Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K974829

    Validate with FDA (Live)

    Date Cleared
    1998-02-12

    (50 days)

    Product Code
    Regulation Number
    880.5965
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INFUSION OF FLUIDS, INCLUDING THOSE CONTAINING MEDICATIONS, FROM A CONTAINER TO A PORT IMPLANTED IN, OR OTHERWISE ATTACHED TO, A PATIENT TO AID IN THE DIAGNOSIS OR TREATMENT OF DISEASE OR OTHER CONDITIONS.

    Device Description

    A NON-PVC PORT ACCESS INFUSION SET IS A STERILE, SINGLE USE DEVICE WITH A NEEDLE, WING, NON-PVC I.D. TUBING AND NON-PVC COMPONENTS USED FOR ACCESSING IMPLANTED MEDICATION PORTS.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the device K974829:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Biocompatibility requirementsDevice meets all biocompatibility test requirements.
    Pyrogenicity requirementsDevice meets all pyrogenicity test requirements.
    Limulus Amebocyte Lysate ("LAL") Test (USP XXII)Device meets the LAL test described at "(85) Bacterial Endotoxins Test", USP XXII, pp 1493-1495.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not contain information regarding the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). The document only mentions that Kawasumi Laboratories "conducted biocompatibility tests on the body fluid contacting material portions of the device."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The provided text does not contain information regarding the number of experts used to establish ground truth or their qualifications. The studies mentioned are laboratory tests (biocompatibility, LAL) and do not appear to involve expert-driven ground truth establishment in a clinical context.

    4. Adjudication Method for the Test Set

    The provided text does not contain information regarding an adjudication method. The tests described are objective laboratory tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not conducted. The document focuses on demonstrating substantial equivalence to a predicate device based on material characteristics and basic safety tests, not on improving human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    The device is an "Infusion Set," which is a physical medical device, not an algorithm or AI system. Therefore, a standalone algorithm study is not applicable and was not performed.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance was established through objective laboratory tests, specifically:

    • Biocompatibility tests: These typically involve in-vitro and sometimes in-vivo toxicology assessments against established ISO standards or similar guidelines.
    • Pyrogenicity testing: This refers to the LAL test, which is a standardized method to detect bacterial endotoxins.

    8. The Sample Size for the Training Set

    The provided text does not contain information regarding a training set sample size. As the device is not an AI/ML algorithm, the concept of a "training set" is not applicable in this context.

    9. How the Ground Truth for the Training Set Was Established

    As the device is not an AI/ML algorithm, the concept of a "training set" and establishing ground truth for it is not applicable. The "ground truth" for the device's safety and performance was established through the aforementioned laboratory tests on the device materials.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1