Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172499

    Validate with FDA (Live)

    Device Name
    NEOSHIELD
    Date Cleared
    2018-06-01

    (287 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NEOSHIELD System is an airtight and leak proof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills.

    The NEOSHIELD system also prevents microbial ingress for 3 days.

    Device Description

    NeoShield is a needle free closed system drug transfer device (CSTD). Neoshield prevents transfer of environmental contaminants into the system and also prevents the escape of drug or vapor from the system.

    The ability to prevent microbial ingress for up to 3 days should not be interpreted as modifying, extending, or superseding a manufacturer labeling recommendations for the storage and expiration dating. Refer to drug manufacturer's recommendations and USP compounding guidelines for shelf life and sterility information.

    AI/ML Overview

    The provide text is a 510(k) summary for a medical device called NEOSHIELD, an intravascular administration set. It does not describe a study involving humans or AI, but rather a set of non-clinical (bench) performance tests to demonstrate substantial equivalence to a predicate device.

    Here's the information extracted and formatted as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test NameAcceptance Criteria (Implied by "No leak" or "Pass")Reported Device Performance
    Leaking Pressure of Pre-Slit SeptumNo leakNo leak
    Pressure Resistant Test (Liquid Tightness)No leakNo leak
    Pressure Resistant Test (Air tightness)No leakNo leak
    Pressure Resistant Test (Connecting Part of Closed System)No leakNo leak
    Airtight ConnectionNo leakNo leak
    Leakproof ConnectionsNo leakNo leak
    Microbial Ingress TestingNo ingress for 3 daysNo ingress for 3 days
    Pyrogen testNon-pyrogenicNon-pyrogenic
    Crack Test for drug compatibilityNo cracks (except for N, N-dimethylacetamide, which is an exclusion)No cracks except for N, N-dimethylacetamide
    Sterility Testing (EO) (ISO 11135)SterileSterile
    Shelf Life TestingSterility kept for 3 yearsSterility kept for 3 years
    Biocompatibility TestingAll materials are biocompatible (Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Systematic Toxicity (Acute), Haemocompatibility, Pyrogen, Particulate Matter)All materials are biocompatible
    Package Testing: Seal Strength/dye penetration testingPass/PassPass/Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes for each non-clinical test. It states that "The following non-clinical tests were conducted," and then lists the tests and their results. The data provenance is from bench testing conducted by the manufacturer, JMS Co., Ltd / JMS Singapore Pte Ltd. This is a retrospective analysis of test results.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    Not applicable. This was a non-clinical (bench) testing study, not a study involving human interpretation or expert ground truth establishment in the traditional sense of clinical studies. The "ground truth" for these tests is based on objective laboratory measurements against defined standards or pass/fail criteria.

    4. Adjudication Method for the Test Set:

    Not applicable, as this was non-clinical bench testing. The results are typically objectively measured against pre-defined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    No, an MRMC comparative effectiveness study was not done. This submission focuses on non-clinical performance and substantial equivalence to a predicate device, not on comparing human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. The device, NEOSHIELD, is an intravascular administration set (hardware), not an AI algorithm.

    7. The Type of Ground Truth Used:

    The ground truth used for these non-clinical tests is based on objective physical and chemical measurements against established engineering and regulatory standards (e.g., "no leak," "sterile," "non-pyrogenic," "biocompatible").

    8. The Sample Size for the Training Set:

    Not applicable. As a hardware device, there is no "training set" in the context of an AI algorithm. The device's design and manufacturing processes are developed through engineering and testing, but not machine learning training.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1