Found 3 results

510(k) Data Aggregation

    K Number
    K093914

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-14

    (478 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Newdeal Compression Plates are intended for fixation of bone fractures or for bone reconstruction.
    Examples include:

    • Arthrodesis in hand or foot surgery .
    • . Fracture management in the foot or hand
    • . Mono or Bi-cortical osteotomies in the foot or hand
    • Distal or proximal metatarsal or metacarpal osteotomies
    • Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc) .
      The Newdeal Uni-CP™ U-Plate is intended for arthrodesis of the second and third cuneo-metatarsal and the inter-cuneiform second and third joints.
      The size and number of the plate(s) used should be adapted to the specific indication.
    Device Description

    The Newdeal Compression Plates will offer the combination of two concepts:

    • By widening the "eye" (diamond shaped opening) on the interaxis of the plate. . mechanical deformation leads to narrowing of the interaxis of the two legs and thus provides compression between the two bone fragments to fuse.
    • . The rigidity of the "legs" is obtained using the Newdeal Locking System including a screw and a washer.
    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device, the "Modified Newdeal® Compression Plates." This type of submission focuses on demonstrating "substantial equivalence" to a predicate device, rather than proving novel effectiveness or meeting specific performance acceptance criteria through clinical studies. Therefore, many of the requested elements for describing acceptance criteria and a study proving device adherence are not directly applicable to this type of regulatory filing.

    Here's an analysis based on the provided text, addressing the points where information is available and explaining why other points are not applicable:

    1. A table of acceptance criteria and the reported device performance

    For a 510(k) submission, the primary acceptance criteria for the modified device are demonstrating mechanical performance at least equivalent to the predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Strength: Modified Newdeal Compression Plates must demonstrate mechanical properties that are substantially equivalent to the predicate Newdeal Compression Plates (K070447)."Static and dynamic 4-point bending mechanical tests have been carried out. Results have shown that the mechanical properties of the modified Newdeal Compression Plates are substantially equivalent to the properties of the predicate devices: unmodified Newdeal Compression Plates, K070447, cleared on March 29, 2007."
    No change in fundamental scientific technology."The proposed device does not change the intended use or fundamental scientific technology of the predicate device..."
    No new issues of safety or effectiveness raised."...and does not raise any new issues of safety or effectiveness."
    Indications for Use remain consistent or comparable to the predicate.The Indications for Use for the modified device are provided and are consistent with bone fixation and reconstruction. The document states the modified device is substantially equivalent, implying the indications are also equivalent.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size: The document does not specify the exact number of plates or tests performed during the static and dynamic 4-point bending mechanical tests. It only states that tests "have been carried out."
    • Data provenance: The document does not explicitly state the country of origin of the mechanical test data. Given the submitter's address is in France (Newdeal SAS), it is likely the testing was conducted in Europe. The testing is prospective in the sense that it was conducted specifically to support this 510(k) submission, not by analyzing past clinical data. This is a pre-clinical (bench-top) study, not a clinical study involving patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This point is not applicable. The "ground truth" concept (e.g., expert consensus on clinical diagnoses) is relevant for studies involving human interpretation or clinical outcomes. This submission is based on mechanical (bench-top) testing, where the "ground truth" is established by engineering standards and measurements, not expert clinical interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This point is not applicable. Adjudication methods are used in clinical trials or studies where human assessment of outcomes or conditions requires agreement or arbitration. Mechanical testing does not involve such human adjudication in the same way.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This point is not applicable. This submission is for a medical implant (bone plate), not an AI-powered diagnostic device. Therefore, MRMC studies and the concept of human readers improving with AI assistance are irrelevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This point is not applicable. This is not an algorithm or software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context refers to the established mechanical properties and performance standards for bone fixation plates, as generally understood in the orthopedic device industry and often guided by international standards (though not explicitly mentioned here) for such testing. The comparison is made against the measured mechanical properties of the predicate device.

    8. The sample size for the training set

    This point is not applicable. There is no "training set" as this is a physical medical device undergoing mechanical testing, not a machine learning model.

    9. How the ground truth for the training set was established

    This point is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091609

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-06-30

    (27 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Newdeal Compression Plate is intended for fixation of bone fractures or for bone reconstruction.

    Examples include:

    • Arthrodesis in hand or foot surgery
    • Fracture management in the foot or hand
    • Mono or Bi-cortical osteotomies in the foot or hand
    • Distal or proximal metatarsal or metacarpal osteotomies
    • Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc)

    The size and number of the plate(s) used should be adapted to the specific indication.

    Device Description

    The Newdeal Compression Plate will offer the combination of two concepts:

    • By widening the "eye" (diamond shaped opening) on the interaxis of the plate, mechanical deformation leads to narrowing of the interaxis of the two legs and thus provides compression between the two bone fragments to fuse. This is the same principle as the Newdeal Compression Plate (K070447), UNI-CLIP (K011716) and Newdeal Large UNI-CLIP (K061594).
    • The rigidity of the "legs" is obtained using the Newdeal Locking System including a screw and a washer. This is the same principle as the Locking System of the following predicate devices: Surfix Knee Osteotomy System (K041601), Newdeal Lisfranc Plates (K060474), Newdeal Lapidus Plates (K060476), Newdeal TTC Plates (K060473 and Newdeal Compression Plate (K070447).
    AI/ML Overview

    The provided text describes a submission for a medical device called the "Newdeal Compression Plate," not an AI/ML device. Therefore, the acceptance criteria and study described are related to the mechanical properties and substantial equivalence of a physical orthopedic implant, not to the performance of a software algorithm.

    Based on the nature of the submission (510(k) for a medical device) and the content, there is no mention of an AI/ML device or any study that would typically be associated with evaluating AI/ML performance (e.g., diagnostic accuracy, reader study, standalone performance against ground truth).

    The document focuses on establishing substantial equivalence to a previously approved predicate device (Newdeal Compression Plate, K070447) based on mechanical testing.

    Here's a breakdown of why the requested information for an AI/ML device cannot be provided from this document:

    • No AI/ML device: The device is a "Newdeal Compression Plate," a physical orthopedic implant.
    • No acceptance criteria for AI/ML performance: The acceptance criteria are implicit in demonstrating substantial equivalence in mechanical properties.
    • No study proving AI/ML performance: The study performed was "Mechanical tests" to show the equivalency of the modified plate to the unmodified predicate device.

    However, I can extract the relevant information from the document regarding the medical device itself:


    Acceptance Criteria and Study for the "Newdeal Compression Plate" (Medical Device - Not AI/ML)

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical properties are equivalent to the predicate device."Results have shown that the mechanical properties of the modified Newdeal Compression Plate are equivalent to the properties of the unmodified device, Newdeal Compression Plate, K070447."
    Does not change intended use or fundamental scientific technology."The modifications do not change the intended use or fundamental scientific technology of the device and do not raise any new issues of safety or effectiveness."

    2. Sample size used for the test set and data provenance:

    • Sample Size: Not explicitly stated for the mechanical tests, but mechanical testing typically involves a set number of samples for various load, fatigue, and other biomechanical assessments.
    • Data Provenance: Not specified, but generally refers to in-vitro laboratory testing of the physical device.

    3. Number of experts used to establish the ground truth for the test set and qualifications:

    • Not applicable. For mechanical testing of a physical device, ground truth is established through standardized engineering test methods and measurements, not expert consensus in the clinical sense.

    4. Adjudication method for the test set:

    • Not applicable. Mechanical testing data is quantitative measurement, not subjective assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No. This is a medical device (implant) and not a diagnostic or screening tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device is a physical implant, not an algorithm.

    7. The type of ground truth used:

    • For the mechanical tests, the ground truth would be the quantitative measurements obtained from destructive and non-destructive mechanical tests (e.g., tensile strength, fatigue life, compression resistance, torsional resistance) against established engineering standards and comparison data from the predicate device.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is a physical device being evaluated for mechanical equivalence, not an AI/ML model being trained.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    In summary, the provided document details a 510(k) submission for a physical orthopedic implant. It does not pertain to an AI/ML device, and therefore, the specific questions related to AI/ML performance evaluation cannot be answered from this text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070447

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-29

    (42 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Newdeal Compression Plate is indicated for fixation of bone fractures or for bone reconstruction.

    Examples include:

    • Arthrodesis in hand or foot surgery
    • Fractures management in the foot or hand
    • Mono or Bi-cortical osteotomies in the foot or hand
    • Distal or proximal metatarsal or metacarpal osteotomies
    • Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.)

    The size and number of plate(s) used should be adapted to the specific indication.

    Device Description

    The Newdeal Compression Plate will offer the combination of two concepts:

    • By widening the "eye" (diamond shaped opening) on the interaxis of the plate, mechanical deformation leads to narrowing of the interaxis of the two legs and thus provides compression between the two bone fragments to fuse. This is the same principle as the UNI-CLIP (K011716) and Newdeal Large UNI-CLIP (K061594).
    • . The rigidity of the "legs" is obtained using the Newdeal Locking System including a screw and a washer. This is the same principle as the Locking system of the
    • . following predicate device: Surfix Knee Osteotomy system (K041601), Newdeal Lisfranc Plates (K060474), Newdeal Lapidus Plates (K060476) and Newdeal TTC Plates (K060473).

    The plate will be available with 2 or 4 holes and with interaxis of 20, 25 or 30 mm.

    Fixation of the Newdeal Compression Plate is provided by two or four Screws & washers, which are the same as for the Lapidus Plate (K060474) and Lisfranc Plate (K060476).

    Plates, screws and washers will be provided either sterile or non-sterile.

    The Locking System (Screw + washer) will be made from Titanium alloy Ti-6Al-4V or 316L Stainless Steel.

    AI/ML Overview

    The provided text describes a medical device, the "Newdeal Compression Plate," and its regulatory submission (510(k)) to the FDA. The document primarily focuses on demonstrating substantial equivalence to existing predicate devices, rather than an AI-powered device performance study.

    Therefore, many of the requested categories about AI device performance will not be applicable.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Device Performance (Not an AI Device)

    Since this document describes a traditional medical device (compression plate) and not an AI-powered one, the concept of "acceptance criteria" for algorithms and their "reported device performance" in terms of metrics like sensitivity, specificity, or AUC is not applicable. The device's performance is assessed through mechanical tests to demonstrate its physical properties are compatible with its intended use.

    Acceptance Criteria (Mechanical)Reported Device Performance
    Mechanical properties compatible with intended use (implicitly, non-inferior to predicate device)Mechanical tests show compatibility with intended uses and substantial equivalence to predicate devices (Large UNI-CLIP, K061594).

    Study Details (Mechanical Device, Not AI)

    1. Sample size used for the test set and the data provenance: Not applicable for an AI device. For the mechanical tests, the specifics of the samples (e.g., number of plates tested, type of mechanical stress applied) are not detailed in this summary. Data provenance is not relevant in the context of mechanical testing of a physical device.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth in this context refers to the physical properties of the device, established through engineering measurements, not expert interpretation.

    3. Adjudication method for the test set: Not applicable.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, this is not an AI device, so MRMC studies involving human readers and AI assistance are not relevant.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI algorithm.

    6. The type of ground truth used: For the mechanical tests, the "ground truth" would be established by the physical and material science properties measured during testing and compared against established engineering standards or the predicate device's performance.

    7. The sample size for the training set: Not applicable. There is no AI model to train.

    8. How the ground truth for the training set was established: Not applicable.


    Summary of Relevant Information for this Traditional Device:

    • Device Type: Newdeal Compression Plate (a metallic bone fixation appliance).
    • Purpose of the Submission: To demonstrate substantial equivalence to legally marketed predicate devices.
    • Demonstration of Performance: Mechanical tests were carried out, and the results showed that the Newdeal Compression Plate has "mechanical properties compatible with their intended uses." These results were compared with a predicate device, the Large UNI-CLIP (K061594).
    • Predicate Devices:
      • Wright Medical Technology Compression Plate, K051908
      • Newdeal Large UNI-CLIP Staple, K061594
    • Conclusion: The device is substantially equivalent to the predicate devices and does not raise new issues of scientific technology, safety, or effectiveness.

    This document clearly pertains to a physical medical device and its mechanical testing, not an AI-powered diagnostic or assistive tool.