510(k) Data Aggregation

    K Number
    K033270

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-02-13

    (126 days)

    Product Code
    Regulation Number
    870.4360
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovoSci ™ Ready System® is indicated for use in surgical procedures requiring extracorporeal circulation and gas exchange support for 6 hours or less.

    Device Description

    The NovoSci" Ready System® is a closed loop extracorporeal system providing circulatory, thermal and gas-exchange support for extracorporeal perfusion. The NovoSci" Ready System® is provided as a sterile system with a non-pyrogenic fluid pathway for single use only and is not to be re-sterilized by the user.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the NovoSci™ Ready System®, an extracorporeal system. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for a new, innovative device. Therefore, much of the requested information regarding AI model performance, human reader studies, and detailed ground truth establishment is not applicable to this submission.

    Here's a breakdown of the information that can be extracted and explanations for what cannot:

    1. A table of acceptance criteria and the reported device performance

    The submission does not explicitly list "acceptance criteria" in the way one would for an AI algorithm's performance metrics (e.g., sensitivity, specificity, AUC). Instead, it relies on demonstrating substantial equivalence to a predicate device through non-clinical testing.

    Acceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance (Non-clinical testing)
    Biocompatibility: Equivalent safety profile regarding interactions with biological systems."Biocompatibility...testing were performed to demonstrate equivalence."
    In-vitro performance: Safe and effective operation in laboratory settings."in-vitro...testing were performed to demonstrate equivalence."
    Performance testing (Mechanical/Functional): Meets functional requirements for extracorporeal circulation and gas exchange."performance testing were performed to demonstrate equivalence. These tests included: blood damage, reservoir pressure drop, minimum operating volume, micro air, vacuum pressure, flow rate vs. vacuum."
    Duration of Use: Capable of supporting extracorporeal circulation and gas exchange for 6 hours or less."The NovoSci™ Ready System® is indicated for use in surgical procedures requiring extracorporeal circulation and gas exchange support for 6 hours or less." (This is an indication for use, which implies performance within this timeframe.)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the given text. Non-clinical testing typically involves bench-top tests, simulations, and material analyses rather than patient-derived "test sets" in the context of an AI device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable/not provided. Ground truth establishment by experts is relevant for diagnostic AI devices where human interpretation is being benchmarked. For an extracorporeal circulation system, "ground truth" is typically defined by engineering specifications, validated test methods, and compliance with industry standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable/not provided. Adjudication methods are used in studies involving human interpretation or agreement, typically for diagnostic accuracy studies of AI.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/not provided. The NovoSci™ Ready System® is a medical device for extracorporeal circulation, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study or AI assistance metrics are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/not provided. As explained above, this is not an AI device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For this type of device, the "ground truth" for non-clinical testing would involve:

    • Engineering specifications and standards: Performance parameters like flow rate, pressure drop, and blood damage are compared against pre-defined, acceptable ranges established through engineering principles and regulatory guidelines for similar devices.
    • Predicate device performance data: The NovoSci™ Ready System®'s performance is compared against the known, established performance of the legally marketed predicate devices (Jostra MECC System, V-bag, V-box). This acts as a de-facto "ground truth" for what constitutes safe and effective performance.

    8. The sample size for the training set

    This information is not applicable/not provided. This is not an AI device, so there is no "training set."

    9. How the ground truth for the training set was established

    This information is not applicable/not provided. As not an AI device, there is no "training set" or ground truth established for it in that context.