Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K231061

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-08

    (239 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (e.g., brain tumor, radiation necrosis, and epileptogenic foci as identified by non-invasive and invasive neurodiagnostic testing, including imaging), through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI-based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides near real-time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis.

    Device Description

    The Monteris NeuroBlate® System is a collection of MRI-compatible laser devices and accessories that create an MRI guided intracranial delivery of precision thermal therapy in the practice of neurosurgery.

    As previously described in K182036 and K222983, the NeuroBlate System is typically used for the ablation of target tissue (tumors, radiation necrosis, epileptic foci) in the brain.

    The NeuroBlate System components consist of:

    • Families of gas-cooled Laser Delivery Probe (SideFire & FullFire) to deliver controlled energy to a target zone.
    • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Probe Drivers and interconnections for the Laser Delivery Probes (e.g., Connector Module).
    • A System Electronics Rack and Components, which includes the laser and necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation.
    • A Control Workstation including the M-Vision™. M-Vision Pro™. M-Vision Fusion™ , or Fusion-STM software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • The Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • The AtamA Stabilization System, MRI head coils, and other optional accessories, . including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.

    There is no change to entire system, with the exception of an additional NB3 (1.6mm) Laser Probe and the corresponding labeling updates.

    AI/ML Overview

    The provided text refers to the FDA's 510(k) premarket notification for the Monteris Medical NeuroBlate System, specifically concerning the addition of a new NB3 (1.6mm) Laser Probe. The document focuses on demonstrating substantial equivalence to existing predicate devices rather than detailing a study that proves the device meets specific acceptance criteria in the context of clinical performance or AI algorithm validation.

    Therefore, many of the requested sections related to acceptance criteria, device performance, sample sizes for test sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth for training sets do not apply to the information provided in this regulatory submission.

    The document primarily describes:

    • Device Description: The NeuroBlate System and the new NB3 probe.
    • Intended Use/Indications for Use: What the device is intended to do (ablate, necrotize, or coagulate intracranial soft tissue) and its role in planning and monitoring thermal therapies under MRI visualization.
    • Comparison to Predicate Devices: How the new NB3 probe is similar to previously cleared NeuroBlate probes and other competitive laser probes.
    • Summary of Supporting Data: A high-level overview of the design verification and validation process, stating that acceptance criteria were defined and met.

    Since this is a 510(k) submission for a modification (adding a new probe size) and relies on substantial equivalence, it does not present a clinical study with detailed performance metrics against specific acceptance criteria in the way an AI/ML device submission would.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "Each verification test protocol incorporated clearly defined acceptance criteria. The corresponding test reports confirmed (and documented) the design output met the design input for the requirements." And "Each validation protocol described the objective, test method and acceptance criteria. The corresponding test reports confirmed (and documented) the modified NeuroBlate System met the user needs and intended use."

    However, the specific quantitative acceptance criteria and the reported numerical device performance results are NOT explicitly provided in the given text. The comparison table (Table 1: Substantial Equivalence; Technical Comparison) focuses on technical characteristics for demonstrating substantial equivalence, not performance against clinical acceptance criteria.

    Based on the provided text, a table cannot be constructed with specific numerical acceptance criteria and reported device performance. The document only states that such criteria were met during design verification and validation.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable/Not provided. The document describes design verification and validation activities, but does not detail a specific "test set" in the context of a clinical performance study with patient data, nor does it mention sample sizes for such tests or data provenance. The focus is on technical equivalence and design control compliance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable/Not provided. This information would typically be relevant for studies evaluating human interpretation or AI algorithm performance. The document focuses on the technical aspects of a medical device (laser probe) and its regulatory clearance based on substantial equivalence.

    4. Adjudication Method for the Test Set

    Not applicable/Not provided. Similar to the above, adjudication methods are associated with studies involving human interpretation or AI performance assessment, which is not the focus of this regulatory submission as presented.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable/Not provided. This device is a surgical tool (laser probe) for thermal therapy, not an AI-assisted diagnostic or interpretative system that would typically be evaluated in an MRMC study concerning human reader improvement. While the NeuroBlate System "provides MRI-based trajectory planning assistance" and "near real-time thermographic analysis," it explicitly states, "Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis." The submission focuses on the new probe's physical properties and safety, not on the comparative effectiveness of an AI component.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable/Not provided. The NeuroBlate System's software (M-Vision, M-Vision Pro, M-Vision Fusion, Fusion-STM) is described as providing "near real-time thermographic analysis," but the submission does not detail a standalone performance evaluation of this analytic component. The overall regulatory focus is on the substantial equivalence of the physical laser probe and system.

    7. The Type of Ground Truth Used

    Not explicitly provided in detail. The document mentions that the design verification and validation process confirmed the device met "design input for the requirements" and "user needs and intended use." This implies a ground truth defined by engineering specifications and intended clinical outcomes, but the specific methodologies for establishing this ground truth (e.g., in vivo pathology, clinical outcomes as defined for a clinical trial) are not described in the given text.

    8. The Sample Size for the Training Set

    Not applicable/Not provided in the context of an AI/ML training set. The document describes the development and testing of a physical medical device (laser probe), not the training of an AI algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable/Not provided. As there is no mention of a training set for an AI/ML algorithm in the provided text, this question is not addressed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201056

    Validate with FDA (Live)

    Date Cleared
    2020-08-21

    (122 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (e.g., brain tumor and epileptic foci as identified by non-invasive and invasive neurodiagnostic testing, including imaging), through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides near real-time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis.

    Device Description

    The Monteris NeuroBlate® System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy that is currently indicated for the practice of neurosurgery.

    As previously described in K171255, K182036, K193375, the NeuroBlate System is typically used for the minimally invasive ablation of neurosurgeon identified target tissue (tumors, epileptic foci) in the brain.

    The NeuroBlate System components consist of:

    • Families of gas-cooled Laser Delivery Probe (SideFire & FullFire) to deliver controlled energy to a target zone.
    • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced and Robotic Probe Drivers and interconnections for the Laser Delivery Probes;
    • A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation,
    • A Control Workstation including the M-Vision, M-Vision Fusion, and Fusion-S software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • The AXiiiS stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.
    AI/ML Overview

    The provided FDA 510(k) summary (K201056) for the NeuroBlate System does not contain a study that proves the device meets specific acceptance criteria in the manner one might expect for a stand-alone AI/software device evaluation.

    Instead, this document describes a modification to the device's "Indications for Use" statement. The core argument for safety and effectiveness is based on the device's substantial equivalence to previously cleared predicate devices (K171255, K182036, K193375). The modifications are described as "minor (non-significant)" and intended to "improve clarity and ease of understanding."

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from a new study, nor details about sample sizes, expert qualifications, or adjudication methods for a test set, because such a study is not presented in this document.

    Here's what can be extracted based on the provided text:

    • No new study data for AI/software performance is presented in this 510(k) for the modifications. The document explicitly states: "As these modifications to the Indications For Use do not create a new intended use or raise new or different questions of safety or efficacy and incorporates previously provided clinical data(4), additional clinical data are not necessary to demonstrate substantial equivalence." Clinical data from previous submissions (which are incorporated by reference) formed the basis for the original clearances.

    Let's address the specific points you asked for, in the context of what is available in this document:

    1. A table of acceptance criteria and the reported device performance: This document does not present such a table because it's a submission for an IFU clarity update, not a new performance study. The performance of the NeuroBlate System was established in prior 510(k) clearances (K171255, K182036, K193375) through "in-vitro (bench) and in-vivo (animal) data." No specific numerical acceptance criteria or performance metrics are reported in this document.

    2. Sample size used for the test set and the data provenance: Not applicable to this specific submission. The data provenance would refer to the previous clearances (K171255, K182036, K193375) which are not detailed here. The submission mentions "Real World Evidence from Monteris' on-going post-market registry" and "scientific clinical literature peer reviewed published data," but these are cited as justification for the clarity of the IFU rather than a new test set for the modifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable to this specific submission.

    4. Adjudication method for the test set: Not applicable to this specific submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable to this specific submission. The device description of the NeuroBlate System as a "neurosurgical tool" that provides "near real-time thermographic analysis" and "information that may be useful in the determination or assessment of therapy" when "interpreted by a trained physician" indicates physician involvement. However, this document does not present an MRMC study comparing physician performance with and without the device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable to this specific submission. The device is explicitly designed for "interpretation by a trained physician" and patient management decisions "should not be made solely on the basis of the NeuroBlate System analysis."

    7. The type of ground truth used: Not applicable to this specific submission. For the previous clearances, it would likely have involved histopathological confirmation for ablation efficacy in animal models and potentially post-treatment imaging or clinical outcomes for human studies (if any were submitted then).

    8. The sample size for the training set: Not applicable to this specific submission. This document pertains to an IFU update, not a new algorithm development or re-training.

    9. How the ground truth for the training set was established: Not applicable to this specific submission.

    In summary: K201056 is a 510(k) submission primarily focused on clarifying the Indications For Use statement of an existing medical device (NeuroBlate System). It does not present new performance data, test sets, or studies comparing the device's performance against acceptance criteria in the context of a new or modified AI/software component, as its justification rests on substantial equivalence to prior FDA clearances. The performance and safety of the device were established in those earlier submissions, which are referenced but not detailed here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193375

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-03

    (29 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through internal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI-based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides near real-time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided intracranial delivery of precision thermal therapy in the practice of neurosurgery.

    As previously described in K182036 and K173305, the NeuroBlate™ System is typically used for the minimally invasive ablation of target tissue (tumors, epileptic foci) in the brain.

    The NeuroBlate™ System components consist of:

    • 0 Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver controlled energy to a target zone.
    • 0 Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position. stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Probe Drivers and interconnections for the Laser Delivery Probes (e.g., Connector Module);
    • . A System Electronics Rack and Components, which includes the laser and necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation,
    • A Control Workstation including the M-Vision™, M-Vision Pro™, or M-Vision ● Fusion™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate™ procedures, and interfaces to the MRI and hardware subsvstems.

    The NeuroBlate™ System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • The AtamA Stabilization System and MRI receive-only head coil, as well as other ● optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, fiducial markers, bone screws, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.

    There is no change to entire system, with the exception of an additional NeuroBlate Fusion-STM Software Package (V3.15).

    AI/ML Overview

    This FDA 510(k) summary for the Monteris Medical NeuroBlate™ System with the Fusion-S Software (V3.15) focuses on demonstrating substantial equivalence to a predicate device, primarily due to software modifications. It does not present a detailed study on device performance against specific acceptance criteria in the manner requested for an AI/ML device.

    The document describes software development processes (design verification and validation) and states that these processes confirmed the device met user needs and intended use, but it does not provide specific performance metrics, sample sizes, or ground truth details common in AI/ML performance studies.

    Therefore, I cannot extract the detailed information requested in the prompt based on the provided text, as the application's focus is on substantial equivalence rather than a new performance study with specific acceptance criteria for a novel AI/ML component.

    Here's a breakdown of why the requested information cannot be provided from this document:

    1. A table of acceptance criteria and the reported device performance: This information is not present. The document states that "Each verification test protocol incorporated clearly defined acceptance criteria" and "Each validation protocol described the objective, test method and acceptance criteria," but it does not list these criteria or the numerical results of meeting them.
    2. Sample sizes used for the test set and the data provenance: This information is not provided. The document mentions "design verification and design validation process" but doesn't specify test set sizes or data origins.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided. Ground truth establishment is not discussed.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not mentioned.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not explicitly mentioned as a separate performance study. The device is described as assisting a "trained physician," indicating a human-in-the-loop context for its intended use, but no specific AI performance study is detailed.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned.
    8. The sample size for the training set: This refers to an AI/ML training set, which is not applicable or discussed here as this is a software modification rather than an AI/ML training process for a new algorithm.
    9. How the ground truth for the training set was established: Not applicable, as detailed AI/ML training is not the subject of this 510(k).

    In summary: The provided FDA 510(k) summary focuses on demonstrating the substantial equivalence of an updated version of an existing device (NeuroBlate System with Fusion-S software) to its predicates, rather than presenting a performance study for a novel AI/ML component against specific acceptance criteria. The modifications are described as "not change the intended use, operating principles, or raise any unaddressed safety concerns." The verification and validation processes mentioned are general quality system activities to ensure the software functions as intended and meets user needs, not a specific clinical performance study with the metrics and details requested for an AI/ML evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182036

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-10-15

    (77 days)

    Product Code
    Regulation Number
    878.4810
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through internal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides near real time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided intracranial delivery of precision thermal therapy in the practice of neurosurgery.

    The NeuroBlate™ System components consist of:

    • Families of gas-cooled Laser Delivery Probe (SideFire & FullFire) to deliver controlled energy to a target zone.
    • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Probe Drivers and interconnections for the Laser Delivery Probes (e.g., Connector Module);
    • A System Electronics Rack and Components, which includes the laser and necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation,
    • A Control Workstation including the M-Vision™ and M-Vision Pro™ FUSION™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate™ procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate™ System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • The AtamA Stabilization System, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, fiducial markers, bone screws, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools).

    This submission's proposed change simply replaces the Probe's existing internal metallic (wire) thermocouple with a non-metallic fiber optic. temperature sensor. The optical fiber temperature sensor has the identical function as the existing metallic (wire) thermocouple, i.e., to measure the probe tip's internal temperature and to transmit the temperature measurement to the NeuroBlate System and associated M-Vision Software.

    Corresponding hardware and software changes are proposed in order to incorporate the new optical fiber component, e.g., (hardware) modified Connector Module, a Signal Conditioner (converts optical fiber signal to electrical temperature signal), and (software) M-Vision Pro Software Package (V3.14), along with associated labeling updates.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Monteris Medical NeuroBlate™ System with the fiber optic temperature sensor.

    It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. Therefore, the level of detail regarding specific acceptance criteria and study results, particularly for standalone performance or comparative effectiveness, is less comprehensive than what might be found in a full clinical study report.


    Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly list specific numerical acceptance criteria for device performance in a table format. Instead, it broadly states that:

    "The Design Verification process utilized protocols to detail the associated tests which confirmed the design output met the design input for the requirements. Each verification test protocol incorporated clearly defined acceptance criteria."

    And for validation:

    "The Design Validation process utilized protocols to detail the associated tests which confirmed the modified NeuroBlate™ System met the user needs and intended use. Each validation protocol described the objective, test method and acceptance criteria."

    The core purpose of this submission is to demonstrate that the functionality of the new fiber optic temperature sensor (FOTS) is identical to the existing metallic (wire) thermocouple in measuring the probe tip's internal temperature and transmitting the data to the NeuroBlate System and M-Vision Software. The key performance aspect is that the FOTS prevents MRI induced unintentional Probe heating, which was the root cause of a recall (Z-0194-2018).

    Given this, the "acceptance criteria" can be inferred as ensuring the new FOTS system performs as reliably and accurately as the previous system, while also actively eliminating the MRI-induced heating issue. Without specific protocols, a table of acceptance criteria and reported device performance cannot be generated with numerical values from this text. The reported device performance is that the new FOTS system successfully prevents the MRI induced unintentional Probe heating by eliminating the coupling of MR energy to the probe umbilical.


    Detailed Information on the Study:

    1. A table of acceptance criteria and the reported device performance

      • As explained above, explicit numerical acceptance criteria are not provided in this 510(k) summary. The primary criteria would revolve around the accurate and reliable measurement of probe tip internal temperature, and the successful prevention of MRI-induced heating. The reported performance is that the new FOTS system achieves this by addressing the root cause of the previous recall.
    2. Sample size used for the test set and the data provenance

      • The document describes "design verification and design validation processes" but does not specify sample sizes for these tests.
      • Data provenance: Not explicitly stated, but given it's a medical device for intracranial use, the testing would generally be conducted in controlled laboratory or simulated environments, primarily in the US (where Monteris Medical is based). The study is retrospective in the sense that it's a modification to an existing device in response to a recall, and testing would validate the new design. It is not a clinical study on patients.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

      • This information is not provided. The ground truth for device performance in this context would likely be established through engineering measurement standards, calibration against established temperature reference standards, and MRI compatibility testing protocols. These would involve engineers and technical experts in fields like laser physics, MRI compatibility, and materials science, rather than medical experts for ground truth establishment for this specific modification.
    4. Adjudication method for the test set

      • Adjudication methods (like 2+1, 3+1) are typically used for interpreting human-derived data like imaging reads. This scenario involves technical performance testing of a physical and software modification. Therefore, an adjudication method in this sense is not applicable and not mentioned.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No, an MRMC comparative effectiveness study was not done. This submission is for a modification to a laser ablation system's internal temperature sensor, not for an AI-assisted diagnostic tool. The device's primary function (laser ablation) and the interpretation of thermographic analysis by physicians remain unchanged.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • The core modification is the replacement of a physical temperature sensor (thermocouple with Fiber Optic Temperature Sensor - FOTS) and associated hardware/software changes. The performance of the FOTS itself, in terms of accurate temperature measurement and signal transmission to the system's software, would have been tested in a standalone capacity (i.e., the sensor's accuracy and behavior independent of a human operator, but within the device system) as part of the "Design Verification" process. The document does not provide details on these specific tests or results.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • The ground truth for the temperature sensor's performance would be based on physical measurement standards. This would involve:
        • Calibration against known temperature references.
        • Evaluation of the sensor's accuracy and precision according to established engineering and metrology standards.
        • Validation that the sensor effectively measures the probe tip's internal temperature and that the system controls cooling appropriately.
        • Crucially, validation that the FOTS does not interact adversely with MRI fields (the root cause of the recall), which would involve specialized MRI compatibility testing.
    8. The sample size for the training set

      • This is not applicable. This device is a hardware and software system, not a machine learning model that requires a "training set" in the conventional sense. The software modifications are to integrate the new sensor signal, not to train an algorithm using data.
    9. How the ground truth for the training set was established

      • This is not applicable as there is no training set mentioned or implied for a machine learning model. The software changes are integration and adaptation for the new sensor's signal.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173305

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-11-17

    (30 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through internal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided intracranial delivery of precision thermal therapy in the practice of neurosurgery.

    As previously described in K172881, the NeuroBlate™ System is typically used for the minimally invasive ablation of target tissue (tumors, epileptic foci) in the brain.

    The NeuroBlate™ System components consist of:

    • Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver . controlled energy to a target zone.
    • . Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Probe Drivers and interconnections for the Laser Delivery Probes (e.g., Connector Module):
    • A System Electronics Rack and Components, which includes the laser and necessary ● umbilicals, cables, penetration panels, and small hardware for system mechanical. electrical, and electronic operation,
    • . A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate™ procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate™ System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • . The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, fiducial markers, bone screws, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools).

    Thus, there is no change to entire system, with the exception of an additional (alternative) NeuroBlate M-Vision/ M-Vision Pro Software Package (V3.13). While the modified software package includes several modifications which do not meet the FDA's "significant change" submission criteria, there is at least one proposed change which modifies an existing risk control measure for a hazardous situation and thus, per FDA 510(k) guidance, requires a new 510(k) submission.

    AI/ML Overview

    The provided text describes the Monteris Medical NeuroBlate™ System (K173305), which is a modification of a previously cleared device (K172881). The submission is for an updated M-Vision/M-Vision Pro Software Package (V3.13). The document primarily focuses on demonstrating substantial equivalence to the predicate device due to a software update that modifies an existing risk control measure.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document mentions that "Each verification test protocol incorporated clearly defined acceptance criteria" and "Each validation protocol described the objective, test method and acceptance criteria." However, it does not provide the specific acceptance criteria or the reported performance data for these tests. It only states that the tests "confirmed (and documented) the design output met the design input for the requirements" and that the "modified NeuroBlate™ System met the user needs and intended use."

    To fill the table below, I must infer the general nature of the criteria based on the device and its intended use, as specific details are not provided. The device performs planning and real-time thermographic analysis for thermal therapy. Therefore, acceptance criteria would likely relate to the accuracy, precision, and reliability of these functions.

    Metric (Inferred)Acceptance Criteria (Not explicitly stated, inferred)Reported Device Performance (Not explicitly stated, inferred)
    Design Verification:
    Accuracy of trajectory planning assistanceTrajectory planning assistance provided by the M-Vision/M-Vision Pro software should meet predefined accuracy specifications to ensure precise placement of probes.The design verification process utilized protocols and corresponding test reports confirmed that the design output met the design input for the requirements, implying that the trajectory planning assistance performs as specified.
    Accuracy of real-time thermographic analysisReal-time thermographic analysis provided by the M-Vision/M-Vision Pro software should accurately reflect thermal changes in tissue within specified tolerances.The design verification process confirmed that the design output met the design input for the requirements, meaning the thermographic analysis functions within its specified accuracy.
    Software functionality and reliabilityAll functions of the M-Vision/M-Vision Pro V3.13 software should perform as intended without errors or crashes, adhering to documented specifications.The design verification process confirmed that the design output met the design input for the requirements, implying that the software is functional and reliable as per its specifications.
    Risk control measure effectiveness (modified)The modified risk control measure in V3.13 software should effectively mitigate the identified hazardous situation to an acceptable level, as per risk assessment.The submission for K173305 was specifically triggered because the updated software modifies an existing risk control measure for a hazardous situation. The design verification and validation process documented the meeting of these requirements, implying the modified risk control measure is effective.
    Design Validation (User Needs & Intended Use):
    System performance in intended useThe NeuroBlate™ System with V3.13 software should effectively ablate, necrotize, or coagulate intracranial soft tissue when used by a trained physician, and information provided should be useful for thermal therapy assessment.The design validation process utilized protocols and corresponding test reports confirmed that the modified NeuroBlate™ System met the user needs and intended use. This implies the system functions as intended and provides useful information for patient management when interpreted by a trained physician, although specific outcomes are not quantified.
    MRI compatibility and visualizationThe system's MRI compatibility and ability to provide MRI-based trajectory planning and real-time thermographic analysis should perform reliably in an MRI environment.The design validation process confirmed the system met user needs and intended use, which would encompass the proper functioning of MRI compatibility and visualization features.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size for the test set used in the design verification or design validation. It only mentions that "The Design Verification process utilized protocols to detail the associated tests. The corresponding test reports which confirmed (and documented) the design output met the design input for the requirements." Similarly, for validation, "The Design Validation process utilized protocols to detail the associated tests. The corresponding test reports which confirmed (and documented) the modified NeuroBlate™ System met the user needs and intended use."

    The data provenance (e.g., country of origin, retrospective or prospective) is also not mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth for any test sets. The indications for use state "When interpreted by a trained physician," implying that qualified medical professionals are involved in the use and interpretation of the system, but this doesn't directly address the establishment of ground truth for testing purposes.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    The provided text does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The submission is for a software update to an existing device, focusing on maintaining substantial equivalence rather than a new clinical effectiveness claim through an MRMC study.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The document describes software (M-Vision/M-Vision Pro) that provides "trajectory planning assistance" and "real-time thermographic analysis." It explicitly states, "Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis," indicating that a human (trained physician) is always intended to be in the loop.

    Therefore, it is highly likely that standalone algorithm-only performance was evaluated as part of the design verification (e.g., measuring the inherent accuracy of planning algorithms or thermographic calculations), but not as a replacement for human-in-the-loop performance in the clinical application. The document indicates that the software provides "information that may be useful" and requires interpretation by a trained physician.

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for either verification or validation. Given the nature of a laser thermal therapy system, potential ground truths could include:

    • Physical measurements/phantoms: For accuracy of laser delivery and thermal profile generation during verification.
    • Pathology/histology: To confirm ablation zones in preclinical or bench testing.
    • Expert consensus/clinical outcomes: To evaluate the system's utility and accuracy in real or simulated clinical scenarios during validation, although this is inferred.

    The document states the purpose of the software update was for a "risk control measure for a hazardous situation," suggesting that the ground truth for evaluating this might involve risk analyses, simulated failure modes, and verification of corrective actions.

    8. The Sample Size for the Training Set

    The document does not mention a training set sample size. This is because the submission concerns a software update (V3.13) for an existing system, and the description focuses on verification and validation of changes rather than the development of a new AI/ML model that typically involves distinct training sets.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is discussed or implied for a new AI/ML model, the establishment of ground truth for a training set is not applicable in this context. The V3.13 software is likely a deterministic software update rather than an adaptive AI system requiring a 'training set' in the machine learning sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172881

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-20

    (29 days)

    Product Code
    Regulation Number
    878.4810
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures, through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

    As previously described in K170724 and K171255, the NeuroBlate™ System is typically used for the minimally invasive ablation of target tissue (tumors, epileptic foci) in the brain.

    The NeuroBlate™ System components consist of:

    • Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver . controlled energy to a target zone.
    • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon . to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides . supporting electronics for the Advanced, and Robotic, Probe Drivers and interconnections for the Laser Delivery Probes (e.g., Connector Module);
    • A System Electronics Rack and Components, which includes necessary umbilicals, . cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation,
    • A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which . includes a user interface for procedure planning, interactive monitoring of NeuroBlate™ procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate™ System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation . components, and
    • The AtamA Stabilization System and MRI receive-only head coil, as well as, other ● optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, fiducial markers, bone screws, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools).

    Thus, there is no change to entire system, with the exception of an additional (alternative) NeuroBlate System modified Laser Probe Umbilicals.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modification to the Monteris Medical NeuroBlate™ System. The primary modification involves "an additional (alternative) NeuroBlate System modified Laser Probe Umbilicals which incorporate a modified Thermocouple assembly manufacturing process and several umbilical related Strain Relief component updates."

    Crucially, the document states that this submission involves no change to the indications for use, intended use, operating principles, or core technology of the NeuroBlate™ system. It asserts substantial equivalence to previous versions (K081509, K120561, K131278, K131955, K141983, K143457, K162762, K170724 and K171255) based on design verification and validation of the modified umbilical components, not a study of the overall clinical performance of the device or its AI components (such as thermographic analysis).

    Therefore, this document does not contain information about a study that measures device performance against acceptance criteria in the way typically expected for an AI/CADe device's clinical efficacy or standalone performance. The "acceptance criteria" discussed are related to engineering and design verification/validation of a modified component, not clinical outcomes or diagnostic accuracy.

    Given this context, I cannot directly provide the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the format you specified, as those types of studies (MRMC, standalone performance, expert ground truth for a test set) are not described for this specific submission. This is a component modification submission, not a new device or a new AI feature submission that would typically require such clinical performance studies.

    However, I can extract the information that is present regarding the validation process for the modified umbilical:


    1. A table of acceptance criteria and the reported device performance

    The document mentions that "Each verification test protocol incorporated clearly defined acceptance criteria" for the design verification process, and "Each validation protocol described the objective, test method and acceptance criteria" for the design validation process concerning the modified Laser Probe Umbilicals. However, the specific acceptance criteria and the reported numeric performance results for these criteria are NOT provided in this document. It generally states that these processes confirmed the design output met design input requirements and that the modified system met user needs and intended use.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The testing described is design verification and validation of a medical device component (umbilicals), not a clinical study with patient data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. Ground truth as typically understood for clinical performance studies (e.g., expert consensus) is not mentioned because this submission focuses on component-level engineering validation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. Adjudication methods are typically relevant for clinical studies involving human interpretation or uncertain diagnoses, which are not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not described in this document. This submission is for a modification to a component of an existing system, not a new AI-assisted diagnostic or therapeutic tool requiring such a comparative effectiveness study. The "real-time thermographic analysis" mentioned in the device's indications for use is part of the existing system, and this submission does not detail new studies on its performance or the effect size of AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study of an algorithm was not described in this document for this submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided in the document. The validation pertains to component functionality rather than clinical ground truth for diagnostic accuracy.

    8. The sample size for the training set

    This information is not provided in the document. This submission describes component modification and validation, not the development or training of an AI algorithm with a training set.

    9. How the ground truth for the training set was established

    This information is not provided in the document. No training set is mentioned as this device modification does not involve an AI model requiring a new training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170724

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-07

    (29 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstital irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real-time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

    As previously described in K143457 and K162762, the NeuroBlate™ System is typically used for the minimally invasive ablation of target tissue (tumors, epileptic foci) in the brain.

    The NeuroBlate™ System components consist of:

    • Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver . controlled energy to a target zone.
    • . Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • . An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced, and Robotic, Probe Drivers and interconnections for the Laser Delivery Probes (e.g., Connector Module);
      • In consideration that the proposed change is an alternative (portable) Connector o Module (PCM) subcomponent, related Connector Module discussion topics are separated out from the overall Interface Platform (IP) description, discussion and comparison purposes.
    • A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation,
    • A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which ● includes a user interface for procedure planning, interactive monitoring of NeuroBlate™ procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate™ System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • . The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • . The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, fiducial markers, bone screws, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools).

    Thus, there is no change to entire system, with the exception of an additional (alternative) Portable Connector Module (PCM). The PCM provides the capability to be mounted in various positions on the Interface Platform (IP) Arms, utilizes an externalized laser fiber connector as well as alternative commercially available thermocouple cable connectors (i.e., connectors which utilize wires soldered to connector pins compared to the commercially available connectors which utilize wires crimped to connector pins).

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Monteris Medical NeuroBlate™ System. It focuses on the substantial equivalence of a modified version of the system (with an alternative Portable Connector Module - PCM) to its predicate devices.

    Crucially, this document does NOT contain the detailed acceptance criteria or the study results proving the device meets those criteria in the context of AI/algorithm performance.

    The document discusses the regulatory approval process (510(k)), device description, indications for use, comparison to predicate device, and a summary of supporting data related to design verification and validation for the specific modification (PCM). It states that the PCM development followed Monteris' Quality System, including design verification and validation processes with defined acceptance criteria. However, it does not provide those specific acceptance criteria, nor does it detail studies related to the clinical performance of an AI/algorithm component (like sensitivity, specificity, accuracy, or human reader improvement with AI assistance).

    The "M-Vision™ and M-Vision Pro™ software" are mentioned as including a user interface for procedure planning and interactive monitoring, but there's no information on them being an AI/algorithm that performs diagnostic or prognostic tasks that would require the kind of performance studies you're asking about. The software provides "real-time thermographic analysis of selected MRI images" and "information that may be useful in the determination or assessment of thermal therapy," explicitly stating that "Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis." This suggests it's a tool providing raw data or analysis assistance, not an AI for automated diagnosis or a "human-in-the-loop" performance study in the way typically discussed for AI/ML diagnostic devices.

    Therefore,Based on the provided text, I cannot answer the questions regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, or MRMC studies because this information is not present. The document focuses on regulatory approval for hardware modifications to an existing system, not the performance validation of a new or modified AI/algorithm component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162762

    Validate with FDA (Live)

    Date Cleared
    2016-10-26

    (26 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

    As previously described in K143457, the NeuroBlate System is typically used for the minimally invasive ablation of target tissue (tumors, epileptic foci) in the brain.

    The NeuroBlate System components consist of:

    • Families of gas-cooled Laser Delivery Probe (SideFire & FullFire) to deliver controlled energy to a target zone.
    • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced and Robotic Probe Drivers and interconnections for the Laser Delivery Probes;
    • A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation.
    • A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.

    The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:

    • The Axiiis stereotaxic mini-frame and the Monteris Cranial Bolt and Mini-Bolt fixation components, and
    • The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.
    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Monteris Medical NeuroBlate™ System. It states that no new performance data or studies were required for this specific submission because the changes are limited to labeling modifications and corrections, primarily due to a Class 1 Recall and FDA 483 observations, and do not involve physical, manufacturing, process, materials, or technology changes to the device itself.

    Therefore, most of the requested information regarding acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert adjudication, and MRMC studies is not applicable to this particular 510(k) submission. The document explicitly states: "Considering the proposed changes are labeling related, the previously provided in-vitro (bench) data remains applicable and no additional bench testing was required."

    However, I can extract the relevant information regarding the device and the nature of this submission:


    1. A table of acceptance criteria and the reported device performance

    Since this submission did not involve new performance studies, specific acceptance criteria for new performance tests are not detailed. The submission focuses on compliance with regulatory requirements for labeling.

    Acceptance Criteria CategoryReported Device Performance Basis (from prior submissions)
    Device FunctionalityThe device's fundamental functional aspects (ablation, necrotization, coagulation of soft tissue via interstitial irradiation/thermal therapy with 1064nm lasers, MRI-based trajectory planning assistance, real-time thermographic analysis) remain unchanged from the predicate device. Prior in-vitro (bench) data still applies.
    Labeling ComplianceUpdates to labeling comply with 21 CFR 1000-1050 (Radiological Health requirements) and FDA Laser Notice #50 requirements, ensuring clarity and correction of information, especially regarding the discontinued 2.2mm SideFire Probes (SFS) and existing warnings/precautions.
    Substantial EquivalenceThe device with modified labeling is substantially equivalent to the predicate Monteris NeuroBlate™ System (K120561, K141983, K143457) in intended use, technology, design, and physician use, as modifications do not change operating principles or raise unaddressed safety concerns.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable to this submission as no new performance studies (and thus no new test sets) were conducted. The submission relies on previously conducted studies for the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable to this submission as no new performance studies were conducted. Clinical interpretation by a "trained physician" is mentioned in the Indications for Use for assessing thermal therapy, but this pertains to clinical use, not a specific study for this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable to this submission as no new performance studies were conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The NeuroBlate System is a thermal therapy device with software for planning and monitoring, not an AI diagnostic tool for which MRMC comparative effectiveness studies with human readers would typically be performed in this context. The document indicates physician interpretation, but not in the context of an MRMC study for this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device's software provides "real time thermographic analysis" and "MRI based trajectory planning assistance." The "Indications for Use" explicitly states: "When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis." This indicates a human-in-the-loop design where the software provides information to a physician, rather than acting as a standalone diagnostic algorithm. No new standalone performance study (algorithm-only) was reported for this submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable to this submission as no new performance studies were conducted. For the underlying efficacy and safety of the thermal ablation, ground truth in previous studies would likely have involved histopathology or clinical outcomes, but this is not detailed in the provided document for this specific filing.

    8. The sample size for the training set

    Not applicable to this submission as no new performance studies were conducted. If the system incorporates machine learning components (which is not explicitly detailed but might be part of "thermographic analysis"), any training data would have been from prior development phases and not detailed in this specific labeling-focused submission.

    9. How the ground truth for the training set was established

    Not applicable to this submission as no new performance studies were conducted.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143457

    Validate with FDA (Live)

    Date Cleared
    2015-02-19

    (78 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

    The NeuroBlate System components consist of:

    • . Families of gas-cooled Laser Delivery Probes (Probe) (SideFire & FullFire) to deliver controlled energy to a target zone. This application includes the smaller diameter SideFire Select and FullFire Select families of Probes;
    • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the . surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
    • An Interface Platform, which attaches to the MRI system patient table and . provides supporting electronics for the Probe Drivers and interconnections for the Laser Delivery Probes;
    • A System Electronics Rack and Components, which includes necessary . umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation, and
    • A Control Workstation including the M-Vision™ software, which includes a user . interface for procedure planning, interactive monitoring of NeuroBlateTM procedures, and interfaces to the MRI and hardware subsystems.

    This submission adds a line of 2.2mm Reduced Diameter Probes to the existing 3.3mm diameter probes. They will be known as the SideFire Select™ and FullFire Select™ Probes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text for the Monteris Medical NeuroBlate™ System (K143457):

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive de novo clinical study with detailed performance metrics and ground truth establishment. Therefore, much of the requested information (like specific acceptance criteria values, sample sizes for test/training sets, number of experts, adjudication methods, MRMC studies, and standalone performance metrics) is not explicitly available in this type of document.

    The document primarily relies on bench testing to demonstrate that the new, smaller diameter probes function equivalently to the previously cleared larger diameter probes.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (from bench testing)
    Electrical RequirementsNew Laser Delivery Probes exhibited compliance to the same electrical requirements as predicate sizes.
    Mechanical Integrity and Operation (e.g., Cooling)New Laser Delivery Probes exhibited compliance to the same mechanical integrity and operation (e.g., cooling) requirements as predicate sizes.
    Laser Emission RequirementsNew Laser Delivery Probes exhibited compliance to the same laser emission requirements as predicate sizes.
    Interface with NeuroBlate SystemNew Laser Delivery Probes interfaced with the NeuroBlate System in an identical manner to predicate sizes.
    Adequate Labeling for Accurate PlacementAdequate labeling information was provided to ensure accurate placement within the target tissue.
    Laser Emission Patterns and Ablation EffectThe application System with the Reduced Diameter Laser Delivery Probes has similar laser emission patterns and laser ablation effect as the predicate Laser Delivery Probes.
    Overall Performance Equivalence to PredicateThe NeuroBlate system works as well with the Reduced Diameter Probe as it does with the larger diameter versions.
    Compliance with Medical Community ExpectationsThe device is in compliance with the medical community's expectations and product labeling and specifications.

    Note: The document states that the new probes are "compliant to the same electrical requirements" or "exhibited compliance to the same electrical requirements" etc. It does not provide the specific numerical values or thresholds for these requirements, but rather asserts that the new device meets whatever those pre-established requirements were for the predicate device.

    2. Sample Sizes and Data Provenance

    • Test Set Sample Size: Not specified. The testing described is bench testing, not a clinical trial with a defined patient test set.
    • Data Provenance: Bench testing data, implies in-house laboratory testing. No country of origin for data specifically is mentioned, but the manufacturer is Monteris Medical Corp. located in Plymouth, MN, USA. The testing is retrospective in the sense that it evaluates a modified device against established performance characteristics of the predicate. It is not prospective clinical data.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Not applicable as the study described is primarily bench testing comparing a modified device to a predicate, not clinical performance requiring expert ground truth for interpretation.
    • The indications for use state: "When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis." This indicates that the system is intended to be used by trained physicians, but this is not part of the ground truth establishment for the reported bench tests.

    4. Adjudication Method (Test Set)

    • Not applicable, as the study described is bench testing and does not involve human interpretation or adjudication in the context of clinical outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study is not mentioned or indicated in the document. The document focuses on demonstrating substantial equivalence to a predicate device through bench testing of physical and functional characteristics. The "AI vs without AI assistance" aspect is not relevant here as the device is a laser thermal therapy system, not an AI diagnostic algorithm.

    6. Standalone Performance Study (Algorithm only without human-in-the-loop performance)

    • The document implies standalone performance through the bench testing described, where the device components (probes, interface, workstation) are tested for their individual and integrated functions. However, "standalone" in the context of AI algorithms typically refers to the algorithm's diagnostic performance without human input. This device is a surgical instrument and monitoring system, meaning its performance by nature is intertwined with human use (e.g., a "trained physician" interpreting thermographic analysis for patient management decisions). Thus, a pure "algorithm only" standalone performance comparison is not directly applicable in the same way it would be for an AI diagnostic tool.

    7. Type of Ground Truth Used

    • For the bench testing, the "ground truth" implicitly refers to the established specifications and performance characteristics of the predicate device, against which the modified probes were compared. This would include measurable physical properties (e.g., laser power, cooling efficiency), mechanical tolerances, and interface compatibilities. It is not expert consensus, pathology, or outcomes data in the clinical sense.

    8. Sample Size for the Training Set

    • Not applicable. The document describes bench testing for substantial equivalence, not the development or training of an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no mention of a "training set" in the context of AI algorithm development.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141983

    Validate with FDA (Live)

    Date Cleared
    2014-08-11

    (21 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

    The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.

    When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

    Device Description

    The Monteris Medical NeuroBlate™ System is a unique collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy. The NeuroBlate System components consist of:

    • . Gas-cooled Laser Delivery Probes (Probes) to deliver controlled energy to a target zone;
    • A Probe Driver which allows the surgeon to precisely position, stabilize and . manipulate a laser probe within the target zone;
    • . A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation; and
    • A Control Workstation including the M Vision™ Software, which includes a user . interface for procedure planning, interactive monitoring of NeuroBlate™ procedures, and interfaces to the MRI and hardware subsystems.

    This submission clears the use of the NeuroBlate System with specific 1.5 & 3.0T Philips Magnetic Resonance Imaging Systems.

    AI/ML Overview

    The provided text describes the Monteris Medical NeuroBlate™ System, its indications for use, and a comparison to a predicate device. However, it does not contain information about acceptance criteria or specific studies proving the device meets those criteria in the context of device performance metrics like accuracy, sensitivity, or specificity.

    The document details regulatory information related to the 510(k) premarket notification process for the NeuroBlate System, specifically its compatibility with additional MRI systems (Philips). The "Summary of Supporting Data" section briefly mentions "Software and bench testing has demonstrated that the System is in compliance with the medical community's expectations and the product labeling" and that it "works as well with the Philips MRI's as it does with the Siemens, IMRIS, and GE MRI's." This is a general statement and does not provide quantified acceptance criteria or detailed study results.

    Therefore, I cannot fulfill your request for the following information based on the provided text:

    1. A table of acceptance criteria and the reported device performance: This information is not present.
    2. Sample size used for the test set and the data provenance: Not mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    4. Adjudication method for the test set: Not mentioned.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned. The device described is a laser thermal therapy system, not an AI-assisted diagnostic tool, so an MRMC study in that context would not be applicable.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not mentioned, and generally not applicable for a device of this type which is an instrument used by a physician.
    7. The type of ground truth used: Not mentioned.
    8. The sample size for the training set: Not mentioned.
    9. How the ground truth for the training set was established: Not mentioned.

    The document focuses on demonstrating substantial equivalence to a predicate device for the purpose of extending compatibility with additional MRI systems, rather than presenting detailed performance efficacy data typically associated with studies proving device meets acceptance criteria related to accuracy or other clinical metrics for a diagnostic or AI-driven device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2