Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K173086

    Validate with FDA (Live)

    Date Cleared
    2017-11-07

    (39 days)

    Product Code
    Regulation Number
    886.5925
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia), and/or presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia), and/or presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.

    The lenses are intended for single-use disposable wear.

    Device Description

    The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibilitytinted, containing a UV blocker and are available in a spheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1.1.1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380nm. Each lens is supplied sterile in a blister pack containing buffered saline solution with Tween 80, Sodium Hyaluronate, and Polyethylene Glycol.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. The submission asserts substantial equivalence to a predicate device (K150385) with the addition of a modified saline solution.

    Here's an analysis of the acceptance criteria and study data provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria presented are in the form of material properties and performance characteristics, demonstrating equivalence to the predicate device.

    Acceptance Criteria / Performance CharacteristicProposed Device (NaturalVue etafilcon A)K150385 Predicate (NaturalVue etafilcon A)
    Production MethodCast-MoldedCast-Molded
    USAN Nameetafilcon Aetafilcon A
    Material ClassificationGroup 4 high water ionicGroup 4 high water ionic
    Water Content (%)58%58%
    Refractive Index1.40231.4023
    Oxygen Permeability (35°C) polarographic method (ISO/Fatt)19.73 x 10⁻¹¹ (cm²/sec)(ml O2/ml-mmHg)19.73 x 10⁻¹¹ (cm²/sec)(ml O2/ml-mmHg)
    Percent Transmittance % T at 593nm> 95%> 95%
    % T at 380-315nm (UVA)< 30%< 30%
    % T at 315-280nm (UVB)< 5%< 5%
    Lens DesignSpherical/Aspherical, Toric, Multifocal, Multifocal ToricSpherical/Aspherical, Toric, Multifocal, Multifocal Toric
    Packaging SolutionBorate buffered saline (with Tween 80, Sodium Hyaluronate, and Polyethylene Glycol)Borate buffered saline

    Additional Performance Tests (Qualitative/Binary):

    • Biocompatibility: Negative responses for cytotoxicity, maximization, and ocular irritation tests. Validity of blister packaging demonstrated by passing extraction tests. (Applies to both proposed and predicate as they share the same material).
    • Microbiology (Sterilization): Steam sterilization validated to deliver minimum SAL of 10⁻⁶. Shelf-life stability supports sterility through expiration.
    • Bacteriostatic Validation: Steam sterilizer effectiveness confirmed. Microorganisms killed under tested conditions. Lenses remained sterilized with no microbial growth for 5 years (accelerated conditions). Package seal remained tight for 5 years (accelerated conditions).
    • Leachability: No leachable monomers and additive residues at detected levels.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical performance tests for material properties, biocompatibility, microbiology, bacteriostatic validation, and leachability. It does not describe a "test set" in the context of an algorithm or diagnostic device. Instead, it refers to the product itself and its components undergoing various testing.

    • Sample Size: The document does not specify exact sample sizes for each non-clinical test (e.g., how many lenses were tested for water content, or how many individual cytotoxicity tests were conducted). It only states that "All tests were conducted in accordance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses."
    • Data Provenance: The tests are described as being performed to demonstrate safety and effectiveness of the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. The data originates from these non-clinical studies conducted by or for Visioneering Technologies, Inc. There is no mention of country of origin for the underlying data, but the submission is to the U.S. FDA. The data is implicitly prospective in the sense that these tests were performed on the device or its components to support this specific 510(k) submission, rather than re-analyzing existing data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This question is not applicable to this submission. The device is a contact lens, not an AI/diagnostic algorithm that requires expert-established ground truth for its performance evaluation in the way a medical image analysis software would. The "ground truth" for contact lenses is established through standardized physical, chemical, and biological testing methods (e.g., measuring water content, oxygen permeability, cytotoxicity, sterility).

    4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

    This question is not applicable as there is no "test set" in the context of human expert review and adjudication for an AI/diagnostic algorithm. The non-clinical tests described have quantifiable outcomes or pass/fail criteria based on established scientific methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device is a contact lens and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a contact lens and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests performed, the "ground truth" is based on:

    • Established scientific methodologies and standards: For material properties (e.g., water content, refractive index, oxygen permeability, transmittance), the ground truth is the measured physical/chemical properties.
    • Validated biological tests: For biocompatibility (cytotoxicity, maximization, ocular irritation), the ground truth is the biological response observed in animal or in-vitro models according to recognized standards.
    • Sterilization validation standards: For microbiology, the ground truth is the demonstration of a specific sterility assurance level (SAL).
    • Analytical chemistry techniques: For leachability, the ground truth is the detection or non-detection of leachable substances.

    8. The sample size for the training set

    This question is not applicable. The device is a contact lens and does not involve a "training set" for an AI algorithm.

    9. How the ground truth for the training set was established

    This question is not applicable. The device is a contact lens and does not involve a "training set" for an AI algorithm.

    Summary of the Study and Conclusion:

    The study presented is a non-clinical evaluation to demonstrate the substantial equivalence of the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses to a previously cleared predicate device (K150385). The central argument for substantial equivalence relies on:

    • Identical material composition (etafilcon A): The USAN name, material classification (Group 4 high water ionic), water content (58%), refractive index (1.4023), oxygen permeability (19.73 x 10⁻¹¹), % T at 593nm (>95%), UVA transmittance (<30%), and UVB transmittance (<5%) are all identical to the predicate device.
    • Identical production method: Cast-molded.
    • Identical range of lens designs.
    • Historical Clinical Use: The submission leverages the well-established safety and effectiveness of etafilcon A lenses, citing several predicate and reference devices, including the K150385.
    • Non-Clinical Performance Tests: Comprehensive testing was conducted in accordance with FDA guidance for daily wear contact lenses, covering:
      • Material Properties: Confirmed the listed characteristics are equivalent to the predicate.
      • Biocompatibility: Demonstrated material safety based on negative responses in cytotoxicity, maximization, and ocular irritation tests.
      • Microbiology: Validated the steam sterilization process to achieve a minimum SAL of 10⁻⁶ and confirmed shelf-life sterility.
      • Bacteriostatic Validation: Confirmed sterilizer effectiveness and lens sterility/package integrity over time (accelerated conditions).
      • Leachability: Showed no detectable leachable monomers or additive residues.
    • Modification: The only significant difference highlighted is the modified saline solution (addition of Tween 80, Sodium Hyaluronate, and Polyethylene Glycol), which was previously cleared under K161739, implying its safety and compatibility.

    The conclusion is that based on these performance and manufacturing verification studies, the proposed device is equivalent to the predicate device with respect to intended use, principles of operation, and technological characteristics, especially given the prior clearance of the modified saline solution.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150385

    Validate with FDA (Live)

    Date Cleared
    2015-04-30

    (71 days)

    Product Code
    Regulation Number
    886.5925
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NaturalVue (etafilcon A) Sphere Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +20.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    NaturalVue (etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

    NaturalVue (etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

    NaturalVue (etafilcon A) Toric Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with non-diseased eyes who may require a reading addition of up to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 10.00 diopters or less.

    NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The lenses are intended for single-use disposable wear.

    Device Description

    The NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are visibility-tinted, containing a UV blocker and are available in a spherical, aspheric, toric, multifocal, and multifocal toric designs. They are manufactured by cast molding method. The device is an ionic hydrogel lens derived from etafilcon A which is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and 1,1,1-trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. The copolymer consists of 42% etafilcon A and 58% water by weight when immersed in normal buffered saline solution. The lenses are tinted blue using C.I. Reactive Blue 19 to make them more visible for handling. The lenses contain a UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of the lenses are less than 5% in the UVB range of 280-315nm and less than 30% in the UVA range of 316-380mm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.

    AI/ML Overview

    The provided text is a 510(k) summary for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses. This document does not describe a study that uses AI or machine learning to determine performance against acceptance criteria. Instead, it describes non-clinical tests to demonstrate the safety and effectiveness of the contact lenses.

    Therefore, many of the requested elements (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable as they relate to AI/ML device evaluations.

    However, I can extract the acceptance criteria and reported device performance from the non-clinical studies described in the document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are based on demonstrating substantial equivalence to a predicate device (K140025 NaturalVue (etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses) and compliance with the May 1994 FDA Guidance Document for Daily Wear Contact Lenses. Where specific numerical acceptance criteria are not explicitly stated, the equivalence to the predicate device serves as the implicit acceptance criterion.

    CharacteristicAcceptance Criteria (Implicitly: Equivalent to Predicate K140025 or regulatory standards)Reported Device Performance (NaturalVue (etafilcon A))Meets Acceptance Criteria?
    Material Properties
    Production MethodCast-Molded (per predicate)Cast-MoldedYes
    USAN Nameetafilcon A (per predicate)etafilcon AYes
    Material ClassificationGroup 4 high water ionic (per predicate)Group 4 high water ionicYes
    Water Content (%)58% (per predicate)58%Yes
    Refractive Index1.4023 (per predicate)1.4023Yes
    Oxygen Permeability19.3 x 10-11 (cm²/sec)(ml O2/ml-mmHg) @ 35°C (per predicate)19.3 x 10-11 (cm²/sec)(ml O2/ml-mmHg) @ 35°CYes
    % Transmittance at 593nm> 95% (per predicate)> 95%Yes
    % Transmittance at 380-315nm< 30% (per predicate)< 30%Yes
    % Transmittance at 315-280nm< 5% (per predicate)< 5%Yes
    BiocompatibilityNegative responses for cytotoxicity, maximization, and ocular irritation (per predicate). Validity of blister packaging (passing standard extraction tests).Negative responses for all tests. Blister packaging validity demonstrated.Yes
    MicrobiologyMinimum SAL of 10-9 for steam sterilization. Sterility maintained through expiration.Validated to deliver SAL of 10-9. Shelf-life stability supports sterility.Yes
    Bacteriostatic ValidationUniformity of temperature in sterilizer; Tested microorganisms killed.Demonstrated uniformity of temperature; Tested microorganisms killed.Yes
    Shelf-LifeLenses remain sterile and packages remain tight for claimed expiration.Lenses remained sterilized with no microbial growth for 4 years (accelerated conditions). Seal remained tight for 4 years (accelerated conditions).Yes
    LeachabilityNo leachable monomers and additive residues at detection levels.No leachable monomers and additive residues at detection levels.Yes

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of an AI/ML device. For non-clinical material property testing, the sample sizes are not explicitly stated. The biocompatibility, microbiology, and leachability studies were conducted on the predicate device (K140025) and applied to the current device due to substantial equivalence, so the specifics of their sample sizes are not in this document.

    Data Provenance: The document refers to the predicate device (K140025), which was submitted by Pegavision Corporation, Taiwan. The current device is from Visioneering Technologies, Inc., and the non-clinical tests were performed to demonstrate equivalence. It's a non-clinical, retrospective comparison to previously established data and standards.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/ML device, and no "ground truth" established by experts for a test set is mentioned.


    4. Adjudication method for the test set

    Not applicable. No test set requiring expert adjudication is described for this type of device and study.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    Not applicable. This is not an AI/ML device, and thus, no MRMC study was conducted.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device or an algorithm.


    7. The type of ground truth used

    Not applicable in the context of AI/ML. For the non-clinical tests, "ground truth" would be the established scientific and regulatory standards for contact lens material properties, biocompatibility, sterility, and leachability. For example:

    • Material Properties: Direct measurements using laboratory techniques.
    • Biocompatibility: Standardized in vitro and in vivo (animal, if applicable, though not specified here) tests.
    • Microbiology: Validation of sterilization processes and sterility testing.
    • Leachability: Analytical chemistry techniques.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."


    9. How the ground truth for the training set was established

    Not applicable. No training set is described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1