Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K013288

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-12-03

    (62 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient.

    Device Description

    Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM-D3578-00 Standard Specification for Latex Examination Gloves for Medical Application.

    AI/ML Overview

    The provided document is a 510(k) summary for "NON-STERILE POWDER FREE GREEN LATEX EXAM GLOVES WITH ALOE VERA & VITAMINE" and does not describe an AI medical device. The information requested in the prompt (acceptance criteria, study details, expert involvement, etc.) is typically relevant for AI/ML-based diagnostic or prognostic devices, not for a medical glove.

    Therefore, I cannot provide the information requested in the format of the prompt, as it is not applicable to the given document.

    The document describes the performance characteristics of medical gloves against ASTM standards and FDA requirements. These are not "AI device performance metrics" in the sense of the prompt.

    However, I can extract the closest analogous information from the document as it relates to the glove's performance against established criteria:


    1. Table of acceptance criteria and the reported device performance:

    CharacteristicAcceptance Criteria (ASTM D3578-00 or FDA)Reported Device Performance (SGMP)
    Dimensions (Width)
    X-Small70 mm +/- 10 mm70 - 75 mm
    Small80 mm +/- 10 mm80 - 85 mm
    Medium95 mm +/- 10 mm90 - 97 mm
    Large111 mm +/- 10 mm105 - 111 mm
    Dimensions (Length)230 mm minimum252 mm
    Thickness
    Finger0.08 mm min0.08 mm min
    Palm0.08 mm min0.08 mm min
    Tensile Strength (Before Aging)
    X-Small14.0 Mpa25.8 Mpa
    Small14.0 Mpa27.8 Mpa
    Medium14.0 Mpa27.9 Mpa
    Large14.0 Mpa27.1 Mpa
    Tensile Strength (After Aging)
    X-Small14.0 Mpa24.1 Mpa
    Small14.0 Mpa27.0 Mpa
    Medium14.0 Mpa28.7 Mpa
    Large14.0 Mpa25.6 Mpa
    Ultimate Elongation (Before Aging)
    X-Small700 %830 %
    Small700 %920 %
    Medium700 %900 %
    Large700 %950 %
    Ultimate Elongation (After Aging)
    X-Small500 %950 %
    Small500 %910 %
    Medium500 %910 %
    Large500 %910 %
    Water Leak Test2.5% AQL (Acceptance Quality Limit)Leaked: X-Small (1), Small (0), Medium (1), Large (2) for UN-AGED. X-Small (0), Small (1), Medium (2), Large (0) for AGED. All within 2.5% AQL.
    BiocompatibilityPassed tests for examination glovesPassed
    Residual Powder Content2 mg/glove max (FDA Internal Requirement)Range: 0.5-0.8 mg/glove; Mean: 0.63 mg/glove
    Presence of CornstarchNegativeNegative
    Residual Protein Level< 50 µg/g (FDA Allowable Level)< 50 µg/g

    The remaining points (2-9) are specific to AI device studies and are not applicable to this 510(k) submission for latex examination gloves, which focuses on physical and chemical properties and safety for an inert medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1