Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K051291

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-09-20

    (125 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The medical/surgical masks listed below are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of microorganism, body fluid, and particulate aerosol transfer.

    Device Description

    The new surgical masks are pleated 3-ply masks. The outer layers are made with 100% spun-bound polypropylene (SBPP). The filter media is composed of 100% melt-blown polypropylene (MBPP). The inner layer is made of either 100% medical grade tissue paper or 100% SBPP. The ear loops are made of flat latex and fiberglass free elastic. The nosepieces are made of malleable aluminum wire. All of the materials used in the construction of the new masks are being used in currently marketed devices.

    AI/ML Overview

    The device in question is a Non-Sterile Surgical Mask manufactured by A.R. Medicom Inc. The filing is a 510(k) Summary, indicating a demonstration of substantial equivalence to predicate devices, rather than a de novo clinical study with specific acceptance criteria in the traditional sense of AI/medical imaging device performance.

    Therefore, the "acceptance criteria" for a surgical mask are typically based on recognized performance standards, which in this case is ASTM F2100-04, "Standard Specification for Performance of Materials Used in Medical Face Masks." The "study" that proves the device meets these criteria would involve various material and barrier property tests, not a clinical trial with human readers or AI.

    Given the nature of the device (a surgical mask) and the provided document (a 510(k) Summary for a medical device cleared in 2005), many of the requested categories (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, AI standalone performance, training set details) are not applicable as they pertain to the evaluation of AI/software as a medical device or diagnostic imaging devices, not a physical barrier device like a surgical mask.

    However, I will address the relevant aspects based on the information provided:


    1. Table of Acceptance Criteria and Reported Device Performance

    For surgical masks, "acceptance criteria" and "reported device performance" are typically defined by recognized consensus standards. In this case, ASTM F2100-04 is listed as the recognized performance standard. This standard categorizes masks into different levels based on their performance in a series of tests. While the exact performance levels for this specific mask are not detailed in the provided 510(k) summary (only the standard itself is mentioned as being applicable), the various tests and their criteria from ASTM F2100-04 are implied to be the 'acceptance criteria'.

    The 510(k) summary states that the device is "substantially equivalent" to predicate devices K955513 and K040333, implying it meets similar performance standards. The device's description (3-ply, spun-bound polypropylene outer layers, melt-blown polypropylene filter media, medical grade tissue paper or SBPP inner layer) strongly suggests the materials are chosen to meet these performance characteristics.

    Test Parameter from ASTM F2100-04 (Implied Acceptance Criteria)Reported Device Performance (Implied from 510(k) based on substantial equivalence to predicate and meeting ASTM F2100-04)
    Bacterial Filtration Efficiency (BFE)Not explicitly stated, but expected to meet ASTM F2100-04 specified level (e.g., ≥95% or ≥98%)
    Particle Filtration Efficiency (PFE)Not explicitly stated, but expected to meet ASTM F2100-04 specified level (e.g., ≥95% or ≥98% at 0.1 micron)
    Fluid ResistanceNot explicitly stated, but expected to meet ASTM F2100-04 specified level (e.g., 80, 120, or 160 mmHg)
    Differential Pressure (Breathability)Not explicitly stated, but expected to meet ASTM F2100-04 specified level (e.g., <5.0, <6.0 mm H2O/cm²)
    FlammabilityNot explicitly stated, but expected to meet ASTM F2100-04 specified level (e.g., Class 1)

    Note: The 510(k) summary explicitly refers to "ASTM 2100-04" as a "Recognized Performance Standard" and states that "Refer to Appendix B for a list of applicable standards." This indicates that the device's performance was evaluated against the requirements of this specification, and thus it meets the stated performance levels defined within ASTM 2100-04 for surgical masks. Without Appendix B or the full testing report, the exact numerical performance values (e.g., a specific BFE percentage) are not detailed in this summary.


    Study that Proves the Device Meets the Acceptance Criteria:

    The filing is a 510(k) summary, which indicates a submission to demonstrate substantial equivalence to existing legally marketed devices, rather than a novel study designed to establish acceptance criteria for an entirely new technology. The "study" here refers to the collection of data and tests (likely in-vitro and material property tests) performed to show that the surgical mask complies with the recognized performance standard ASTM F2100-04 and is substantially equivalent to the predicate devices.

    The evidence for meeting acceptance criteria is typically derived from laboratory testing of the mask materials and design according to the methods prescribed by the ASTM F2100-04 standard. These tests would evaluate parameters such as bacterial filtration efficiency, particle filtration efficiency, fluid resistance, breathability, and flammability. The results of these tests, when compared against the thresholds defined in ASTM F2100-04, demonstrate compliance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the 510(k) summary. For material performance tests like those required for surgical masks, sample sizes are typically determined by the testing methodology outlined in the relevant ASTM or ISO standards to ensure statistical validity for the specific tests (e.g., multiple masks or multiple samples from a mask might be tested for each parameter).
    • Data Provenance: Not explicitly stated, but the manufacturer is A.R. Medicom Inc., located in Lachine, Quebec, Canada. Testing would likely be conducted at certified laboratories, either internal or third-party, which may be in Canada or elsewhere. The nature of these tests is not "retrospective" or "prospective" in the clinical trial sense, but rather laboratory-based characterization.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For a physical barrier device like a surgical mask, "ground truth" is not established by clinical experts in the way it is for diagnostic imaging or AI devices. The "ground truth" is based on objective, quantifiable measurements from standardized laboratory tests (e.g., the measured bacterial filtration efficiency percentage). The "experts" involved would be laboratory technicians and engineers experienced in conducting ISO/ASTM standard tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 refer to human expert consensus for interpreting medical images or clinical data. This is not relevant for the performance testing of a surgical mask.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are used to evaluate the diagnostic performance of human readers, often with and without AI assistance, typically in medical imaging. This is not relevant for a surgical mask.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This refers to the evaluation of an AI algorithm's performance without human interaction. This device is a physical product, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for a surgical mask's performance is established by objective measurements from physical property tests conducted according to recognized standards (e.g., ASTM F2100-04). Examples include:

    • Bacterial Filtration Efficiency (measured percentage)
    • Particle Filtration Efficiency (measured percentage)
    • Fluid penetration resistance (measured pressure in mmHg)
    • Differential pressure (measured value in mm H2O/cm²)
    • Flammability (classification per standard test)

    These are empirical facts determined by standardized test methods, not subjective consensus or clinical outcomes from patients.

    8. The sample size for the training set

    Not applicable. This device is a physical product, not an AI model that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As above, there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1