510(k) Data Aggregation

    K Number
    K251567

    Validate with FDA (Live)

    Device Name
    NovoSorb BTM
    Date Cleared
    2025-06-20

    (29 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NovoSorb BTM is indicated for use in the management of wounds including: partial and full thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic and vascular ulcers, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) and draining wounds.

    Device Description

    The NovoSorb BTM device has been modified to expand the available thickness of the foam component from 2 mm to 2-6 mm. Sizes are limited to a maximum volume of 160 cm³. The device continues to be a porous, biodegradable, polyurethane foam, adhered to a fenestrated transparent sealing membrane. The sealing membrane is designed to physiologically close the wound limiting evaporative moisture loss during integration of the foam. The sealing membrane is to be removed and discarded when appropriate, leaving only the foam layer to biodegrade in patients.

    NovoSorb BTM is a terminally sterilized, single use device intended for deep partial and full thickness wounds. Each NovoSorb BTM is packed in a transparent polymer pouch, enclosed in an aluminized pouch. A product label is placed on the pouch. The aluminized pouch together with an instruction for use is individually packaged in a cardboard envelope and a product label is placed on the envelope.

    AI/ML Overview

    It appears there might be a misunderstanding of the provided FDA 510(k) clearance letter. The document is for a medical device called "NovoSorb BTM," which is a wound dressing. This device is a physical medical device (a porous, biodegradable, polyurethane foam), not an AI/software-based medical device.

    Therefore, the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance) is not applicable to this submission.

    The 510(k) summary clearly states:

    • "No clinical tests were performed; determination of substantial equivalence was not based on an evaluation of clinical performance data."
    • "PolyNovo believes the modified NovoSorb BTM device is substantially equivalent to the legally marketed predicate devices NovoSorb BTM (K172140) and NovoSorb MTX (K242149), because no differences in technological characteristics exist between different foam thicknesses as there is no change to the foam density for different foam thicknesses."

    The "acceptance criteria" for this device, as described in the non-clinical tests section, relate to physical and mechanical properties of the wound dressing, not diagnostic performance based on AI.

    Below is a brief summary of how the device met its acceptance criteria, based only on the provided document, without attempting to fit it into an AI study framework:


    Acceptance Criteria and Study for NovoSorb BTM (Physical Medical Device)

    The NovoSorb BTM is a physical wound dressing device, not an AI or software-based diagnostic tool. Therefore, the acceptance criteria and supporting studies are based on the physical, mechanical, and biological properties of the material and its substantial equivalence to previously cleared predicate devices, rather than AI performance metrics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicitly Met by Substantial Equivalence Claim)Reported Device Performance (Summary from Document)
    Physical CharacteristicsConsistency in material properties (e.g., polyurethane composition, porosity, density, foam structure)."modified to expand the available thickness of the foam component... Device continues to be a porous, biodegradable, polyurethane foam... properties of the foam component, including average pore size, porosity, and density, remain unchanged compared to both predicate devices."
    "Sealing membrane adhered to the foam; constructed from the same synthetic biodegradable polyurethane material.""Same as Predicate Device"
    Mechanical StrengthAbility to withstand typical handling and application, maintaining structural integrity (e.g., tensile strength, tear resistance)."Verification performance testing demonstrates that the proposed NovoSorb BTM will consistently meet established functional and performance requirements. These requirements include physical characteristics, mechanical strength, and durability."
    DurabilityResistance to degradation beyond intended bioabsorption timeframe, maintenance of integrity during use."Verification performance testing demonstrates that the proposed NovoSorb BTM will consistently meet established functional and performance requirements... include physical characteristics, mechanical strength, and durability."
    BiocompatibilityNon-toxic and safe for contact with human tissue.(Implied by substantial equivalence to predicate devices using same materials and manufacturing processes)
    SterilityAchievement and maintenance of desired sterility assurance level."sterilized using gamma radiation at a dose of 25–40 kGy, achieving a sterility assurance level (SAL) of 10⁻⁶".
    Packaging & Shelf-lifeMaintenance of sterility and device integrity throughout shelf-life."packed in a transparent polymer pouch, enclosed in an aluminized pouch... consistent with both predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The verification testing would involve a sufficient number of device samples to demonstrate consistency in manufacturing and material properties.
    • Data Provenance: The data are from non-clinical bench testing. No geographical or temporal provenance is specified, as the tests are against material and mechanical standards for the device itself rather than patient data. The tests are prospective in the sense that they are performed on newly manufactured devices to demonstrate compliance.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • This concept is not applicable. "Ground truth" for this device relates to established engineering and material science standards for physical device properties (e.g., material composition confirmed by analytical chemistry, mechanical properties confirmed by engineering tests). It does not involve expert consensus on medical image interpretation or clinical outcomes.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods are typically used in clinical studies or AI evaluations to resolve discrepancies among human readers or algorithm outputs. For physical device testing, the results are objective measurements against predefined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (often with AI assistance). The submission explicitly states: "No clinical tests were performed; determination of substantial equivalence was not based on an evaluation of clinical performance data."

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Not applicable. This device is not an algorithm or software. It is a physical wound dressing.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's acceptance is based on established engineering specifications, material science standards, and performance characteristics of the predicate devices. This includes:
      • Chemical composition (e.g., polyurethane).
      • Physical properties (e.g., density, porosity, pore size).
      • Mechanical properties (e.g., strength, durability).
      • Sterility assurance level (SAL).
      • Packaging integrity.

    8. The Sample Size for the Training Set

    • Not applicable. This device is not an AI model, and therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no AI model or training set, the concept of establishing ground truth for it does not apply.

    In summary, the FDA 510(k) for NovoSorb BTM is a clearance based on substantial equivalence to existing predicate devices, supported by non-clinical bench testing that confirms the modified device maintains the same physical, material, and performance characteristics. The detailed questions about AI study methodologies are not relevant to this specific device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172140

    Validate with FDA (Live)

    Date Cleared
    2017-08-11

    (25 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovoSorb™ BTM Wound Dressing is indicated for use in the management of wounds including: partial and full thickness wounds, pressure ulcers, diabetic ulcers, chronic and vascular ulcers, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, second-degree burns, and skin tears) and draining wounds. The device is for single use only.

    Device Description

    The NovoSorb™ BTM Wound Dressing device is a 2mm thick, porous, white biodegradable polyurethane foam bonded with a polyurethane adhesive layer to a fenestrated one-sided transparent sealing membrane. The sealing membrane is designed to physiologically close the wound limiting evaporative water loss during integration of the foam. The adhesive layer and sealing membrane are to be removed and discarded when appropriate leaving only the foam layer to biodegrade in patients.

    NovoSorb™ BTM Wound Dressing device is supplied in various sizes, ranging from 4cm² to 800cm². The NovoSorb™ BTM Wound Dressing is a single use, terminally sterilized device, individually packed in a transparent polymer pouch enclosed in a white aluminized pouch contained in a cardboard envelope.

    AI/ML Overview

    This FDA 510(k) letter ([K172140](https://510k.innolitics.com/device/K172140)) is a notification of an administrative change in product code for the NovoSorb™ BTM Wound Dressing and explicitly states that the submission was not re-reviewed (referencing document {0} and {1}). The original substantial equivalence determination was made on August 11, 2017.

    The 510(k) summary (documents {4} and {5}) states that no clinical tests were performed, and the determination of substantial equivalence was not based on an evaluation of clinical performance data. Therefore, the provided documents do not contain information related to acceptance criteria, device performance, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies that would be typically associated with clinical studies for AI/ML devices.

    The NovoSorb™ BTM Wound Dressing is described as a medical device for wound management, not an AI/ML device. The provided text refers to it as a "2mm thick, porous, white biodegradable polyurethane foam bonded with a polyurethane adhesive layer to a fenestrated one-sided transparent sealing membrane." (document {5}). The core of the 510(k) submission was a comparison to a predicate device (BTM Wound Dressing, [K142879](https://510k.innolitics.com/device/K142879)) based on technological characteristics and mechanical testing.

    Thus, none of the requested information regarding acceptance criteria, study details, sample sizes, expert ground truth, or multi-reader studies can be extracted from the provided text for an AI/ML device, as this is a traditional medical device submission.