Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K243150

    Validate with FDA (Live)

    Date Cleared
    2025-01-31

    (123 days)

    Product Code
    Regulation Number
    874.3335
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nuance Audio Hearing Aid Software is a software-only mobile medical application that is intended to be used with compatible wearable electronic products. The feature is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The software utilizes a preset fitting strategy and is adjusted by the user to meet their hearing needs without the assistance of a hearing healthcare professional. The device is intended for Over-the-Counter use.

    Device Description

    The Nuance Audio Hearing Aid Software is a software-only device that is comprised of a pair of software modules which operate on two separate required products: (1) Nuance Audio Application on a compatible iOS or Android product, and (2) Nuance Audio Hearing Aid Software Firmware (FW) using the compatible wearable electronic computing platform, the Nuance Audio Glasses (smart glasses). The Nuance Audio Hearing Aid Software amplifies sounds for adult users, 18 years of age and older, who have a perceived mild to moderate hearing loss. The Nuance Audio™ App, iOS or Android mobile application, quides the user through the initial onboarding and set up process providing step-by-step instructions to guide the users can select from four preset options, which are generated from typical mild to moderate hearing loss configurations (Urbanski et al., 2021) with NAL-NL2 gain settings applied. The user may also adjust the overall volume, noise reduction, and microphone directionality. The device incorporates a feedback control algorithm to reduce the likelihood of the user experiencing acoustic feedback. The same amplification settings will be applied to both the left and right ear. This device can be fitted by a user without assistance from a hearing healthcare professional.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Nuance Audio Hearing Aid Software, based on the provided text:

    Nuance Audio Hearing Aid Software: Acceptance Criteria and Study Information

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Metric/StandardAcceptance Criteria (Target)Reported Device Performance (Nuance Audio Hearing Aid Software)
    Electro-AcousticsLatency ClauseMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)8.8 ms (Predicate: 3.15 ms – performance testing evaluated differences and supports substantial equivalence)
    Frequency ResponseMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)200Hz - 6300 Hz (Predicate: 100 — 10,000 Hz – performance testing evaluated differences and supports substantial equivalence)
    Self-Generated NoiseMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)29.9 dBA (Predicate: 28 dBA – performance testing evaluated differences and supports substantial equivalence)
    Harmonic DistortionLess than or equal to 1%Less than or equal to 1%
    Maximum Output ValueMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)112.7 dB SPL (Predicate: 106 dB SPL – performance testing evaluated differences and supports substantial equivalence)
    Insertion GainWithin 5 dB of predicate device at most frequencies; within 5 dB of NAL-NL2 targets except at one frequency (5 kHz, within 8 dB).Within 5 dB of predicate device at most frequencies; within 5 dB of NAL-NL2 targets except at one frequency (5 kHz, within 8 dB).
    Usability TestingCritical ErrorsZero critical errors during simulated use scenariosAll tasks completed without critical errors.
    User SatisfactionOverall positive satisfaction on usabilityOverall positive satisfaction on the usability of the device expressed by users.
    Acceptance CriteriaAll defined acceptance criteria for EN 62366-1:2015 metAll defined acceptance criteria were met (EN 62366-1:2015).
    Clinical PerformanceReal-Ear Aided Response (REAR)Each preset (A, B, C, D) within ±6 dB of NAL-NL2 targets at 250, 500, 1000, 2000, and 4000Hz.Each Nuance Audio Hearing Aid Software preset (A, B, C, and D) profile was within the pre-specified criterion range of NAL-NL2 targets (±6 dB).
    Speech-in-Noise (SNR-50)Improvement in SNR (threshold not explicitly stated, but typically a positive improvement is expected)Average unaided vs. aided improvement was 3.49 dB.
    IOI-HA Total ScoreClinically meaningful outcome score for the intended population (threshold not explicitly stated)Mean IOI-HA total score was 29.04 (SD = 4.62), considered clinically meaningful for individuals with perceived mild to moderate hearing loss to achieve hearing benefit.
    General ComplianceSpecial ControlsMeets special controls for 874.3335 (including addressing clinical performance for fitting strategy)Meets special controls for 874.3335, and clinical data provided to address differences in technology (preset vs. self-fitting) to support substantial equivalence.
    OTC Hearing AidsMeets specifications associated with over-the-counter hearing aids (21 CFR 800.30)Nuance Audio Hearing Aid Software was shown to meet the specifications associated with over-the-counter hearing aids (21 CFR 800.30).

    2. Sample Size Used for the Test Set and Data Provenance

    • Usability Testing:
      • Sample Size: 15 representative adult users.
      • Data Provenance: Not explicitly stated, but implies prospective data collection from participants in a usability validation study.
    • Real-ear measurements:
      • Sample Size: 29 participant ears.
      • Data Provenance: Not explicitly stated, but implies prospective data collection from participants.
    • Speech in noise clinical performance:
      • Sample Size: 19 adults with mild to moderate hearing loss.
      • Data Provenance: Not explicitly stated, but implies prospective data collection from participants.
    • Patient Reported Outcome Measure (IOI-HA):
      • Sample Size: 23 adult subjects identified with self-perceived mild-to-moderate sensorineural hearing loss.
      • Data Provenance: Not explicitly stated, but implies prospective data collection after a 14-day trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly mention the use of experts to establish ground truth for the test set from these clinical studies. Instead, the studies compare the device's performance against:

    • Objective measures: NAL-NL2 targets (a commonly used clinical fitting formula).
    • Standardized clinical outcomes: SNR-50 (speech-in-noise ratio) and IOI-HA scores (patient-reported outcome).
    • Usability standards: IEC 62366-1:2015.

    The ground truth for the "typical mild to moderate hearing loss configurations" used to generate the presets (Urbanski et al., 2021) suggests a reliance on established audiological research rather than individual expert consensus for a specific test set.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned for any of the test sets. The studies appear to rely on direct measurements, standardized clinical tests, or user feedback.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is reported in relation to human readers improving with AI vs. without AI assistance. The device is a "hearing aid software" rather than an interpretative AI for medical imaging, for example, where MRMC studies are common. The clinical performance studies directly evaluate the device's efficacy.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies described for "Performance Testing" (Electro-Acoustics, Real-ear measurements, Speech in Noise, IOI-HA) represent standalone performance of the algorithm integrated into the compatible hardware (Nuance Audio Glasses) without human-in-the-loop clinical decision-making. The usability testing assesses the user's ability to operate the device effectively, which is human-in-the-loop operation, but the core performance metrics are evaluating the device's output.

    7. Type of Ground Truth Used for the Test Set

    The ground truth for the test sets consisted of:

    • Objective Audiological Targets: NAL-NL2 (National Acoustics Laboratories-Non-Linear 2) fitting formula.
    • Standardized Clinical Metrics: SNR-50 (signal-to-noise ratio at 50% word recognition) and International Outcome Inventory for Hearing Aids (IOI-HA) scores.
    • Usability Standards: Compliance with IEC 62366:2015 and completion of tasks without critical errors.
    • Bench Test Standards: Adherence to ISO, ANSI/ASA, CTA, IEC standards for biocompatibility, electrical safety, EMC, battery safety, and software.

    8. Sample Size for the Training Set

    The document does not specify a sample size for a training set. The device uses "four preset options, which are generated from typical mild to moderate hearing loss configurations (Urbanski et al., 2021) with NAL-NL2 gain settings applied." This suggests that the "training" or development of these presets was based on established audiological research and fitting formulas, rather than a data-driven machine learning training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the "training" (i.e., the generation of the preset fitting strategies) for the Nuance Audio Hearing Aid Software was established based on:

    • Published Audiological Research: "Typical mild to moderate hearing loss configurations (Urbanski et al., 2021)."
    • Standardized Fitting Formula: Application of the NAL-NL2 (National Acoustics Laboratories-Non-Linear 2) gain settings to these configurations.

    This indicates that the presets are derived from established audiological science and validated hearing loss profiles, rather than being "learned" from a distinct training dataset with a separate ground truth determination process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162967

    Validate with FDA (Live)

    Device Name
    nuance UNIVERSAL
    Date Cleared
    2017-03-24

    (151 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Direct restorations of anterior or posterior teeth (Class I - Class V)
    2. Diastema closures
    3. Intraoral repairs of fractured crowns/bridges
    Device Description

    nuance® UNIVERSAL is a Class II medical device, under Classification 21 CFR 872.3690. Product Code EBF. intended to be used as a composite tooth restorative. The device falls within the scope of guidance document, "Dental Composite Resin Devices -Premarket Notification [510(k)] Submissions. According to the applicable FDA recognized consensus standard, ISO 4049 - "Dentistry - Polymer-based restorative materials, this device is classified as Class 2: materials whose setting is affected by the application of energy from an external source, such as blue light or heat (Group 2). nuance® UNIVERSAL is used in direct restorations of anterior or posterior teeth. They are polymerizable dental monomer resins that are chemically-cured when exposed to dental curing lights. It is used to form a durable, aesthetic restoration. It is a higher viscosity material to be used where mechanical strength is a primary importance. nuance® UNIVERSAL may be appropriate for dental Class I, II, III, IV and V type restorations. nuance® UNIVERSAL is packaged in a 4 gram single-barrel syringe, a 0.35 gram singules, and offered in 11 shades listed below:

    AI/ML Overview

    This document details the 510(k) submission for Nuance UNIVERSAL, a light-curable dental restorative material. The submission aims to prove substantial equivalence to a predicate device, CLEARFIL MAJESTY Posterior, rather than establishing acceptance criteria and proving performance through a clinical study as would be done for novel or high-risk devices.

    Therefore, many of the requested details, such as sample size for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, or standalone algorithm performance, are not applicable to this type of device submission, which relies on bench testing against recognized standards and comparison to a legally marketed predicate device with a history of safe use.

    Here's the information that is available based on the provided text, structured as closely as possible to your request, with explicit notes for inapplicable sections:


    Device: Nuance UNIVERSAL (Light-Curable Dental Restorative Material)

    Study Focus: Substantial Equivalence to Predicate Device (CLEARFIL MAJESTY Posterior)

    Type of Study Conducted: Bench Testing against ISO 4049 (Dental - Polymer-based restorative materials) and biocompatibility testing. This is not a clinical study involving human patients or an AI/algorithm-based device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from the ISO 4049 standard, and the reported performance is compared against these and the predicate device.

    Acceptance Criteria (ISO 4049 Requirement)Nuance UNIVERSAL Reported PerformancePredicate Device (CLEARFIL MAJESTY Posterior) Reported Performance
    Sensitivity to ambient light: >60 seconds, no curePassPass
    Depth of cure: > 0.5 mm5.588 mm5.550 mm
    Flexural strength: >80 Mpa98.7 Mpa83.4 Mpa
    Water sorption/solubility: ≤ 40 micrograms/mm^323.3 / 2.8 micrograms/mm^324.9 / 2.7 micrograms/mm^3
    Shade/color stability: match standardMatch/PassMatch/Pass
    Radio opacity: ≥ 1.0 mm aluminum2 mm2 mm

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in terms of the number of samples for each test (e.g., how many specimens for flexural strength). The testing was conducted according to the ISO 4049 standard, which specifies the number of samples required for each test. This is bench testing, not a clinical test set from patient data.
    • Data Provenance: The data comes from in vitro (bench) testing performed by the manufacturer, Den-Mat Holdings, LLC. The location of the testing facility is not specified, but the company is based in Lompoc, California, USA. The data is prospective in the sense that the tests were performed to generate data for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. This device is a material, and its performance is evaluated through physical and chemical property testing against an international standard (ISO 4049) and comparison to a predicate device. Ground truth is established by objective measurements based on the standard, not expert consensus on interpretations of images or clinical outcomes.

    4. Adjudication Method for the Test Set

    • Not Applicable. See point 3. Testing involves objective measurements, not subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size.

    • No. This is not an AI/imaging device requiring an MRMC study. It is a dental restorative material.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device does not involve an algorithm or AI.

    7. The type of ground truth used

    • Objective measurement against an international standard (ISO 4049) and comparative data to a legally marketed predicate device.
    • Additionally, Biocompatibility was demonstrated through:
      • Substantial equivalence to the predicate device, citing materials with a long history of safe use in dentistry.
      • Specific in vitro cytotoxicity testing (ISO 10993 Part 5), with the conclusion "the test article was not considered to have a cytotoxic effect (no reactivity)."

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/machine learning device that requires a training set. The device formulation is based on established dental material science principles and minor modifications to an existing predicate device.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143679

    Validate with FDA (Live)

    Device Name
    Nuance FLOW
    Date Cleared
    2015-05-01

    (128 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    nuance® FLOW is intended to be used as a composite tooth restorative.

    nuance® FLOW is recommended for the following types of applications:

    1. Direct restorations of anterior or posterior teeth
    2. Cavity base/liner
    3. Intraoral repairs of fracture crowns/bridges
    Device Description

    nuance® FLOW is a Class II medical device intended to be used as a composite tooth restorative. According to the applicable FDA recognized consensus standard, ISO 4049-"Dentistry- Polymer-based restorative materials", this device is classified as Class 2: materials whose setting is affected by the application of energy from an external source, such as blue light or heat (Group 2). nuance® FLOW is DenMat's Flowable for use in direct restorations of anterior or posterior teeth. They are polymerizable dental monomer resins that are chemically-cured when exposed to dental curing lights. It is used to form a durable, esthetic restoration. It is a lower viscosity material that allows it to be used when ease of application and thinner layers are needed. nuance® Flow may be appropriate for dental Class I, III, IV and V type restorations where maximum mechanical strength and wear resistance is not required. A small amount of fluoride (.5-.7%) is included to minimize likelihood of secondary caries. nuance® FLOW is packaged in a 1 mL single• barrel syringe and offered in 8 shades: A1, A2, A3, A3.5, B1, B2, BL1 and BL2.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the nuance® FLOW dental restorative material, and its substantial equivalence to the predicate device LC FLOWFILL.

    Here's an analysis based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test (ISO 4049 Std.)Acceptance Criteria (Requirement)nuance® FLOW PerformanceLC FLOWFILL Performance (Predicate)
    Film thickness< 50 micron37.8 micron40.5 micron
    Sensitivity to ambient light> 60 seconds, no curepassPass
    Depth of cure> 0.5mm2.49 mm2.55 mm
    Flexural strength> 50 Mpa85.9 Mpa79.0 Mpa
    Water sorption/solubility< 40 micrograms/mm³27.2 micrograms/mm³25.9 micrograms/mm³
    Shade/color stabilitymatch standardmatch/passmatch/pass
    Radio opacity> 1.0mm aluminum3mm aluminum2mm aluminum

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each test performed on nuance® FLOW or LC FLOWFILL. It only states that "Testing results indicate that nuance® FLOW was as effective and performs as good as the predicate device."

    The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective. However, given that it's a 510(k) submission to the FDA, it is expected that the tests were conducted according to recognized international standards (ISO 4049) which implies controlled laboratory settings. The document also mentions "Scientific literature have been evaluated to support the use of similar products for the same indications," suggesting some reliance on existing published data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests are bench tests based on an international standard (ISO 4049), not clinical evaluations requiring expert interpretation of results for ground truth.

    4. Adjudication method for the test set

    This information is not applicable as the tests are objective, quantitative measurements against a standard, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There was no MRMC comparative effectiveness study and no mention of AI assistance. This device is a dental restorative material, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable as the device is a physical dental material and not an algorithm.

    7. The type of ground truth used

    The ground truth used for performance evaluation is based on the objective measurement criteria defined in the ISO 4049 standard for "Dentistry – Polymer-based restorative materials." This standard sets specific quantitative thresholds for various material properties.

    8. The sample size for the training set

    This information is not applicable as the device is a chemical material, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as the device is a chemical material, not a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130995

    Validate with FDA (Live)

    Date Cleared
    2013-06-14

    (65 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    1,1,1,2-tetrafluoroethane, pentafluoroethane, and 1,1,1-trifluoroethane is to be used for the treatment of verruca (warts) including plantar warts, seborrheic keratoses, actinic keratosis, achrochorodon, molluscum contagiosum, age spots, dermatofibroma, small keltoids, granuloma annulare, Porokeratosis Plantaris, Angiomas, Keratoacanthoma, chrondrodermatitis, epithelial nevus, Leukoplakia, granuloma pyogenicum, and pyogenic granuloma.

    Device Description

    The Nuance Freeze Spray System is used in the practice of dermatology in the treatment of skin lesions using a cryogen spray system. This methodology is an accepted practice used by physicians for decades using accepted procedures and techniques. It utilizes a standard cryogen composition profile to freeze common skin lesions.

    AI/ML Overview

    This 510(k) summary for the Nuance Freeze Spray System does not contain a study that proves the device meets specific acceptance criteria in the way a traditional clinical study with defined endpoints and statistical analysis would. Instead, the submission relies on the concept of substantial equivalence to previously cleared predicate devices.

    Therefore, many of the requested elements regarding acceptance criteria and study design are not applicable or cannot be extracted from the provided text.

    Here is an analysis based on the information provided:


    1. Table of Acceptance Criteria and Reported Device Performance

    Not applicable. The submission does not define specific acceptance criteria (e.g., a specific percentage reduction in lesion size, a cure rate, or a predefined safety profile) nor does it report performance metrics from a dedicated study against such criteria for the Nuance Freeze Spray System itself. The device's "performance" is implicitly assumed to be equivalent to the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. There was no specific test set or clinical study conducted for the Nuance Freeze Spray System described in this document. The device relies on comparing its characteristics to predicate devices that have already demonstrated safety and effectiveness through their past clearances.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable, as no dedicated test set or ground truth establishment process is described for this specific submission. The established practices and efficacy of cryogen spray for dermatological conditions, as supported by the predicate devices, serve as the foundational "ground truth."

    4. Adjudication Method for the Test Set

    Not applicable, as no test set or adjudication process is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (cryogen spray system), not an AI/imaging device. Therefore, MRMC studies and AI effectiveness are irrelevant to this submission.

    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    Not applicable. This is a medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" implicitly relied upon is the established clinical efficacy and safety of cryogen spray systems, particularly those using the specified chemical composition, for the treatment of various skin lesions, as evidenced by the predicate devices. This can be considered accepted clinical practice and outcomes data from previously cleared devices and general medical knowledge.

    8. The Sample Size for the Training Set

    Not applicable. The Nuance Freeze Spray System is a physical device, and its clearance is based on substantial equivalence rather than a machine learning model requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.


    Summary of the Basis for Clearance:

    The Nuance Freeze Spray System gained 510(k) clearance based on Substantial Equivalence to legally marketed predicate devices:

    • Primary Predicate: Cryosurgery, Inc. Verruca-Freeze Cryosurgery Delivery System (K982506)
    • Secondary Predicates: Cryosurgery, Inc. Verruca-Freeze Cryosurgery Delivery System (K881349) and (K944221)

    The "study" that proves the device meets the acceptance criteria is effectively this comparison demonstrating substantial equivalence. The core arguments for substantial equivalence are:

    • Identical Indications for Use: The Nuance Freeze Spray System is intended for the treatment of the same specific skin lesions as the predicate devices.
    • Identical Technical Characteristics (Cryogen Composition): The cryogen used (1,1,1,2-tetrafluoroethane, pentafluoroethane, and 1,1,1-trifluoroethane, also known as R-404a) is stated to be the "exact same chemical composition and formulation as the predicate device."
    • Similar Delivery Method: The methods for delivering the cryogen are described as "similar to the predicate."
    • No New Issues of Safety or Efficacy: The statement explicitly says, "Nuance Freeze Spray System and its predicate devices do not raise any new issues of safety or efficacy."

    Therefore, the "acceptance criteria" for the Nuance Freeze Spray System's clearance were met by demonstrating that it is fundamentally the same in terms of composition, intended use, and operational principles as already cleared devices, thereby inferring equivalent safety and effectiveness without requiring a new clinical performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101215

    Validate with FDA (Live)

    Device Name
    NUANCE
    Manufacturer
    Date Cleared
    2010-05-12

    (12 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nuance Universal Composite:

    • Direct restorations of anterior or posterior teeth (Class I Class V) .
    • Diastema closures .
    • Intraoral repairs of fractured crowns/bridges ●

    Nuance Posterior Composite:

    • Direct restorations of Class I and Class II cavities .

    Nuance Flowable Composite:

    • Direct restorations of anterior or posterior teeth .
    • Cavity base/liner .
    • Intraoral repairs of fractured crowns/bridges .
    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA to Discus Dental, LLC concerning dental composite materials. It is a regulatory document and does not contain any information regarding acceptance criteria or a study proving the device meets those criteria.

    Therefore, I cannot extract the requested information to fill in the table or answer the questions. The document is strictly about the FDA's decision on the substantial equivalence of the dental composites to legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1