Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163119

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-01

    (174 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NaviSuite SSI Edition is used to support a clinical ultra-sound session and follow percutaneous procedure by means of providing additional real-time image information from a 2nd modality like CT or MR or US or its combination with the possibility to plan the ideal treatment procedure. The second modality or their combinations are acquired and 3D is generated inside the software permitting the anatomical correlation with the patient position.

    "The second modality image is not intended to be used as a standalone diagnostic image since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always been seen as an additional source of information."

    "The 2nd modality image only should not be used for diagnostic purposes."

    Device Description

    NaviSuite SSI Edition 0.9.10 is a medical product to support a clinical ultrasound session and follow percutaneous procedure by means of providing additional realtime image information from a 2nd modality like CT or MR or US or its combination with the possibility to plan the ideal treatment procedure. The second modality or their combinations are acquired and a 3D is generated inside the software permitting the anatomical correlation with the patient position.

    By using the 2nd modality image the user gains security in assessing the morphology of the ultrasound image and obtains a virtual feed-back on the procedure provided by the virtual percutaneous tool position enabled by a dedicated receiver, its treatment region delineated by custom tables and a dedicated computation in 3D of the target region.

    The 2nd modality image is not intended to be used as a standalone diagnostic image since it represents information of a patient probably not congruent with the current (real) patient situation and shall therefore always been seen as an additional source of information.

    NaviSuite SSI Edition is a software designed for use together with an ultrasound device: AIXPLORER (K142100).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the NaviSuite SSI Edition, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    System Accuracy (2nd Modality Image Display)An "overlapping" accuracy of at least 5mm can be achieved between the ultrasound image and the 2nd modality image, under the presumption that the registration has been done with an overall error better than 5mm. This accuracy is highly dependent on the registration process and the patient's condition (e.g., breathing) compared to when the 2nd modality scan (e.g., CT) was acquired.
    Needle Navigation AccuracyThe software projects the virtual needle position over the real-time US image. The system passed tests to confirm this function using a phantom. The software itself "cannot determine the accuracy of the treatment procedure," as the therapist/physician is responsible for verification.
    Tracking Quality DetectionThe system can detect low sensor tracking quality and indicate if tracking is distorted (e.g., by metallic objects). Tests for verifying tracking accuracy and "sensor in range status" were carried out.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for a test set in terms of patient data or specific image sets.

    • Needle Navigation Test: A CIRS Abdominal Phantom was used for testing. This is a non-human, simulated anatomical model.
    • Tracking Quality Test: No specific sample size or type of data (beyond stating "tests... have been carried out") is mentioned for this.

    The data provenance for any testing mentioned (phantom-based) is not from human patients, and therefore the concepts of "country of origin" or "retrospective/prospective" do not apply in the context of clinical data for the reported tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The document does not mention using experts to establish ground truth for any of the reported tests. The tests performed were primarily technical verifications (phantom-based needle navigation, tracking quality) rather than diagnostic performance evaluations requiring expert adjudication.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    No adjudication method is mentioned, as the tests performed do not involve human diagnostic interpretation that would require such a process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No MRMC comparative effectiveness study is mentioned. The device is intended to support a clinical ultrasound session by providing additional real-time image information from a 2nd modality; it is explicitly stated that the 2nd modality image is not intended to be used as a standalone diagnostic image and should not be used for diagnostic purposes. This phrasing suggests the device is an assistive tool rather than a standalone diagnostic AI that would be subject to MRMC studies comparing human performance with and without its assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    The NaviSuite SSI Edition, as described, is not a standalone diagnostic algorithm. It is a software designed to be used "together with an ultrasound device" and provides "additional real-time image information." The provided texts highlight that the "2nd modality image is not intended to be used as a standalone diagnostic image" and "The 2nd modality image only should not be used for diagnostic purposes." Therefore, a standalone performance study in the context of diagnostic accuracy would not be applicable or appropriate for this device's intended use. The technical tests performed (system accuracy, tracking quality) are standalone in the sense that they evaluate the software's functional performance, but not its diagnostic capability.

    7. The Type of Ground Truth Used

    • System Accuracy (2nd Modality Image Display): The ground truth for the 5mm accuracy claim relies on the "presumption that the registration has been done with an overall error better than 5mm." This implies a technical measurement or calculation of registration error.
    • Needle Navigation Testing: The ground truth for this appears to be the physical location of the needle within the CIRS Abdominal Phantom compared to the software's projected virtual needle position. This would be a physical phantom-based ground truth.
    • Tracking Quality Test: The ground truth would be the known presence or absence of factors causing "low sensor tracking quality" or "distorted tracking" (e.g., metallic objects) and the software's ability to correctly detect and report these. This is also a technical ground truth.

    8. The Sample Size for the Training Set

    The document does not mention any training set size. This device appears to be primarily an image registration and visualization tool rather than a machine learning model that would require a dedicated training set of images with established ground truths for learning.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned, information on how its ground truth was established is not provided.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1