Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221518

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-11-03

    (162 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NP Medical Extension Set with stabilizing base is a sterile single patient use device.

    The NP Medical Extension Set is intended for direction, intermittent infusion, continuous infusion or aspiration for all patients for which IV therapy is prescribed.

    The NP Medical Extension Set incorporates a normally closed bidirectional pressure-activated valve that prevents low pressure retrograde flow and is intended for power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

    Device Description

    The NP Medical Extension Set with Stabilizing Base is a sterile, single use, nonpyrogenic intravenous administration extension set comprised of a side repositionable tube bonded to a connector hub on one end and a female luer on the other end with a clamp in between. The connector hub has a locking male luer and a rotating tube connection joint. The hub houses a bidirectional Pressure-Activated Valve (PAV) and an integral stabilizer under the connector hub. The stabilizing base component (along with IV tape / adhesive) is intended to help anchor the extension set to the patient. This device is not made with plasticizers Diethylhexylphthalate (DEHP). The NP Medical Extension Set with Stabilizing Base may be used with compatible Vascular Access Devices (VAD), Needle Free Connectors, and other commercially available devices.

    AI/ML Overview

    Based on the provided text, the device in question is the "NP Medical nCompass Extension Set with Stabilizing Base," which is an intravascular administration set. The document is a 510(k) summary for premarket notification to the FDA, demonstrating substantial equivalence to a predicate device.

    The acceptance criteria and study proving the device meets these criteria are primarily based on bench performance testing and biocompatibility testing, as no clinical data was required or provided.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance:

    The document doesn't provide specific numerical acceptance criteria for each test with their corresponding reported performance values. Instead, it states that the device "met all requirements" and "satisfy the acceptance criteria specified by the above applicable standards."

    However, we can infer some criteria and the general reported performance:

    Performance Test / Acceptance Criteria CategoryReported Device Performance (as stated in the document)Inferable Acceptance Criteria (based on predicate comparison and industry standards)
    Device Flow Rate (gravity)"Equal to or greater than 3 L/hr."≥ 3 L/hr. (matching predicate)
    Device Flow Rate (pressurized)"Equal to or greater than 10ml/second from the device distal end at a pressure at or below 325psi."≥ 10 ml/second at ≤ 325 psi (matching predicate)
    Retrograde Pressure ResistanceMet requirementsPrevent low pressure retrograde flow (as per Indications for Use)
    Dynamic HemolysisMet requirementsDemonstrate substantial equivalence (per USP <788>)
    Positive PressureMet requirementsWithstand positive pressure without failure
    325psi Burst Resistance (rotating tube connection and other joints)Met requirementsNo burst at 325 psi
    Extension tube bond integrityMet requirementsMaintain integrity/no leaks
    Stabilization pad bond integrityMet requirementsMaintain integrity/proper adhesion
    Extension tubing port (rotating connection) interfaces, sub-atm. pressure air leakageMet requirementsNo significant air leakage
    Resistance to overridingMet requirementsFunction correctly when components are connected/operated
    ESC resistance to lipidsMet requirementsMaintain integrity/functionality when exposed to lipids
    PAV RobustnessMet requirementsMaintain functionality over repeated use
    Particulate Contamination (sub-visible particles)Met requirementsConform to USP <788> for particulate limits
    Biocompatibility (Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemocompatibility, Chemical Characterization, Allowable Limits Establishment)"All the Biocompatibility study results satisfy the acceptance criteria specified by the above applicable standards and proposed device has been found biocompatible for its intended use/indications for use."Conformance with ISO 10993-1 and specific endpoints.
    Sterilization (SAL)"Sterility Assurance Level (SAL) of 10-6"Achieve an SAL of 10-6 (as per ISO 11135-2014)
    Sterile Barrier PackagingMet requirementsConform to ISO 11607-1:2019 for maintaining sterility

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document does not specify the sample size for any of the performance tests. It states that tests were conducted using "appropriate methodology" and "NP Medical established test protocols."
    • Data Provenance: The data is generated from bench testing conducted by NP Medical. The country of origin for the testing itself is not explicitly stated, but NP Medical is based in Clinton, Massachusetts, USA. The studies appear to be prospective in nature, as they were conducted specifically for the purpose of demonstrating substantial equivalence for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device, being a physical medical device (extension set), does not involve diagnostic interpretations or algorithms that would require expert human review for ground truth establishment in the traditional sense of AI/imaging studies.
    • The "ground truth" for its performance is established through adherence to recognized consensus standards (e.g., ISO, USP) and validated bench testing protocols. The experts involved would be technical professionals (engineers, quality assurance personnel, lab technicians) performing the tests and interpreting results against the objective criteria of these standards, rather than clinical experts establishing a "truth" from patient data. Their qualifications would be in metrology, engineering, and quality systems for medical devices. The document does not specify the number or specific qualifications of these individuals.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or AI performance evaluations where multiple human readers interpret data, and a consensus or tie-breaking mechanism is needed for ground truth. For bench testing of physical device performance against objective criteria, such methods are not used. The results are quantitative measurements against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. An MRMC comparative effectiveness study is not applicable as this is a physical medical device (extension set), not an AI algorithm for diagnosis or interpretation. The document explicitly states: "VIII. CLINICAL DATA: Not applicable."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This relates to AI/software as a medical device. This product is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the performance of this device is established by:
      • Objective measurements against physical and chemical standards (e.g., flow rate, burst pressure, particulate count, material composition).
      • Adherence to recognized consensus standards (e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization, ISO 11607-1 for packaging) and FDA guidance documents.
      • Comparison to the performance characteristics of the legally marketed predicate device.

    8. The sample size for the training set:

    • Not applicable. This device is a physical product and does not involve machine learning or AI, and therefore no "training set" in the context of data for algorithm development.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1