510(k) Data Aggregation

    K Number
    K251006

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-11-10

    (223 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NeVasc Aspiration Catheter:
    As part of the NeVasc Aspiration System, the NeVasc Aspiration Catheter with the Vesalio Aspiration Tubing Set and a compatible aspiration pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    Vesalio Aspiration Tubing Set:
    As part of the NeVasc Aspiration System, the Vesalio Aspiration Tubing Set is indicated to connect the NeVasc Aspiration Catheter to a compatible aspiration pump.

    Device Description

    The NeVasc Aspiration System consists of 1) NeVasc Aspiration Catheter, 2) Peel Away Introducer Sheath and 3) Vesalio Aspiration Tubing Set.

    The NeVasc Aspiration Catheter is a single-lumen, reinforced, flexible, variable stiffness composite catheter that facilitates the removal of thrombus from the neurovasculature when connected to a compatible aspiration pump and the Vesalio Aspiration Tubing Set. The catheter distal shaft has an external hydrophilic coating aimed at reducing friction during use. The distal end of the catheter shaft is radiopaque for fluoroscopic visualization, and the proximal end contains a luer hub that allows attachment of accessories for flushing and aspiration.

    The peel-away introducer sheath is provided in the package to facilitate the insertion of the NeVasc Aspiration Catheter's distal tip into an appropriate vascular sheath.

    The Vesalio Aspiration Tubing Set is provided in a separate package and is made of common medical grade polymers. It comprises of a hollow cylindrical tube that is bonded to a standard male rotator fitting on one end. The male rotator allows the tubing to connect to a rotating hemostasis valve (RHV) or the female luer hub of the NeVasc Aspiration Catheter. The other end of the tubing consists of a hose fitting to enable connection with an aspiration pump. A flow switch is connected in line to provide vacuum control.

    The NeVasc Aspiration System is provided sterile, non-pyrogenic, and is intended for single use only.