Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162408

    Validate with FDA (Live)

    Date Cleared
    2018-04-05

    (584 days)

    Product Code
    Regulation Number
    872.3310
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Varnish for final coating for resin temporaries to produce a smooth and shiny surface.

    Dental sealing lacquer for coating resin temporaries

    Device Description

    The light-curing one-component dental sealing lacquers NanoVarnish, Plaquit and Lightpaint on Surface are designed for coating resin parts of complete and partial dental prosthesis and for provisional crowns and bridges. The lacquer is applied to a dental restoration and cured with the help of a dental light-curing unit.

    AI/ML Overview

    This document is a 510(k) premarket notification for a dental sealing lacquer. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of AI-powered device performance. Therefore, many of the requested criteria regarding AI acceptance, human readers, and ground truth for AI model training cannot be extracted from this document.

    Here's what can be extracted based on the provided text:

    Acceptance Criteria and Device Performance

    For this type of device (a dental material), "acceptance criteria" are based on demonstrating comparable physical and chemical properties to a legally marketed predicate device, rather than accuracy metrics for an AI system. The study performed is a comparative study of the new device against a predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Predicate: VITA ENAMIC® Glaze K123761)Reported Device Performance (NanoVarnish, Plaquit, Lightpaint on Surface)
    Intended UseSealing lacquer for characterization of dental restorationsDental sealing lacquer for coating resin temporaries
    Indications for UseFinal coating for shade customization/characterization, protection from abrasion wearVarnish for final coating for resin temporaries to produce a smooth and shiny surface
    DesignSupplied in glass bottlesSupplied in glass bottles
    Chemical DescriptionOne component UV-light curing (meth)acrylic lacquer radical polymerizationOne component UV-light curing (meth)acrylic lacquer radical polymerization
    Appearance (visual)Colorless, liquidColorless, liquid
    Viscosity at 23°C≤ 0.15≤ 0.15 - 0.02
    Pencil hardness (acc. to Wolff Wilborn)2 H2 H
    Working / Processing time< 1 min.< 1 min.
    Curing timeapprox. 3 min.approx. 3 - 5 min.
    TemperatureRoom temperatureRoom temperature
    Shelf life2 years2 years
    Conformance to consensus standardsEN ISO 13485: 2012 + AC:2012, ISO 7405: 2008, ISO 14971: 2007, ISO 10993-1:2009EN ISO 13485: 2012 + AC:2012, ISO 7405: 2008, ISO 14971: 2007, ISO 10993-1: 2009

    Explanation of Differences (as stated in the document): The document explicitly states that "The difference of the Intended Use and the Indications for Use of the Dreve dental sealing lacquers (Plaquit, NanoVarnish and Lightpaint on Surface) and the predicate are due to the availability of the predicate K123761 VITA ENAMIC® Glaze in a set of devices." It clarifies that the Glaze material within the predicate's set is used "for coating purpose only" and provides "protection of the stains from abrasion wear." The document concludes that "The technological characteristics of the subject device and the predicate are virtually identical and both devices have generically the same intended use."


    Information NOT available in the document (due to the nature of the device and the submission):

    This document describes a conventional medical device (a dental varnish), not an AI/ML-powered device. Therefore, the following information, which is relevant for AI acceptance criteria, is not present and cannot be extracted:

    1. Sample sizes for test set and data provenance: Not applicable. The "study" here is a set of comparable physical and chemical tests, not a clinical trial with patient data or an AI model's test set.
    2. Number of experts and their qualifications for ground truth: Not applicable for a physical material. Ground truth in this context refers to standard measurements/tests.
    3. Adjudication method for the test set: Not applicable.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not applicable. This is for AI-assisted human performance studies.
    5. Standalone (algorithm only without human-in-the loop performance): Not applicable as there is no algorithm.
    6. Type of ground truth used: For a physical material, the "ground truth" is established through standardized laboratory testing and material characterization against defined specifications and recognized consensus standards (e.g., ISO standards for dental materials). The document states: "Testing has confirmed these devices meet their product specification."
    7. Sample size for the training set: Not applicable. This document is not about an AI model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1