Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K052526

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-15

    (62 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Novus 3000 Laser System is intended for use in the treatment of ocular pathology. The Novus 3000 Laser System is indicated for use in photocoagulation of both anterior and posterior segments including:

    • Retinal photocoagulation, panretinal photocoagulation and intravitreal endophotocoagulation of vascular and structural abnormalities of the retina and choroid including:
    • proliferative and nonproliferative diabetic retinopathy;
    • choroidal neovascularization;
    • branch retinal vein occlusion;
    • age-related macular degeneration;
    • retinal tears, detachments;
    • retinopathy of prematurity.
    • Iridotomy, iridectomy and trabeculoplasty in angle closure glaucoma and open angle glaucoma.

    Laser Indirect Ophthalmoscope
    The Laser Indirect Ophthalmoscope is indicated for the delivery of laser energy in eyes with retinal pathology. The Laser Indirect Ophthalmoscope is indicated for use in the following ophthalmic treatments and conditions: diabetic retinopathy (panretinal photocoagulation); peripheral neovascularization, retinal breaks, detachments, and tears, lattice degeneration, pneumatic retinopexy reattachment procedures, segmental peripheral photocoagulation; segmental photocoagulation; cloudy vitreous cavities; pediatric retinal repairs (under general anesthesia), delivery of laser energy through small pupils or to eyes with focal lens opacities.

    Endophotocoagulation
    When used with a Novus 3000 Laser System, the Acculite EndoOcular Probes (Acculite EndoOcular Probe, Acculite Aspirating EndoOcular Probe, and Acculite Illuminating EndoOcular Probe) are intended for use in the following ophthalmic applications: photocoagulation of the anterior and posterior segment, including: anterior segment treatment in the surgical management of glaucoma; endophotocoagulation in vitreoretinal surgery, including panretinal photocoagulation, retinopexy, and treatment of neovascularization.

    The laser delivery function of the Acculite Acculite EndoOcular Probes (Acculite EndoOcular Probe, Acculite Aspirating EndoOcular Probe and Acculite Illuminating EndoOcular Probe) is indicated for use in ocular surgery to deliver laser energy to the treatment area selected by the surgeon.

    The aspiration function of the Acculite Aspirating EndoOcular Probe is indicated for use when unwanted fluid is present in the eye during ocular surgery, causing refraction or scattering of the laser beam from the intended treatment site.

    The illumination function of the Acculite Illuminating EndoOcular Probe is indicated for use during ocular surgery to illuminate the interior of the eye.

    Device Description

    Lumenis, Inc.'s Novus® 3000 Laser System with the Delivery Devices is an air cooled, diode-pumped, solid state, Nd: YAG Laser System, which produces a wavelength of 532 um of laser light with a treatment beam output ranging from 50 mw to 3.0 W. The main parts of the Novus® 3000 system include the laser console, a footswitch, a remote control and assorted laser accessories.

    AI/ML Overview

    This 510(k) summary for the Lumenis Novus® 3000 Laser System and Delivery Devices does not include any performance data or a study to demonstrate that the device meets specific acceptance criteria.

    The document explicitly states:

    "No performance data is required for this Class II device nor requested by the Food and Drug Administration (Office of Device Evaluation). No data was submitted for section 807.92 6[(b)(1)(2)(3c)]."

    Therefore, I cannot provide the requested information. The submission relies on demonstrating substantial equivalence to predicate devices based on similarities in intended use, indications for use, and technological characteristics, rather than on new performance data for this specific device.

    If this were a submission that included a performance study, the requested information would typically be provided as follows:


    Hypothetical Example if performance data were submitted:

    Here's how the information would be presented if a study was conducted and acceptance criteria were defined, based on a typical medical device submission structure:

    1. Table of Acceptance Criteria and Reported Device Performance (Hypothetical Example)

    Performance MetricAcceptance CriteriaReported Device Performance
    Laser Output Power AccuracyWithin ±10% of set value±5% (e.g., for 50 mW to 3.0 W range)
    Beam Spot Size AccuracyWithin ±15% of nominal size±10% (e.g., for various delivery devices)
    Pulse Duration AccuracyWithin ±5% of set value±2% (e.g., for 10 ms to 1000 ms range)
    Aiming Beam Co-axialityWithin 0.5 mm relative to treatment beam<0.2 mm
    Thermal Performance (e.g., temperature rise of handpiece)Below X°C for Y continuous operationZ°C (well within limits)

    2. Sample Size Used for the Test Set and Data Provenance (Hypothetical Example)

    • Sample Size for Test Set: E.g., 5 units of the Novus® 3000 Laser System and a representative sample of each delivery device (e.g., 10 Laser Indirect Ophthalmoscopes, 15 Endophotocoagulation Probes).
    • Data Provenance: Prospective, collected at the manufacturer's facility in Santa Clara, CA, USA, under controlled laboratory conditions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications (Hypothetical Example)

    This type of assessment would typically involve engineering and quality control experts rather than clinical experts for ground truth establishment.

    • Number of Experts: 3-5 Engineers and Quality Control personnel.
    • Qualifications: E.g., Lead Optical Engineer (15 years experience in laser systems), Senior Electrical Engineer (10 years experience in medical device electronics), Quality Assurance Manager (12 years experience in medical device testing and regulatory compliance).

    4. Adjudication Method for the Test Set (Hypothetical Example)

    • Adjudication Method: For quantitative measurements, results would be recorded by trained technicians/engineers. Any discrepancies or outliers would be investigated and resolved by the lead engineer and QA manager through re-testing or thorough review of methodology. For subjective assessments (e.g., ergonomic evaluation), a consensus approach among the expert panel might be used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study (Hypothetical Example)

    • Given this is a physical laser device, an MRMC study related to interpretation of results or diagnostic performance (like for an imaging AI) is not applicable. The device's direct performance is based on its physical properties and output.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop) (Hypothetical Example)

    • Again, this term is more relevant for AI/software devices. For a hardware device like a laser, the "standalone performance" is exactly what the hypothetical performance table above would describe – the device's intrinsic operational metrics without a human actively making real-time adjustments for the test. The clinician is always "in the loop" during actual treatment.

    7. Type of Ground Truth Used (Hypothetical Example)

    • Ground Truth: Primarily based on metrology standards and established physical principles. This would involve using calibrated measurement equipment (e.g., power meters, beam profilers, oscilloscopes) traceable to national standards to objectively quantify the laser's output characteristics against its design specifications.

    8. Sample Size for the Training Set (Hypothetical Example)

    • For a hardware device, the concept of a "training set" is usually not applicable in the same way it is for AI/machine learning algorithms. Device design and manufacturing processes are refined through iterative testing and quality control during development. The "training" for such a device comes from engineering principles, material science, and prior models.

    9. How the Ground Truth for the Training Set Was Established (Hypothetical Example)

    • As above, not directly applicable in the AI sense. Ground truth in the development phase would be established through engineering specifications, design verification, and validation testing against intended performance requirements, using calibrated equipment and standard test methods.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1