510(k) Data Aggregation

    K Number
    K130318

    Validate with FDA (Live)

    Date Cleared
    2013-04-22

    (73 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InfiMed i?™ Digital X-ray Imaging System is a high resolution digital imaging system intended to replace conventional film techniques, or existing digital systems, in multipurpose or dedicated applications specified below. The i-TM Digital X-ray Imaging System enables an operator to acquire, display, process, export images to portable media, send images over a network for long term storage and distribute hardcopy images with a laser printer. Image processing algorithms enable the operator to bring out diagnostic details difficult to see using conventional imaging techniques. Images can be stored locally for temporary storage. The i Digital X-ray Imaging System has the ability to interface with a variety of image receptors from CCD cameras to commercially available flat panel detectors. The major system components include an image receptor, computer, monitor and imaging software.

    For the DR application, the InfiMed is TM Digital X-ray Imaging System is intended for use in general radiographic examinations and applications (excluding fluoroscopy, angiography, and mammography).

    For the RF/DSA application, the InfiMed i2m Digital X-ray Imaging System is intended for use where general fluoroscopy, interventional fluoroscopy or angiography imaging procedures are performed.

    Device Description

    The InfiMed if TM Digital X-ray Imaging System is a high resolution digital imaging system designed for digital X-ray imaging through the use of an X-ray detector. The InfiMed i " Digital X-ray Imaging System is designed to support general radiography (excluding mammography), fluoroscopy, interventional fluoroscopy or angiography imaging procedures through a single common imaging platform.

    The modified InfiMed i TM Digital X-ray Imaging System consists of an X-ray imaging receptor (any of the following: CCD Camera, Trixell Pixium 3543, Trixell Pixium 4600, Varian PaxScan 4336R, Varian PaxScan 4343R, Carestream Health Detector, Samsung LTX240AA01-A, Toshiba FDX 4343R, Trixell Pixium RF4343. Varian PaxScan 4343CB), computer, monitor, and the digital imaging system.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Varian Medical Systems Nexus DRF Digital X-ray Imaging System (with PaxScan 4343CB). This submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed study proving the device meets specific performance acceptance criteria in terms of diagnostic accuracy or impact on human readers.

    Here's a breakdown of the requested information based on the provided text, along with notes on what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Equivalent or better image quality than predicate device"delivers equivalent or better image quality as the predicate device."
    Performance of functions performed by predicate device"The comparison chart reveals that functions performed by the predicate device are performed by the modified i^2TM Digital X-ray Imaging System."
    No new indications for use"has no new indications for use"
    No significant technological differences"has no significant technological differences"
    As safe and effective as predicate"is as safe and effective as the predicate device"
    Does not raise different questions of safety and effectiveness"does not raise different questions of safety and effectiveness"
    All release criteria met (from validation protocols)"all release criteria have been met"

    Note: The document states that "Validation was completed in accordance with the Validation Protocols included with this submission" and that "Protocols were designed, executed and documented according to the Design Validation process with predetermined test methods and corresponding acceptance criteria." However, the specific acceptance criteria and quantitative results of these validation protocols are not provided in the summary. The acceptance criteria above are implied from the statements about "substantial equivalence."

    2. Sample size used for the test set and the data provenance

    The document does not provide details about a specific "test set" in the context of a clinical performance study with human readers or diagnostic accuracy. It mentions "Clinical Data submitted is consistent with FDA guidance document 'Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance: Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices'." This suggests that any clinical data submitted would be in line with general requirements for imaging devices, but specific details about a test set, sample size, or data provenance (e.g., country of origin, retrospective/prospective) are not included in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The submission focuses on demonstrating technical equivalence and safety/effectiveness relative to a predicate, not on a new clinical study involving expert interpretation for ground truth establishment.

    4. Adjudication method for the test set

    This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC comparative effectiveness study was done, especially one involving "AI assistance." The device is described as a "Digital X-ray Imaging System," and the "Image processing algorithms enable the operator to bring out diagnostic details difficult to see using conventional imaging techniques." This refers to standard image processing, not necessarily AI in the modern sense, and no comparative effectiveness study results are given.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    There is no indication that a standalone algorithm performance study was conducted or reported in this summary. The device is an imaging system, not a diagnostic AI algorithm in itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not specify the type of ground truth used for any clinical performance evaluation. Given the nature of a 510(k) for an imaging system upgrade (adding a new receptor), the clinical data likely involves demonstrating image quality and diagnostic utility in comparison to the predicate, rather than establishing new ground truth for specific pathologies.

    8. The sample size for the training set

    This information is not provided in the document. The device is an imaging system, and while it has image processing algorithms, the concept of a "training set" in the context of modern machine learning algorithms used for diagnostic purposes is not explicitly discussed or applicable here.

    9. How the ground truth for the training set was established

    This information is not provided in the document, and the concept of a training set ground truth is not applicable given the details provided.

    In summary:

    This 510(k) summary is primarily focused on demonstrating substantial equivalence of a modified digital X-ray imaging system to an existing predicate device, largely based on technological characteristics and the claim of equivalent or better image quality. It does not contain the detailed clinical study information (sample sizes, expert qualifications, ground truth methods, MRMC studies, or AI performance metrics) that would be expected for a submission involving a novel diagnostic AI algorithm or a device requiring new clinical performance claims beyond equivalence to a predicate. The "clinical data" referenced is likely related to general image quality and functional performance as per the FDA guidance for solid-state X-ray imaging devices, rather than a specific study comparing diagnostic accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103416

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-03-17

    (115 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InfiMed i one Digital X-ray Imaging System is a high resolution digital imaging system intended to replace conventional film techniques, or existing digital systems, in multipurpose or dedicated applications specified below. The i2M Digital X-ray Imaging System enables an operator to acquire, display, process, export images to portable media, send images over a network for long term storage and distribute hardcopy images with a laser printer. Image processing algorithms enable the operator to bring out diagnostic details difficult to see using conventional imaging techniques. Images can be stored locally for temporary storage. The i2m Digital X-ray Imaging System has the ability to interface with a variety of image receptors from CCD cameras to commercially available flat panel detectors. The major system components include an image receptor, computer, monitor and imaging software.

    For the DR application, the InfiMed i-714 Digital X-ray Imaging System is intended for use in general radiographic examinations and applications (excluding fluoroscopy, angiography, and mammography),

    For the RF/DSA application, the InfiMed i TM Digital X-ray Imaging System is intended for use where general fluoroscopy, interventional fluoroscopy or angiography imaging procedures are performed.

    Device Description

    The InfiMed i-TM Digital X-ray Imaging System is a high resolution digital imaging system designed for digital X-ray imaging through the use of an X-ray detector. The InfiMed i?™ Digital X-ray Imaging System is designed to support general radiography (excluding mammography), fluoroscopy, interventional fluoroscopy or angiography imaging procedures through a single common imaging platform.

    The modified InfiMed ism Digital X-ray Imaging System consists of an X-ray imaging receptor (any of the following: CCD Camera, Trixell Pixium 3543, Trixell Pixium 4600, Varian PaxScan 4336R, Varian PaxScan 4343R, Carestream Health Detector, Samsung LTX240AA01-A, Toshiba FDX 4343R, Trixell Pixium RF4343), computer, monitor, and the digital imaging system.

    AI/ML Overview

    The provided text describes the Infimed Nexus DRF Digital X-ray Imaging System and its substantial equivalence to a predicate device. However, it does not contain specific details about acceptance criteria, a dedicated study proving device performance against those criteria, or the methodology typically associated with such studies (sample sizes, expert qualifications, adjudication, MRMC studies, ground truth details, training set information).

    The document mentions "Validation Protocols" and "predetermined test methods and corresponding acceptance criteria" in its Non-clinical Tests Discussion, and states that "all release criteria have been met." It also refers to a "Clinical Data submitted is consistent with FDA guidance document... for Solid State X-ray Imaging Devices." However, it does not provide the actual acceptance criteria, the specific results against them, or the details of the clinical study itself.

    Therefore, I cannot fully answer your request for specific details regarding acceptance criteria and the study proving the device meets them, as that information is not present in the provided text.

    Based on the available information, here's what can be extracted and what information is missing:

    Information Not Found in the Provided Document:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for the test set.
    • Data provenance (country of origin, retrospective/prospective) for any test data.
    • Number of experts used to establish ground truth for the test set.
    • Qualifications of those experts.
    • Adjudication method for the test set.
    • Results of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, including effect size.
    • Results of a standalone (algorithm-only) performance study.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for either training or testing.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    Information Extracted (Limited):

    The document broadly states that:

    • "Validation Protocols were designed, executed and documented according to the Design Validation process with predetermined test methods and corresponding acceptance criteria."
    • "In conclusion, all release criteria have been met and the modified i-TM Digital Xray Imaging System is as safe and effective as the predicate device."
    • "Clinical Data submitted is consistent with FDA guidance document 'Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance: Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices'."

    Essentially, the document confirms that non-clinical and clinical tests were performed, and acceptance criteria were met, but it does not specify what those criteria were or the detailed methodology and results of the studies. These details would typically be found in the actual validation protocols and study reports referenced, but they are not included in this 510(k) summary.