510(k) Data Aggregation

    K Number
    K212840

    Validate with FDA (Live)

    Date Cleared
    2021-12-11

    (95 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    The proposed devices are powder-free nitrile examination gloves, provided as non-sterile and disposable devices. The proposed devices are mainly made from nitrile and there are four sizes, includes small (S), medium (M), large (L), X-large (XL) for optional. The gloves are provided with blue color, the colorant is NBR BLUE 7214 (which composed of Phthalocyanine blue (CAS No.147-14-8) and Propylene glycol (CAS No.57-55-6). The examination glove is smooth surface with textured fingertips and a rolled rim atthe cuff edge.

    The examination gloves meet the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for "Nitrile Gloves (Model: NE01)". It does not describe an AI medical device. Therefore, many of the requested fields are not applicable.

    Here's the information extracted from the provided text, focusing on the device's acceptance criteria and the studies performed:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodTest PurposeAcceptance CriteriaReported Device PerformanceConclusion
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions TestTo determine the width, length, and thickness of the glovesWidth: For M: 95 ± 10mm For L: 110 ± 10mm For XL: 120 ± 10mm Length: For M: ≥230mm For L: ≥230mm For XL: ≥230mm Thickness: Finger: ≥0.05mm Palm: ≥0.05mmWidth: For M: 9396mm For L: 104-107mm For XL: 112116mm Length: For M: 236239mm For L: 236239mm For XL: 255258mm Thickness: For M: Finger: 0.060.09mm, Palm: 0.060.08mm For L: Finger: 0.060.08mm, Palm: 0.060.08mm For XL: Finger: 0.080.10mm, Palm: 0.06~0.09mmPassed
    ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Properties TestTo determine the tensile strength and ultimate elongation before and after acceleration agingBefore Aging: Tensile Strength: ≥14Mpa, Ultimate Elongation: ≥500% After Aging: Tensile Strength: ≥14Mpa, Ultimate Elongation: ≥400%Before Aging: Tensile Strength: ≥14Mpa, Ultimate Elongation: ≥500% After Aging: Tensile Strength: ≥14Mpa, Ultimate Elongation: ≥400%Passed
    ASTM D6319-19 (ASTM D5151-11) Standard Test Method for Detection of Holes in Medical GlovesTo determine the holes in the glovesAQL 2.5Pass at AQL 2.5Passed
    ASMT D6319-19 (ASTM D6124-11) Standard Test Method for Residual Powder on Medical GlovesTo determine the residual powder in the gloves< 2.0 mg/gloveFor M: Pass at 0.46 mg/glove For L: Pass at 0.52 mg/glove For XL: Pass at 0.14 mg/glovePassed
    ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin irritation and skin sensitizationTo evaluate the potential intracutaneous reactivity caused by intracutaneously inject the extract to rabbitsUnder the conditions of study not an irritationUnder the conditions of the study not a sensitizationPassed
    ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin irritation and skin sensitizationTo determine the skin sensitization potential in guinea pigs.Under the conditions of the study not a sensitizationUnder the conditions of study not an irritationPassed
    ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for acute systemic toxicityThe test item was evaluated for acute systemic toxicity in ICR mouseUnder the conditions of the study no systemic toxicityUnder the conditions of the study no systemic toxicityPassed

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for dimensions, tensile strength, or holes). However, the tests reference ASTM and ISO standards, which typically specify sample sizes for their methodologies. The data provenance is implied to be from laboratory testing performed on the "Nitrile Gloves (Model: NE01)" manufactured by BYD Auto Industry Company Limited, based on the context of a 510(k) submission. It is a retrospective study since the tests were conducted prior to submission. The country of origin of the data is not explicitly stated beyond the manufacturing location in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the device is a medical glove, not an AI-powered diagnostic device. The "ground truth" for these tests is based on objective measurements and established standards (ASTM, ISO), not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable. The tests involve objective measurements against predefined criteria, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a non-AI medical device (nitrile gloves).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a non-AI medical device (nitrile gloves).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance tests on the nitrile gloves is based on:

    • Physical measurements: Dimensions (width, length, thickness).
    • Mechanical properties: Tensile strength and ultimate elongation before and after aging.
    • Defect detection: Freedom from holes (water tight test).
    • Chemical properties: Residual powder.
    • Biocompatibility results: Skin sensitization, skin irritation, and acute systemic toxicity results (determined through biological assays).

    These are objective, quantifiable standards as defined by ASTM D6319-19 and ISO 10993 series.

    8. The sample size for the training set:

    Not applicable. This is a non-AI medical device; there is no "training set."

    9. How the ground truth for the training set was established:

    Not applicable. This is a non-AI medical device; there is no "training set" or ground truth associated with it in the context of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211252

    Validate with FDA (Live)

    Device Name
    Nitrile Glove
    Date Cleared
    2021-06-25

    (60 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a Nitrile Glove, which is a Class I medical device. Therefore, the "device" in this context is a physical product (a medical glove), not a software or AI-driven diagnostic tool. The questions you've asked are highly relevant to the evaluation of AI/ML-based medical devices, specifically regarding their performance on test sets, ground truth establishment, and clinical studies.

    Since the document describes a physical medical device (nitrile glove) and not an AI or software device, the concepts of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context relate to physical properties, performance standards, and biocompatibility rather than diagnostic accuracy, reader studies, or training sets for AI models.

    Therefore, I cannot extract the information requested in your prompt because it pertains to a different type of medical device (AI/ML) than what is described in the provided text.

    However, I can extract the acceptance criteria and performance data relevant to this specific device (Nitrile Glove) as described in the document.

    Acceptance Criteria and Device Performance for a Nitrile Glove

    The document details performance testing for the Nitrile Glove based on established ASTM and ISO standards for examination gloves.

    1. A table of acceptance criteria and the reported device performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength (mm):- S: ≥220- M/L/XL: ≥230Width (mm):- S: 80±10- M: 95±10- L: 110±10- XL: 120±10Thickness (mm):- Finger: ≥0.05- Palm: ≥0.05Length:- S: >220/Pass- M/L/XL: >230/PassWidth:- S: 81-85 /Pass- M: 91-96/ Pass- L: 102-106/ Pass- XL: 114-116/PassThickness:- Finger: 0.06-0.08/Pass- Palm: 0.07-0.08/Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaks / Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.02mg/Pass
    ASTM D412Physical Properties (Before Aging)Tensile Strength: ≥14MPaUltimate Elongation: ≥500%Tensile Strength: 15-18/PassUltimate Elongation: 530-560/Pass
    ASTM D412Physical Properties (After Aging)Tensile Strength: ≥14MPaUltimate Elongation: ≥400%Tensile Strength: 15-18/PassUltimate Elongation: 530-555/Pass
    ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells./Pass
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant/Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer./Pass

    Since the requested questions (2-9) are specific to AI/ML device evaluations and are not applicable to the physical medical device (Nitrile Glove) described in the provided FDA submission, I cannot provide answers for those. The document explicitly states: "Clinical testing is not needed for this device." and discusses non-clinical and performance testing related to material integrity, physical dimensions, and biological safety for a physical product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210145

    Validate with FDA (Live)

    Date Cleared
    2021-04-21

    (90 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Glove Powder Free Blue is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is powder free, non-sterile nitrile examination gloves. Following FDA Code LZA, a "nitrile (or polymer) patient examination glove is a disposable device made of nitrile rubber or synthetic polymers that may or may not bear a trace amount of residual powder, and is intended to be worn on the hand for medical purposes to provide a barrier against potentially infectious materials and other contaminants."

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device: Nitrile Glove Powder Free Blue. It seeks to demonstrate substantial equivalence to a predicate device.

    The provided text describes the regulatory process, device information, and a comparison of the proposed device with a predicate device, focusing on non-clinical performance testing. It does not contain information about an AI/ML device, nor does it describe a study involving human readers or comparative effectiveness for an AI model.

    Therefore, I cannot provide the detailed information requested regarding acceptance criteria and a study proving an AI/ML device meets those criteria, specifically concerning:

    • A table of acceptance criteria and reported device performance for an AI/ML device.
    • Sample sizes, data provenance, and ground truth establishment for AI/ML test and training sets.
    • Details on expert involvement (number, qualifications, adjudication) for an AI/ML ground truth.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies for AI assistance.
    • Stand-alone AI algorithm performance.

    The document is entirely focused on the physical and chemical properties and safety of a medical glove, not an AI/ML system.

    However, I can extract the acceptance criteria and performance data for the Nitrile Glove Powder Free Blue from the provided document, which pertains to its non-clinical testing.

    Here's a breakdown of the acceptance criteria and results for the medical glove as presented in the document:

    1. A table of acceptance criteria and the reported device performance (for the physical glove)

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    ISO 10993-5 Biological evaluation of medical devices Test for in vitro cytotoxicityTo determine the potential cytotoxicity.Non-cytotoxicPASS
    ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for Systemic ToxicityTo determine the potential systemic toxicity.Does not cause systemic toxicityPASS
    ISO 10993-10 Biological evaluation on medical device Part 10: Test for Irritation and Skin IrritationTo determine the potential for irritation and skin irritation.Non-irritant, and Non skin irritantPASS
    ISO 10993-10 Biological evaluation on medical device Part 10: Test Skin SensitizationTo determine the potential skin sensitization.Does not cause skin sensitizationPASS
    ASTM D6319, Standard Specification for Nitrile Examination Gloves for Medical ApplicationTo test for: (1) freedom from holes (2) physical dimensions (3) Aging(1) Shall not leak (2) For size M (mm): Width: 110 +/- 10, Length: ≥ 230, Finger and palm thickness: median value ≥ 0.05 (3) After Aging: Tensile Strength: ≥ 14 MPa, Ultimate elongation: ≥ 400%PASS
    EN 455-1, EN 455-2, Medical Gloves For Single UseTo test for: (1) freedom from holes (2) dimensions and strength(1) Shall not leak (2) For size M (mm): Width: 110, Length: ≥ 240, Strength: Force at Break: ≥ 6N, Force at break after challenge testing ≥ 6N at 7 daysPASS

    2. Sample size used for the test set and the data provenance:

    The document does not specify the exact sample sizes (number of gloves) used for each non-clinical test. It only states the tests were conducted. The data provenance is implied to be from the manufacturer's internal testing or accredited labs in China (Changzhou Universal Medical Equipment Co. Ltd. is located in Changzhou, Jiangsu Province, China). The tests are retrospective in the sense that they were performed on finished products.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable to this type of non-clinical testing for a medical glove. The "ground truth" for these tests is established by standardized laboratory methods (e.g., measuring dimensions, performing leak tests, chemical assays for cytotoxicity) according to the specified ISO, ASTM, and EN standards. These are objective measures, not requiring expert consensus or interpretation in the same way an AI model's output would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As noted above, these are objective physical and biological tests, not requiring adjudication of human performance or AI output.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This document is for a medical glove, not an AI/ML device. No human reader studies were conducted or relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI algorithm. The performance described is the standalone performance of the physical glove.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the glove's performance is based on the results of standardized laboratory engineering and biological tests as defined by ISO, ASTM, and EN standards (e.g., presence/absence of leaks, specific measurements (length, width, thickness), tensile strength, elongation, cytotoxicity, irritation, sensitization).

    8. The sample size for the training set:

    Not applicable. There is no training set for a physically manufactured medical glove in the context of an AI/ML model.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183068

    Validate with FDA (Live)

    Date Cleared
    2018-12-21

    (46 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Glove Powder Free White is a disposable device intended purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is white. The proposed device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) summary for the "Nitrile Glove Powder Free White." This type of document is a premarket notification to the FDA to demonstrate that a device is substantially equivalent to a predicate device already on the market. It primarily focuses on comparing the new device's characteristics and performance to the predicate device and relevant standards, rather than describing a traditional clinical study with human subjects, AI involvement, or ground truth establishment as might be found for more complex diagnostic devices.

    Therefore, many of the requested categories are not directly applicable or are not detailed in this document.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are established by various ASTM and ISO standards, and the reported device performance indicates compliance with these standards.

    Acceptance Criteria (Standard/Requirement)Reported Device Performance (Nitrile Glove Powder Free White)
    Physical Properties (Before Aging):
    Tensile Strength (ASTM D6319)14MPa, min
    Ultimate Elongation (ASTM D6319)500% min
    Physical Properties (After Aging):
    Tensile Strength (ASTM D6319)14MPa, min
    Ultimate Elongation (ASTM D6319)400% min
    Compliance with ASTM D6319Complies
    Freedom from Holes (ASTM D5151)Be free from holes when tested in accordance with ASTMD5151 AQL=2.5
    Powder Content (ASTM D6124)0.47 mg per glove (meets requirements)
    Biocompatibility:
    Irritation (ISO 10993-10)Under the conditions of the study, not an irritant
    Sensitization (ISO 10993-10)Under conditions of the study, not a sensitizer. Complies with ISO10993-10
    Cytotoxicity (ISO 10993-5)Under conditions of the study, device extract is not cytotoxic

    2. Sample sized used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each individual test (e.g., number of gloves tested for tensile strength or holes). It refers to compliance with standards like ASTM D5151 and D6319, which would typically define such sample sizes. The data provenance is implied to be from laboratory testing performed by "Haining Medical Products Co., Ltd." or their designated testing facilities in China (as per the submitter's address). The tests are non-clinical tests, not studies with human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of device and submission. The "ground truth" for medical gloves is established by objective physical and chemical testing against recognized international and national standards (e.g., ASTM, ISO) for properties like tensile strength, elongation, freedom from holes, powder content, and biocompatibility. There are no human "experts" establishing a "ground truth" in the clinical sense for this product's performance.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic devices where subjective interpretation of images or other data by multiple readers is involved. For medical glove testing, objective measurements against pre-defined thresholds from standards are used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This document concerns a physical medical device (nitrile gloves), not an AI-powered diagnostic tool. No MRMC study or AI assistance is mentioned or relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. This document concerns a physical medical device (nitrile gloves), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is established by objective measurements and laboratory tests against the specific requirements and methodologies outlined in the referenced standards (ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993-10, ISO 10993-5). For example, "freedom from holes" is determined by a specific test method in ASTM D5151, not by expert consensus or pathology. Biocompatibility is assessed through standardized in-vitro and in-vivo tests as per ISO 10993.

    8. The sample size for the training set

    This is not applicable. There is no "training set" in the context of this device. Training sets are relevant for machine learning algorithms.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set mentioned or relevant to this device's submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182528

    Validate with FDA (Live)

    Date Cleared
    2018-11-15

    (62 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nitrile Glove Powder Free Black is a disposable device intended purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is black. The proposed device is non-sterile.

    AI/ML Overview

    This document is a 510(k) summary for a Nitrile Glove Powder Free Black, classifying it as a Class I medical device. The submission aims to demonstrate substantial equivalence to a legally marketed predicate device (K171422).

    Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria & Reported Device Performance:

    The document presents a "Technological Characteristic Comparison Table" (Table 1), "Device Dimensions Comparison" (Table 2), "Performance Comparison" (Table 3), and "Safety Comparison" (Table 4) where the proposed device's performance is compared against the predicate device and relevant standards. This serves as the primary method to demonstrate that the proposed device meets the acceptance criteria (substantial equivalence to the predicate and compliance with recognized standards).

    Acceptance Criteria TypeAcceptance CriteriaProposed Device Performance (K182528)Predicate Device Performance (K171422)Comparison/Outcome
    General Characteristics
    Product CodeLZALZALZASame
    Regulation No.21 CFR 880.625021 CFR 880.625021 CFR 880.6250Same
    ClassIIISame
    Intended UseTo prevent contamination between patient and examiner.Same as predicate (quoted directly)Same as proposed (quoted directly)Same
    Powdered StatusPowdered freePowdered freePowdered freeSame
    Design FeatureAmbidextrousAmbidextrousAmbidextrousSame
    Labeling InformationSingle-use indication, powder free, device color, device name, glove size and quantity, Non-SterileNitrile Glove Powder Free Black, Non-SterileDisposable Powder Free Nitrile Examination Glove, Non-SterileSame (description-wise)
    DimensionsAccording to the predicate device's specifications
    Length, mm (all sizes)230 min230 min230 minSame
    Width, mm (XS, S, M, L, XL)75, 85, 95, 105, 115 ±575, 85, 95, 105, 115 ±575, 85, 95, 105, 115 ±5Same
    Thickness, mm (Finger, Palm)0.05 min0.05 min0.05 minSame
    Physical Properties (Before Aging)
    Tensile Strength14MPa, min14MPa, min14MPa, minSame
    Ultimate Elongation500% min500% min500% minSame
    Physical Properties (After Aging)
    Tensile Strength14MPa, min14MPa, min14MPa, minSame
    Ultimate Elongation400% min400% min400% minSame
    Compliance with StandardComply with ASTM D6319Comply with ASTM D6319Comply with ASTM D6319Same
    Freedom from HolesBe free from holes when tested in accordance with ASTMD5151, AQL=2.5Be free from holes when tested in accordance with ASTMD5151, AQL=2.5Be free from holes when tested in accordance with ASTMD5151, AQL=2.5Same
    Powder ContentMeet the requirements of ASTM D61240.48 mg per gloveMeet the requirements of ASTM D6124Similar (Specific value versus general compliance statement for predicate; assumed to meet the same standard)
    Material Safety
    MaterialNitrileNitrileNitrileSame
    Biocompatibility: Irritation (ISO 10993-10:2010)Not an irritantUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritantSame
    Biocompatibility: Sensitization (ISO 10993-10:2010)Not a sensitizerUnder conditions of the study, not a sensitizer.Under conditions of the study, not a sensitizer.Same
    Biocompatibility: Cytotoxicity (ISO 10993-5:2009)No potential toxicity to L-929 cellsUnder conditions of the study, did not show potential toxicity to L-929 cells.N/A (indicated by '/')Similar (predicate doesn't explicitly state cytotoxicity but proposed device meets standard)

    2. Sample Sizes and Data Provenance (for test set, if applicable):

    The document states, "Non-clinical tests were conducted to verify that the proposed device met all design specifications." However, it does not specify the sample sizes used for these non-clinical tests (e.g., how many gloves were tested for Freedom from Holes, Tensile Strength, etc.).

    The data provenance is not explicitly stated as retrospective or prospective, nor does it mention a country of origin for the data, other than the submitting company being located in China. Given the nature of these tests (bench testing for physical and chemical properties), they would typically be prospective tests performed on samples of the manufactured device.

    3. Number of Experts and Qualifications (for ground truth of test set):

    This information is not applicable to this submission type. Ground truth in this context typically refers to clinical diagnosis or pathology readouts. For a Class I medical device like a glove, the "ground truth" is adherence to physical/chemical standards and biocompatibility tests, not expert clinical interpretation. The "experts" would be the personnel in the testing laboratories conducting the ASTM and ISO standard tests. Their qualifications are inherent in their ability to perform these accredited tests.

    4. Adjudication Method (for test set):

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies for diagnostic accuracy, where multiple human readers interpret data, and a consensus mechanism is employed to establish ground truth or resolve discrepancies. For physical and biocompatibility testing of a medical glove, objective laboratory test results are obtained, and no such adjudication is necessary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC study was not done and is not relevant for this type of device. MRMC studies are typically performed for diagnostic imaging devices to assess the impact of AI on reader performance.

    6. Standalone (Algorithm Only) Performance Study:

    A standalone performance study was not done and is not applicable. This is a physical product (a glove), not an algorithm or AI-powered device.

    7. Type of Ground Truth Used:

    The "ground truth" for the device's performance is established through:

    • Compliance with recognized international and national standards: ISO 10993-10, ISO 10993-5, ASTM D6124-06, ASTM D5151-06, and ASTM D6319-10. These standards define the acceptable performance characteristics for medical gloves.
    • Comparison to a legally marketed predicate device (K171422): Demonstrating that the proposed device's characteristics (dimensions, physical properties, material, biocompatibility) are substantially equivalent to a device already cleared by the FDA.

    8. Sample Size for the Training Set:

    This information is not applicable. This device is a physical product, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970135

    Validate with FDA (Live)

    Device Name
    NITRILE GLOVES
    Date Cleared
    1997-08-25

    (222 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A medical glove worn on the hand of a healthcare worker and similar personnel to prevent contamination between healthcare personnel and patient. In addition, this glove is worn for the protection of the wearer against exposure to chemotherapeutic agents.

    Device Description

    8 mil thick, Non Sterile, fitted or ambidextrous, Nitrile gloves with beaded cuff extending beyond the wrist.

    AI/ML Overview

    This document does not contain information about an AI/ML device or a study comparing AI with human readers. The document is a 510(k) summary for Sage Nitrile Gloves, a medical device (examination glove), submitted to the FDA in 1997. It describes the device, its intended use, a substantial equivalence comparison to other gloves, and lists performance tests conducted according to various ASTM and FDA standards for physical properties and biocompatibility.

    Therefore, I cannot provide the requested information regarding acceptance criteria and studies for an AI/ML device, comparative effectiveness, standalone performance, ground truth, or training set details as this information is not present in the provided text.