Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120182

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-04-19

    (90 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended for short-term use < 30 days in neonatal and pediatric patients to provide nutrition via nasal or oral gastric placement.

    Device Description

    The NeoMed Gastrointestinal Tube and Accessories is a PVC single lumen catheter that is used to deliver liquid nutritional media. This NeoMed PVC feeding tube is designed to connect with other NeoMed accessories such as the NeoMed Extension tubing under FDA clearance K100288 and NeoMed oral syringe under FDA clearance K092908.

    The device consists of the following main components: a feeding tube single lumen catheter and a one-piece molded enteral only connector hub with integral tethered connection closure plug. The catheter tubing has an orange radiopaque stripe of barium sulfate embedded in the tubing wall which can be visualized on x-ray, for exact placement of the tip. The tip is closed. The single lumen catheter tubing has side holes for better flow and to provide multiple openings for aspiration.

    AI/ML Overview

    The provided text describes a 510(k) submission for the NeoMed Gastrointestinal Tube and Accessories, which seeks substantial equivalence to existing predicate devices. The study conducted to demonstrate this equivalence is primarily an engineering bench test study rather than a clinical study involving human patients or complex AI algorithms.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (General Purpose of Test)Reported Device Performance
    BiocompatibilityMaterials have a long history of use in feeding tube manufacturing, are biocompatible, and testing in accordance with ISO 10993-1: Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing passed.
    Tensile StrengthDesign verification functional test results, including tensile strength, demonstrate that the NeoMed Gastrointestinal Tube and Accessories performs its intended use.
    DimensionsDesign verification functional test results, including dimensions, demonstrate that the NeoMed Gastrointestinal Tube and Accessories performs its intended use.
    Non-InterconnectabilityDesign verification functional test results, including non-interconnectability, demonstrate that the NeoMed Gastrointestinal Tube and Accessories performs its intended use.
    Priming VolumeDesign verification functional test results, including priming volume, demonstrate that the NeoMed Gastrointestinal Tube and Accessories performs its intended use.
    Substantial Equivalence (Overall)Bench testing has demonstrated that the NeoMed Gastrointestinal Tube and Accessories is functionally equivalent to the proposed predicate feeding tubes. The device is concluded to be equivalent to the Vygon Nutrisafe 2 (KQ60944) and Kendall Argyle PVC Feeding Tube (K820441) with respect to intended use, materials, design, and technological characteristics.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes "bench testing" and "design verification functional test results." It does not specify a "test set" in the context of patient data or algorithm performance. The testing involved physical tubes and components. No information is provided regarding the specific number of units tested for each functional test (tensile strength, dimensions, etc.), nor is there any mention of data provenance (e.g., country of origin) as the tests are laboratory-based engineering evaluations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The study is a bench test, and there is no mention of experts establishing a "ground truth" for a test set in the context of clinical or image-based evaluations. The "ground truth" for these engineering tests would be the established engineering specifications and performance standards.

    4. Adjudication Method for the Test Set

    Not applicable. There is no mention of adjudication, as this study does not involve human interpretation or subjective assessments that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This study is an engineering bench test for physical device characteristics and does not involve human readers, AI assistance, or comparative effectiveness in a clinical setting.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    No. This device is a physical medical tube, not an AI algorithm. Therefore, no standalone algorithm performance study was conducted.

    7. Type of Ground Truth Used

    The ground truth used for this study is based on engineering specifications, material properties, and functional performance standards derived from the intended use and comparison with predicate devices. For example, for dimensions, the ground truth would be the specified dimensions in the design; for tensile strength, it would be the minimum required strength; for biocompatibility, it would be successful compliance with ISO 10993-1.

    8. Sample Size for the Training Set

    Not applicable. This study does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This study does not involve a "training set" for which ground truth would need to be established.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1