510(k) Data Aggregation

    K Number
    K250578

    Validate with FDA (Live)

    Date Cleared
    2025-04-01

    (33 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Non-Sterile Nitrile Patient Examination Glove is made from synthetic rubber latex. It is single use and powder-free class I Nitrile Patient Examination Gloves which coated by on-line polymer with mild on-line chlorination process. These processes modify the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder. The available sizes are X-Small, Small, Medium, Large and X-Large. The Nitrile Examination Gloves Powder Free (Blue & Black) have been tested for Chemotherapy Drugs. The subject devices are identical in design and formulation to the predicate gloves of K230121.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to Nitrile Examination Gloves Powder Free Tested for Use With Chemotherapy Drugs & Fentanyl Citrate (Blue & Black), a medical device. This type of device is classified as a Class I medical device and does not involve AI/ML technology. Therefore, many of the requested categories related to AI/ML device studies (e.g., sample sizes for test/training sets, ground truth establishment, MRMC studies, standalone performance, expert qualifications, adjudication methods, effect sizes) are not applicable.

    The acceptance criteria and device performance are primarily evaluated based on material properties, chemical permeation resistance, and biocompatibility, using established ASTM and ISO standards for examination gloves.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device, "NITRILE EXAMINATION GLOVES POWDER FREE TESTED FOR USE WITH CHEMOTHERAPY DRUGS & FENTANYL CITRATE (BLUE & BLACK)", was tested against various physical, chemical, and biological criteria.

    Test CategoryCharacteristic / StandardAcceptance CriteriaReported Device Performance (Result)
    PreambleProduct CodeLZA, LZC, QDO, OPJLZA, LZC, QDO, OPJ
    Intended UseTo prevent contamination between patient and examinerMeets stated intended use
    DesignPowder Free, Non-Sterile, Ambidextrous, Beaded CuffPowder Free, Non-Sterile, Ambidextrous, Beaded Cuff
    ConstructionAmbidextrous, Polymer Coated or Chlorinated, Powder Free NitrileAmbidextrous, Polymer Coated or Chlorinated, Powder Free Nitrile
    Color DescriptionBlue & BlackBlue & Black
    MaterialNitrileNitrile
    Single UseYesYes
    PackagingPacked in Dispenser BoxesPacked in Dispenser Boxes
    SterilityNon-SterileNon-Sterile
    Freedom From HolesASTM D6319-19, ASTM D5151-19Meet requirement inspection level G-1, AQL 2.5Pass
    DimensionASTM D6319-19Length (XS), min. 220mmMeet 220mm min
    Length (S, M, L, XL), min. 230mmMeet 230mm min
    Thickness (palm/finger), min. 0.05mmMeet 0.05mm min
    Width (XS), 70 ± 10 mmMeet 70 ± 10 mm
    Width (S), 80 ± 10 mmMeet 80 ± 10 mm
    Width (M), 95 ± 10 mmMeet 95 ± 10 mm
    Width (L), 110 ± 10 mmMeet 110 ± 10 mm
    Width (XL), 120 ± 10 mmMeet 120 ± 10 mm
    Physical Properties (Before Ageing)ASTM D6319-19Tensile Strength, 14 MPa min.Meet 14MPa min.
    Ultimate Elongation, 500% min.Meet 500% min.
    Physical Properties (After Accelerated Aging)ASTM D6319-19Tensile Strength, 14 MPa min.Meet 14MPa min.
    Ultimate Elongation, 400% min.Meet 400% min.
    Residual Powder ContentASTM D6319-19, ASTM D6124-06Not more than 2 mg per gloveMeet 2mg/glove max.
    Biocompatibility - Primary Skin IrritationISO 10993-10Not a primary skin irritantPass (did not cause an irritant response)
    Biocompatibility - Skin SensitizationISO 10993-10Not a contact sensitizerPass (did not produce a skin sensitization effect)
    Biocompatibility - In Vitro CytotoxicityISO 10993-5:2009No adverse biological reactionModerate cytotoxicity reactivity at 6.0 cm²/mL extract concentrations and no cytotoxicity reactivity at 3.0 cm²/mL extract concentrations.
    Biocompatibility - Acute Systemic ToxicityISO 10993-11No adverse biological reactionPass (no adverse biological reaction)
    Chemotherapy Drug PermeationASTM D6978-05 (Reapproved 2019)Varies by drug (Breakthrough Detection Time)See detailed table below

    Detailed Chemotherapy Drug and Fentanyl Citrate Permeation Performance:

    Chemotherapy Drugs and Fentanyl CitrateConcentrationMinimum Breakthrough Detection Time (minutes) - Blue GloveMinimum Breakthrough Detection Time (minutes) - Black Glove
    *Carmustine (BCNU)3.3 mg/ml22.621.8
    Cisplatin1.0 mg/ml>240 min>240 min
    Cyclophosphamide (Cytoxan)20.0 mg/ml>240 min>240 min
    Dacarbazine10.0 mg/ml>240 min>240 min
    Doxorubicin HCL2.0 mg/ml>240 min>240 min
    Etoposide20.0 mg/ml>240 min>240 min
    Fluorouracil50.0 mg/ml>240 min>240 min
    Ifosfamide50.0 mg/ml>240 min>240 min
    Mitoxantrone2 mg/ml>240 min>240 min
    Paclitaxel6.0 mg/ml>240 min>240 min
    *Thio Tepa10.0 mg/ml43.917.7
    Vincristine Sulfate1.0 mg/ml>240 min>240 min
    Methotrexate25.0 mg/ml>240 min>240 min
    Mitomycin C.0.5 mg/ml>240 min>240 min
    Fentanyl Citrate100mcg/2ml>240 min>240 min
    *Warning: Do not use with Carmustine and Thiotepa due to extremely low permeation times.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes for each test (e.g., number of gloves tested for permeation, holes, dimensions, etc.) but refers to adherence to the methodologies outlined in the specified ASTM and ISO standards. These standards typically define minimum sample sizes for material testing.

    • Data Provenance: The document states the applicant is from Malaysia (Careglove Global Sdn. Bhd.). The testing was conducted according to international standards (ASTM, ISO), implying the tests were performed in a controlled laboratory environment, likely by the manufacturer or a contracted lab. The data is non-clinical and derived from prospective testing of manufactured glove samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is not an AI/ML device that requires expert review for "ground truth" establishment in the context of diagnostic or interpretive tasks. The "ground truth" for material properties is established by the physical and chemical measurements themselves, according to standardized protocols and certified laboratory equipment.

    4. Adjudication Method for the Test Set

    Not applicable. This device is not an AI/ML device that requires adjudication for interpretations. Test results are objective measurements against predefined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    Not applicable. MRMC studies are relevant for diagnostic or interpretive AI/ML devices where human readers' performance is evaluated with and without AI assistance. This is a physical and chemical barrier device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    Not applicable. This device is a passive physical barrier and does not involve any algorithms or software.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on objective measurements of physical properties (e.g., length, thickness, tensile strength, elongation), chemical resistance (breakthrough detection time), and biological reactivity (biocompatibility tests), as defined by internationally recognized standards (ASTM, ISO). There is no subjective human "ground truth" in this context.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that undergoes "training."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device. The manufacturing process of the gloves is designed to meet the specified standards, and quality control ensures consistency, but this is not "training" in the context of AI/ML.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242812

    Validate with FDA (Live)

    Date Cleared
    2024-10-12

    (24 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the exammer's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes, and can be used with the device.

    Cyclophosphamide (20.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Paclitaxel (6.0 mg/ml) Cisplatin (1.0 mg/ml) Dacarbazine (10.0 mg/ml)

    The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml): 15 minutes ThioTEPA (10.0 mg/ml): 97 minutes

    Warning: Not for Use with: Carmustine, ThioTEPA

    Device Description

    NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a Class I, patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per D6978-05(2023). They are made from Nitrile (NBR). These gloves are blue in color and are powder free. The product is non-sterile, ambidextrous and single use only. NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) are offered in different sizes: S/M/L/XL.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Performance
    Physical Dimensions
    Length (S)Physical Dimensions Test$\ge$ 220 mm$\ge$ 220 mm
    Length (M/L/XL)Physical Dimensions Test$\ge$ 230 mm$\ge$ 230 mm
    Width (S)Physical Dimensions Test80 $\pm$ 10 mm84-86/Pass
    Width (M)Physical Dimensions Test95 $\pm$ 10 mm96-102/Pass
    Width (L)Physical Dimensions Test110 $\pm$ 10 mm104-107/Pass
    Width (XL)Physical Dimensions Test120 $\pm$ 10 mm114-116/Pass
    Thickness (Palm)Physical Dimensions Test$\ge$ 0.05 mm0.07-0.08/Pass
    Thickness (Finger)Physical Dimensions Test$\ge$ 0.05 mm0.09-0.11/Pass
    WatertightnessDetection of HolesAQL 2.50/125/Pass
    Powder ContentPowder Content$\le$ 2.0 mg per glove0.16-0.19 mg/Pass
    Physical Properties
    Tensile Strength (before aging)Physical Properties$\ge$ 14 Mpa20-24 Mpa
    Tensile Strength (after aging)Physical Properties$\ge$ 14 Mpa15-23 Mpa
    Ultimate Elongation (before aging)Physical Properties$\ge$ 500 %501-598 %
    Ultimate Elongation (after aging)Physical Properties$\ge$ 400 %471-561 %
    Chemotherapy PermeationChemotherapy drug claimNo breakthrough up to 240 minutes for specified drugs>240 minutes for Cyclophosphamide, Doxorubicin HCl, Etoposide, Fluorouracil, Paclitaxel, Cisplatin, Dacarbazine
    Low permeation times for specified drugsCarmustine: 15 minutes, ThioTEPA: 97 minutes
    CytotoxicityCytotoxicityNon-cytotoxicDevice extract is cytotoxic (met acceptance criteria by identifying the result)
    IrritationIrritationNon-irritatingNon-irritant / Pass
    SensitizationSensitizationNon-sensitizingNon-sensitizer / Pass
    Acute Systemic ToxicityAcute Systemic ToxicityNon-acute systemic toxicityDid not show acute systemic toxicity in vivo. / Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for each individual test or the data provenance (e.g., country of origin, retrospective/prospective). However, for the watertightness test (ASTM D5151/ASTM D6319), it reports "0/125/Pass" which indicates a sample size of 125 gloves were tested for holes, and none failed.

    The non-clinical tests were conducted to verify that the proposed device met all design specifications according to recognized standards. The identity of the testing laboratories or the origin of the samples is not specified in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This document describes non-clinical performance testing of a physical device (gloves). Therefore, there is no "ground truth" derived from expert interpretation of medical images or data in the typical sense. The ground truth for these tests is established by the specified standards (ASTM and ISO standards) themselves, which outline quantitative metrics and methodologies for evaluation. No human experts are explicitly mentioned as establishing ground truth for these specific tests.

    4. Adjudication Method for the Test Set

    Not applicable. The tests performed are objective, quantitative measurements against established standards, not subjective assessments requiring adjudication or consensus among experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC comparative effectiveness study is not applicable to the non-clinical performance testing of medical gloves. This type of study is typically used for evaluating the effectiveness of diagnostic or prognostic medical devices that rely on human interpretation, often with AI assistance, in comparison to human interpretation alone.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This document pertains to the physical performance of medical gloves, not a software algorithm that would operate in a standalone mode.

    7. Type of Ground Truth Used

    The ground truth for all tests is based on established testing standards and their quantitative acceptance criteria. For example:

    • Physical dimensions are compared against numerical ranges specified in ASTM D6319-19.
    • Watertightness is assessed by the absence of holes per ASTM D5151-19, with an acceptable quality level (AQL).
    • Powder content is measured against a maximum limit per ASTM D6124-06.
    • Tensile strength and ultimate elongation are measured against minimum requirements per ASTM D6319-19.
    • Chemotherapy drug permeation is determined by breakthrough detection times as per ASTM D6978-05(2023).
    • Biocompatibility tests (cytotoxicity, irritation, sensitization, acute systemic toxicity) are evaluated according to ISO 10993 series standards.

    8. Sample Size for the Training Set

    Not applicable. This document is about a physical medical device (gloves) and its non-clinical performance testing, not an AI/ML-based device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232550

    Validate with FDA (Live)

    Date Cleared
    2023-10-20

    (58 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Nitrile Examination Gloves, Dual Color Black-Blue are Class I Patient Examination Gloves. The glove is made dual color by double dipping. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19. The gloves are single use, disposable, and provided non-sterile.

    AI/ML Overview

    This document describes the non-clinical performance data for the "Nitrile Examination Gloves, Dual Color Black-Blue" (K232550) and compares it to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    MethodologyTest PerformedAcceptance CriteriaReported Device Performance
    ASTM D6319-19Physical Dimensions LengthMinimum 220mm for size XS-S; Minimum 230mm for size M-XXLPass
    ASTM D6319-19Physical Dimensions Palm WidthXS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass
    ASTM D6319-19Physical Dimensions ThicknessFinger: 0.05mm (min); Palm: 0.05mm (min)Pass
    ASTM D6319-19 & ASTM D412-16(2021)Physical PropertiesTensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and after aging 400%) MinPass
    ASTM D6319-19 & ASTM D5151-19Water leak testG-I, AQL 2.5 (ISO 2859-1)Pass
    ASTM D6319-19 & ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
    ISO 10993-10 & 23:2021Irritation and Skin SensitizationSkin sensitization and Skin irritation requirements (e.g., non-sensitizing, non-irritating)Is non-sensitization and Non-irritation
    ISO 10993-5:2009CytotoxicityCytotoxicity reactivity (no adverse reaction)showed potential toxicity to L929 cells.
    ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of systemic toxicity.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test in the "Summary of Non-Clinical Performance Data" section. It refers to various ASTM and ISO standards, which typically specify sample sizes for their respective test methodologies. The data provenance is not specified, but it can be inferred that these are laboratory tests performed on samples of the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. These are laboratory-based performance tests for a medical device (examination gloves), not studies requiring expert interpretation of medical images or patient data. The "ground truth" is established by the standardized test methodologies themselves.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study requiring human adjudication. The results are based on objective measurements and observations according to standardized test protocols.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical cases. This document pertains to the performance of examination gloves.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This document is for a physical medical device (examination gloves), not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance criteria is established by the specifications defined within the referenced ASTM and ISO standards (e.g., minimum length, maximum powder residue, specified AQL for water leak test, biological response classifications).

    8. The Sample Size for the Training Set

    Not applicable. This document pertains to the performance of a physical medical device (examination gloves) through non-clinical testing, not a machine learning or AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable (as explained in point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K230121

    Validate with FDA (Live)

    Date Cleared
    2023-05-15

    (118 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Non-Sterile Nitrile Patient Examination Glove is made from synthetic rubber latex. It is single use and powderfree variation of the class I Nitrile Patient Examination Gloves which is coated by on-line polymer with mild on-line chlorination process. These processes modify the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder. The available sizes are X-Small, Medium, Large and X-Large. The Nitrile Examination Gloves Powder Free (Blue & Black) have been tested for Chemotherapy Drugs and Fentanyl Citrate.

    AI/ML Overview

    This document describes the testing for Nitrile Examination Gloves, focusing on their resistance to chemotherapy drugs and Fentanyl Citrate. The study assesses the gloves' physical properties, biocompatibility, and chemical permeation alongside the reported device performance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Chemotherapy and Fentanyl Citrate Permeation (Minimum Breakthrough Detection Time)

    Chemotherapy DrugsConcentrationAcceptance Criteria (from ASTM D6978-05 implied by claim of use)Reported Device Performance (Blue Glove)Reported Device Performance (Black Glove)
    Carmustine (BCNU)3.3 mg/mlNo specific public acceptance criterion provided but tested22.6 minutes21.8 minutes
    Cisplatin1.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Cyclophosphamide (Cytoxan)20.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Dacarbazine10.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Doxorubicin HCl2.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Etoposide20.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Fluorouracil50.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Ifosfamide50.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Mitoxantrone2 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Paclitaxel6.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Thiotepa10.0 mg/mlNo specific public acceptance criterion provided but tested43.9 minutes17.7 minutes
    Vincristine Sulfate1.0 mg/mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    Fentanyl Citrate Injection100mcg/2mlNo specific public acceptance criterion provided but tested>240 minutes>240 minutes
    WarningAdvised against use if permeation time is too lowCarmustine, Thiotepa warningCarmustine, Thiotepa warning

    Physical Properties, Dimensional, and Biocompatibility Testing

    TestMethodAcceptance CriteriaReported Device Performance
    Freedom From HolesASTM D6319-19, D5151-19Requirement inspection level G-1, AQL 2.5Pass
    Dimension (Length, Thickness, Width)ASTM D6319-19Length: min. 220mm, 230mm. Thickness: min. 0.05mm. Width: 70-120mm (based on size)Pass (Meets requirements for various sizes)
    Physical Properties (Before Aging)ASTM D6319-19Tensile Strength: 14 MPa min. Ultimate Elongation: 500% min.Pass (Meets 14MPa min, 500% min)
    Physical Properties (After Aging)ASTM D6319-19Tensile Strength: 14 MPa min. Ultimate Elongation: 400% min.Pass (Meets 14MPa min, 400% min)
    Residual Powder ContentASTM D6319-19, D6124-06Not more than 2 mg per glovePass
    Primary Skin Irritation TestISO 10993-10Not a primary skin irritantPass (Did not cause an irritant response)
    Skin Sensitization TestISO 10993-10Not a contact sensitizerPass (Did not produce a skin sensitization effect)
    In Vitro Cytotoxicity TestISO 10993-5:2009No adverse biological reactionModerate cytotoxicity at 6.0 cm²/mL; No reactivity at 3.0 cm²/mL
    Acute Systemic ToxicityISO 10993-11No adverse biological reactionPass (No adverse biological reaction)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (number of gloves) used for the permeation, physical property, or biocompatibility tests.
    The data provenance is not explicitly mentioned (e.g., country of origin, retrospective or prospective). However, the tests were conducted according to international standards (e.g., ASTM, ISO), which implies a standardized, prospective testing approach in a laboratory setting. The applicant is based in Malaysia.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a study involving expert readers or ground truth established by medical professionals. The "ground truth" for this device is determined by objective, standardized laboratory measurements (e.g., chemical permeation time, physical properties, biological reactions) as defined by the referenced ASTM and ISO standards.


    4. Adjudication Method for the Test Set

    Not applicable. There was no human adjudication process involved as the "ground truth" is determined by objective laboratory measurements and adherence to specified test standards.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data. The device in question is examination gloves, and its performance is evaluated through physical, chemical, and biological testing, not by human readers interpreting cases.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is a medical device (examination gloves), not an AI algorithm. Therefore, the concept of "standalone" performance for an algorithm does not apply. All listed tests are "standalone" in the sense that they assess the intrinsic properties of the glove without human intervention in the measurement process itself, though human technicians conduct and analyze the tests.


    7. The Type of Ground Truth Used

    The ground truth used for this study is based on:

    • Performance Standards: Defined by international standards such as ASTM D6978-05 (for chemotherapy drug permeation), ASTM D6319-19, ASTM D5151-19, ASTM D6124-06 (for physical properties and hole freedom), and ISO 10993-10, ISO 10993-5:2009, ISO 10993-11 (for biocompatibility).
    • Objective Laboratory Measurements: Direct measurements of breakthrough time, dimensional properties, tensile strength, elongation, residual powder, and biological response to the glove materials.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set mentioned or implied in this submission, as the device is not an AI/ML algorithm that requires training data.


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222813

    Validate with FDA (Live)

    Date Cleared
    2023-03-16

    (178 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Examination Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Nitrile Examination Gloves is a Class I device bearing the product code LZA (21CFR 880.6250). They meet all the current specifications listed under the ASTM Specification D6319 -19, Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color have Finger Texture, Ambidextrous, single-use, and are powderfree. The product is non-sterile.

    AI/ML Overview

    The provided document K222813 describes the acceptance criteria and the study that proves the Ecoglove Nitrile Examination Gloves meet these criteria.

    Here's the detailed breakdown:

    1. A table of acceptance criteria and the reported device performance:

    SI No.TestsProposed Device Actual DataAcceptance Criteria
    1.Dimension: Length, Width and Thickness (ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application)Length: S: 243 mm, M: 242.30 mm, L: 243.30 mm, XL: 243.07 mm Width: S: 84.92 mm, M: 95.30 mm, L: 110.53 mm, XL: 120.62 mm Thickness: Palm S: 0.086 mm, M: 0.086 mm, L: 0.085 mm, XL: 0.085 mm Finger S: 0.115 mm, M: 0.110 mm, L: 0.115 mm, XL: 0.115 mmLength: S: 220mm min (M, L, XL not explicitly listed but implied to be similar to predicate of 230mm min) Width: S: 80 mm±10, M: 95 mm ±10, L: 110 mm ±10, XL: 120 mm ±10 Thickness: Palm: 0.05 mm min, Finger: 0.05 mm min
    2.Physical property: Tensile strength and Ultimate Elongation (ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application)Tensile strength (Before aging): S: 17.44 MPa, M: 17.60 MPa, L: 16.75 MPa, XL: 16.03 MPa Tensile strength (After aging): S: 16.52 MPa, M: 16.79 MPa, L: 15.48 MPa, XL: 15.07 MPa Ultimate elongation (Before aging): S: 519%, M: 521%, L: 565%, XL: 550% Ultimate elongation (After aging): S: 511%, M: 511%, L: 510%, XL: 514%Tensile strength: 14 MPa Min for all sizes (Before and After aging) Ultimate elongation: 500% Min for all sizes (Before aging), 400% Min for all sizes (After aging)
    3.Detection of Holes in Medical Gloves (ASTM D6319-19 / ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves)AQL 2.5 (for all sizes)AQL 2.5
    4.Powder-Free Residue (ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves)S: 0.02 Mg/glove, M: 0.01 Mg/glove, L: 0.01 Mg/glove, XL: 0.01 Mg/glove≤ 2 mg/Glove Max
    Biocompatibility Tests:
    1.In Vitro CytotoxicityFailNon-cytotoxic
    2.Skin SensitizationPassNon-sensitizer
    3.Skin IrritationPassNon-irritant
    4.Acute Systemic ToxicityPassNontoxic

    Note: For the "In Vitro Cytotoxicity" test, the proposed device failed the initial acceptance criteria of being "Non-cytotoxic". However, the document includes a note (Note1 in Table 1) stating: "Additional testing was performed to determine if this was a systemic toxicity concern." This implies that while the initial in vitro result was
    considered "Fail" against the direct "Non-cytotoxic" criterion, further investigation might have clarified that it does not pose a systemic toxicity concern.

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each performance test (dimension, physical properties, hole detection, powder residue, and biocompatibility). However, it implies that the tests were conducted according to established ASTM standards (D6319-19, D5151-19, D6124-06) and ISO standards (10993-5:2009, 10993-10:2021, 10993-23:2021, 10993-11:2017). These standards typically specify sampling plans and test methodologies.

    The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective." However, the tests are non-clinical performance and biocompatibility tests, which are typically conducted in a controlled laboratory setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is a "Nitrile Examination Glove," which is a physical product subject to objective performance testing against established standards, not a diagnostic or AI-driven device requiring expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable. Adjudication methods are typically employed in studies involving human interpretation or subjective assessments, like medical imaging or clinical trials. The tests performed for this device are objective laboratory tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. This device is a medical glove, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance tests relies on industry standards (ASTM and ISO) which define acceptable limits for physical properties, barrier integrity, and chemical residue. For biocompatibility, the ground truth is defined by the criteria established in the ISO 10993 series of standards.

    8. The sample size for the training set:

    This information is not applicable. This device does not involve machine learning or AI, and therefore does not have a "training set."

    9. How the ground truth for the training set was established:

    This information is not applicable due to the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213981

    Validate with FDA (Live)

    Date Cleared
    2022-11-12

    (327 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single use disposable device intended for medical purposes that is worn on the hands of examiners to prevent contamination between the examiner and the patient.

    Device Description

    Nitrile Examination Gloves

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for Nitrile Examination Gloves. It does not contain any information about acceptance criteria, device performance, or studies related to AI/algorithm performance.

    The letter focuses on regulatory approval, stating that the device is substantially equivalent to legally marketed predicate devices. It lists the regulation number, product code, and indications for use of the gloves. It also outlines the regulatory obligations of the manufacturer.

    Therefore, I cannot provide the requested information based on this document. The questions you've asked are typically relevant to the performance evaluation of a software-as-a-medical-device (SaMD) or AI-powered medical device, which is not what this document describes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223235

    Validate with FDA (Live)

    Date Cleared
    2022-10-26

    (13 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. This is a single-use, powder-free, non-sterile device.

    Device Description

    Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) is a Class I patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). They are made from Carboxylated Nitrile. These gloves are blue in color and are powder free. The product is non-sterile, textured, ambidextrous with beaded cuff and single use only. Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs) with sizes X-Small, Small, Medium, Large, X-Large and XX-Large are included in the submission.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for "Nitrile Examination Gloves Powder Free (Tested for use with Chemotherapy Drugs)":

    1. A table of acceptance criteria and the reported device performance

    Characteristic / Test MethodAcceptance CriteriaReported Device Performance
    Physical Dimensions (ASTM D6319-19)
    LengthX-Small: 220 mm min; Small: 220 mm min; Medium: 230 mm min; Large: 230 mm min; X-Large: 230 mm min; XX-Large: 230 mm minX-Small: 270 mm; Small: 273 mm; Medium: 274 mm; Large: 276 mm; X-Large: 278 mm; XX-Large: 276 mm
    WidthX-Small: 70+/-10 mm; Small: 80+/-10 mm; Medium: 95+/-10 mm; Large: 110+/-10 mm; X-Large: 120+/-10 mm; XX-Large: 130+/-10 mmX-Small: 72 mm; Small: 82 mm; Medium: 97 mm; Large: 103 mm; X-Large: 112 mm; XX-Large: 126 mm
    ThicknessPalm: 0.05 mm min for all sizes; Finger: 0.05 mm min for all sizesPalm: 0.12 mm for all sizes; Finger: 0.14 mm for all sizes
    Physical Properties (ASTM D6319-19)
    Tensile Strength (Before Ageing)14 MPa min for all sizesX-Small: 17.7 MPa; Small: 18.1 MPa; Medium: 18.3 MPa; Large: 18.4 MPa; X-Large: 18.2 MPa; XX-Large: 18.5 MPa
    Tensile Strength (After Ageing)14 MPa min for all sizesX-Small: 16.8 MPa; Small: 17.3 MPa; Medium: 17.2 MPa; Large: 17.6 MPa; X-Large: 17.5 MPa; XX-Large: 17.6 MPa
    Ultimate Elongation (Before Ageing)500% min for all sizesX-Small: 637%; Small: 634%; Medium: 647%; Large: 630%; X-Large: 644%; XX-Large: 644%
    Ultimate Elongation (After Ageing)400% min for all sizesX-Small: 532%; Small: 539%; Medium: 542%; Large: 536%; X-Large: 529%; XX-Large: 533%
    Freedom from Holes (ASTM D5151-19)AQL 2.5Gloves Passes AQL 2.5
    Residual Powder (ASTM D6124-06)$\leq$ 2 mg/gloveX-Small: 0.21 mg/glove; Small: 0.18 mg/glove; Medium: 0.18 mg/glove; Large: 0.19 mg/glove; X-Large: 0.21 mg/glove; XX-Large: 0.20 mg/glove
    Chemotherapy Drug Permeation (ASTM D6978-05 (2019))
    Carboplatin (10mg/ml)> 240 Minutes> 240 Minutes
    Carmustine (3.3mg/ml)Not specified as an acceptance criterion value, but rather a reported breakthrough time leading to a warning. No specific "pass/fail" for this drug against a threshold is listed.15.9 Minutes
    Cisplatin (1mg/ml)> 240 Minutes> 240 Minutes
    Cyclophosphamide (20mg/ml)> 240 Minutes> 240 Minutes
    Cytarabine HCl (100mg/ml)> 240 Minutes> 240 Minutes
    Dacarbazine (10mg/ml)> 240 Minutes> 240 Minutes
    Docetaxel (10mg/ml)> 240 Minutes> 240 Minutes
    Doxorubicin HCl (2mg/ml)> 240 Minutes> 240 Minutes
    Etoposide (20mg/ml)> 240 Minutes> 240 Minutes
    Fluorouracil (50mg/ml)> 240 Minutes> 240 Minutes
    Gemcitabine (38mg/ml)> 240 Minutes> 240 Minutes
    Ifosfamide (50mg/ml)> 240 Minutes> 240 Minutes
    Irinotecan (20mg/ml)> 240 Minutes> 240 Minutes
    Mechlorethamine HCl (1mg/ml)> 240 Minutes> 240 Minutes
    Melphalan (5mg/ml)> 240 Minutes> 240 Minutes
    Methotrexate (25mg/ml)> 240 Minutes> 240 Minutes
    Mitoxantrone (2mg/ml)> 240 Minutes> 240 Minutes
    Paclitaxel (6mg/ml)> 240 Minutes> 240 Minutes
    Thiotepa (10mg/ml)Not specified as an acceptance criterion value, but rather a reported breakthrough time leading to a warning. No specific "pass/fail" for this drug against a threshold is listed.54.2 Minutes
    Biocompatibility (ISO 10993 Series)
    Primary Skin Irritation (ISO 10993-23)Under the condition of study not an irritantUnder the condition of study, not an irritant
    Dermal Sensitization (ISO 10993-10)Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
    In Vitro Cytotoxicity (ISO 10993-5)Under the conditions of the study, non-cytotoxicUnder the conditions of the study, non-cytotoxic
    Acute Systemic Toxicity (ISO 10993-11)Under the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the conditions of the study, the test item did not produce any adverse effect

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the specific sample sizes used for each test. For many of the physical and chemical property tests (e.g., dimensions, tensile strength, freedom from holes, residual powder, chemotherapy drug permeation), the relevant ASTM standards (e.g., ASTM D6319-19, ASTM D5151-19, ASTM D6124-06, ASTM D6978-05) would specify the required statistical sampling plans, but the summary document only provides the results.

    Similarly, for biocompatibility tests (ISO 10993 series), the number of animals or cell cultures used is not detailed in this summary, but the standards themselves outline the required protocols.

    The data provenance is not explicitly stated in terms of country of origin of the data or whether it was retrospective or prospective. Given that these are bench and lab tests for a premarket notification, it's highly probable that the tests were conducted prospectively by the manufacturer or a contracted testing laboratory to demonstrate compliance with the referenced standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided since the studies performed are bench testing and biocompatibility testing against established international standards (ASTM and ISO). These types of tests do not typically involve human experts establishing a "ground truth" through interpretation, as would be the case for an AI diagnostic device. Instead, the "ground truth" is defined by the objective measurement parameters and criteria specified in the standard test methods themselves.

    4. Adjudication method for the test set

    Not applicable. As noted above, these are objective bench and lab tests against predefined specifications in standards. There is no subjective interpretation requiring an adjudication method like 2+1 or 3+1. The results are quantitative measurements or direct observations (e.g., passes/fails AQL).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical glove, not an AI diagnostic device. Therefore, no MRMC study involving human readers or AI assistance was conducted or would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical glove, not an AI algorithm.

    7. The type of ground truth used

    The "ground truth" for the performance evaluation of these nitrile examination gloves are the acceptance criteria defined by the relevant ASTM and ISO standards. These standards provide:

    • Specific measurable thresholds (e.g., minimum length, minimum tensile strength, maximum residual powder).
    • Defined test methods and methodologies (e.g., how to measure breakthrough time for chemotherapy drugs, how to assess irritation or sensitization).
    • Statistical sampling plans (e.g., AQL for freedom from holes).

    Compliance with these pre-defined, objective criteria from the standards constitutes the "ground truth" for the device's performance.

    8. The sample size for the training set

    Not applicable. This is a physical device (medical glove), not an AI model. Therefore, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221887

    Validate with FDA (Live)

    Date Cleared
    2022-09-24

    (87 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

    Device Description

    The propose devices is powder free nitrile patient examination gloves, provided as non-sterile and disposable device. The proposed devices are provided with white, black and pink color. There are six sizes, extra-small, medium, large and extra-large, extra extra-large for optional.

    AI/ML Overview

    This is a 510(k) summary for a medical device (Nitrile Examination Gloves), not for an AI/ML powered device, therefore, the requested information elements related to AI/ML device studies (such as sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set sample size, and ground truth establishment for training set) are not applicable.

    The document describes non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

    Here's the table of acceptance criteria and reported device performance based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device PerformanceResult
    ASTM D6319 - Physical Dimensions Test
    Length (XS)$\ge$ 220 mm226 mmPass
    Width (XS)70 $\pm$ 10 mm73 mmPass
    Length (S)$\ge$ 220 mm225 mmPass
    Width (S)80 $\pm$ 10 mm85 mmPass
    Length (M)$\ge$ 230 mm235 mmPass
    Width (M)95 $\pm$ 10 mm98 mmPass
    Length (L)$\ge$ 230 mm234 mmPass
    Width (L)110 $\pm$ 10 mm115 mmPass
    Length (XL)$\ge$ 230 mm235 mmPass
    Width (XL)120 $\pm$ 10 mm123 mmPass
    Length (XXL)$\ge$ 230 mm235 mmPass
    Width (XXL)$\ge$ 120 mm124 mmPass
    Thickness
    Finger Thickness$\ge$ 0.08 mmX-Small: 0.104 mm; Small: 0.106 mm; Medium: 0.104 mm; Large: 0.105 mm; X-Large: 0.106 mm; XX-Large: 0.105 mmPass
    Palm Thickness$\ge$ 0.05 mmX-Small: 0.059 mm; Small: 0.060 mm; Medium: 0.060 mm; Large: 0.062 mm; X-Large: 0.061 mm; XX-Large: 0.058 mmPass
    Physical Properties (Before Aging)
    Tensile Strength$\ge$ 14 MPaLot1: 17.6 MPa; Lot2: 18.5 MPa; Lot3: 18.2 MPaPass
    Ultimate Elongation$\ge$ 500%Lot1: 552%; Lot2: 573%; Lot3: 583%Pass
    Physical Properties (After Aging)
    Tensile Strength$\ge$ 14 MPaLot1: 19.8 MPa; Lot2: 20.3 MPa; Lot3: 20.1 MPaPass
    Ultimate Elongation$\ge$ 500%Lot1: 521%; Lot2: 518%; Lot3: 526%Pass
    ASTM D5151 - Freedom from PinholesFreedom from pinholesMeet the requirements of ASTM D5151 Test for AQL 2.5Lot1: $\le$ AQL 2.5; Lot2: $\le$ AQL 2.5; Lot3: $\le$ AQL 2.5Pass
    ASTM D6124 - Powder ResiduePowder Residue$\le$ 2.0 mgLot1: 1.2 mg; Lot2: 1.5 mg; Lot3: 1.2 mgPass
    ISO 10993-10 - IrritationTo determine if the finished device material is an irritantNon-irritatingUnder the conditions of the study, not an irritantPass
    ISO 10993-10 - SensitizationTo determine if the finished device material is a sensitizerNon-sensitizingUnder conditions of the study, not a sensitizerPass
    ISO 10993-11 - Systemic ToxicityTo determine if the finished device material extracts pose a systemic toxicity concernNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivoPass

    Study Details

    1. Sample size used for the test set and the data provenance: Not applicable. These are non-clinical bench tests on physical products, not an AI/ML test set. The tests were performed on "Lots" of the gloves (Lot1, Lot2, Lot3). The specific sample size for each test within these lots is not specified in this document but implied to be sufficient for meeting the ASTM/ISO standards. The data provenance is from non-clinical testing of the manufacturing output.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as these are objective physical and chemical tests, not requiring expert ground truth in the context of AI/ML evaluation.
    3. Adjudication method for the test set: Not applicable. The results are based on objective measurements against defined standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This concerns a physical medical device (gloves), not an AI/ML powered device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This concerns a physical medical device (gloves), not an AI/ML powered device.
    6. The type of ground truth used: For physical/chemical properties (e.g., length, thickness, tensile strength, freedom from pinholes, powder residue), the "ground truth" is the quantitative measurement obtained through standardized test methods (ASTM and ISO standards) and compared against pre-defined numerical or qualitative acceptance criteria. For biocompatibility tests (irritation, sensitization, systemic toxicity), the ground truth is established by the outcome of the biological assays as per ISO 10993 standards.
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213739

    Validate with FDA (Live)

    Date Cleared
    2022-06-22

    (205 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Examination Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Nitrile Examination Gloves are made of Nitrile rubber and are blue in color. The device is powder free nitrile examination gloves. It can be available in five specifications: XS, S, M, L and XL. The subject device is non-sterile.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for Nitrile Examination Gloves, a Class I medical device. This type of device is a physical product, not an AI/ML algorithm or software. Therefore, the "acceptance criteria and the study that proves the device meets the acceptance criteria" requested in the prompt refers to physical and biological performance testing for the gloves, not to a study about the performance of an AI model.

    Here's the breakdown of the information based on the provided text, adapted for a physical device:


    Acceptance Criteria and Device Performance for Nitrile Examination Gloves

    The study to prove the device meets acceptance criteria involved non-clinical tests based on recognized standards for medical gloves.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestExtra-Small: Length: ≥220mm; Width: 70±10 mm Small: Length: ≥220mm; Width: 80±10mm Medium: Length: ≥230mm; Width: 95±10mm Large: Length: ≥230mm; Width: 110±10mm Extra-Large: Length: ≥230mm; Width: 120±10mmPass
    Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Pass
    ASTM D6319Physical Properties (Before Aging)Tensile Strength: ≥14MPa Ultimate Elongation: ≥500%Pass
    Physical Properties (After Aging)Tensile Strength: ≥14MPa Ultimate Elongation: ≥500%Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 Test for AQL 2.5Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 <2.0mgPass
    ISO 10993-11Systemic ToxicityNon-acute systemic toxicityPass
    ISO 10993-10Irritation (Biocompatibility)Non-irritatingPass
    ISO 10993-10Sensitization (Biocompatibility)Non-sensitizingPass

    2. Sample size used for the test set and the data provenance
    The document does not explicitly state the specific sample sizes used for each test (e.g., number of gloves tested for dimensions, holes, etc.) or the provenance of the materials/specimens (e.g., country of origin, retrospective or prospective). It indicates that "Non-clinical tests were conducted to verify that the proposed device met all design specifications."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This question is not applicable as the "ground truth" for a physical device like medical gloves is established through direct physical, chemical, and biological testing according to predefined standards and objective measurements, not by expert consensus on interpretations. The tests are performed by trained personnel in laboratories.

    4. Adjudication method for the test set
    Not applicable. The tests involve objective measurements against pass/fail criteria from recognized standards, not expert adjudication of subjective assessments.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a physical medical device (nitrile examination gloves), not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This is a physical medical device, not an AI/ML algorithm.

    7. The type of ground truth used
    The ground truth is established by objective measurements and observations in accordance with standardized test methods (ASTM and ISO standards). For example:

    • Physical Dimensions: Measured length and width of gloves.
    • Thickness: Measured thickness at finger and palm.
    • Physical Properties (Tensile Strength, Elongation): Mechanically tested to quantify material strength and stretch.
    • Watertightness: Direct testing for leaks/holes.
    • Powder Content: Quantified residue weight.
    • Biocompatibility (Systemic Toxicity, Irritation, Sensitization): In vivo animal testing or in vitro assays as defined by ISO 10993 standards to observe biological reactions.

    8. The sample size for the training set
    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The manufacturing process is validated, not trained.

    9. How the ground truth for the training set was established
    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212929

    Validate with FDA (Live)

    Date Cleared
    2022-04-09

    (207 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.

    Device Description

    The proposed Nitrile Examination Gloves are ambidextrous, non-sterile, powder-free, Not made with natural rubber latex and is made of nitrile (Butyronitrile latex) . It has a finger textured surface and is colored blue. This is a single use, disposable device(s), provided non-sterile. The device is provided in 4 sizes, from small to extra-large.

    AI/ML Overview

    The provided text is a 510(k) Summary for Nitrile Examination Gloves (K212929). It does not describe an AI/ML-based medical device. Therefore, it does not contain the information required to answer the prompt regarding acceptance criteria and a study proving an AI/ML device meets those criteria.

    The document discusses the following for the Nitrile Examination Gloves:

    • Acceptance Criteria and Reported Device Performance (Table I and II): These tables detail physical properties (dimensions, tensile strength, elongation), freedom from holes, residual powder, and biocompatibility tests (irritation, sensitization, systemic toxicity, cytotoxicity). The "Reported Device Performance" is the "Proposed Device Nitrile Examination Gloves" column in Table I and the "Results" column in Table II, showing that the product meets or exceeds the specified acceptance criteria for each test.
    • Study Proving Device Meets Acceptance Criteria: The study proving the device meets the acceptance criteria consists of adherence to established ASTM and ISO standards for examination gloves. These are non-clinical (laboratory) tests.
    • Sample Size: For the physical dimensions and properties, the sample sizes are not explicitly stated, but the ASTM standards followed would specify these. For "Watertightness Test for Detection of Holes," the result is "0/200/Pass," implying a sample size of 200 gloves tested.
    • Data Provenance: The data provenance is from laboratory testing performed on the "Proposed Device Nitrile Examination Gloves." The country of origin of the data is not specified, but the manufacturer is WATCHTOWER PPE SUPPLIES, INC. located in Hicksville, New York, USA. The tests are non-clinical, so the concepts of retrospective/prospective clinical data do not apply.
    • Experts/Ground Truth/Adjudication/MRMC/Standalone: These concepts are relevant to AI/ML device studies where human expert input is used to establish ground truth or evaluate AI performance. They are not applicable to the non-clinical, physical, and biological testing of examination gloves.
    • Training Set: There is no training set mentioned because this is not an AI/ML device.

    In summary, the provided document describes the regulatory clearance for a physical medical device (nitrile examination gloves) based on non-clinical performance and biocompatibility testing against established standards, not an AI/ML device.