patients/studies) for each test set is not explicitly provided in the given text. The provenance is retrospective and data was collected from U.S. dental clinics (for bruxism) and a national hospital (for other sleep studies).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for the test set was established by "qualified polysomnographic technologists" who "followed the American Academy of Sleep Medicine (AASM) scoring guidelines."

    The number of experts is not specified. Their qualification is: "qualified polysomnographic technologists." No details on years of experience are provided, but adherence to AASM guidelines suggests a standardized and recognized level of expertise.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document states, "All scorers were qualified polysomnographic technologists and followed the American Academy of Sleep Medicine (AASM) scoring guidelines." It then describes comparison of the automatic scoring results to "the results of the manually scored data."

    This implies that the manual scoring by the polysomnographic technologists served as the ground truth. There is no explicit mention of an adjudication method (like 2+1 or 3+1) among multiple human scorers for establishing the ground truth. It seems to rely on single-expert scoring as the reference standard.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was described where human readers' performance with and without AI assistance was evaluated. The study focused on the standalone performance of the automatic scoring algorithms against human-scored ground truth. The device is intended as an aid for professionals, requiring manual verification.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, standalone performance was done. The clinical performance testing directly compared the output of the automatic scoring algorithms to the manually scored ground truth data. The results (sensitivity, specificity, PPV, NPV, ICC, Cohen's Kappa, and accuracy) reported are measures of the algorithm's performance on its own.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used was expert consensus (manual scoring by qualified polysomnographic technologists) following the American Academy of Sleep Medicine (AASM) scoring guidelines. This is a form of expert consensus, even if the "consensus" here refers to the adherence to established scoring rules rather than explicit multi-reader agreement on individual cases.

    8. The sample size for the training set

    The document details the "clinical performance testing," which involved "retrospectively analyzing pre-existing clinical data." This section describes the test set used for validation.

    The sample size for the training set is not provided or mentioned in this document. Information about the training data used to develop the automatic scoring algorithms is absent.

    9. How the ground truth for the training set was established

    Since the document does not discuss the training set, there is no information provided on how its ground truth was established.