Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151332

    Validate with FDA (Live)

    Date Cleared
    2015-08-17

    (90 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nexus Universal/Nexus Universal Chroma is indicated for cementation of all indirect restorations including ceramic, resin, and metal-based inlays/onlays, crowns, bridges, posts, and veneers*. Additional indications include core build-ups and cementation of crowns to implants.

    • Adhesive application on the preparation is required for veneer cementation using Nexus Universal Chroma.
    Device Description

    Nexus Universal / Nexus Universal Chroma is indicated for cementation of all indirect restorations including ceramic, resin, and metal-based inlays/onlays, crowns, bridges, posts, and veneers*. Additional indications include core build-ups and cementation of crowns to implants.

    • Adhesive application on the preparation is required for veneer cementation using Nexus Universal and Nexus Universal Chroma.

    Nexus Universal and Nexus Universal Chroma are offered in paste/paste formulations for use as self-etch, self-adhesive resin cements or as bonded resin cements. These cements are considered permanent and not temporary.

    The dual-cure material is packaged in dual barrel syringes with single-use automix tips and optional curved dispensing tips to allow the user to deliver the desired volume of cement directly into the restoration and/or tooth preparation.

    Nexus Universal and Nexus Universal Chroma are offered in multiple shades. The difference between Nexus Universal and Nexus Universal Chroma is that Nexus Universal Chroma offers a gel-state color indicator, which visually displays the optimal time to remove the excess cement.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study to demonstrate substantial equivalence for the dental cement "Nexus Universal" and "Nexus Universal Chroma."

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document uses the predicate device (Maxcem 2, K073209) as a benchmark for acceptance criteria. The Nexus Universal/Chroma is considered substantially equivalent if its performance is comparable or improved.

    ElementAcceptance Criteria (Predicate: Maxcem 2)Reported Device Performance (Nexus Universal/Chroma)
    Gel/Set Time (fresh ambient/fresh oral)≥ 2'00" / ≤ 3'30"≥ 2'00" / ≤ 3'30"
    Bond Strength to Dentin20.35 MPa23.05 MPa (Improved)
    Bonding strength for Post Cementation59.68 lbs62.34 lbs (Improved)
    Radiopacity, % aluminum280%283% (Improved)
    Film Thickness (μm)22 μm24 μm
    Diametral Strength, Mpa (Dual-Cure mode)54.8 MPa58.1 MPa (Improved)
    Consistency3.00 cm3.39 cm
    Maximum Temperature During Self CuringN/A (Predicate did not report)26.89°C
    Water sorptionN/A (Predicate did not report)33.5 µg/mm³
    Color at set time (for Chroma version)N/A (Predicate did not have a color indicator)Pass (Visually displays optimal time to remove excess)
    BiocompatibilityMeets requirementsMeets requirements
    Shelf Life18 months based on real-time data18 months based on accelerated data

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes for each specific test mentioned (e.g., bond strength, film thickness). It globally refers to the data as "Non-Clinical Performance Data" and "internal performance testing."

    • Sample Size: Not explicitly stated for individual tests.
    • Data Provenance: The data is described as "internal performance testing." This implies the tests were conducted by the manufacturer, Sybron Dental Specialties. The country of origin is not specified but given the submitter information, it is likely the US. The data is retrospective as it refers to completed tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests are non-clinical (laboratory-based) and don't involve human expert assessment for ground truth in the traditional sense of medical image analysis or clinical diagnosis. The "ground truth" for these tests would be the measured physical and chemical properties of the material.

    4. Adjudication method for the test set

    This information is not applicable as the study is based on non-clinical laboratory tests, not subjective expert assessment requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study done. This device is a dental cement, and the evaluation is based on material properties and non-clinical performance, not on AI-assisted human reading of medical cases.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a dental cement, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the non-clinical performance tests would be the measured physical, chemical, and mechanical properties of the dental cement, determined through standardized laboratory testing methods (e.g., ISO standards, FDA guidance). For example, a bond strength measurement has a direct numerical ground truth.

    8. The sample size for the training set

    This information is not applicable. The device is a physical material (dental cement), not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1