510(k) Data Aggregation

    K Number
    K233251

    Validate with FDA (Live)

    Date Cleared
    2024-05-23

    (238 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).

    The NOxBOXi Nitric Oxide Delivery System includes:

    • The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.
    • Continuous monitoring and alarms for NO, O2, and NO2.
    • The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O₂, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.

    The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling (currently neonates). Refer to this material prior to use.

    Device Description

    The NOxBOXi Nitric Oxide Delivery System (NOxBOXi) simultaneously delivers Nitric Oxide (NO) medical gas, while monitoring Nitric Oxide, Nitrogen Dioxide (NO2), and Oxygen (O2) levels in the inspiratory limb of a ventilator for patients undergoing inhaled Nitric Oxide Therapy.

    The system is designed for use by healthcare professionals to administer treatment to patients undergoing inhaled Nitric Oxide (iNO) therapy. The NOxBOXi will deliver nitric oxide in a synchronous manner to a single patient.

    An integrated component to the NOxBOXi, the NOxMixer is intended to deliver a continuous flow of Nitric Oxide from the NOxBOXi, mixed in line with O₂ for use in iNO therapy. The NOxMixer will be used in conjunction with manually bagging a patient.

    The NOxBOXi includes the NOxBOXi Head Unit, a NOxFLOW sample line, two NO feed hoses, two regulators (connector type dependent on the gas supplier), a test circuit, NO, O2, and NO2 monitors, power supply, drainage syringe, Operating Manual & Technical Guide.

    This submission is for the introduction of new compatible ventilators including the addition of pediatric categories for existing ventilators, an additional optional software mode which disables the "Vent Flow Idle" alarm to reduce this alarm which may not be necessary and is considered a "nuisance" alarm in certain situations. Alarm initiations are still recorded in the log file. Additionally, language choices other than English have been disabled for this mode. There are no changes to the indications for use of the product, patient population of neonates, and there are no significant design changes.

    AI/ML Overview

    The provided text is a 510(k) summary for the NOxBOXi Nitric Oxide Delivery System, focusing on changes to compatible ventilators and an optional software mode. It does not present a study proving the device meets acceptance criteria in the manner typically associated with AI/ML-enabled devices, which often involve performance metrics like sensitivity, specificity, or AUC against a defined ground truth.

    Instead, this document describes a modification to an already cleared medical device (NOxBOXi Nitric Oxide Delivery System). The "study" here refers to non-clinical performance testing to demonstrate that these changes do not alter the substantial equivalence of the modified device to its predicate. The acceptance criteria are therefore related to the safety and performance parameters of the delivery system itself, rather than diagnostic or analytical accuracy of an AI model.

    Therefore, many of the requested points related to AI/ML study design (like sample size for test/training sets, expert ground truth establishment, MRMC studies, or standalone performance for an algorithm) are not applicable to this document's content.

    However, I can extract information related to the device's technical specifications and the testing performed for this submission.

    Here's an interpretation based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the existing performance characteristics of the predicate device (K231823) and regulatory standards. The reported "performance" for this submission is that the device, with the new ventilator compatibility and software mode, continues to meet these original performance specifications and does not raise new questions of safety or effectiveness.

    Performance Parameter / Acceptance CriteriaReported Device Performance (with new changes)
    NO & NO2 Monitoring Accuracy+/- 2% or 0.2ppm (No Change)
    NO Dosing Accuracy (Manual Mode)+/- 20% or 2 ppm (5-80 ppm NO); +/- 40% or 4 ppm (0-<5 ppm or >80-185 ppm NO) (No Change)
    Battery Backup Capability4 hours without AC power (No Change)
    Ventilator CompatibilityVarious models from listed manufacturers, including new additions and pediatric categories for existing models.
    Optional Software Mode FunctionalityDisables "Vent Flow Idle" alarm; alarm initiations still recorded. Language choices (other than English) disabled for this mode.
    Safety and EffectivenessNo new questions of safety or effectiveness raised. Passed all testing.
    Compliance with Standards/GuidanceVerified and validated to comply with ISO 14971, ISO 10993 (various parts), IEC 60601-1, IEC 60601-1-2, IEC 62366, ISO 80601-2-55, IEC 62304, ISO 15223-1, and relevant FDA guidance documents.
    VOC & Particulate Matter in delivered gasesVOC levels three orders of magnitude below OSHA permissible exposure levels. Particulate levels well below EPA's maximum limits.

    2. Sample size used for the test set and the data provenance:

    • Test Set Description: The "test set" here refers to the specific modifications being evaluated: the compatibility with additional ventilators and an optional software mode.
    • Sample Size: Not quantified in terms of a "sample size" of patients or images, as this is a hardware/software modification submission. The testing involves specific ventilator models and configurations, and the software mode itself.
    • Data Provenance: Not applicable in the context of clinical data or patient-derived data for an AI/ML model. The testing is non-clinical, likely bench testing, and usability testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable in the context of this 510(k) summary. Ground truth as typically understood for AI/ML diagnostic devices (e.g., expert radiological reads) is not established here. The "truth" is based on the engineering specifications and performance of the device under various conditions and its compliance with regulatory standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This relates to clinical endpoint adjudication, which is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-enabled diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a medical device for gas delivery and monitoring, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance, as demonstrated in this submission, is based on:
      • Engineering specifications and design requirements.
      • Validated test methods conforming to international standards (e.g., ISO, IEC).
      • Compliance with FDA guidance documents related to nitric oxide delivery apparatus and software.
      • Bench testing results for parameters like accuracy, flow rates, alarm functionality, and gas purity.

    8. The sample size for the training set:

    • Not applicable. This document does not describe an AI/ML model with a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This document does not describe an AI/ML model with a training set.

    Summary of the study conducted:

    The "study" or testing performed for this 510(k) submission was non-clinical performance testing and usability testing. The purpose was to demonstrate that adding compatibility with new ventilators (including pediatric categories for existing ones) and an optional software mode does not adversely impact the safety or effectiveness of the NOxBOXi Nitric Oxide Delivery System, and that it remains substantially equivalent to its predicate device (K231823). The testing focused on validating that the changes did not introduce new safety concerns or alter the specified performance characteristics of the device. This involved testing against international standards and FDA guidance relevant to medical devices, particularly those for nitric oxide delivery systems. No clinical testing was required for this specific submission given the nature of the changes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231823

    Validate with FDA (Live)

    Date Cleared
    2023-08-11

    (51 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspirator lines of a patient undergoing nitric oxide therapy (iNO).

    The NOxBOXi Nitric Oxide Delivery System includes:

    · The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.

    · Continuous monitoring and alarms for NO, O2 and NO2.

    • The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.

    The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging (currently neonates). Refer to this material prior to use.

    Device Description

    The NOxBOXi Nitric Oxide Delivery System (NOxBOXi) simultaneously delivers Nitric Oxide (NO) medical gas, while monitoring Nitric Oxide, Nitrogen Dioxide and Oxygen levels in the inspiratory limb of a ventilator for patients undergoing inhaled Nitric Oxide Therapy.

    The system is designed for use by healthcare professionals to administer treatment to patients undergoing inhaled Nitric Oxide (iNO) therapy. The NOxBOXi will deliver nitric oxide in a synchronous manner to a single patient.

    An integrated component to the NOxBOXi, the NOxMixer is intended to deliver a continuous flow of Nitric Oxide from the NOxBOXi, mixed in line with Oxygen (O₂) for use in iNO therapy. The NOxMixer will be used in conjunction with manually bagging a patient.

    The NOxBOXi includes the NOxBOXi Head Unit, a NOxFLOW sample line, two NO feed hoses, two regulators (connector type dependent on the gas supplier), a test circuit, NO, O₂ & NO₂ monitors, power supply, drainage syringe, Operating Manual & Technical Guide.

    This submission is for the addition of compatibility claims for specific ventilators and for the ventilators previously cleared for pediation. It is not related to product changes. There are no changes to the indications for use of the product, patient population of neonates, and there are no significant design changes.

    AI/ML Overview

    The provided document is a 510(k) summary for the NOxBOXi Nitric Oxide Delivery System, seeking clearance for additional ventilator compatibility. It is important to note that this document does not contain the detailed methodology and results of a clinical study or a multi-reader multi-case (MRMC) comparative effectiveness study. Instead, it focuses on non-clinical performance data and adherence to various standards and guidance documents to demonstrate substantial equivalence to a previously cleared device.

    Therefore, many of the requested items (e.g., ground truth establishment, expert qualifications, adjudication methods, MRMC study effect sizes, standalone performance, training set details) are not explicitly present in this type of regulatory submission, as they would typically be part of a full clinical trial report or a more comprehensive performance study dossier.

    However, I can extract and infer information about the acceptance criteria and the type of study conducted to prove the device meets these criteria based on the provided text, particularly from section "9. Non-Clinical and Usability Performance Data" and the comparison table in section "8. Substantial Equivalence Discussion".

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this 510(k) submission are implicitly defined by the technical specifications of the predicate device (K220898) and the accuracy requirements outlined in relevant FDA guidance documents. The "study" in this context is primarily non-clinical testing to ensure the device maintains its performance and safety profile, especially with the added ventilator compatibilities.

    Acceptance Criterion (Implicit)Reported Device Performance (as stated by "No Change" in comparison, and tested via non-clinical data)
    Functional/Performance Criteria (derived from "Technical" table)
    NO administration methodNO blended with O2 in the patient's inhalation circuit (No Change)
    NO flow rate (sample flow rate)225 ml/min (No Change)
    NO concentration provided0.0 to 80ppm (No Change)
    Presence of NO monitorYes (No Change)
    Presence of O2 monitorYes (No Change)
    Gas Monitoring accuracy (NO & NO2)+/- 2% or 0.2ppm (No Change, per comparison table; confirmed by testing to FDA guidance "Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer")
    Presence of NO2 monitor & alarmYes (No Change)
    Battery Backup capability4 hours without AC power (No Change)
    Manual bagging & back up systemNOxMIXER® (No Change)
    NO dosing range in manual mode0 - 185ppm on 800ppm cylinders (No Change)
    NO dosing Accuracy in manual mode± 20% or 2 ppm (greater) for NO doses from 5 - 80 ppm (800 ppm cylinder) and O2 flow rates of 5 - 14 L/min; +/-40% or 4 ppm (greater) for NO doses from 0 to < 5 ppm or > 80 to 185 ppm (800 ppm cylinder) and O2 flow rates of 2 to < 5 L/min or > 14 to 25 L/min (No Change, per comparison table)
    Ventilator CompatibilityEquivalent; testing shows no new questions raised regarding safety and effectiveness for various models from several manufacturers (Bio-Med Devices, Bunnell, Carefusion, Drägerwerk, Fisher & Paykel Healthcare, General Electric, Hamilton Medical, IMT Medical (Vyaire), Maquet (Getinge), Newport (Covidien), Nihon Kohden, Percussionaire, Philips Respironics, Puritan Bennett (Covidien), Smiths Medical) compared to the predicate's list.
    VOC & Particulate matter in delivered gasesVOC concentrations three orders of magnitude below OSHA permissible exposure levels. Particulate levels well below EPA's maximum limits for total suspended particulates.
    Safety and Compliance Criteria
    BiocompatibilityCompliance with ISO 10993-1, -5, -10 (K171696 references indicate prior clearance met these).
    Electrical Safety & EMCCompliance with IEC 60601-1 and IEC 60601-1-2 (K171696 references indicate prior clearance met these).
    Usability/Human FactorsCompliance with IEC 62366 and FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices." (K171696 references indicate prior clearance met these). No additional usability testing was conducted for this submission as it was for ventilator compatibility, not product changes.
    Risk ManagementCompliance with ISO 14971 (K171696 references indicate prior clearance met these).
    Software Life Cycle ProcessesCompliance with IEC 62304 (K171696 references indicate prior clearance met these).
    Respiratory Gas Monitor Specific RequirementsCompliance with ISO 80601-2-55 (K171696 references indicate prior clearance met these).
    Overall Safety & EffectivenessThe device "passed all testing and no different questions of safety or effectiveness were raised." The submission supports substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document indicates that testing was "limited to the aspects that could be affected by including compatibility with additional ventilators and ventilator categories." This implies bench testing and performance verification on the device itself and its interaction with a sample of the listed ventilators. The specific number of ventilators tested or the number of test cycles is not provided in this summary.
    • Data Provenance: The data provenance is from non-clinical (bench) testing performed by the manufacturer, Linde Gas & Equipment Inc. and NOxBOX Ltd. The country of origin of the data is not specified, but the submission is to the U.S. FDA. The testing conducted for this submission (ventilator compatibility) was of a prospective nature (i.e., new tests performed for this specific submission).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This type of information (experts, qualifications) is not applicable or not provided in this 510(k) summary because the study proving the device meets criteria is based on:

    • Comparison to an existing predicate device's established performance.
    • Direct technical measurements and verification testing (bench testing) against defined engineering specifications and regulatory standards.
    • The criteria are objective performance metrics (e.g., accuracy of gas delivery, flow rates, alarm functionality, electrical safety), not subjective interpretations requiring expert consensus.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods are relevant for studies involving subjective human interpretation (e.g., image reading by radiologists). This submission relies on objective engineering and performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. A MRMC comparative effectiveness study was not done. This type of study is used to evaluate how human readers' performance with a medical device (e.g., AI assistance) compares to their performance without it. The NOxBOXi is a medical device for gas delivery and monitoring, not an AI-assisted diagnostic tool that would typically involve human readers interpreting output. The submission explicitly states, "No clinical testing was required to support substantial equivalency of this medical device."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable for this device type. The NOxBOXi is a physical gas delivery and monitoring system, not an algorithm. Its "performance" is its ability to accurately deliver and monitor gases, which is assessed through bench testing, not algorithmic standalone performance.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance claims are:

    • For the technical specifications, it is the design specifications and measured performance against those specifications as verified in non-clinical testing.
    • For regulatory compliance, it is the requirements of various international standards (e.g., ISO, IEC) and FDA guidance documents.
    • For substantial equivalence, it is the previously cleared predicate device's established performance and safety profile.

    8. The Sample Size for the Training Set

    Not applicable. This submission describes a physical medical device and its non-clinical performance verification, not a machine learning model that would require a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As stated above, this is not an AI/ML device that uses training data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220898

    Validate with FDA (Live)

    Date Cleared
    2022-04-27

    (30 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).

    The NOxBOXi Nitric Oxide Delivery System includes:

    • The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.
    • Continuous monitoring and alarms for NO, O2 and NO2.
    • The integrated NOxMixer which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.

    The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling (currently neonates). Refer to this material prior to use.

    Device Description

    The NOxBOXi Nitric Oxide Delivery System (NOxBOXi) simultaneously delivers Nitric Oxide (NO) medical gas, while monitoring Nitric Oxide, Nitrogen Dioxide and Oxygen levels in the inspiratory limb of a ventilator for patients undergoing inhaled Nitric Oxide Therapy.

    The system is designed for use by healthcare professionals to administer treatment to patients undergoing inhaled Nitric Oxide (iNO) therapy. The NOxBOXi will deliver nitric oxide in a synchronous manner to a single patient.

    An integrated component to the NOxBOXi, the NOxMixer is intended to deliver a continuous flow of Nitric Oxide from the NOxBOXi, mixed in line with Oxygen (O₂) for use in iNO therapy. The NOxMixer will be used in conjunction with manually bagging a patient.

    The NOxBOXi includes the NOxBOXi Head Unit, a NOxFLOW sample line, two NO feed hoses, two regulators (connector type dependent on the gas supplier), a test circuit, NO, O3 & NO2 monitors, power supply, drainage syringe, Operating Manual & Technical Guide.

    Optional accessories include 4 separate NOxKITs (22mm, 12 mm, & 10mm), one way valve for HFOV (High Frequency Oscillatory Ventilation), bagging kits (hyperinflation & selfinflating) and circuit reducers (22f - 15m).

    This submission is for the addition of compatibility claims for specific ventilators and is not related to product changes. It also introduces a new accessory, a modified sample line. There are no changes to the indications for use of the product and there are no significant design changes.

    AI/ML Overview

    This document is a 510(k) premarket notification for the NOxBOXi Nitric Oxide Delivery System, specifically for adding compatibility claims for additional ventilators and introducing a new accessory. The submission states that there are no changes to the product itself, its indications for use, or its general design. Therefore, the device specifications and acceptance criteria are those of the previously cleared predicate device, K201339.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Since this submission is for adding compatibility claims and an accessory without product changes, the acceptance criteria and reported device performance are identical to the predicate device (K201339) and relate to its general operational specifications. The document compares the subject device (K220898) with the predicate (K201339) and explicitly states "No Change" for all performance parameters. Therefore, the acceptance criteria and performance are as listed below:

    Feature/ParameterAcceptance Criteria (from predicate K201339)Reported Device Performance (K220898)
    Monitoring AccuracyNO & NO2 - +/- 2% or 0.2ppmNO & NO2 - +/- 2% or 0.2ppm (No Change)
    NO concentration provided0.0 to 80ppm0.0 to 80ppm (No Change)
    NO flow rate (sample flow rate)225 ml/min225 ml/min (No Change)
    Battery Backup capability4 hours without AC power4 hours without AC power (No Change)
    NO Dosing Accuracy in Manual Mode (800ppm drug cylinder)±20% or 2 ppm (whichever is greater) for NO doses from 5-80 ppm and O2 flow rates of 5-14 L/min±20% or 2 ppm (whichever is greater) for NO doses from 5-80 ppm and O2 flow rates of 5-14 L/min (No Change)
    NO Dosing Accuracy in Manual Mode (800ppm drug cylinder)±40% or 4 ppm (whichever is greater) for NO doses from 0 to <5 ppm or >80 to 185 ppm and O2 flow rates of 2 to <5 L/min or >14 to 25 L/min±40% or 4 ppm (whichever is greater) for NO doses from 0 to <5 ppm or >80 to 185 ppm and O2 flow rates of 2 to <5 L/min or >14 to 25 L/min (No Change)
    NO flow in manual modeAdjustable 50 – 600 mL/min of NO/N2Adjustable 50 – 600 mL/min of NO/N2 (No Change)
    O2 flow range in manual bagging mode2 to 25 L/min of O22 to 25 L/min of O2 (No Change)
    Oxygen inlet pressure3.5 – 4.5 bar3.5 – 4.5 bar (No Change)
    NO delivery pressure1.65 bar from manual control valve1.65 bar from manual control valve (No Change)
    Monitoring during manual baggingYesYes (No Change)
    Alarms active during baggingYesYes (No Change)
    Back-up accuracySame as NO dosing accuracy in manual modeSame as NO dosing accuracy in manual mode (No Change)

    The study that proves the device meets these acceptance criteria is referenced as a series of non-clinical performance data and compliance to international standards and FDA guidance documents. The submission states, "Testing for this submission was limited to the aspects that could be affected by including compatibility with additional ventilators and introduction of a new accessory. No additional usability testing was conducted for this submission. No clinical testing was required to support substantial equivalency of this medical device." The previous clearances (K171696 and K201339) would have established the initial performance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily refers to "non-clinical performance data" and testing for compatibility with additional ventilators and a new accessory. It does not specify a "test set" in the context of clinical trials or expert review of cases. The testing conducted appears to be engineering and validation testing rather than human subject testing.

    • Sample size for compatibility testing: Not explicitly stated as a number of specific ventilators, but a list of manufacturers is provided (e.g., Bio-Med Devices, Bunnel, Carefusion, Drägerwerk, Hamilton Medical, Philips Respironics, etc.). It implies that various models from these manufacturers were tested for compatibility.
    • Data provenance: Not explicitly stated. The manufacturer is "NOxBOX Ltd.", suggesting the testing could have occurred in the UK or other locations where the manufacturer operates. The submission correspondent is in Austin, Texas, USA. Given the context of FDA submission, the data would need to be acceptable to the US regulatory body, but the origin itself isn't specified.
    • Retrospective or prospective: Not applicable for this type of non-clinical engineering and validation testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the submission describes non-clinical engineering and validation testing, not a study involving expert review of cases for ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable as the submission describes non-clinical engineering and validation testing, not a study involving expert review of cases for ground truth establishment.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The NOxBOXi Nitric Oxide Delivery System is a medical device for delivering and monitoring nitric oxide, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable in the context of an "algorithm only" performance. The device is hardware with integrated software for delivery and monitoring, not an AI algorithm performing a standalone task. Its performance is inherent to its design and validated through engineering tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance evaluation, the "ground truth" would be established by:

    • Reference standards/equipment: Calibration gases, calibrated flow meters, pressure sensors, and gas analyzers that provide highly accurate measurements of NO, NO2, and O2 concentrations, flow rates, and pressures.
    • Engineering specifications: The device's design specifications for accuracy, delivery ranges, and alarm thresholds serve as the targets for "ground truth".
    • Standardized test methods: Compliance with international standards (e.g., ISO 80601-2-55 for respiratory gas monitors) and FDA guidance documents implies validated test methodologies where the "true" values are known or traceable to national/international standards.

    8. The sample size for the training set

    This section is not applicable. The document does not describe a machine learning algorithm that requires a "training set" of data. The device's functionality is based on established engineering principles and calibrated sensors/actuators.

    9. How the ground truth for the training set was established

    This section is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171696

    Validate with FDA (Live)

    Date Cleared
    2018-10-02

    (482 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).

    The NOxBOXi Nitric Oxide Delivery System Includes:

    • · The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.
    • · Continuous monitoring and alarms for NO, O2, and NO2.
    • The integrated NOxMixer, which provides a backup NO delivery capability that is intended to deliver a continuous flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer conditions within hospitals.

    The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging (currently neonates). Refer to this material prior to use.

    Device Description

    The NOxBOXi Nitric Oxide Delivery System (NOxBOXi) simultaneously delivers Nitric Oxide (NO) medical gas, while monitoring Nitric Oxide, Nitrogen Dioxide and Oxygen levels in the inspiratory limb of a ventilator for patients undergoing inhaled Nitric Oxide Therapy.

    The system is designed for use by healthcare professionals to administer treatment to patients undergoing inhaled Nitric Oxide (iNO) therapy. The NOxBOXi will deliver nitric oxide in a synchronous manner to a single patient.

    An integrated component to the NOxBOXi, the NOxMixer is intended to deliver a continuous flow of Nitric Oxide from the NOxBOXi, mixed in line with Oxygen (O₂) for use in iNO therapy. The NOxMixer will be used in conjunction with manually bagging a patient.

    The NOxBOXi includes the NOxBOXi Head Unit, a NOxFLOW sample line, two NO feed hoses, two regulators (connector type dependent on the gas supplier), a test circuit, NO, O₂ & NO2 monitors, power supply, drainage syringe, Operating Manual & Technical Guide.

    Optional accessories include 4 separate NOxKITs (22mm, 12 mm, & 10mm), one way valve for HFOV (High Frequency Oscillatory Ventilation), bagging kits (hyperinflation & selfinflating) and circuit reducers (22f – 15m).

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the NOxBOXi Nitric Oxide Delivery System, based on the provided text:

    Important Note: The provided document is a 510(k) summary for a medical device. This type of submission relies on demonstrating "substantial equivalence" to a predicate device, rather than proving efficacy and safety through extensive clinical trials as seen with PMAs. Therefore, most of the studies mentioned are non-clinical (bench testing, software validation, usability testing), and a multi-reader multi-case (MRMC) comparative effectiveness study or standalone algorithm-only performance is not applicable in this context. The acceptance criteria are largely based on engineering and safety standards rather than clinical performance metrics like sensitivity and specificity.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) summary for a device delivering and monitoring a gas, the "acceptance criteria" and "device performance" are primarily related to the accuracy of gas delivery, monitoring, and safety standards.

    Acceptance Criteria / Performance AspectReported Device PerformanceComments
    NOxBOXi (General)
    Indication for UseDelivers and monitors constant user-set NO concentration, monitors NO2 and O2 in inspiratory ventilator lines.Directly meets the stated indications for use.
    Compliance with Regulatory StandardsPassed electrical safety, EMC, biocompatibility, and software validation.Complies with ISO 60601-1, ISO 60601-1-2, ISO 80601-2-55, ISO 10993-1, ISO 62304, ISO 62366.
    Battery Backup Capability4 hours without AC power.Different from predicate (6 hours), but deemed acceptable for emergency backup situations due to hospital generator availability.
    Pre-use Setup TimeInstant set-up.Reduced from predicate (4-6 mins), indicating improved usability.
    NO/NO2/O2 Monitoring & Delivery
    NO concentration provided0.0 to 80 ppmIdentical to predicate.
    NO monitorYesIdentical to predicate.
    O2 monitorYesIdentical to predicate.
    NO & NO2 Monitoring Accuracy+/- 2% or 0.2 ppm (whichever is greater)Improved compared to predicate (+/-20% or 2ppm). The NOxBOXi uses a closed-loop system for automatic adjustment.
    NO2 monitor & alarmYesIdentical to predicate.
    NO flow rate (sample flow rate)225 ml/minMinor difference from predicate (250 ml/min), but still meets measuring accuracy requirements.
    Purity of NO drug deliveryTestedAcceptable.
    Acceptable/minimal production of NO2TestedAcceptable.
    NO analyzer accuracyTestedAcceptable (as above: +/- 2% or 0.2 ppm).
    NO2 analyzer accuracyTestedAcceptable (as above: +/- 2% or 0.2 ppm).
    Compatibility testing of ventilatorsTestedCompatible with listed ventilators.
    NOxMixer (Backup / Manual Mode)
    Manual bagging & backup systemNOxMIXER®Integrated with NOxBOXi system, provides a continuous treatment option during transit and transfer.
    NO dosing range in manual mode0 - 185 ppm (800 ppm cylinders)Expanded user-selectable dose range compared to predicate (5-80ppm or 2.5-40ppm).
    NO dosing accuracy in manual mode+/- 20% or 2 ppm (for 5-80 ppm, 800 ppm cylinder, and 5-14 L/min O2 flow); +/- 40% or 4 ppm (for 0 to <5 ppm or >80 to 185 ppm, or 2 to <5 L/min or >14 to 25 L/min O2 flow)Similar accuracy for the limited range of the predicate, but with an expanded user-selectable dose range.
    NO flow in manual modeAdjustable 50 – 600 mL/min of NO/N2User adjustable flow range; predicate was fixed 250 mL/min.
    O2 flow range in manual bagging mode2 to 25 L/min of O2Expanded user-selectable flow range compared to predicate (5 to 14 L/min).
    Oxygen inlet pressure3.5 – 4.5 barDifferent from predicate (3-5 bar), but deemed substantially equivalent as this pressure is typically available in clinical settings.
    NO delivery pressure1.65 bar from manual control valveDifferent from predicate (1.7 to 2.4 bar), but deemed substantially equivalent for neonates.
    Monitoring during manual baggingYesEquivalent to predicate.
    Alarms active during baggingYesEquivalent to predicate.
    Backup accuracy+/- 20% or 2 ppm (for 5-80 ppm, 800 ppm cylinder); +/- 40% or 4 ppm (for 0 to <5 ppm or >80 to 185 ppm, 800 ppm cylinder)Equivalent to predicate.
    Volatile organic compounds (VOC)Three orders of magnitude below OSHA permissible exposure levels.Acceptable.
    Particulate matterWell below EPA's maximum limits for total suspended particulates.Acceptable.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated in terms of patient numbers or specific data points. The testing described focuses on device performance, electrical safety, EMC, software validation, biocompatibility, and usability, which are typically performed on the device itself and its components, or with simulated data/environments.
    • Data Provenance: The studies are primarily non-clinical bench testing and simulations. No patient data (retrospective or prospective) is mentioned as being used for the substantial equivalence determination for the device's technical performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. The "ground truth" for non-clinical performance and safety tests is established by published engineering standards (e.g., ISO, IEC), regulatory guidelines, and internal specifications, rather than expert interpretation of clinical data.
    • Qualifications of Experts: N/A for establishing ground truth in this context. Test engineers and qualified personnel would have performed and validated the tests according to established protocols.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Adjudication methods (like 2+1 or 3+1 consensus) are typically used in clinical studies where multiple human readers interpret medical images or data and their interpretations need to be reconciled to form a robust "ground truth." For non-clinical device testing, results are usually compared directly against pre-defined performance specifications from recognized standards or internal design requirements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images, and the goal is to assess how a new technology (e.g., AI) impacts their diagnostic performance. The NOxBOXi is a gas delivery and monitoring system, not an imaging diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance Study: Yes, in essence, the "non-clinical performance data" and "software verification and validation" sections describe standalone testing of the device's functional integrity. The device's ability to precisely deliver and monitor gases according to its specifications, and the functionality of its software, were tested independently of human intervention during the gas delivery process after initial user setup. However, it's important to differentiate this from "standalone AI performance" often discussed for diagnostic AI, as this device primarily involves control systems and sensor-based monitoring rather than complex AI interpretation.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: The ground truth for the non-clinical tests involved:
      • Engineering Standards and Specifications: For electrical safety (ISO 60601-1), EMC (ISO 60601-1-2), respiratory gas monitors (ISO 80601-2-55), biocompatibility (ISO 10993-1), and software (FDA Guidance, ISO 62304).
      • Reference Measurements/Calibrated Equipment: For accuracy of NO delivery, NO/NO2 analyzer accuracy, purity of gas, and NO2 production.
      • Usability Feedback: For usability testing (ISO 62366, FDA guidance), user feedback identifies issues and informs design improvements, rather than establishing a "truth" about a medical condition.
      • Predicate Device Performance: Used as a benchmark for comparison to demonstrate substantial equivalence, rather than strict "ground truth."

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is a hardware system with embedded software, not a machine learning or AI model that requires a "training set" in the conventional sense of data-driven learning. Its operating parameters are based on physical laws, engineering principles, and pre-programmed algorithms, not learned from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no "training set" for an AI model. The "ground truth" for the device's design and operational parameters is established through scientific principles, engineering design, and adherence to relevant national and international standards for medical gas delivery and monitoring systems.